K043262 · GE Healthcare · LNH · Jan 12, 2005 · Radiology
Device Facts
Record ID
K043262
Device Name
GE SIGNA EXCITE MR SURGICAL OPTION
Applicant
GE Healthcare
Product Code
LNH · Radiology
Decision Date
Jan 12, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The MR Surgical Option when integrated with the GE Signa® 3.0T or 1.5T magnetic resonance scanner (previously cleared K041476) can be used to produce head and whole body magnetic resonance images that are high resolution, high signal-to-noise ratio, with short scan times. The Signa® Excite MR system with Surgical Option is indicated for use as a diagnostic imaging device to produce axial sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the Signa® Excite system with Surgical Option reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. The GE Signa® Excite system with Surgical Option can provide MR images at various stages of a surgical procedure, and provides the capability to transfer a patient to the MR system table directly from a surgical table, or patient transport system. Images can be obtained using the MR scanner body coil, or local surface coils, and the surface coils used with the integrated system may / can accommodate sterile draping for surgical procedures.
Device Story
Device is a modified MR table cradle and transfer board system for GE Signa 1.5T/3.0T MR scanners. Enables patient transfer from surgical table or transport system to MR table without repositioning. Used in integrated Surgical/MR suites. Operated by clinical staff. Facilitates intraoperative MR imaging. Benefits include seamless patient movement between OR and MR, reducing risk of repositioning during surgery. Output is MR images/spectra for physician interpretation to assist in diagnosis.
Clinical Evidence
Bench testing only. Evaluated for compliance with IEC 60601-1 (Medical Equipment Safety) and IEC 60601-2-33 (Particular Requirements for Safety of Magnetic Resonance Equipment). No clinical data provided.
Technological Characteristics
Modified table cradle and transfer board for GE Signa 1.5T/3.0T MR systems. Integrates with existing MR scanner hardware. Supports sterile draping for surgical use. Complies with IEC 60601-1 and IEC 60601-2-33 standards.
Indications for Use
Indicated for diagnostic imaging of head and whole body (including neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, musculoskeletal) in patients requiring MR imaging, including during surgical procedures. Contraindications not specified.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
# JAN 1 2 2005
# 510(K) SUMMARY
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.92(c).
| Submitter: | GE Healthcare<br>PO Box 414<br>Milwaukee, WI 53201 |
|-----------------|-------------------------------------------------------------|
| Contact Person: | Larry A. Kroger Ph.D.<br>Senior Regulatory Programs Manager |
| Telephone: | 262-544-3894 |
| Fax: | 262-548-4768 |
| Date Prepared: | November 12, 2004 |
#### Device Name:
GE Signa® Excite MR Surgical Option
#### Marketed Device:
The GE Signa® Excite MR Surgical Option is substantially equivalent to the currently marketed Signa" 1.5T and 3.0T MR Excite System (K041476). The main difference is a modification to the table cradle that allows a separate transfer board to be transferred onto the MR Signa® table or patient transporter to enable traditional MR scanning of the patient.
#### Device Description:
The GE Signa® Excite MR Surgical Option is a MR table and transfer board that allows the patient to be moved from a surgical table to the MR scanner without changing the patient's position. It can be used in conjunction with a surgical table and other MR compatible equipment, such as but not limited to, patient monitoring equipment, IV pole, patient restraints, patient positioning accessories, and skull clamps that can be mounted to the patient transfer board, to facilitate ease of movement between the MR and The GE Signa® Excite MR Surgical Option can be used with single, or multiple surgical tables. operating rooms. The option will seamlessly integrate patient transfer from a surgical table and facilitate imaging within the GE Signa® Excite MR system. The patient, once transferred onto the MR system table, can then be scanned in the same manner as with traditional routine MR imaging. This device allows the GE Signa® Excite MR system to be utilized in a configuration of an integrated Surgical / MR suite.
