K050562 · Iridex Corp · GEX · Apr 14, 2005 · General, Plastic Surgery
Device Facts
Record ID
K050562
Device Name
OCULIGHT GL/GLX
Applicant
Iridex Corp
Product Code
GEX · General, Plastic Surgery
Decision Date
Apr 14, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The OcuLight GL/GLx is indicated for retinal photocoagulation, laser trabeculoplasty, iridotomy, iridoplasty, the treatment of vascular and pigmented skin lesions, stapedotomy, stapedectomy, and other ear, nose, and throat (ENT) applications. The following are examples of applications for the OcuLight GL/GLx laser systems. Condition Treatment Diabetic Retinopathy Retinal Photocoagulation (RPC); Focal and Grid Laser Treatments Nonproliferative Retinopathy Macular Edema Proliferative Retinopathy Glaucoma Laser Trabeculoplasty; Iridotomy; Iridoplasty Primary Open Angle Closed Angle Refractory Glaucoma Retinal Tears and Detachments RPC; Focal and Grid Laser Treatments Lattice Degeneration RPC; Focal and Grid Laser Treatments Age-related Macular Degeneration (AMD) RPC; Focal and Grid Laser Treatments Intra-Ocular Tumors RPC; Focal and Grid Laser Treatments Choroidal Hemangioma Choroidal Melanoma Retinoblastoma Retinopathy of Prematurity RPC; Focal and Grid Laser Treatments Sub-Retinal (choroidal) Neovascularization RPC; Focal and Grid Laser Treatments Central and Branch Retinal Vein Occlusion RPC; Focal and Grid Laser Treatments Dermatology Focal Laser Treatments Pigmented Skin Lesions Vascular lesions Ear, Nose and Throat Otosclerotic hearing loss and/or diseases of the inner ear Stapedectomy Stapedotomy Myringotomies Lysis of Adhesions Control of Bleeding Removal of Acoustic Neuromas Soft tissue Adhesion in Micro/Macro Otologic Procedures
Device Story
Semiconductor-based laser system delivering continuous wave green (532 nm) laser light for photocoagulation and surgical applications; visible red (630-650 nm) diode laser used for aiming. Operated by clinicians in ophthalmic, dermatological, and ENT settings. Device transforms electrical energy into laser output for tissue interaction (photocoagulation, ablation, or hemostasis). Healthcare providers use the laser to treat ocular conditions, skin lesions, and ENT pathologies. Output affects clinical decision-making by providing a therapeutic tool for tissue management, potentially benefiting patients by addressing vision-threatening conditions, skin lesions, and otologic diseases.
Clinical Evidence
No clinical data.
Technological Characteristics
Semiconductor-based laser system; 532 nm green laser for treatment; 630-650 nm red diode laser for aiming; continuous wave delivery; hardware and firmware unchanged from previous versions.
Indications for Use
Indicated for patients requiring retinal photocoagulation, laser trabeculoplasty, iridotomy, iridoplasty, treatment of vascular/pigmented skin lesions, or ENT procedures including stapedotomy, stapedectomy, myringotomies, lysis of adhesions, bleeding control, and removal of acoustic neuromas. Applicable to conditions including diabetic retinopathy, glaucoma, retinal tears/detachments, lattice degeneration, AMD, intra-ocular tumors, retinopathy of prematurity, sub-retinal neovascularization, and retinal vein occlusion.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
# APR 1 4 2005
# 510(k) Summary IRIDEX Corporation IRIS Medical® OcuLight® GL/GLx
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
John Jossy IRIDEX Corporation 1212 Terra Bella Avenue Mountain View, CA 94043 (650) 962-8848 ext. 3016
Contact Person: (same as above)
Date Prepared: May 23, 2003
## Name of Device and Name/Address of Sponsor
IRIS Medical OcuLight GL/GLx Laser Systems
IRIDEX Corporation 1212 Terra Bella Avenue Mountain View, CA 94043
### Classification Name
Laser Instrument, Surgical, Powered CFR Section: 878.4810 and 886.4390 Product Code: GEX and HQF
### Predicate Devices
The OcuLight GL/GLx laser systems are substantially equivalent to other currently legally marketed ophthalmology laser devices including Ocutiight SLASLASE systems and the Lumenis Novus Spectra (K022327).
