VERICIS CARDIOVASCULAR IMAGE AND INFORMATION SYSTEM, MODEL 4.0
K051649 · Camtronics Medical Systems · LLZ · Aug 2, 2005 · Radiology
Device Facts
Record ID
K051649
Device Name
VERICIS CARDIOVASCULAR IMAGE AND INFORMATION SYSTEM, MODEL 4.0
Applicant
Camtronics Medical Systems
Product Code
LLZ · Radiology
Decision Date
Aug 2, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
VERICIS is a system intended to be used to acquire, store, print, transfer, and archive clinical information from Camtronics and other vendors systems including images, Hemodynamic studies and reports, measurements (via import from DICOM Structured Reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (waveform) data. VERICIS is intended to allow users to review diagnostic and non-diagnostic quality images, annotate studies, perform digital subtraction on images, to perform quantitative measurements on images (including but not limited to quantitative coronary analysis, left ventricular analysis, time, area, length, velocity, angle, volume, and velocity-time integrals), to generate physiciangenerated clinical reports (via structure reporting and template based tools), and to store this information in a database. VERICIS is software comprised of modules that perform under standard "offthe-shelf' personal computers and servers running the Microsoft Windows 2000/2003/XP operating systems. VERICIS is image data storage and display software that accepts DICOM (Digital Imaging and Communications in Medicine) image data files from multiple modalities. It accepts text data using other standards-based formats including but not limited to HL7 and XML. VERICIS is an Internet/Intranet network system that is designed for small and large, multi-user environments. The VERICIS network structure (including server and workstations) provides for the system's database management, storage, printing, and all DICOM/HL-7 interface services.
Device Story
VERICIS is a cardiovascular image and information system (PACS) for acquiring, storing, and reviewing multi-modality clinical data. Inputs include DICOM images, hemodynamic studies, cardiology waveforms, and text data (HL7/XML/OCR). System operates on standard off-the-shelf PCs and servers running Windows 2000/2003/XP. Users (clinicians) review images, perform digital subtraction, and conduct quantitative measurements (e.g., length, area, volume, velocity). Output includes annotated images and physician-generated clinical reports via template-based tools. System supports multi-user network environments for database management and printing. Facilitates clinical decision-making by centralizing diagnostic data and automating report generation.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Software-based PACS running on standard IBM PC-compatible hardware and servers. Operating systems: Microsoft Windows 2000, 2003, and XP. Connectivity: DICOM, HL7, XML, and Internet/Intranet networking. Modular architecture for image processing, storage, and reporting.
Indications for Use
Indicated for clinical use in cardiovascular environments to acquire, store, archive, and review diagnostic/non-diagnostic images, hemodynamic studies, and cardiology waveform data. Used by clinicians to perform quantitative measurements (e.g., coronary/ventricular analysis) and generate clinical reports. Prescription use only.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K051649
Image /page/0/Picture/1 description: The image shows the logo for Camtronics Medical Systems. The logo consists of a stylized "C" shape on the left, followed by the word "Camtronics" in a larger, bold font. Below "Camtronics" is the phrase "MEDICAL SYSTEMS" in a smaller font. The logo is simple and professional, and it is likely used to represent the company in its marketing and branding materials.
AUG 2 - 2005
## Summary of Safety and Effectiveness for VERICIS Cardiovascular Image and Information System
#### Submitter A.
Camtronics Medical Systems 900 Walnut Ridge Drive Hartland, Wisconsin 53029 262-367-0700 Telephone: Contact person: Steve Krueger Date prepared: June 1, 2005
#### Device B.
| Trade Name: | VERICIS Cardiovascular Image and Information System |
|--------------------|---------------------------------------------------------------|
| Common Name: | Picture Archiving and Communications Systems and Workstations |
| Regulation Number: | 21CFR 892.2050 |
| Regulation Name: | System, Image Processing, Radiological |
| Regulation Class: | II |
| Product Code: | LZZ |
#### C. Predicate Devices
McKesson Horizon Medical Imaging (K043146)
#### D. Device Description
The VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system. The VERICIS is a modification of the Camtronics Analytical Review Station (K955519) and the Echocardiography System (K992259) to accept multiple modalities. The VERICIS combines modular software applications and third party "off-the-shell" software on standard computer workstations and servers running the specified Microsoft Windows operating systems.
