The Porous Titanium Acetabular Shells are indicated for cemented or non-cemented use in total hip replacement in cases of: 1. Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis. 2. Rheumatoid arthritis. 3. Correction of functional deformity. 4. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques. 5. Revision of previously failed total hip arthroplasty.
Device Story
Porous Titanium Acetabular Shells are single-use orthopedic implants for hip joint reconstruction. Constructed from porous titanium alloy (Ti-6Al-4V) with a solid titanium rim for reinforcement; sizes range from 52mm to 80mm. Designed for cemented or non-cemented fixation, with optional screw fixation to native bone. Used in clinical settings by orthopedic surgeons during primary or revision total hip arthroplasty. The shell houses commercially available polyethylene acetabular liners, which are cemented into the shell. The device provides a stable acetabular component to restore hip function, reduce pain, and improve mobility in patients with degenerative joint disease or trauma.
Clinical Evidence
No clinical data provided as a basis for substantial equivalence. Bench testing only.
Technological Characteristics
Constructed of porous titanium alloy (Ti-6Al-4V). Features a solid titanium rim for reinforcement and mating geometry. Designed for cemented or non-cemented use with optional screw fixation. Sizes range from 52mm to 80mm in 2mm increments. Compatible with polyethylene acetabular liners.
Indications for Use
Indicated for skeletally mature patients requiring total hip replacement due to noninflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, proximal femur fractures (non-union, femoral neck, trochanteric), or revision of failed total hip arthroplasty.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
K052996
Image /page/0/Picture/1 description: The image shows the word BIOMET in a bold, sans-serif font. The letters are white with a thick black outline. The letters are all capitalized and evenly spaced. The overall impression is one of strength and stability.
178 - J
# 510(k) Summary
Preparation Date: April 26, 2006
Applicant/Sponsor: Blomet Manufacturing Corp.
Contact Person: Tracy Bickel Johnson, RAC
Proprietary Name: Porous Titanium Acetabular Shells
Common Name: Acetabular shells
Classification Name: Hip joint, Metal, semi-constrained, porous-coated uncemented prosthesis (888.3358); Prosthesis, Hip, Semi-constrained, Metal/Polymer, Cemented (888.3350); Hip foint metal/polymer constrained cemented or uncemented prosthesis (888.3310); Hip foint metal/ceramic/polymer semi-constrained cemented or non porous uncemented prosthesis (888.3353)
Legally Marketed Devices To Which Substantlal Equivalence Is Claimed: Hedrocel® Revision Shells (K001759), Hedrocel® Acetabular Cages (K983837)
Device Description: Porous Titanium Acetabular Shells are Intended for use in reconstruction of the hip joint due to disease, deformly, or trauma. The devices are intended for general use in skeletally mature individuals undergoing primary and/or secondary revision surgery. The device is a single use implant.
This submission includes acetabular shells for use as reconstructive components. The acetabular shells range In size (OD) from 52mm to 80mm in 2mm Increments. The shell is constructed of a porous titanlum allov (Ti-6Al-4V) and has been designed for cemented or non-cemented use with the option of using screws for added fixation to the native bone. The wall of the shell is composed entirely of the porous material. The rim of the implant is solld titanlum for reinforcement and provides the mating geometry necessary to mate with the instrumentation. The acetabular shells are designed for use with commercially available all polyethylene acetabular liners. These polyethylene ilners are to be cemented into the acetabular shells.
#### Intended Use:
Please note that the indications for the compatible product listing have not been changed and/or modified for this submission.
The Porous Titanium Acetabular Shells are indicated for cemented use in total hip replacement in cases of:
- 1. Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- 2. Rheumatoid arthritis.
- 3. Correction of functional deformity.
- বে Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques.
- ร์.
บก Revision of previously failed total hip arthroplasty.
MAILING ADDRESS P.O. Box 587 Warsaw, IN 46581-0587
.
SHIPPING ADDRESS 56 11. Bell Drive Warsaw. IN 46582
.
