SYNTHES MANDIBLE EXTERNAL FIXATOR

K050378 · Synthes (Usa) · MQN · Mar 15, 2005 · Dental

Device Facts

Record IDK050378
Device NameSYNTHES MANDIBLE EXTERNAL FIXATOR
ApplicantSynthes (Usa)
Product CodeMQN · Dental
Decision DateMar 15, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.4760
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Synthes Mandible External Fixator is intended to stabilize and provide treatment for fractures of the maxillofacial area, including severe open mandibular fractures, highly comminuted closed fractures, nonunions and delayed unions (especially associated with infection), fractures associated with infections, tumor resections, facial deformity corrections, gunshot wounds, pan facial fractures, burn maintenance, and bone grafting defects.

Device Story

Synthes Mandible External Fixator is a mechanical orthopedic system used for stabilization of maxillofacial fractures. Components include adjustable parallel pin clamps, 4.0/2.5 Schanz screws, 2.0 mm K-wires, and 4.0 mm titanium rods. The device is designed for use in the MR environment. It is applied by surgeons in a clinical or surgical setting to provide rigid fixation for bone segments. The system stabilizes fractures, nonunions, or defects, facilitating healing and structural support during recovery.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

System consists of adjustable parallel pin clamps (MR safe), 4.0/2.5 Schanz screws, 2.0 mm K-wires, and 4.0 mm titanium rods. Mechanical fixation device; no energy source or software.

Indications for Use

Indicated for patients requiring stabilization and treatment of maxillofacial fractures, including severe open mandibular fractures, highly comminuted closed fractures, nonunions, delayed unions, infection-associated fractures, tumor resections, facial deformity corrections, gunshot wounds, pan facial fractures, burn maintenance, and bone grafting defects.

Regulatory Classification

Identification

A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a logo that appears to be a stylized representation of a human figure inside of a circle. There is a horizontal line underneath the word "SYNTHES". | 1<br>ﺮ<br>------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| 3.0 510(k) Summary . : ﺴﻪ इ · : : : : : : | Sponsor: | Synthes (USA)<br>1302 Wrights Lane East<br>West Chester, PA 19380<br>(610) 719-5000 | MAR 1 5 2005 | |-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------| | Device Name: | Synthes Mandible External Fixator | | | Classification: | Class II, 21 CFR §872.4760<br>Bone Plate | | | Predicate Device: | Synthes Mandible External Fixator | | | Device Description: | Synthes Mandible External Fixator consists of the following<br>components: An adjustable parallel pin clamp - MR safe; 4.0/2.5<br>Schanz Screws, 2.0 mm K-wires; and 4.0 mm Titanium Rods. The<br>components of this system are designed for use in the MR<br>environment. The rods and Schanz Screws are available in various<br>lengths. | | | Intended Use: | Synthes Mandible External Fixator is intended to stabilize and<br>provide treatment for fractures of the maxillofacial area, including<br>severe open mandibular fractures, highly comminuted closed<br>fractures, nonunions and delayed unions (especially associated with<br>infection), fractures associated with infections, tumor resections,<br>facial deformity corrections, gunshot wounds, pan facial fractures,<br>burn maintenance, and bone grafting defects. | | | Substantial<br>Equivalence: | Documentation is provided which demonstrates that Synthes<br>Mandible External Fixator is substantially equivalent to other<br>legally marketed Synthes devices. | | : 000005 : . {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized graphic of a caduceus, a symbol often associated with healthcare, featuring a staff with a serpent entwined around it. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Sheri L. Musgnung Senior Regulatory Affairs Specialist Synthes (USA) 1690 Russell Road Paoli, Pennsylvania 19301 MAR 1 5 2005 Re: K050378 Trade/Device Name: Synthes Mandible External Fixator Regulation Number: 872.4760 Regulation Name: Bone Plate Regulatory Class: II Product Code: MQN Dated: February 14, 2005 Received: February 15, 2005 Dear Ms. Musgnung: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Page 2 – Ms. Musgnung Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours, Chiu Lin, Ph.D. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a circular logo with a design inside. A registered trademark symbol is located to the right of the word. ## Indications for Use | 510(k) Number (if known): | K050378 | |---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name: | Synthes Mandible External Fixator | | Indications: | Synthes Mandible External Fixator is intended to stabilize and<br>provide treatment for fractures of the maxillofacial area,<br>including severe open mandibular fractures, highly comminuted<br>closed fractures, nonunions and delayed unions (especially<br>associated with infection), fractures associated with infections,<br>tumor resections, facial deformity corrections, gunshot wounds,<br>pan facial fractures, burn maintenance, and bone grafting defects. | Prescription Use _ X (Per 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use_ (21 CFR 807 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Susan Quann . , at Anacthesiology, General Hospital onon Control. Dental Devices Number: L050875 000004
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...