GE SIGNA EXCITE 1.5T MR SYSTEM, GE SIGNA EXCITE 3.0T MR SYSTEM
K041476 · Ge Medical Systems, LLC · LNH · Jun 17, 2004 · Radiology
Device Facts
Record ID
K041476
Device Name
GE SIGNA EXCITE 1.5T MR SYSTEM, GE SIGNA EXCITE 3.0T MR SYSTEM
Applicant
Ge Medical Systems, LLC
Product Code
LNH · Radiology
Decision Date
Jun 17, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The GE Signa® Excite MR system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. The Signa Excite MR system is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, and musculoskeletal regions of the body. The images produced by the Signa Excite system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Device Story
The GE Signa® Excite 1.5T and 3.0T MR systems are superconducting magnetic resonance scanners. They acquire 2D and 3D images and spectra by detecting the spatial distribution of protons (hydrogen nuclei) using magnetic resonance. The systems support 1, 4, 8, and 16 independent receive channels, with modular expansion capability in 16-channel increments. The device is operated by trained professionals in clinical settings to produce high-resolution anatomical images and spectroscopic data. These outputs are interpreted by physicians to assist in clinical diagnosis. The systems are designed for high signal-to-noise ratios and short scan times.
Clinical Evidence
Bench testing only. The systems were evaluated against IEC 60601-1 International Medical Equipment Safety Standard and IEC 60601-2-33 Particular Requirements for Safety of Magnetic Resonance Equipment for Medical Diagnosis.
Technological Characteristics
Superconducting magnet operating at 1.5T and 3.0T. Supports 1, 4, 8, 16 independent receive channels with modular expansion. Acquires 2D/3D images in sagittal, coronal, axial, oblique, and double oblique planes. Supports multinuclear spectroscopy. Complies with IEC 60601-1 and IEC 60601-2-33 standards.
Indications for Use
Indicated for use as a diagnostic imaging device for the entire body, including head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, and musculoskeletal regions. Produces axial, sagittal, coronal, and oblique images and spectra to assist in diagnosis. For use by trained physicians.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Signa® 3.0T Infinity with Excite Technology MR System (K030874)
Submission Summary (Full Text)
{0}------------------------------------------------
K041476
Image /page/0/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circular shape. The letters are stylized and connected, and the entire logo is in black against a white background.
GE Healthcare Technologies
P.O. Box 414, Milwaukee, WI 53201
# JUN 1 7 2004
## 510(k) Summary
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.92(c).
| Submitter: | GE Healthcare Technologies<br>PO Box 414<br>Milwaukee, WI 53201 |
|-----------------|-----------------------------------------------------------------|
| Contact Person: | Larry A. Kroger Ph.D.<br>Manager, Regulatory Programs |
| Telephone: | 262- 544-3894 |
| Fax: | 262- 548-4768 |
| Date Prepared: | May 28, 2004 |
## Device Name:
GE Signa® Excite 1.5T MR System, and
GE Signa® Excite 3.0T MR System GL Signa "Exons C.o. P.M. System, 21 CFR 892.1000, 90-LNH
### Marketed Device:
The Signa® Excite 1.5T MR System is substantially equivalent to the currently marketed The Signa® 1.5T Infinity MR system (K013636) with the main differences being a change to the Signa - 1.51 millily Mil Dyoton (10 receive channels, and allows for future expansion in 16 channel increments.
The Signa® Excite 3.0T MR System is substantially equivalent to the currently marketed The Signa "Excite U.S.T. MIT System (K030874) with the main differences Sigra 5.01 minity with Exono Toonnelogy the Stateen independent receive channels, and allows for future expansion in 16 channel increments.
## Device Description:
The Signa® Excite 1.5T and 3.0T Magnetic Resonance Systems are a modification to the The Oight - Exole 110 Pana 2007 Mass 6, and K030874 which utilizes a superconducting previously cleared Mr Systems No roooo, and Roover images. The Signa® Excite magnet to acquire 2D Single and make and make a superconducting magnet operating at 1.51 and 0.0 r Magnolis (1600)and institution system supports 1, 4, 8, 16 independent receive channels and multiple independent coil elements per channel during a single acquisition channolo and manification is designed for expansion in 16 channel Sencs. The system can image in the sagittal, coronal, axial, oblique and double oblique increments. The system our mage in the sage are acquired and reconstructed using 2D and planes, asing ranous palos objaction. The system is intended for high-resolution anatomical applications, short scan times, and multinuclear spectroscopy.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized font, enclosed within a circular shape. The circle has a swirling pattern around the letters, giving it a dynamic and recognizable appearance.
