K043146 · Mckesson Medical Imaging Company · LLZ · Jan 4, 2005 · Radiology
Device Facts
Record ID
K043146
Device Name
HORIZON MEDICAL IMAGING
Applicant
Mckesson Medical Imaging Company
Product Code
LLZ · Radiology
Decision Date
Jan 4, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
Horizon Medical Imaging is a medical image and information management system that is intended to receive, transmit, store, retrieve, display, print and process digital medical images, digital medical video, and associated medical information from various medical imaging systems. The medical modalities of these medical imaging systems include, but are not limited to, all modalities supported by ACR/NEMA DICOM 3.0 (specifically including mammographic images). Horizon Medical Imaging is intended to connect to a variety of storage systems and printers via DICOM and other computer industry standard interfaces and protocols. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 megapixel resolution and meets other technical specifications reviewed and approved by FDA. Horizon Medical Imaging will simply perform normal image manipulations for grayscale and image contrast on mammographic images and will not perform image processing on mammographic images. Horizon Medical Imaging is indicated for use by trained medical professionals including, but not limited to, radiologists, physicians, and medical technologists. Horizon Medical Imaging is also indicated for use in soft-copy diagnostic interpretation of medical images and video by physicians trained in such practice (specifically including soft-copy diagnostic interpretation of mammographic images).
Device Story
Horizon Medical Imaging is a PACS software system running on Windows 2000/XP PCs. It receives, stores, retrieves, displays, and prints digital medical images and video from various modalities via DICOM 3.0. Used by radiologists, physicians, and technologists in clinical settings for diagnostic interpretation and administrative review. Features include image manipulation (zoom, pan, rotate, window/level), measurements, annotations, and audio clip management. For mammography, it performs basic contrast/grayscale adjustments but no advanced image processing; requires FDA-cleared 5MP monitors for diagnostic interpretation. Output is displayed on high-resolution workstations to support clinical decision-making. System integrates with storage servers and printers via standard interfaces.
Clinical Evidence
Bench testing only. No clinical data provided. The device relies on established DICOM standards and performance characteristics of previously cleared PACS systems.
Technological Characteristics
Software-based PACS; runs on Windows 2000/XP. Supports DICOM 3.0. Connectivity via standard network interfaces. Image manipulation tools include zoom, pan, rotate, flip, and non-linear Look Up Tables. Mammographic interpretation requires 5MP resolution monitors. No proprietary hardware; utilizes commercially available IBM PC compatible hardware.
Indications for Use
Indicated for trained medical professionals (radiologists, physicians, medical technologists) for receiving, transmitting, storing, retrieving, displaying, printing, and processing digital medical images/video (all DICOM 3.0 modalities, including mammography). Indicated for soft-copy diagnostic interpretation of medical images/video by trained physicians. Contraindications: Lossy compressed mammographic images and digitized film screen images must not be used for primary interpretation. Mammographic images require FDA-approved 5MP monitors.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Picture Archiving and Communication Systems (K925965)
{0}------------------------------------------------
K043146
# Summary of Safety and Effectiveness for Horizon Medical Imaging
## Demographic Information 5.1
## Date Prepared 5.1.1
October 8, 2004
## Submitter 5.1.2
McKesson Medical Imaging Company 130 - 10711 Cambie Road Richmond, B.C. Canada, V6X 3G5 Tel: 604-279-5422 Fax: 604-279-5468
## 5.1.3 Contact
Robert MacNeil P.Eng. Manager, Quality and Regulatory Affairs
## Device Name 5.2
## Trade or Proprietary Name 5.2.1
Horizon Medical Imaging
## 5.2.2 Common Name
Horizon Medical Imaging
## Classification Name 5.2.3
Picture Archiving and Communications System (per 21 CFR 892.2050).
## Devices to which Substantial Equivalence is being Claimed 5.3
| 510(k) Number | Device Name | Manufacturer |
|---------------|------------------------------------------------|---------------------------------------------------------------------------------------|
| K925965 | Picture Archiving and<br>Communication Systems | Advanced Light Imaging Technologies<br>Inc. (now McKesson Medical Imaging<br>Company) |
| K973959 | O2 Workstation and/or<br>PACSView software | Olicon Imaging Systems Inc. (now<br>McKesson Medical Imaging Company) |
| K023557 | Centricity PACS Plus | GE Medical Systems |
Horizon Medical Imaging is based on the above identified FDA cleared PACS software of A.L.I. Technologies Inc and Olicon Imaging Systems inc (which A.L.I Technologies Inc acquired in June 1998). A.L.I Technologies Inc. became McKesson Medical Imaging Company in July 2002.
## Device Description 5.4
## 5.4.1 Function
Horizon Medical Imaging is software which when installed and run on Microsoft Windows 2000 and XP operating systems on commercially available IBM PC compatible computers, hardware components and peripherals, forms a medical image and information management system that
{1}------------------------------------------------
receives, transmits, stores, retrieves, displays, prints and processes digital medical images, digital receives, transmits, stored, routers, alippy, a promotion medical imaging systems.
Its core components are:
- components are:
High resolution color and grayscale workstations for primary diagnostic interpretation and . secondary review of the medical images, video and related information.
- Standard workstations for performing administrative functions of the system .
- Otandard worked in the corporation for short-term and long-term storage of system . data
- Servers for managing the distribution of system data .
- Oct For handing the components for providing the communication channels between the t system's core components and the various medical imaging systems
- system o our ochiponent for producing hardcopy of medical images and related information. .
## Scientific Concept 5.4.2
Horizon Medical Imaging is not based upon particular scientific concepts.
## Significant Physical and Performance Characteristics 5.4.3
Medical Image and Video Modalities Supported
- All DICOM 3.0 recognized modalities (specifically including mammographic images) .
