DRAEGER MEDICAL INFORMATION BUS (MIB, MIB II, MIB DUO) PROTOCOL CONVERTERS
K050974 · Draeger Medical Systems, Inc. · DRQ · Apr 29, 2005 · Cardiovascular
Device Facts
Record ID
K050974
Device Name
DRAEGER MEDICAL INFORMATION BUS (MIB, MIB II, MIB DUO) PROTOCOL CONVERTERS
Applicant
Draeger Medical Systems, Inc.
Product Code
DRQ · Cardiovascular
Decision Date
Apr 29, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2060
Device Class
Class 2
Indications for Use
The Draeger Medical Information Bus Protocol Converters are intended for use in an environment where patient care is provided by healthcare professionals (Physician, Nurse, Technician) when the professional determines that a third party medical device should be connected to an Infinity Modular Monitor for display of data. Connectable devices include: Maquet SV 900, SV 300, Servoi ventilators, Baxter Vigilance blood gas/continuous cardiac output monitor. Draeger Evita II, VI, EvitaXL, Babylog, & Savina ventilators, Puritan Bennett 7200 & 840 ventilators, Hamilton Galileo ventilator, Draeger Narkomed 6000 & 6400 / Narkomed II & VI Anesthesia Systems, Draeger Julian Anesthesia machine, Ohmeda 7900 Anesthesia Machine, Abbott Oximetric 3 Blood Gas Analyzer, AVL Medical Instruments: Opti Critical Care Analyzer Portable Blood Gas Analyzer, Optical Sensors Inc .: OSI-Optical CAM, VIA Medical: VIA V-ABG1 Blood Gas Chemistry Monitor, and Aspect A-2000 BIS, Abbott Q2, Sensormedics Micro Gas 7650, Draeger Fabius Tiro, Draeger Primus, and Viasys Bear 1000 ventilator.
Device Story
Infinity Medical Information Bus (MIB) Protocol Converters facilitate connectivity between third-party medical devices and Draeger Infinity Modular Monitors. Device inputs include digital data streams from external ventilators, anesthesia machines, and blood gas/cardiac output monitors. The converter transforms these proprietary device signals into a format compatible with the Infinity monitor network. Output consists of device-specific data displayed on the Infinity modular monitor and alarm annunciation via the VentCentral application. Used in clinical settings by physicians, nurses, and technicians to centralize patient monitoring data. Enables healthcare providers to view integrated data from multiple sources on a single display, supporting clinical decision-making by providing comprehensive patient status information.
Clinical Evidence
Bench testing only. No clinical data was required or provided for this submission.
Technological Characteristics
Protocol converter for medical device connectivity. Operates via IEEE 1073.3.1 and 1073.3.2 standards for medical device communications. Software version VF5.1. Not MRI compatible.
Indications for Use
Indicated for use by healthcare professionals (physicians, nurses, technicians) to connect third-party medical devices (ventilators, anesthesia systems, blood gas/cardiac output monitors) to Draeger Infinity Modular Monitors for data display in clinical care environments.
Regulatory Classification
Identification
A transducer signal amplifier and conditioner is a device used to provide the excitation energy for the transducer and to amplify or condition the signal emitted by the transducer.
Predicate Devices
Infinity Medical Information Bus Protocol Converter (K050974)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image contains the logo for Dräger medical. The logo is in black and features the word "Dräger" in a bold, stylized font, followed by the word "medical" in a smaller, less bold font. Below the logo, there is some smaller text, and to the right of the logo, there is some additional text that is partially visible.
A Dräger and Siem
APR 2 9 2005
# 510(k) SUMMARY
as required per 807.92(c)
| Submitter's Name and Address: | Draeger Medical Systems, Inc.<br>16 Electronics Avenue<br>Danvers, MA 01923 |
|-----------------------------------------|-------------------------------------------------------------------------------------|
| Contact Person: | Penelope H. Greco<br>Regulatory Affairs Manager<br>(978) 907-7503<br>(978) 750-6879 |
| Date submission was prepared: | April 12, 2005 |
| Device Name: | |
| Common Name: | Amplifier and Signal Conditioner,<br>Transducer Signal |
| Classification Name: | DRQ |
| Regulation Number: | 21 CFR 870.2060 |
| Class: | 2 |
| Legally Marketed Device Identification: | Infinity Medical Information Bus<br>Protocol Converter |
#### Device Description:
Draeger's Infinity Medical Information Bus Protocol Converters have received numerous 510(k) clearances for connectivity to third party devices. The release of MIB VF5.1 software enables connectivity of the Viasys Bear 1000 ventilator to the Infinity modular monitors.
This connection enables the display of device specific data on an Infinity modular monitor. Data from the Viasys Bear 1000 ventilator can also be displayed and alarms annunciated on the VentCentral application (K003246) of the MultiView.
Page 1 of 2
### COMPANY CONFIDENTIAL
Draeger Medical Systems, Inc. 16 Electronics Avenue Danvers. MA 01923 Tel: (978) 907-7500 Fax: (978) 750-6879
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for Dräger medical. The logo is in black and white, with the word "Dräger" in a bold, sans-serif font. The word "medical" is in a thinner, sans-serif font. Below the logo, it says "A Dräger and Siemens Company" in a smaller font.
