K013251 · Caire, Inc. · BYJ · Dec 18, 2001 · Anesthesiology
Device Facts
Record ID
K013251
Device Name
SPIRIT 300
Applicant
Caire, Inc.
Product Code
BYJ · Anesthesiology
Decision Date
Dec 18, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5655
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The CAIRE Spirit 300 will provide a source of supplemental oxygen for ambulatory home healthcare patients, by vaporizing medical grade liquid oxygen and then dispensing it to the patient via an integral electronic conserving device. The Spirit 300 is neither a life sustaining nor life supporting device.
Device Story
Portable liquid oxygen unit; stores 0.3L liquid oxygen in vacuum-insulated cryogenic dewar. Vaporizer coil warms oxygen; economizer regulator manages gas/liquid draw based on internal pressure (20 PSI threshold). Electronic conserving device delivers oxygen in pulse doses at onset of inhalation (approx. 15 ml/LPM setting) or continuous flow (2 LPM). Used by ambulatory home patients. Conserving device extends duration compared to continuous flow. Healthcare providers prescribe flow settings (1, 1.5, 2, 3, 4, 5 LPM). Device improves patient mobility by providing portable supplemental oxygen.
Clinical Evidence
Bench testing only. Non-clinical testing compared oxygenation performance of Spirit 300 against Caire Stroller, Puritan-Bennett HELiOS, and DeVilbiss LP05PP. Results indicated Spirit 300 oxygenates similarly to predicates, with higher sensitivity noted compared to the DeVilbiss unit.
Indicated for ambulatory home healthcare patients requiring supplemental oxygen. Not for life-sustaining or life-supporting use.
Regulatory Classification
Identification
A portable liquid oxygen unit is a portable, thermally insulated container of liquid oxygen that is intended to supplement gases to be inhaled by a patient, is sometimes accompanied by tubing and an oxygen mask. An empty portable liquid oxygen unit is a device, while the oxygen contained therein is a drug.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "CHART" in white letters inside of a black oval. The word is written in a bold, sans-serif font. The oval has a thin white border around it. The image is simple and eye-catching.
CAIRE. MVE
# DEC 1 8 2001
K013251
# 510(k) Summary
| Applicant: | CAIRE Inc. |
|-------------------|-------------------------------------------------------|
| Address: | 3505 County Road 42 West<br>Burnsville, MN 55306-3803 |
| Contact: | Roger Briese<br>Bio-Medical Engineering Manage |
| Phone Number: | (952) 882-5071 |
| Facsimile Number: | (952) 882-5175 |
| E-Mail Address: | roger.briese@chart-ind.com |
| Date: | September 20, 2001 |
| Device Name: | Spirit 300 |
| Device Class: | II |
Classification Panel: Anesthesiology 868.5655 CFR Section: BYJ Product Code:
# Legally Marked Device to Which Substantially Equivalency is Claimed:
Caire Stroller [FDA 510(k) #K800742] Puritan Bennett HELiOS [FDA 510(k) #K993220] DeVilbiss LP05PP [FDA 510(k) #K833994]
#### Description of the Device:
The Spirit 300 is a small, lightweight, liquid oxygen portable unit. The unit consists of a vacuuminsulated cryogenic dewar, a vaporizer coil, an economizer regulator, two pressure safety relief valves, a manifold, an electronic printed circuit board conserving device and a protective case. The dewar has a capacity and capability to store 0.3 liters of liquid oxygen. The vaporizer coil warms the oxygen gas to a suitable temperature, as it exits the dewar. The economizer regulator riams the ony gas from the dewar head space, if the internal pressure is above 20 PSI. If the internal pressure is below 20 PSI, the economizer regulator will draw liquid oxygen from the bottom of the dewar, which must then have to pass through a long vaporizing coil. The economizer regulator assures a longer duration time for the unit, because it allows the gas in the dewar head regulator be used for oxygen delivery. The gas outlet of the economizer regulator connects with the manifold. Dependant on the flow selection setting, the manifold either delivers the gas to the the manifolial Deportually diverts a 2 LPM continuos flow directly to the oxygen outlet. The continuos flow setting is labeled CF on the flow selector. The conserving device has equivalent settings for 1, 1.5, 2, 3, 4 and 5 LPM prescriptions. At the various conserving device settings, the outlet gas is delivered in quick pulse dosages just at the onset of patient inhalation. The amount of gas delivered (with each breath) is approximately 15 ml/LPM setting.