Significant Software Features
- User roles and levels of access to patient records and studies .
- Studies/images management .
- Patient information management .
- Actions on studies (e.g. find, open, report, close) .
- Images and videos displays to user defined display protocols .
- Image manipulation (e.g. zoom, pan, rotate, flip, window, level, non-linear Look Up Tables) .
- Image measurements (e.g. distance, area, angle, intensity) .
- Annotations on images .
- Receiving images .
- . Printing images
- Sending images .
- Importing/Exporting images .
- Storing and archiving studies and patient information .
- Recording, playing, storing and retrieving audio clips .
- Scanning hardcopy documents; displaying, storing and retrieving electronic versions .
- Creating, displaying, storing, retrieving accessory information for studies (e.g. notes, diagrams) .
- DICOM standard approved lossy compression algorithms and file formats (lossy compression not . indicated for mammographic images)
## Hardware Reguirements
- Minimum hardware configuration required by Microsoft Windows 2000 and XP, and .
- Minimum hardware configuration indicated by American College of Radiology (in technical . standards on "Digital Image Data Management" and "Teleradiology") for the official diagnostic interpretation of images.
- Mammographic images may only be interpreted using an FDA approved monitor that offers at . least 5 megapixel resolution and meets other technical specifications reviewed and accepted by FDA.
- Only hard copy devices specifically FDA cleared for printing diagnostic quality copies of medical . images are indicated for the official diagnostic interpretation of hard copy medical images printed from Horizon Medical Imaging.
{2}------------------------------------------------
## 5.5 Statement of Intended Use
Horizon Medical Imaging is a medical image and information management system that is intended to rionizon Medical Imaging is a medically, print and process digital medical images, digital medical video, and associated medical information from various medical imaging systems.
The medical modalities of these medical imaging systems include, but are not limited to, all modalities supported by ACR/NEMA DICOM 3.0 (specifically including mammographic images).
Horizon Medical Imaging is intended to connect to a variety of storage systems and printers via DICOM and other computer industry standard interfaces and protocols.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for orimary image interpretations. Mammographic images may only be interpreted using an FDA primaly interprocations: Maint 5 megapixel resolution and meets other technical specifications reviewed and approved by FDA. Horizon Medical Imaging will simply perform normal image manipulations for grayscale and image contrast on mammographic images and will not perform image processing on mammographic images.
Horizon Medical Imaging is indicated for use by trained medical professionals including, but not limited to, radiologists, physicians, and medical technologists. Horizon Medical Imaging is also indicated for use in soft-copy diagnostic interpretation of medical images and video by physicians trained in such practice (specifically including soft-copy diagnostic interpretation of mammographic images).
## Comparison of Technological Characteristics 5.6
Horizon Medical Imaging maintains the functionality of the A.L.I. Technologies Picture Archiving and Communication System (K925965) and Olicon Imaging Systems O2 Workstation and/or PACSView software (K973959).
Horizon Medical Imaging adds the intended use of soft-copy official diagnostic interpretation of mammographic images, which is an intended use of the GE Medical Systems Centricity PACS Plus.
In conclusion, Horizon Medical Imaging is substantially equivalent to the GE Medical Systems Centricity PACS Plus for the intended use of soft-copy official diagnostic interpretation of mammographic images.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" around the edge. Inside the circle is a stylized image of an eagle or bird-like figure. The image is black and white and appears to be a scan or photocopy.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN - 4 2005
McKesson Medical Imaging Company % Mr. N. E. Devine, Jr. Responsible Third Party Official Entela, Inc. 3033 Madison Ave., SE GRAND RAPIDS MI 49548
Re: K043146 Trade/Device Name: Horizon Medical Imaging Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: December 17, 2004 Received: December 20, 2004
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced we have leviewed your been a verice is substantially equivalent (for the indications for use stated in above and have decorminivarketed predicate devices marketed in interstate commerce prior to the cherosate) to regally manding of the Medical Device Amendments, or to devices that have been May 20, 1970, the enature with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the do not require approval or a provisions of the Act. The general controls provisions of the Act device, subject to the generaliation, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket If your de roo is classinia ( to such additional controls. Existing major regulations affecting your Apploval), It may oc babyer to ratiFederal Regulations, Title 21, Parts 800 to 898. In addition, FDA as roo blish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that I lease of advisod that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the I caraft statues and regulating, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) (21 CFR Part 8017), and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmadsmamain.html
Sincerely yours,
Nancy C. brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: Horizon Medical Imaging
Indications For Use:
Horizon Medical Imaging is a medical image and information management system that is intended to Honzon Medical Inflaging is a medical intage und firstmes. I mages, digital medical images, digital medical receive, transmit, store, rothers, acopial in from various medical imaging systems.
The medical modalities of these medical imaging systems include, but are not limited to, all modalities The medical modalities of these mounding mammographic images).
Horizon Medical Imaging is intended to connect to a variety of storage systems and printers via Florizon Medican imaging to industry standard interfaces and protocols.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for Lossy compressed mammographic images and only be interpreted using an FDA primary mage interpretations. Manimographia intagos may only of hereformed in the specifications
approved monitor that offers at least 5 megapixel resolution and meets other approved monition that offers at lead. of mogalities reaging will simply perform normal image reviewed and approved by PDF. Hollen moated images and will not perform image processing on mammographic images.
Horizon Medical Imaging is indicated for use by trained medical professionals including, but not Honzon wedical maging is malbated for aco by frainologists. Horizon Medical Imaging is also ilmitted to, radiologists, physiodins, and notical images and video by physicians Inficated in such sole-copy diagnoule interpretation of mostic interpretation of mammographic images).
Prescription Use
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David Ch. Leygon
(Division Sign-Off) Division of Reproductive, Abdor and Radiological Device 510kr Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.