#### E. Intended Use
VERICIS is a system intended to be used to acquire, store, print, transfer, and archive clinical information from Camtronics and other vendors systems including images, Hemodynamic studies and reports, measurements (via import
{1}------------------------------------------------
from DICOM Structured Reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (waveform) data. VERICIS is intended to allow users to review diagnostic and non-diagnostic quality images, annotate studies, perform digital subtraction on images, to perform quantitative measurements on images (including but not limited to quantitative coronary analysis, left ventricular analysis, time, area, length, velocity, angle, volume, and velocity-time integrals), to generate physiciangenerated clinical reports (via structure reporting and template based tools), and to store this information in a database.
· VERICIS is software comprised of modules that perform under standard "offthe-shelf" personal computers and servers running the Microsoft Windows 2000/2003/XP operating systems.
• VERICIS is image data storage and display software that accepts DICOM (Digital Imaging and Communications in Medicine) image data files from multiple modalities. It accepts text data using other standards-based formats including but not limited to HL7 and XML.
• VERICIS is an Internet/Intranet network system that is designed for small and large, multi-user environments. The VERICIS network structure (including server and workstations) provides for the system's database management, storage, printing, and all DICOM/HL-7 interface services.
#### ட். Substantial Equivalence
The VERICIS Cardiovascular Image and Information system employs the same fundamental scientific technology as the McKesson Medical Imaging Company, Horizon Medical Imaging (K043146), the predicate device.
#### Technological Characteristics G.
The VERICIS combines modular software applications and third party off-theshelf software on standard computer workstations and servers running the Microsoft Windows 2000/2003 and XP operating systems. When the VERICIS software is loaded on the specified IBM PC compatible computers, hardware and peripherals the resulting system will act and perform similar to the McKesson Horizon Imaging system to receive, transmit, store, retrieve, display print and process digital medical images, digital medical video, and associated medical information from various medical imaging systems.
#### Conclusions H.
Camtronics Medical Systems has demonstrated through its comparison of performance with the predicate devices that the VERICIS is equivalent to the Horizon Medical Imaging system.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three birds in flight, arranged one behind the other.
AUG 2 - 2005
Received: June 21, 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Steve Krueger Director of Regulatory Affairs/Quality Assurance Camtronics Medical System 900 Walnut Ridge Drive HARTLAND WI 53029
Re: K051649
Trade/Device Name: VERICIS Cardiovascular Image and Information System Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: June 20, 2005
Dear Mr. Krueger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stied in the enclosure) to legally marketed predicate devices marketed in interstate commons on use to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal requirements of the Accoln and Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); abeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# ×05 16 49
## Indications for Use
### 510(k) Number (if known):
- Camtronics Medical Systems VERICIS Cardiovascular Image and Information Device Name: System
VERICIS is a system intended to be used to acquire, store, print, transfer, and Indications For Use: archive clinical information from Camtronics and other vendors systems including images, Hemodynamic studies and reports, measurements (via import from DICOM Structured Reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (waveform) data. VERICIS is intended to allow users to review diagnostic and non-diagnostic quality images, annotate studies, perform digital subtraction on images, to perform quantitative measurements on images (including but not limited to quantitative coronary analysis, left ventricular analysis, time, area, length, velocity, angle, volume, and velocity-time integrals), to generate physiciangenerated clinical reports (via structure reporting and template based tools), and to store this information in a database.
> · VERICIS is software comprised of modules that perform under standard "offthe-shelf' personal computers and servers running the Microsoft Windows 2000/2003/XP operating systems.
• VERICIS is image data storage and display software that accepts DICOM (Digital Imaging and Communications in Medicine) image data files from multiple modalities. It accepts text data using other standards-based formats including but not limited to HL7 and XML.
• VERICIS is an Internet/Intranet network system that is designed for small and large, multi-user environments. The VERICIS network structure (including server and workstations) provides for the system's database management, storage, printing, and all DICOM/HL-7 interface services.
X Prescription Use (Part 21 CFR 801 Subpart D)
Radiological Devic
()(k) Number
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IFNEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy Brogdon
r →
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.