#### Indications for Use:
The MR Surgical Option when integrated with the GE Signa® 3.0T or 1.5T magnetic resonance scanner (previously cleared K041476) can be used to produce head and whole body magnetic resonance images that are high resolution, high signal-to-noise ratio, with short scan times. The Signa® Excite MR system with Surgical Option is indicated for use as a diagnostic imaging device to produce axial sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by
{1}------------------------------------------------
the Signa" Excite system with Surgical Option reflect the spatial distribution of protons (hydrogen nucle) exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained musician yield information that may assist in diagnosis. The GE Signa" Excite system with Surgical phynelan fretarine MR images at various stages of a surgical procedure, and provides the capability to transfer a patient to the MR system table directly from a surgical table, or patient transport system. Images can be obtained using the MR scamer body coil, or local surface coils, and the surface coils used with the integrated system may / can accommodate sterile draping for surgical procedures.
### Comparison with Predicate Device:
The GE Signa® Excite MR Surgical Option utilized the same of the previously cleared Signa " 1.5T and 3.0T Excite MR System (K041476) with the main difference being a modification to the cradle that allows a separate transferred onto the MR Signa" table from a surgical table or patient transporter to enable traditional MR scanning of the patient.
### Summary of Studies:
The GE Signa® Excite MR Surgical Option has been evaluated to the appropriate IEC 60601-1 International Medical Equipment Safety standard and IEC 60061-2-33 Particular Requirements for Safety of Magnetic Resonance Equipment for Medical Diagnosis.
The GE Signa® Excite MR Surgical Option utilizes the currently marketed Signa® 1.5T and 3.0T Excite MR System (K041476). The modified table cradle in the Surgical Option adds the capability to allow a separate transfer board to be transferred onto the MR Signa® table from an operating table or patient transporter.
### Conclusion:
It is the opinion of GE that the GE Signa® Excite MR Surgical Option is substantially equivalent to the Signa® Excite 1.5T and Signa® Excite 3.0T Magnetic Resonance System Tables. Usage of the GE Signa® Excite MR Surgical Option and any potential hazards of the modified table cradle have been addressed and do not present any safety concerns.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the department's name around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or a stylized human figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 1 2 2005
Larry A. Kroger, Ph.D. Senior Regulatory Programs Manager GE Healthcare W-400, P.O. Box 414 MILWAUKEE WI 53201
Re: K043262
Trade/Device Name: GE Signa® Excite MR Surgical Option Regulation Number: 21 CFR §892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 LNH Dated: November 23, 2004 Received: November 24, 2004
Dear Dr. Kroger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Nancy C. brogdon
Nancy C. Brogdon
Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# 510(k) Number (if known): _
Device Name: GE Signa Excite MR Surgical Option
#### Indications For Use:
The MR Surgical Option when integrated with the GE Signa® 3.0T or 1.5T magnetic resonance scanner (previously cleared K041476) can be used to produce head and whole body magnetic seanner (previously viat are high resolution, high signal-to-noise ratio, with short scan times. The Signal Excite MR system with Surgical Option is indicated for use as a diagnostic imaging device to produce axial sagittal, coronal and oblique images, spectroscopic images, and/or de rece to produce anni if the internal structures and organs of the entire body, including, but .. peetin, a finance mage of, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskelal regions of the body. The images produced by the Signa® Excite vessem with Surgical Option reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. The GE Signa® Excite system with Surgical Ontion can provide MR images at various stages of a surgical procedure, and provides Burgear Spiren a transfer a patient to the MR system table directly from a surgical table, or patient transport system. Images can be obtained using the MR scanner body coil, or local surface coils, and the surface coils used with the integrated system may / can accommodate sterile draping for surgical procedures.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|--|
| <i>David G. Severson</i> | |
| (Division Sign-Off) | |
| Division of Reproductive, Abdominal, | |
| and Radiological Devices | |
| K043262 | |
| Prescription Use<br>21 CFR 801.109) | <div style="display:inline-block;"> ✓ </div> <div style="display:inline-block;">OR Over-The-Counter Use</div> <div style="display:inline-block;">(Per</div> |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.