K050199 · Siemens Medical Solution · LNH · Feb 18, 2005 · Radiology
Device Facts
Record ID
K050199
Device Name
MAGNETOM SYMPHONY A TIM SYSTEM
Applicant
Siemens Medical Solution
Product Code
LNH · Radiology
Decision Date
Feb 18, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The MAGNETOM Symphony a Tim System is intended for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. These images and/or spectra, when interpreted by a trained physician, yield information that may assist in diagnosis. The MAGNETOM Symphony a Tim System may also be used for imaging during interventional procedures performed with MR compatible devices such as in room display and MR safe biopsy needles.
Device Story
MAGNETOM Symphony a Tim System; magnetic resonance diagnostic device (MRDD). Inputs: magnetic resonance signals from patient anatomy. Transformation: system hardware processes signals to generate transverse, sagittal, coronal, and oblique cross-sectional images and spectroscopic data. Output: visual images/spectra for physician interpretation. Usage: clinical environment; operated by trained physicians/technicians. Supports interventional procedures using MR-compatible accessories (e.g., biopsy needles, in-room displays). Clinical benefit: provides internal structural/functional visualization to assist in medical diagnosis.
Clinical Evidence
Bench testing only. Device performance and safety parameters verified against FDA-recognized NEMA standards and international IEC standards for Magnetic Resonance Imaging Devices.
Technological Characteristics
1.5T Magnetic Resonance Diagnostic Device. Hardware components consistent with predicate MAGNETOM Avanto and Symphony systems. Supports cross-sectional imaging and spectroscopy. Compatible with MR-safe interventional accessories. Conforms to NEMA and IEC safety/performance standards.
Indications for Use
Indicated for use as a magnetic resonance diagnostic device (MRDD) to produce cross-sectional images and spectra of the head, body, or extremities to assist in diagnosis. May be used during interventional procedures with MR-compatible devices.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
# 12 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in This summary of 910(R) barry and of Safe Medical Device Act 1990 and 21 CFR § 807.92.
I. General Information
| Establishment | Siemens Medical Solutions. Inc.<br>51 Valley Stream Parkway<br>Malvern. PA 19355 |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Registration Number | 2240869 |
| Manufacturer | Siemens AG. Bereich Med<br>Henkestrasse 127<br>D-91052 Erlangen. Germany |
| Registration Number | 8010024 |
| Contact Person | Ms. Ana Ladino<br>Technical Specialist, Regulatory Submissions<br>51 Valley Stream Parkway<br>Malvern. PA 19355<br>Phone: (610)448-1785<br>Fax: (610) 448-1787 |
| Device Name | Trade Name: MAGNETOM Symphony a Tim System |
| | Classification Name: Magnetic Resonance Diagnostic Device |
| | CFR Code: 21 CFR § 892.1000 |
| | Classification: Class II |
#### Performance Standards
None established under Section 514 the Food, Drug, and Cosmetic Act.
Classification:
CONFIDENTIAL
{1}------------------------------------------------
### III. Safety and Effectiveness Information Supporting Substantial Equivalence.
#### Intended Use
The MAGNETOM Symphony a Tim System is intended for use as a magnetic resonance The MANDI on MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. These images and/or spectra, when interpreted by a trained physician, yield information that may assist in diagnosis.
The MAGNETOM Symphony a Tim System may also be used for imaging during interventional procedures performed with MR compatible devices such as in room display and MR safe biopsy needles.
#### Device Description
Device Deseripati designed scanner. It consists of the same types of hardware that are currently available with the MAGNETOM Avanto and Symphony systems.
#### Substantial Equivalence
The system is substantially equivalent to the following cleared medical devices:
| Predicate Device Name | FDA Clearance<br>Number | FDA Clearance Date |
|---------------------------------|-------------------------|--------------------|
| Siemens MAGNETOM 1.5 T Avanto | K032428 | 10/16/03 |
| Siemens MAGNETOM 1.5 T Symphony | K971684 | 08/05/97 |
#### General Safety and Effectiveness Concerns:
Operation of the MAGNETOM Symphony a Tim System is substantially equivalent to the commercially available MAGNETOM 1.5 T Avanto System and 1.5 T Symphony System.
The MAGNETOM Symphony a Tim System will conform to the FDA recognized NEMA Standards for the measurement of performance and safety parameters and the international IEC standard for safety issues with Magnetic Resonance Imaging Devices. This will assure that the performance of this device can be considered safe and effective with respect to the currently available MAGNETOM Avanto and Symphony systems.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation.
FEB 1 8 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Ana Ladino Technical Specialist, Regulatory Submissions Siemens Medical Solutions, USA, Inc. 51 Valley Stream Parkway MALVERN PA 19355
Re: K050199 Trade/Device Name: MAGNETOM Symphony a Tim System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 LNH Dated: January 26, 2005 Received: January 28, 2005
Dear Ms. Ladino:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced we have teviewed your becaren be (a) p.
above and have determined the device is substantially equivalent (for the indications for use stated in above and have decemininarketed predicate devices marketed in interstate commerce prior to the chelosure) to regary mannent date of the Medical Device Amendments, or to devices that have been May 20, 1 770, the Chactinent dato of the Federal Food, Drug, and Cosmetic Act (Act) that recrassmen in accordantes internet approval application (PMA). You may, therefore, market the do not require approval of a promance appear. The Act. The general controls provisions of the Act uceries, subject to the generation, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket n your de rice is classinia (coo acch additional controls. Existing major regulations affecting your Apploval), it may of bacycer to and Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA de roce but of total announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that I lease be advised mat rest in that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the I cacan statues and regalative, but not limited to registration and listing (21 CFR Part 807); labeling Act 3 requirements, moractice requirements as set forth in the quality systems (QS) (21 CFR Part 8017, good x20); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to begin hanceing of substantial equivalence of your device to a legally premarked notheation: The PDA maing of basisification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your cerros of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Also, picase note the regulation entitled, formation on your responsibilities under the Act from the 607.77). Tou may obtain outer general and Consumer Assistance at its toll-free number (800) DVISion of 301) 443-6597 or at its Internet address http://www.fda.gov/cdrb/dsmadsmam.html
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## 3 Indications for Use Statement
Koro199 510(k) Number (if known)
Device Name: MAGNETOM Symphony a Tim System
Indications for Use:
The "MAGNETOM Symphony a Tim System" is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. These images and/or spectra, when interpreted by a trained physician, yield information that may assist in diagnosis.
The "MAGNETOM Symphony a Tim System" may also be used for imaging during interventional procedures performed with MR compatible devices such as, in room display and MR safe biopsy needles.
(please do not write below this line- continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation Prescription Use OR Over-The-Counter Use Division Sinn-Of Division of Reproductive. Ab and Radiological Device 510(k) Number
Siemens 510(k) Premarket Notification MAGNETOM Symphony a Tim System
January 26, 2005
Page 14
CONFIDENTIAL
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.