K053268 · Boston Scientific - Precision Vascular · DQX · Mar 13, 2006 · Cardiovascular
Device Facts
Record ID
K053268
Device Name
PVS 2000 SYNCHRO2 GUIDEWIRE
Applicant
Boston Scientific - Precision Vascular
Product Code
DQX · Cardiovascular
Decision Date
Mar 13, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1330
Device Class
Class 2
Indications for Use
The PV 2000 Synchro®2 Guidewire series of products is intended for peripheral and neurovascular use. The device can be used to selectively introduce and position catheters and other interventional devices within the peripheral and neurovasculature. This device should be used only by physicians trained in percutaneous, intravascular techniques and procedures.
Device Story
PV 2000 Synchro®2 Guidewire is a sterile, single-use, 0.014" diameter guidewire designed for navigating small, tortuous peripheral and neurovasculature. It features a shapeable distal tip, Nitinol construction, and a lubricious coating. A platinum marker coil at the distal tip enables fluoroscopic visualization. The device is manually operated by physicians during percutaneous, intravascular procedures to facilitate catheter exchange and positioning. It is supplied with a torque device and an introducer/shaping mandrel to assist in navigation and insertion. The guidewire provides steerability and torque transmission, allowing clinicians to reach distal vascular sites. By enabling precise catheter placement, the device facilitates diagnostic and interventional procedures, potentially improving patient outcomes in complex vascular anatomy.
Clinical Evidence
Bench testing only. Testing conducted in accordance with ISO 11070:1998. Performance evaluations included dimensional inspection, fatigue assessment, tip flexibility, tip shaping, tensile strength, guidewire compatibility, torsional strength, torqueability, and EO sterilization validation. Biocompatibility verified per ISO 10993-1 for external communicating, blood contact, short duration (<24 hours) devices. All results were acceptable.
Technological Characteristics
0.014" diameter guidewire; Nitinol distal tip; platinum marker coil for fluoroscopic visualization; lubricious coating on distal portion. Materials verified per ISO 10993-1. Performance tested per ISO 11070:1998. Supplied with torque device and introducer/shaping mandrel. Sterilized via Ethylene Oxide (EO).
Indications for Use
Indicated for general intravascular use, including neuro and peripheral vasculature, to selectively introduce and position catheters and other interventional devices. For use by physicians trained in percutaneous, intravascular techniques.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
B. Braun guidewire introducer/shaping mandrel (K760389)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a date and a number. The date at the top of the image is July 7, 2026. Below the date is the number 1053268. The text is black against a white background.
page 1 of 2
## Summary of Safety and Effectiveness
| Prepared November 22, 2005<br>Revised March 2, 2006 | | | | | |
|--------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|-------------------|
| General<br>Provisions | Submitter of 510(k)<br>Premarket Notification: | | Boston Scientific - Precision Vascular<br>2405 West Orton Circle<br>West Valley City, UT 84119<br>Phone: 801.974.1700<br>Fax: 801.974.1740 | | |
| | Contact Person: | | Rick Gaykowski<br>VP, Regulatory/Clinical Affairs<br>Quality Systems & Operations | | |
| | Device Trade Name:<br>Device Generic Name: | | PV 2000 Synchro®2 Guidewire<br>Guidewire | | |
| | The predicate devices are listed in the table below. | | | | |
| Predicate<br>Devices | Device | Manufacturer | | 510(k) Number/Date | Pro Code |
| | Synchro™ .014" | Precision Vascular | | K032146, 12 August 2003 | DQX |
| | Synchro™ .010"<br>Synchro™ .016"<br>Transend | Precision Vascular<br>Precision Vascular<br>Target Therapeutics | | K032146, 12 August 2003<br>K050964, 13 June 2005<br>K971254, 01 July 1997* | DQX<br>DQX<br>DQX |
| * This information is assumed based upon our best knowledge. | | | | | |
| Classification | Class II, 21 CFR 870.1330, Wire, Guide, Catheter 74DQX | | | | |
| Performance<br>Standards | Performance standards have not been established by FDA under section 514 of the<br>Federal Food, Drug and Cosmetic Act. | | | | |
| Intended Use | The PV 2000 Synchro®2 Guidewire series of products is intended for peripheral and<br>neurovascular use. The device can be used to selectively introduce and position<br>catheters and other interventional devices within the peripheral and neurovasculature.<br>This device should be used only by physicians trained in percutaneous, intravascular<br>techniques and procedures. | | | | |
| Device<br>Description | The PV 2000 is a member of the Synchro® Guidewire family of products having a<br>0.014" outside diameter, being a sterile, single use/disposable product, with a<br>shapeable tip which is used to gain intravascular access to and facilitate the positioning<br>and exchange of interventional devices in small diameter, tortuous vasculature for<br>peripheral and neuro diagnostic and interventional procedures. The guidewire can be<br>torqued to facilitate navigation through the vasculature. A torque device, (Merit Medical<br>(K936032)) is supplied with the wire to facilitate deployment & positioning. A guidewire | | | | |
Page 11 of 64
{1}------------------------------------------------
K05 3268
page 2042
introducer/shaping mandrel (B. Braun (K760389)) is also supplied and may be used to aid introduction of the guidewire into the catheter hub and/or hemostasis valve and to gently shape the guidewire's distal flexible tip, if desired, according to standard practice. Neither the guidewire introducer nor the torque device are intended to enter the body. The product is projected to be provided in a 100cm - 300cm length range, with 200cm being nominal. The Nitinol tip length is projected to be presented in a 35cm length. A traditional range of flexibility profiles shall also be provided, ranging from support (stiff) to flex (soft). The device is coated on the outer diameter with a lubricious coating over the distal portion of the device. The marker coil is platinum wire at the distal tip of the device to aid visualization under fluoroscopy. The subject device has the ability to access distal, tortuous vasculature, with steerability and torque transmission properties.