## GE Healthcare Technologies
P.O. Box 414, Milwaukee, Wi 53201
## Indications for Use:
The GE Signa® Excite MR system is a whole body magnetic resonance scanner designed to The GE Signa - Excite Mrs System is a wholo bout magnolo rean times. The Signal Excite MR
support high resolution, high signal-to-noise ratio, and short scan times. Costition system is indicated for use as a diagnostic imaging device to produce axial sagittal, coronal system is indicated for use as a diagnocto imaging spectra, dynamic images of the internal and oblique images, spectroscopic intagos, and or special, head, nead, neck, TMJ, spine, a structures and organs of the entire body, modale, blood vessels, and much as and only and breast, theat, abuonten, perio, jointo, proclare, and excite system reflect the spatial distribution of the body. The inhages produced by the elgha- Enence. These images and/or spectra when profons (rydrogon hadler) extraining may assist in diagnosis.
and the country of the work of work a work a work a want
## Comparison with Predicate Device:
The Signa® Excite 1.5T and 3.0T MR Systems are a modification of the previously cleared MR The Signa Excite 1.51 and 6.61 with the main differences being the change to the receive systems K015000, and 100007 exten independent receive channels, and allows for future expansion in 16 channel increments.
## Summary of Studies:
The Signa® Excite 1.5T and 3.0T Magnetic Resonance Systems were evaluated to the The Signa - Excite 1.0 Parks of Philaghts as well as the IEC 60601-1 International Medical appropriate NEWs porformance etanceres in 33 Particular Requirements for Safety of Equipment Salety Standard and 120 0000 - 12 Signal "Excite 1.5T Magnetic
Magnetic Resonance Equipment for Medical Diagnosis. The Signal "Excite 1.5T Magnetic Floonance System. The Signa® Excite 3.0T Magnetic Resonance System is comparable to the currently marketed Signa® 3.0T Infinity with Excite Technology Magnetic Resonance System.
## Conclusion:
It is the opinion of GE that the Signa® Excite 1.5T Magnetic Resonance System is substantially it is the Signa® 1.5T Infinity Magnetic Resonance System. Usage of the Signa equivalorit to the organice System does not result in any new potential hazards.
It is the opinion of GE that the Signa® Excite 3.0T Magnetic Resonance System is substantially it is the Signa" 3.0T Infinity with Excite Technology Magnetic Resonance System. Usage of the Signa® Excite 3.0T Magnetic Resonance System does not result in any new potential hazards.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 9200 Gorporate Boulevard Rockville MD 20850
# JUN 1 7 2004
Larry A. Kroger, Ph.D. Senior Regulatory Programs Manager GE Healthcare Technologies P.O. Box 414 W-400 MILWAUKEE WI 53201
Re: K041476
Trade/Device Name: GE Signa® Excite 1.5T MR System, and GE Signa® Excite 3.0 MR Systen Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 LNH
Dated: May 28, 2004 Received: June 3, 2004
Dear Dr. Kroger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for associated in the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, there, manns of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r that FDA has made a determination that your device complies with other requirements of the Act that I Dri has statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brigdon
Nancv C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized font, enclosed within a circular shape. The circular shape has a swirling pattern around the letters. The logo is black and white.
Ko41476
GE Healthcare Technologies P.O. Box 414, Milwaukee, WI 53201
# STATEMENT OF INTENDED USE
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
#### GE Signa® Excite 1.5T MR System, and Device Name: GE Signa® Excite 3.0T MR System
# Indications for Use
The GE Signa® Excite MR system is a whole body magnetic resonance scanner designed The GE Signa Excile Mrs System is a whole boar light scan times. The Signal
to support high resolution, high signal-to-noise ratio, and short so produce axial to support high resolution, ngil signal-lo-rane, androing device to produce axial
Excite MR system is indicated for use as a diagnostic images, and/or spectra, dynamic Excite MH System IS indicated for use as a diagnose innages, and/or spectra, dynamic
sagittal, coronal and oblique images, spectroscopic inqueling, but not limi sagittal, coronal and oblique images, spectrosophy, including, but not limited to,
images of the internal structures and organs of the inite, inited vesels, and images of the internal structures and organs of the original , prostale , head, messels, and head, neck, TMJ, spine, breast, neat, abdoment, portly, positions
musculoskeletal regions of the body. The images produced by the Signal excites system musculoskeletal regions of the body. The integoe proclass as hibiting magnetic resonance.
reflect the spatial distribution of protons (hydrogen nucled information has reflect the spatial distribution of protons (hydrogen hastory only in of the mation that
These images and/or spectra when interpreted by a trained physician yield information may assist in diagnosis.
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801-109)
OR
Over-The-Counter Use_
David A. Anderson
5 . Xk) Number
E-2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.