### Device Description
The OcuLight GL/GLx is a semiconductor-based laser that delivers true continuous wave green laser (532 nm) light for the indication of retinal photocoagulation, laser trabeculoplasty, the treatment of vascular and pigmented skin lesions, and other laser treatments. Visible red (630-650 nm) semiconductor diode laser is used for aiming.
{1}------------------------------------------------
## Intended Use
The OcuLight GL/GLx is indicated for retinal photocoagulation, laser trabeculoplasty, iridotomy, iridoplasty, the treatment of vascular and pigmented skin lesions, stapedotomy, stapedectomy, and other ear, nose, and throat (ENT) applications. The following are examples of applications for the OcuLight GL/GLx laser systems.
| Condition | Treatment |
|----------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Diabetic Retinopathy | Retinal Photocoagulation (RPC); Focal<br>and Grid Laser Treatments |
| Nonproliferative Retinopathy Macular Edema Proliferative Retinopathy | |
| Glaucoma | Laser Trabeculoplasty; Iridotomy;<br>Iridoplasty |
| Primary Open Angle Closed Angle Refractory Glaucoma | |
| Retinal Tears and Detachments | RPC; Focal and Grid Laser Treatments |
| Lattice Degeneration | RPC; Focal and Grid Laser Treatments |
| Age-related Macular Degeneration (AMD) | RPC; Focal and Grid Laser Treatments |
| Intra-Ocular Tumors | RPC; Focal and Grid Laser Treatments |
| Choroidal Hemangioma Choroidal Melanoma Retinoblastoma | |
| Retinopathy of Prematurity | RPC; Focal and Grid Laser Treatments |
| Sub-Retinal (choroidal) Neovascularization | RPC; Focal and Grid Laser Treatments |
| Central and Branch Retinal Vein Occlusion | RPC; Focal and Grid Laser Treatments |
| Dermatology | Focal Laser Treatments |
| Pigmented Skin Lesions Vascular lesions | |
| Ear, Nose and Throat | |
| Otosclerotic hearing loss and/or diseases<br>of the inner ear | Stapedectomy<br>Stapedotomy<br>Myringotomies<br>Lysis of Adhesions<br>Control of Bleeding<br>Removal of Acoustic Neuromas<br>Soft tissue Adhesion in Micro/Macro<br>Otologic<br>Procedures |
{2}------------------------------------------------
## Technological Characteristics and Substantial Equivalence
The OcuLight GL/GLx is indicated for retinal photocoagulation, laser trabeculoplasty, iridotomy, iridoplasty, the treatment of vascular and pigmented skin lesions, stapedotomy, stapedectomy, and other ENT applications. The expansion of the indications for use for the proposed OcuLight does not result in a change to the hardware or firmware for the currently marketed OcuLight GL/GLx.
The OcuLight SL/SLx Laser Systems are indicated for Retinal Photocoagulation, Laser Trabeculoplasty, Transscleral Cyclophotocoagulation (TSCPC), Transscleral Retinal Photocoagulation (TSRPC), Iridotomy, and Iridoplasty. The OcuLight SL/SLx diode laser systems feature a combination of pulsed diode laser and optical fiber technology to deliver the correct balance of 810 nm wavelength, spot size, and pulse duration for effective laser photocoagulation.
The Lumenis Novus Spectra Diode Laser System is indicated for many Ophthalmic, Ears, Nose and Throat (including stapedectomy), Dermatological and Dentistry applications. The Lumenis Novus Spectra delivers the same 532nm wavelength, pulses of equivalent duration, treatment spots of equivalent size, and energy densities equivalent to the OcuLight GL/GLx.