Technological Characteristics Technological similarities between the subject PV 2000 Synchro®2 Guidewire and the PVS 1300 Synchro™ predicate remain identical. This is also true for competitive predicate device features including the basal design and dimensions, generic materials & construction, and hydrophilic coating. There are no new questions raised regarding safety or efficacy of the PVS 1300 Synchro® 0.014" Guidewire.
Biocompatibility of the PV 2000 Synchro®2 Guidewire materials have been verified in accordance with ISO 10993-1, Biological Evaluation of Medical Devices - Part 1. Materials test results confirmed biocompatibility of the subject device when tested as an external communicating, blood contact, short duration (<24 hours) device.
Safety and Performance Tests
Performance testing of materials comprising the PV 2000 Synchro®2 Guidewire was conducted in accordance with ISO 11070:1998, Sterile, Single-Use Intravascular Catheter Introducers. Ventication testing for the subject device included dimensional inspection, fatique assessment, tip flexibility, tip shaping, tensile strength, guidewire compatibility testing and performance under simulated conditions. Subject product testing has yielded acceptable safety & performance outcomes.
In addition, torsional strength, torqueability, and EO sterilization adoption tests also The results of these tests, in conjunction with the vielded acceptable results. substantial equivalence claims as outlined in the premarket notification, effectively demonstrate the PV 2000 Synchro®2 Guidewires' substantial equivalence to the cited predicate devices.
Summary of Substantial Equivalence
Based on the indications for use, technological characteristics, and safety and performance testing, the subject PV 2000 Synchro@2 Guidewire meets the minimum requirements that are considered adequate for its intended use and is substantially equivalent in design, materials, sterilization, principles of operation and indications for use to current commercially available quidewires/cited predicates.
Page 12 of 64
{2}------------------------------------------------
Food and Drug Administration
Image /page/2/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or bird-like figure.
9200 Corporate Boulevard Rockville MD 20850 MAR 1 3 2006
Boston Scientific - Precision Vascular c/o Mr. Rick Gaykowski V.P. Regulatory/Clinical Affairs Quality Systems and Operations 2405 West Orton Circle West Valley City, UT 84119
Re: K053268
PVS 2000 Synchro®2 Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Wire, Guide, Catheter Regulatory Class: Class II Product Code: DQX Dated: February 15, 2006 Received: February 16, 2006
Dear Mr. Gaykowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality
{3}------------------------------------------------
Page 2 - Mr. Rick Gaykowski
systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation 335this (Q) 70g visions 531-542 of the Act); 21 CFR 1000-1050. Control provisions (Soctions 551 - 12 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 This letter with anow you're ough maning of substantial equivalence of your device to a legally premarket notification: "The PDF maing sification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific at notiance at (240) 276-0120. Also, please note the regulation entitled, Connate the Office of Central of Centralion" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small general intermational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Blymmman for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): K053268
Device Name: PV 2000 Synchro®2 Guidewire
Indications For Use: The PV 2000 Synchro®2 Guidewire series of products are intended for general intravascular use, including the neuro and peripheral vasculature. The device can be used to selectively introduce and position catheters and other interventional devices within the peripheral and neurovasculature. This device should be used only by physicians trained in percutaneous, intravascular techniques and procedures.
Prescription Use ﮯ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
ivision Sigh-O ivision of Clardiova
Page 1 of
**510(k) Number** k053268
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.