K052423 · Siemens Medi Cal Solutions, Inc. · LNH · Jan 13, 2006 · Radiology
Device Facts
Record ID
K052423
Device Name
MAGNETOM SYSTEMS WITH SYNGO EXPERT - I OPTION
Applicant
Siemens Medi Cal Solutions, Inc.
Product Code
LNH · Radiology
Decision Date
Jan 13, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The following MAGNETOM systems: MAGNETOM 1.5 T Avanto MAGNETOM 1.5 T Espree MAGNETOM 1.5 T Symphony a Tim System MAGNETOM 3.0 T Trio a Tim System MAGNETOM 1.0 T Harmony MAGNETOM 1.5 T Symphony MAGNETOM 1.5 T Sonata MAGNETOM 0.2 T Concerto MAGNETOM 3.0 T Trio with syngo Expert—i option described in this premarket notification are indicated for use as magnetic resonance diagnostic devices (MRDD) that produce transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that display the internal structure and/or function of the head, body, or extremities. These images and/or spectra, when interpreted by a trained physician, yield information that may assist in diagnosis. They may also be used for imaging during interventional procedures performed with MR compatible devices such as, in room display and MR safe biopsy needles.
Device Story
The syngo Expert-i option is a software feature for Siemens MAGNETOM MRI systems; it enables remote access to the MRI main or satellite console by authorized personnel within the local radiology department network. The system allows a remote user (e.g., physician or technician) to assist the local operator (e.g., technician) in real-time during scanning. This collaborative workflow aims to improve scan speed and image quality by facilitating expert guidance. The device does not alter the fundamental MR imaging physics or hardware of the predicate systems; it functions as a communication and remote-control interface for the existing console software. Healthcare providers use the remote access to provide guidance, which may influence scan parameters or procedural decisions, potentially benefiting patients through optimized imaging protocols and reduced scan times.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing and risk analysis regarding the safety of the remote access software feature.
Technological Characteristics
Software option for existing MAGNETOM MRI systems. Connectivity: Local area network (LAN) integration for remote console access. Hardware: Unchanged from predicate systems. Safety: Risk analysis performed for remote access functionality.
Indications for Use
Indicated for use as MR diagnostic devices to produce cross-sectional images and spectra of the head, body, or extremities for diagnostic assistance. Applicable to patients requiring MR imaging, including during interventional procedures using MR-compatible devices. No specific age or gender contraindications stated.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
## JAN I 3 2006 (052423 Section 1 - 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92.
## I. General Information
| Establishment | Siemens Medical Solutions. Inc.<br>51 Valley Stream Parkway<br>Malvern. PA 19355 |
|---------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Registration Number | 2240869 |
| Manufacturer | Siemens Medical Solutions<br>Henkestrasse 127<br>D91052 Erlangen, Germany |
| Registration Number | 8010024 |
| Contact Person | Ms. Judy Campbell<br>Technical Specialist, Regulatory Submissions<br>51 Valley Stream Parkway<br>Malvern. PA 19355<br>Phone: (610)448-4918<br>Fax: (610) 448-1787 |
| Device Name | Trade Name:<br>MAGNETOM Systems with <i>syngo</i> Expert—i option |
| Classification Name:<br>CFR Section:<br>Classification: | Magnetic Resonance Diagnostic Device<br>21 CFR § 892.1000<br>Class II |
Performance Standards None established under Section 514 the Food. Drug, and Cosmetic Act.
H. Safety and Effectiveness Information Supporting Substantial Equivalence.
Intended Use The following MAGNETOM systems:
MAGNETOM 1.5 T Avanto MAGNETOM 1.5 T Espree MAGNETOM 1.5 T Symphony a Tim System MAGNETOM 3.0 T Trio a Tim System
Siemens 510(k) Premarket Notiification 1 Syngo Expert-I Option for MAGNETOM Systerns CONFIDENTIAL
{1}------------------------------------------------
K05 2423
MAGNETOM 1.0 T Harmony MAGNETOM 1.5 T Symphony MAGNETOM 1.5 T Sonata MAGNETOM 0.2 T Concerto MAGNETOM 3.0 T Trio
with syngo Expert—i option described in this premarket notification are indicated for use as magnetic resonance diagnostic devices (MRDD) that produce transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that display the internal structure and/or function of the head, body, or extremities. These images and/or spectra, when interpreted by a trained physician, yield information that may assist in diagnosis.
They may also be used for imaging during interventional procedures performed with MR compatible devices such as, in room display and MR safe biopsy needles.
Siemens Medical Solutions, Inc., intends to offer a software option called syngo Expert—i. The indications for use will stay exactly the same, with respect to the previous > oftware versions mentioned in the comparison matrix.