{1}------------------------------------------------
#### MVE CAIRE.
Image /page/1/Picture/1 description: The image shows the word "CHART" in white letters inside of a black oval. The word is written in a bold, sans-serif font. The oval has a white border around it.
# KO 13251
#### Intended Use of the Device:
The CAIRE Spirit 300 will provide a source of supplemental oxygen for ambulatory home healthcare patients, by vaporizing medical grade liquid oxygen and then dispensing it to the patient via an integral electronic conserving device. The Spirit 300 is neither a life sustaining nor life supporting device.
### Technological Characteristics: CAIRE Spirit 300 vs. CAIRE Stroller
The main technical difference between these devices is that the Spirit 300 incorporates an integral electronic conserving device to deliver oxygen doses with patient inhalation, while the Stroller delivers oxygen in a continuous flow. The Spirit 300 is also smaller in both size and weight than the Stroller. However, the dewar design, plumbing style, fill connectors, relief safety valves and vent valves of these devices are either quite similar, if not common. With the method of oxygen delivery being the core technical difference, the non-clinical testing clearly shows that the Spirit 300 oxygenates similarly to the Stroller.
#### Technological Characteristics: CAIRE Spirit 300 vs. Puritan-Bennett HELiOS
The main technical difference between these devices is that the Spirit 300 incorporates an integral electronic conserving device to deliver oxygen doses with patient inhalation, while the HELiOS incorporates a pneumatic conserving device to deliver the oxygen doses. Other than this technical point, the Spirit 300 and HELiOS are quite similar in other technical/physical aspects. Again, the non-clinical testing clearly shows that the Spirit 300 oxygenates similarly (if not significantly better) than HELiOS.
# Technological Characteristics: CAIRE Spirit 300 vs. DeVilbiss #LP05PP
The technical design of the Spirit 300 and the DeVilbiss #LP05PP are quite similar. The physical size of the two devices is different. The Spirit 300 is a 0.34L capacity unit, while the DeVilbiss #LP05PP is a 0.49L capacity unit. Specific to the conserving devices are very similar in technical operation. Again, the non-clinical testing clearly shows that the Spirit 300 oxygenates similarly to the DeVilbiss product. Note the non-clinical testing indicated that the Spirit 300 is more sensitive than this DeVilbiss product. This explains why the Spirit 300 oxygenated a little better at the higher flow settings.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the department's name around the perimeter. Inside the circle, there is a symbol consisting of three stylized, overlapping shapes that resemble birds in flight or abstract human figures.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 8 2001
Caire, Inc. c/o Mr. Mark Job TÜV Product Service, Inc. 1775 Old Highway 8 NW, Suite #104 New Brighton, MN 55112-1891
Re: K013251 Spirit 300 Regulation Number: 868.5655 Regulation Name: Portable Liquid Oxygen Unit Regulatory Class: Class II (two) Product Code: 73 BYJ Dated: December 7, 2001 Received: December 10, 2001
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may ontrols provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 - Mr. Mark Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
N. Dallas Teller
Bram D. Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1
510(k) Number (if known):
SPIRIT 300 Device Name:_
Indications For Use:
The CAIRE Spirit 300 will provide a source of supple for ambulatory home healthcare patients by raporizing medical grade liquia axygen and then dispensing it to the patient via an integral electronic conserving anserving dev
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription
(Optional Format 3-10-98)
Division of Cardiovascular & Respiratory Devices
510(k) Number K013251
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.