Image /page/1/Picture/2 description: The image shows handwritten text that appears to be a combination of letters and numbers. The first line reads 'K050974', and the second line reads 'P 3/2'. The handwriting is somewhat rough, but the characters are generally legible. The text may represent a code, identifier, or some other form of notation.
### Intended Use:
The Draeger Medical Information Bus Protocol Converters are intended for use in an environment where patient care is provided by healthcare professionals (Physician, Nurse, Technician) when the professional determines that a third party medical device should be connected to an Infinity Modular Monitor for display of data.
Connectable devices include: Maquet SV 900, SV 300, Servoi ventilators, Baxter Vigilance blood gas/continuous cardiac output monitor. Draeger Evita II, VI, EvitaXL, Babylog, & Savina ventilators, Puritan Bennett 7200 & 840 ventilators, Hamilton Galileo ventilator, Draeger Narkomed 6000 & 6400 / Narkomed II & VI Anesthesia Systems, Draeger Julian Anesthesia machine, Ohmeda 7900 Anesthesia Machine, Abbott Oximetric 3 Blood Gas Analyzer, AVL Medical Instruments: Opti Critical Care Analyzer Portable Blood Gas Analyzer, Optical Sensors Inc .: OSI-Optical CAM, VIA Medical: VIA V-ABG1 Blood Gas Chemistry Monitor, and Aspect A-2000 BIS, Abbott Q2, Sensormedics Micro Gas 7650, Draeger Fabius Tiro, Draeger Primus, and Viasys Bear 1000 ventilator.
Substantial Equivalence:
## Assessment of non-clinical performance data for equivalence:
Verification and validation testing performed indicates that the modifications implemented with software version VF5.1 are as safe and effective as previous versions and have not altered the fundamental technology of the device(s).
Assessment of clinical performance data for equivalence: Not applicable
Biocompatibility: Not applicable
- Sterilization: Not applicable
Standards and Guidance:
1073.3.1 Medical Device Communications-Transport Profile-Connection Mode 1073.3.2 - 2000 IEEE Standard for Medical Communications Transport Profile - IrDA Based - Cable Connected
Page 2 of 2
# COMPANY CONFIDENTIAL
Draeger Medical Systems, Inc. 16 Electronics Avenue Danvers, MA 01923 Tel: (978) 907-7500 Fax: (978) 750-6879
{2}------------------------------------------------
APR 2 9 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Draeger Medical Systems, Inc. c/o Ms. Penelope H. Greco Regulatory Affairs Manager 16 Electronics Avenue Danvers, MA 01923
Re: K050974
Trade Name: Infinity Medical Information Bus Protocol Converter Regulation Number: 21 CFR 870.2060 Regulation Name: Transducer Signal Amplifier and Signal Conditioner Regulatory Class: II (two) Product Code: DRQ Dated: April 12, 2005 Received: April 18, 2005
Dear Ms. Greco:
We have reviewed your Section 510(k) premarket notification of intent to market the device we neversioned your betermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for ass stated in the encreases 9776, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, do noos mat have been require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may a news provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 -- Ms. Penelope H. Greco
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Errt in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Oixision of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Blymmmoe for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K050974
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Infinity Medical Information Bus Protocol Converter Indications for use
The Infinity Medical Information Bus (MIB) Protocol Converters (MIB, 11 & MIB Duo) are indicated for use in an environment where patient care is provided by healthcare professionals (Physician, Nurse, Technician) when the professional determines that third party medical devices that provide data should be connected to a Draeger Infinity Modular Monitor for display. Such devices include:
> Maquet SV 300 ventilator Maquet Servoi Ventilator Maquet SV900 ventilator Draeger Evita 2 ventilator Draeger Evita 4 ventilator Draeger EvitaXL ventilator Draeger Savina ventilator Draeger Babylog ventilator Draeger FabiusGS Anesthesia System Draeger Narkomed 2 Anesthesia System Draeger Narkomed 4 Anesthesia System Draeger Narkomed 6000 / 6400 Anesthesia Systems Draeger Julian Anesthesia Machine Puritan Bennett 7200 ventilator Puritan Bennett 840 ventilator Hamilton Galileo ventilator Ohmeda 7900 Anesthesia Machine Abbott Oximetrix 3 Blood Gas Analyzer Abbott Q2 CCO monitor AVL Medical Instruments: Opti Critical Care Analyzer, Portable Blood Gas Analyzer Baxter Vigilance blood gas/continuous cardiac output monitor Optical Sensors Inc .: OSI - Optical CAM VIA Medical: VIA V-ABG1 Blood Gas Chemistry Monitor Aspect A-2000 BIS Monitor* Sensormedics Micro Gas 7650 Draeger Fabius Tiro Draeger Primus Viasys Bear 1000
Note: * The SC 9000 does not support communication with the Aspect BIS Monitor
#### MRI Compatibility Statement:
The MIB, MIB II and MIB DUO Protocol Converters are not compatible for use in a MRI magnetic field.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
BELAST DONES BEDEBEL BED AND AND A DENALLED ANGELE PAGE OF NEW DI
o (, DRH, Olfice of Dance I wildather (OD)
Blymmman
sign-on
f Cardiovascular Devices
Number K050974
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.