K053049 · Cardinal Health, Alaris Products · FPA · Jan 12, 2006 · General Hospital
Device Facts
Record ID
K053049
Device Name
ALARIS SAFETY MALE LUER
Applicant
Cardinal Health, Alaris Products
Product Code
FPA · General Hospital
Decision Date
Jan 12, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The Safety Male Luer is intended to provide leak-free handling of potentially hazardous fluids, such as chemotherapy, radioactive isotopes, and blood products, during all three phases of drug handling when the SML is used in conjunction with the SmartSite® Needle Free Valve port: the preparation of the drug, the administration of the drug to the patient, and waste handling.. The Safety Male Luer is intended to reduce environmental surface contamination of the work area.
Device Story
Passive, sealed male luer connector; prevents fluid dripping/spillage during disconnection. Used for handling hazardous/non-hazardous fluids (chemotherapy, isotopes, blood). Operates by internal mechanism that automatically seals upon disconnection from SmartSite® Needle Free Valve or standard female luer. Used by clinicians in clinical settings during drug preparation, administration, and waste disposal. Reduces environmental surface contamination; improves safety for healthcare workers handling hazardous drugs.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Passive male luer connector; internal sealing mechanism; no external cap required for sealing. Compatible with SmartSite® Needle Free Valve and standard female luers. Includes optional female transport cap and priming cap.
Indications for Use
Indicated for the reconstitution, transfer, administration, and disposal of hazardous (chemotherapy, radioactive isotopes, blood products) and non-hazardous fluids. Intended for use with SmartSite® Needle Free Valve ports or standard open female luers.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
K053049
ORIGINAL PREMARKET 510
# SUMMARY OF SAFETY AND EFFECTIVENESS Cardinal Health, Alaris® Products®
## SAFETY MALE LUER
# JAN 1 2 2006
## SUBMITTER INFORMATION
| A. | Company Name: | Cardinal Health, Alaris® Products |
|----|--------------------------------|------------------------------------------------------------------------------------------|
| B. | Company Address: | 10221 Wateridge Circle<br>San Diego, CA 92121-2733 |
| C. | Company Phone:<br>Company Fax: | (858) 458-7830<br>(858) 458-6114 |
| D. | Contact Person: | Stacy L. Lewis<br>Sr. Regulatory Affairs Specialist<br>Cardinal Health, Alaris® Products |
| E. | Date Summary Prepared: | October 25, 2005 |
## DEVICE IDENTIFICATION
| A. | Generic Device Name: | Intravascular Administration Set |
|----|-------------------------|---------------------------------------|
| B. | Trade/Proprietary Name: | Alaris® Safety Male Luer |
| C. | Classification: | Class II |
| D. | Product Code: | FPA, Intravascular Administration Set |
# DEVICE DESCRIPTION
The SML is a sealed male luer. The SML is designed to promote safe handling of fluids and medications, particularly hazardous or cytotoxic drugs. Leakage of drug into the environments is effectively avoided during all phases of drug handling when the SML is used in conjunction with the SmartSite® Needle Free Valve port: the preparation of the drug, the administration of the drug to the patient, and waste handling.
{1}------------------------------------------------
## SUMMARY OF SAFETY AND EFFECTIVENESS Cardinal Health, Alaris® Products SAFETY MALE LUER Page 2 of 3
# DEVICE DESCRIPTION (Continued)
The SML is a passive device – it requires no cap and automatically seals upon disconnection. An optional female cap will be provided for transport and an optional priming cap will be provided to allow the user to prime as necessary.
The Safety Male Luer (SML) is a unique male luer connector that is intended to be used with the currently marketed SmartSite® Needle Free Valve port and standard open female luers. The SML design includes an internal mechanism that causes the luer to seal when disconnected. In doing so, it prevents the dripping or accidental spillage of fluids that otherwise occur when using a standard, unsealed male luer. Thus, the SML is intended to provide leak-free handling of potentially hazardous fluids, such as chemotherapy, radioactive isotopes, and blood products. Furthermore, the SML reduces environmental surface contamination of the work surface when used to reconstitute, dispense/transfer, administer, and dispose of hazardous or non-hazardous fluids when used in conjunction with SmarSite® Needle-Free Valve port.
## SUBSTANTIAL EQUIVALENCE
The Cardinal Health, Alaris® Products Safety Male Luer is of comparable type and is substantially equivalent to the following predicate devices:
| Predicate Device | Manufacturer | 510(k) No. | Date Cleared |
|------------------|-------------------|------------|--------------|
| PhaSeal System | Carmel Pharma AB | K972527 | 9/18/97 |
| CML 1000 | ICU Medical, Inc. | K051437 | 8/3/05 |
### INTENDED USE
The Safety Male Luer is intended to provide leak-free handling of potentially hazardous fluids, such as chemotherapy, radioactive isotopes, and blood products, during all three phases of drug handling when the SML is used in conjunction with the SmartSite®
{2}------------------------------------------------
KOS-304-9
ORIGINAL PREMARK
## SUMMARY OF SAFETY AND EFFECTIVENESS Cardinal Health, Alaris® Products SAFETY MALE LUER Page 3 of 3
## INTENDED USE (Continued)
Needle Free Valve port: the preparation of the drug, the administration of the drug to the patient, and waste handling.. The Safety Male Luer is intended to reduce environmental surface contamination of the work area.
# TECHNOLOGICAL CHARACTERISTICS
A comparison of the technological characteristics of the Alaris® Safety Male Luer and the predicate devices has been performed. The results of this comparison demonstrate that the Alaris® Safety Male Luer is equivalent to the marketed predicate devices in technological characteristics.
## PERFORMANCE DATA
The performance data indicate that the Alaris® Safety Male Luer meets specified requirements and is substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized eagle or bird in flight, with three curved lines representing the wings and a human profile in the negative space.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## JAN 1 2 2006
Ms. Stacy L. Lewis Senior Regulatory Affairs Specialist Cardinal Health, Alaris® Products 10221 Wateridge Circle San Diego, California 92121-2772
Re: K053049
Trade/Device Name: Alaris® Safety Male Luer Regulation Number: 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: October 25, 2005 Received: October 28, 2005
Dear Ms. Lewis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the cnactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Fxisting major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Lewis
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE
510(k) Number:
K053049 (To Be Assigned By FDA)
Device Trade Name:
## Indications For Use:
The Safety Male Luer is indicated for use when reconstituting, dispensing/ transferring, administering, and disposal of potentially hazardous fluids, such as chemotherapy, radioactive isotopes, and blood products, as well as non-hazardous fluids. The Safety Male Luer is intended for use with the SmartSite® Needle Free Valve port or standard open female luers.
Prescription Use X (Per 21 CFR 801.109)
Over-The-Counter Use
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anthony D. mc
tion, and :nanges
K45↓49
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.