K022327 · Lumenis, Inc. · GEX · Oct 16, 2002 · General, Plastic Surgery
Device Facts
Record ID
K022327
Device Name
NOVUS SPECTRA
Applicant
Lumenis, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 16, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Ophthalmic Applications; Ear, Nose and Throat Applications; Dentistry; and Dermatologic Applications. A complete list is contained in the Indications for Use Statement.
Device Story
Novus Spectra is a solid-state, frequency-doubled, green Nd:YAG surgical laser system (532nm) for photothermolysis/photocoagulation of soft tissue. Compatible with slit lamps, LIOs, microfilters, collimated handpieces, fibers, and endocular/endooto probes. Operated by physicians in clinical settings. Features a digital 'closed-loop' control system to manage energy output, pulse duration, and functional parameters, replacing analog systems found in predicate devices. System monitors safety interlocks and displays for user intervention. Digital environment provides improved hazard mitigation and risk management. Output allows precise tissue coagulation, vaporization, or hemostasis, benefiting patients by enabling minimally invasive surgical and therapeutic interventions across multiple specialties.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological comparison, functional specifications, and improved hazard mitigation hardware compared to predicate devices.
Technological Characteristics
Solid-state, frequency-doubled Nd:YAG laser; 532nm emission wavelength. Digital closed-loop control system for energy output and pulse duration management. Compatible with various delivery optics (slit lamps, LIOs, fibers, probes). Spot sizes 200-1200 microns. Class II device.
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Elite Solid State Green Diode Pumped 532nm
Elite Ultra Solid State Green Diode Pumped 532nm
Corium 200 Solid State Green Diode Pumped 532nm
Corium 400 Solid State Green Diode Pumped 532nm
Submission Summary (Full Text)
{0}------------------------------------------------
OCT 1 6 2002
022322
1/2
## Attachment 5 510(k) Summary Statement for the Lumenis Novus Spectra Laser System
## General Information:
| Submitter: | Lumenis<br>2400 Condensa Street<br>Santa Clara, California, U. S. A.<br>95051-0901 |
|------------------------------|------------------------------------------------------------------------------------|
| Contact Person: | Karen Baker |
| Summary Preparation Date: | October 10, 2002 |
| Names: | |
| Device Names: | Novus Spectra Laser System |
| Primary Classification Name: | 79 GEX, Laser Instrument, Surgical Powered |
## Predicate Devices:
Elite Solid State Green Dlode Pumped 532nm Elite Ultra Solid State Green Diode Pumped 532nm Corium 200 Solid State Green Diode Pumped 532nm Corium 400 Solid State Green Diode Pumped 532nm
## Ratlonale for Substantial Equivalence:
The Novus Spectra Laser System has the same indications for use as the Elite family of lasers and the Corium series of lasers. They have similar functional elements such as treatment wavelengths, pulse rates, treatment power, spot size and cooling system. Control systems such as the door interlock, and the safety systems and displays are constantly monitored in these systems for user intervention during a procedure or maintenance.
## Description of Submitted Device:
The Novus Spectra solid state, frequency-doubled, green Nd:YAG surgical laser system is an instrument used in the photothermolysis (photocoagulation) of soft tissue at en emission wavelength of 532nm.
Compatible delivery devices include: slit lamps, slit lamp adapters/attachments, laser indirect ophthalmoscopes (LIO), microfilters, collimated handpieces with spot sizes ranging from 200-1200microns, fibers and endocular and endooto probes.
{1}------------------------------------------------
022327 2/2
### Indications for Use:
Ophthalmic Applications; Ear, Nose and Throat Applications; Dentistry; and Dermatologic Applications. A complete list is contained in the Indications for Use Statement.
#### Technological Characteristics and Substantial Equivalence
The Novus Spectra Laser System, the Elite family of lasers, and the Corium series deliver the same frequency-doubled Nd:YAG wavelength. Both deliver a 532nm wavelength of similar power and pulses having similar duration in single and repeat modes.
The Novus Spectra Laser System is identical to the Elite family of lasers, and the Corium series in terms of its intended uses and functionality. The Novus Spectra improves on the industry-standard digital operating environment present in the Elite to manage energy output, pulse duration and other functional aspects. A digital "closed loop" control system in the Novus Spectra replaces the analog "closed loop" control system in the Elite and Corium. In addition the digital environment improves the overall safety of the system through improved and more rigorously controlled hazard mitigation hardware
#### Conclusion
The Novus Spectra Laser System is substantially equivalent to the Elite and Corium laser systems that are now in commercial distribution and have been successful in meeting the intended uses. The Novus Spectra has an improved hazard mitigation and risk management operating environment that reduces the level of concern regarding its uses per the Intended Uses document.
The Lumenis Novys Spectra shares the same intended use, indications for use, and similar technological characteristics as the predicate laser systems.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines suggesting movement or connection.
Public Health Service
OCT 1 6 2002
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Karen Baker Regulatory Consultant to Lumenis, Inc. Lumenis, Inc. 2400 Condensa Street Santa Clara, California 95051-0901
#### Re: K022327
Trade/Device Name: Novus Spectra Laser System Regulation Number: 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: July 10, 2002 Received: July 18, 2002
Dear Ms. Baker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as setforth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Karen Baker
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Muriam C. Provost
for Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Attachment 2 Indications for Use Statement as Requested by FDA
510(K) Number (if Known):
## Device Name: Novus Spectra Laser System
## Indications for Use:
#### Ophthalmology
Retinal Photocoagulation Trabeculoplasty Iridotomy Diabetic retinopathy Peripheral Iridectomy Posterior and Anterior procedures Senile macular degeneration
#### Ear, Nose and Throat (ENT) Stapedectomy Stapedotomy Myringotomies Lysis of Adhesions Control of Bleeding Removal of Acoustic Neuromas Soft Tissue Adhesion in Micro/Macro Otologic Procedures
#### Dermatology
Pigmented lesions, including soar lentgine Vascular lesions, including cherry hemangiomas and angiokeratomas Extremeties telangiectasias, including facial and leg telangiectasias Cutaneous lesions Flat warts Dermatosis Papulosa Nigra
#### Dentistry
- Frenectomy .
- Treatment of Oral Mucous Cyst
- Treatment of Benign Vascular Lesions: Capillary hemangioma Hemorrhagic hereditary telangiectasia Capillary/cavernous hemangiomas Lymphangioma AV malformation of the tongue Hemangiolymphangiomas
- Photocoagulation of superficial vessels .
- Vaporization of superficial blood or lymph containing vessels
- Treatment of superficial tongue lesions .
- Tissue management and hemostasis for crown and bridge impressions
- Incision and drainage for abscess .
- Gingivoplasty/ gingivectomy Operative procedures Crown and bridge, gingival reduction Crown lengthening
- Hyperplasia (Drug, Irritation, Epulus, ... )
- Hemostasis during dental procedures ●
- Operculectomy (Operculotomy) ●
- Excisional biopsy .
- Free Ginvical Graft (Adjunct): . Hemostasis of donor site Hemostasis of graft site
- Vestibuloplasty .
- Soften Gutta Percha .
- Treatment of canker sores, herpetic lesions, and aphthous ulcers
- Laser-assisted bleaching/whitening
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
◆
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use: *✓*
(Per 21 CFR 801.109) ,
sion of General, Restorative Neurological Devices
510(k) Number K022327
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.