## Non-Clinical performance Data
None
## Clinical performance Data
None
## Conclusion
The OcuLight GL/GLx is substantially equivalent to the predicate devices currently legally marketed for stapedotomy, stapedectomy, and other ENT applications as shown in the Indications for Use statement.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 1 4 2005
IRIDEX Corporation c/o John Jossy Director of Regulatory Affairs and Quality Assurance 1212 Terra Bella Avenue Mountain View, CA 04043
Re: K050562
Trade/Device Name: IRIS Medical OcuLight GL/GLx Laser Systems Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: March 4, 2005 Received: March 8, 2005
Dear Mr. Jossy:
We have reviewed your Section 510(k) premarket notification of intent to market the device we have and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device to such additional controls. Existing major regulations affecting your device can may be subject to back at Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r that FDA has made a determination that your device complies with other requirements of the Act that I D. I has intacted and regulations administered by other Federal agencies. You must or any I cuttar statutes and registanents, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set OFF in the quality systems (21 CFR Part 820); and if applicable, the electronic forth in the quants of byevelle (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 827-8910. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
David M. Whipple
Acting Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
Kosos62 ----------------------------------------------------------------------------------------------------------------------------------------------------------------------510(k) Number (if known): Pending_____________________________________________________________________________________________________________________________________________
Device Name: IRIS Medical® OcuLight® GL/GLx
Indications For Use:
The IRIDEX OcuLight GL/GLx is used in ophthalmic, dermatology, and ear nose and throat (ENT) applications. Specific indications are listed below:
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
| Condition | Treatment |
|----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|
| Diabetic Retinopathy<br>• Nonproliferative Retinopathy<br>• Macular Edema<br>• Proliferative Retinopathy | Retinal Photocoagulation (RPC); Focal and Grid Laser Treatments |
| Glaucoma<br>• Primary Open Angle<br>• Closed Angle | Laser Trabeculoplasty; Iridotomy, Iridoplasty |
| Retinal Tears and Detachments | RPC; Focal and Grid Laser Treatments |
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
OR (Per 21 CFR 801.109) Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
{6}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known): Pending_____________________________________________________________________________________________________________________________________________
Device Name: IRIS Medical® OcuLight® GL/GLx __________________________________________________________________________________________________________________________________
Indications For Use:
| Condition | Treatment |
|-------------------------------------------------------------------------------------------|--------------------------------------|
| Lattice Degeneration | RPC; Focal and Grid Laser Treatments |
| Age-related Macular Degeneration (AMD) | RPC; Focal and Grid Laser Treatments |
| Intra-Ocular Tumors<br>• Choroidal Hemangioma<br>• Choroidal Melanoma<br>• Retinoblastoma | RPC; Focal and Grid Laser Treatments |
| Retinopathy of Prematurity | RPC; Focal and Grid Laser Treatments |
| Sub-Retinal (choroidal) Neovascularization | RPC; Focal and Grid Laser Treatments |
| Central and Branch Retinal Vein Occlusion | RPC; Focal and Grid Laser Treatments |
# (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use _____________________________________________________________________________________________________________________________________________________________
OR (Per 21 CFR 801.109) Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
{7}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known): Bending Kosos 62 2 2 2 2 2 2 2 2 2 2 ----------
Device Name: IRIS Medical® OcuLight® GL/GLx___________________________________________________________________________________________________________________________________
Indications For Use:
| Condition | Treatment |
|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Dermatology<br>• Pigmented Skin Lesions<br>• Vascular lesions | Focal Laser Treatments |
| Ear, Nose and Throat<br>• Otosclerotic hearing loss and/or diseases of the inner ear | Stapedectomy<br>Stapedotomy<br>Myringotomies<br>Lysis of Adhesions<br>Control of Bleeding<br>Removal of Acoustic Neuromas<br>Soft tissue Adhesion in Micro/Macro Otologic Procedures |
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | <div> <span style="text-decoration: overline;">✓</span> </div> |
|------------------|----------------------------------------------------------------|
| | OR |
| | (Per 21 CFR 801.109) |
forex & to hom
(Division Sign-Off) Division of Ophthalmic Ear, Nose and Throat Devises
510(k) Number_________________________________________________________________________________________________________________________________________________________________
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.