## Substantial Equivalence
Within the definition of the Safe Medical Devices Act of 1990, the
MAGNETOM 1.5 T Avanto MAGNETOM 1.5 T Esprec MAGNETOM 1.5 T Symphony a Tim System MAGNETOM 3.0 T Trio a Tim System MAGNETOM 1.0 T Harmony MAGNETOM 1.5 T Symphony MAGNETOM 1.5 T Sonata MAGNETOM 0.2 T Concerto MAGNETOM 3.0 T Trio
with syngo Expert-i Option are substantially equivalent to the following cleared medical devices:
{2}------------------------------------------------
Section 2: Required Forms and Documents
| Predicate Device Name | FDA Clearance Number | FDA Clearance Date |
|--------------------------------------|----------------------|--------------------|
| MAGNETOM 1.5 T Avanto | K032428 | October 16, 2003 |
| MAGNETOM 1.5 T Espree | K041112 | July 21, 2004 |
| MAGNETOM 1.5 T Symphony a Tim System | K050199 | February 18, 2005 |
| MAGNETOM 3.0 T Trio a Tim System | K050200 | February 28, 2005 |
| MAGNETOM 1.0 T Harmony | K970852 | June 5, 1997 |
| MAGNETOM 1.5 T Symphony | K971684 | August 5, 1997 |
| MAGNETOM 1.5 T Sonata | K993731 | December 23, 1999 |
| MAGNETOM 0.2 T Concerto | K003192 | December 21, 2000 |
| MAGNETOM 3.0 T Trio | K013586 | December 28, 2001 |
KOSZ
General Safety and Effectiveness Concerns:
The introduction of the syngo Expert—i Option has effect on the MR safety. Therefore a risk analysis has been performed including new safety measures to reduce the residual risks to an acceptable level.
The MRI systems are exactly the same as what was described and cleared in the predicate premarket notifications. Safety testing will be completed before commercial introduction to the market.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a logo with a stylized eagle and text around it. The eagle is depicted with three curved lines forming its body and wings. The text is arranged in a circular pattern around the eagle, following the curve of the circle. The text is not fully legible, but it appears to be part of a government agency or organization's name.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 1 3 2006
Ms. Judith Campbell Technical Specialist, Regulatory Submissions Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway MALVERN PA 19355
Re: K052423
Trade/Device Name: MAGNETOM Systems with syngo Expert-i option Regulation Number: 21 CFR §892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: December 22, 2005 Received: December 30, 2005
Dear Ms. Campbell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device (o proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Kos 2423 Section 3 - Indications for Use
| Device Name | FDA Clearance Number |
|--------------------------------------|----------------------|
| MAGNETOM 1.5 T Avanto | K032428 |
| MAGNETOM 1.5 T Espree | K0411-12 |
| MAGNETOM 1.5 T Symphony a Tim System | K050199 |
| MAGNETOM 3.0 T Trio a Tim System | K050200 |
| MAGNETOM 1.0 T Harmony | K970852 |
| MAGNETOM 1.5 T Symphony | K971684 |
| MAGNETOM 1.5 T Sonata | K993731 |
| MAGNETOM 0.2 T Concerto | K003192 |
| MAGNETOM 3.0 T Trio | K013586 |
The above listed MAGNETOM systems are indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. These images and/or spectra, when interpreted by a trained physician, yield information that may assist in diagnosis.
They may also be used for imaging during interventional procedures performed with MR compatible devices such as, in room display and MR safe biopsy needles.
The syngo Expert-i feature allows the local user of the MRI (e.g. tech) to get help and assistance from other personnel of the radiology department (e.g. other tech or physician) to perform scans faster and with better quality. For this purpose, a remote user within the local network of the MRI (i.e. the network of the radiology) can log onto the MR main or satellite console.
(please do not write below this line- continue on another page if needed)
| | Concurrence of CDRH, Office of Device Evaluation |
|--|--------------------------------------------------|
|--|--------------------------------------------------|
| Prescription Use | |
|----------------------|--|
| OR | |
| Over-The-Counter Use | |
| | <div align="center">(Division Sign-Off)</div> |
|--|-----------------------------------------------|
|--|-----------------------------------------------|
| Siemens 510(k) Premarket Notification | Division of Reproductive, Abdominal, and Radiological Devices |
|--------------------------------------------|---------------------------------------------------------------|
| Syngo Expert-I Option for MAGNETOM Systems | (k) Number K052473 CONFIDENTIAL |
Syngo Expert-I Option for MAGNETOM Systems(k) Number _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.