SPIRIT, MODEL HAS

K050153 · Caire, Inc. · BYJ · Apr 15, 2005 · Anesthesiology

Device Facts

Record IDK050153
Device NameSPIRIT, MODEL HAS
ApplicantCaire, Inc.
Product CodeBYJ · Anesthesiology
Decision DateApr 15, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5655
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The CARE Spirit 300 HAS will provide a source of supplemental oxygen for ambulatory home healthcare patients, by vaporizing 93% liquid oxygen and then dispensing it to the patient via an integral electronic conserving device. The Spirit 300 HAS is intended to be transfilled only by the In-X Corporation's Home-Away System. The Spirit 300 HAS is neither a life sustaining nor life supporting device.

Device Story

Spirit 300 HAS is a portable liquid oxygen unit for ambulatory home use. Device consists of a vacuum-insulated cryogenic dewar (0.3L capacity), vaporizer coil, economizer regulator, pressure relief valves, manifold, and electronic conserving device PCB. Unit is filled via In-X Corporation Home-Away System. Vaporizer warms liquid oxygen to gas. Manifold directs gas to either conserving device or 2 LPM continuous flow outlet. Conserving device delivers pulse doses at onset of inhalation (approx. 15 ml/LPM setting) for settings 1, 1.5, 2, 3, 4, and 5 LPM. Patient uses device to receive supplemental oxygen; healthcare provider prescribes flow settings. Device improves patient mobility by providing portable oxygen source.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Portable liquid oxygen unit. Components: vacuum-insulated cryogenic dewar, vaporizer coil, economizer regulator, pressure relief valves, manifold, electronic conserving device PCB. Energy source: electronic PCB for conserving function. Form factor: small, lightweight portable unit. Connectivity: none. Sterilization: not specified.

Indications for Use

Indicated for ambulatory home healthcare patients requiring supplemental oxygen. Not for life-sustaining or life-supporting use.

Regulatory Classification

Identification

A portable liquid oxygen unit is a portable, thermally insulated container of liquid oxygen that is intended to supplement gases to be inhaled by a patient, is sometimes accompanied by tubing and an oxygen mask. An empty portable liquid oxygen unit is a device, while the oxygen contained therein is a drug.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K050153 ### APR 1 5 2005 510(k) Summary Caire, Inc Submitter: 1800 Sandy Plains Industrial Parkway Suite 316 Marietta, GA 30066 Contact Person: Dan Chlan BioMedical Quality System / Compliance Manager (770) 257-1279 Phone Number: (770) 425-4740 Facsimile Number: dan.chlan@chart-ind.com E-Mail Address: December 28, 2004 Date Prepared: Spirit 300 HAS Device Name: Device Class: II Anesthesiology Classification Panel: Portable Liquid Oxygen Unit Classification Name: 868.5655 CFR Section: BYI Product Code: Spirit 300 [FDA 510(k) #K013251] Predicate Device: ### Device Description: The Spirit 300 HAS is a small, lightweight, liquid oxygen portable unit. The unit consists of a vacuum-insulated cryogenic dewar, a vaporizer coil, an economizer regulator, two pressure safety relief valves, a manifold, an electronic printed circuit board conserving device and a protective case. The dewar has a capacity and capability to store 0.3 liters of liquid oxygen. The vaporizer coil warms the oxygen gas to a suitable temperature, as it exits the dewar. Dependant on the flow selection setting, the manifold either delivers the gas to the conserving device PCB or delivers a 2 LPM continuous flow directly to the oxygen outlet. The continuous flow setting is labeled CF on the flow selector. The conserving device has equivalent setting for 1, 1.5, 2, 3, 4, and 5 LPM prescriptions. At the various conserving device settings, the outlet gas is delivered in quick pulse dosages just at the onset of patient inhalation. The amount of gas delivered (with each breath) is approximately 15 ml/LPM setting. The Spirit 300 HAS is filled from a concentrator liquefaction device. ### Indications For Use: The CARE Spirit 300 HAS will provide a source of supplemental oxygen for ambulatory home healthcare patients, by vaporizing 93% liquid oxygen and then dispensing it to the patient via an integral electronic conserving device. The Spirit 300 HAS is intended to be transfilled only by the In-X Corporation's Home-Away System. The Spirit 300 HAS is neither a life sustaining nor life supporting device. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized design of three overlapping human profiles facing to the right, creating a sense of movement and unity. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR 1 5 2005 CAIRA, Incorporated Mr. Dan Chlan Quality System / Compliance Manager BioMedical Division 1800 Sandy Plains Industrial Parkway, Suite 316 Marietta, Georgia 30066 Re: K050153 Trade/Device Name: Spirit 300 HAS Regulation Number: 868.5655 Regulation Name: Portable Liquid Oxygen Unit Regulatory Class: II Product Code: BYJ Dated: March 21, 2005 Received: March 29, 2005 Dear Mr. Chlan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting, your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Page 2 - Mr. Chlan Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Chiu Lin, Ph.D. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use # 510(k) Number (if known): K () 50 1 5 3 Spirit 300 HAS Device Name: The Spirit 300 HAS will provide a source of Indications for use: supplemental oxygen for ambulatory home healthcare patients, by vaporizing 93% liquid oxygen and then dispensing it to the patient via an integral electronic conserving device. The Spirit 300 HAS is intended to be transfilled only by the In-X Corporation's Home-Away System. The Spirit 300 HAS is neither a life-sustaining nor life-supporting device. Prescription Use X (21 CFR 801 Subpart D) AND/OR Over-the-Counter Use (21 CFR 801 Subpart C) ### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Clus ission Sign-Off) Number KOS0153 <sup>.</sup>sion of Anesthos.cology, Ganeral Hospital. John Control. Dental Devices
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...