OPPICE 574.267.6639
ドヘズ 574.267.8157
N 46552
E-MAIL
b>t
E-MAIL blomet@blomet.com
{1}------------------------------------------------
# Page 2 of 3 510(k) Summary - Porous Titanium Acetabular Shells Blomet Manufacturing Corp.
Specific Indications for compatible components (femorals, heads, liners, and screws) that can be used with the Regenerex™ Hip Products include:
Kor7996
### Ereedom® Constrained Liners (K030047)
The Blomet Freedom® Constrained liner is indicated for use as a component of a total hip prosthesis in primary and/or revision patients at high risk of history of prior dislocation, bone loss. joint or soft tissue laxity, neuromuscular disease, or intra-operative instability and for whom all other options to constrained acetabular components have been considered.
## Tri-Polar System (K991990)
Indication #5 (Revision of previously failed total hip arthroplasty) is further specified to include recurrent distocation.
OSS/Salvage Systems/Total Femur (K974558, K002757, K021380, K033871) Salvage/Oncology Hip and Total Femur components are also indicated for cases of Ilgament deficiency, tumor resection, trauma and revision of unsuccessful osteotomy or arthrodesis.
#### Interiocking Stems (K990830. K042774)
Interlocking hip stems are indicated for non-cemented application in cases of revision, trauma, fracture, oncology or other situations where severe proximal bone loss may compromise the fixation and stability of a standard type hip replacement prosthesis.
#### HA PMI Hip Femoral Stems (K030048)
Non-cemented use for skeletally mature patients undergoing primary hip replacement surgery as a result of non-Inflammatory degenerative folnt disease.
#### Gross Hip Femoral (K001580) and Bohn Hip Femoral (K000262)
These femoral components are also indicated for treatment in confunction with tumor resection and trauma.
#### Taper 2™ Porous Femoral (K050441) and Balance Microplasty (K050251)
Indication #5 (Revision of previously failed total hip arthroplasty) is not applicable for these components. These stems are indicated for the revision of previously failed femoral head resurfacing component.
## Color Buffed Femoral (K992903). CB DDH Femoral (K012019), Bi-Metric Hip (K992058), CB Answer Femoral (K991987), Reach Femoral (K982367),
Indication #5 (Revision of previously failed total hip arthroplasty) instead reads. "Revisions of hip replacement components".
#### 150mm Bi-Metric Hip Stem (K983710)
Indication #5 (Revision of previously failed total hip arthroplasty) is not applicable for these components. They are not Indicated for revision procedures.
#### Zirconia Ceramic Heads (K964431 and K991708)
Indication #4 (Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques) Is not applicable to these Zirconia Ceramic Heads,
{2}------------------------------------------------
# Page 3 of 3 510(k) Summary - Porous Titanium Acetabular Shells Blomet Manufacturing Corp.
Biolox® detta Ceramic Heads (K042091), Rx90™ Standard/Lateralized Femoral (K023085), Portrait Femoral (K010560), Mallory/Head Smooth Femoral (K994007), Generation 4 Pollshed Femoral (K031734), Modular Reach Femoral (K994038),
Kos 2996
Indication #5 (Revision of previously failed total hip arthroplasty) Instead reads, "Revision procedures where other treatment or devices have failed".
Medallion Modular Hip System (K041850) Same list of five indications but cleared for uncemented use only.
Summary of Technologies: The technological characteristics (material modification, design, sizing, Indications) of the Porous Titanium Acetabular Shells are similar to or identical to the predicate devices.
Non-Clinical Testing: Non-cilnical laboratory testing was performed to determine substantial equivalence. The results indicated that the device was functional within its Intended use.
Clinical Testing: None provided as a basis for substantial equivalence.
All trademarks are property of Blomet, Inc. except for Hedrocel®, which is property of Zimmer Holdings
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services, USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" around the perimeter. In the center of the seal is a stylized representation of a caduceus, a symbol often associated with medicine and healthcare.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 7 2 2006
Biomet Manufacturing Corp. c/o Ms. Tracy Bickel Johnson, RAC Manager, Regulatory Affairs P.O. Box 587 Warsaw, Indiana 46581-0587
Re: K052996
Trade/Device Name: Porous Titanium Acetabular Shells Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, LZO, JDI, KWZ, MEH, MAY Dated: April 17, 2006 Received: April 18, 2006
Dear Ms. Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Ms. Tracy Bickel Johnson, RAC
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Neilkr Ogden
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
## 510(k) Number (if known): K052996
#### Device Name: Porous Titanium Acetabular Shells
#### Indications for Use:
The Porous Titanium Acetabular Shells are indicated for cemented or non-cemented use in total hip replacement in cases of:
- 1. Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- 2. Rheumatoid arthritis.
- Correction of functional deformity. 3.
- Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal ধ femur with head involvement, unmanageable using other techniques.
- 5. Revision of previously failed total hip arthroplasty.
Specific indications for compatible components (femorals, heads, liners, and screws) that can be used with the Regenerex™ Hip Products include:
#### Freedom® Constrained Liners (K030047)
The Biomet Freedom® Constrained liner is indicated for use as a component of a total hip prosthesis in primary and/or revision patients at hip dislocation due to history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intra-operative instability and for whom all other options to constrained acetabular components have been considered.
#### Tri-Polar System (K991990)
Indication #5 (Revision of previously failed total hip arthroplasty) is further specified to include recurrent dislocation.
# OSS/Salvage Systems/Total Femur (K974558. K002757, K021380, K033871)
Salvage/Oncology Hip and Total Femur components are also indicated for cases of figament deficiency, tumor resection, trauma and revision of unsuccessful osteotomy or arthrodesis.
#### Interlocking Stems (K990830, K042774)
Interlocking hip stems are indicated for non-cemented application in cases of revision, trauma, fracture, oncology or other situations where severe proximal bone loss may compromise the fixation and stability of a standard type hip replacement prosthesis.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use No (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nigh for mem
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
Page 1 of 2
Kc52996
{6}------------------------------------------------
## Porous Titanium Acetabular Shells K052996 Indications for Use
#### HA PMI Hip Femoral Stems (K030048)
Non-cemented use for skeletally mature patients undergoing primary hip replacement surgery as a result of non-inflammatory degenerative joint disease.
# Gross Hip Femoral (K001580) and Bohn Hip Femoral (K000262)
These femoral components are also indicated for treatment in conjunction with tumor resection and trauma.
Taper 2 ™ Porous Femoral (K050441) and Balance Microplasty (K050251) Indication #5 (Revision of previously failed total hip arthroplasty) is not applicable for these components. These stems are indicated for the revision of previously failed femoral head resurfacing component.
# Color Buffed Femoral (K992903), CB DDH Femoral (K012019), Bi-Metric Hip (K992058), CB Answer Femoral (K991987), Reach Femoral (K982367)
Indication #5 (Revision of previously failed total hip arthroplasty) is not applicable for these components. The statement should read, "Revisions of hip replacement components".
#### 150mm Bi-Metric Hip Stem (K983710)
Indication #5 (Revision of previously failed total hip arthroplasty) is not applicable for these components. They are not indicated for revision procedures.
# Zirconia Ceramic Heads (K964431 and K991708)
Indication #4 (Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques) is not applicable to these Zirconia Ceramic Heads.
Biolox® delta Ceramic Heads (K042091), Rx90™ Standard/Lateralized Femoral (K023085), Portrait Femoral (K010560), MalloryHead Smooth Femoral (K994007), Generation 4 Polished Femoral (K031734), Modular Reach Femoral (K994038),
Indication #5 (Revision of previously failed total hip arthroplasty) is not applicable for these components. The statement should read, "Revision procedures where other treatment or devices have failed".
Medallion Modular Hip System (K041850) Same list of five indications but cleared for uncemented use only.
Mikre R. John faimitin
and Neurological Devices
510(k) Number_ K- C52 99 k
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.