K042781 · Realtimeimage, Inc. · LLZ · Oct 19, 2004 · Radiology
Device Facts
Record ID
K042781
Device Name
IPACS CARDIO
Applicant
Realtimeimage, Inc.
Product Code
LLZ · Radiology
Decision Date
Oct 19, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
iPACS CARDIO™is intended for the manipulation, display, and distribution of medical images. It can show images from different modalities and interfaces to various image storage and printing devices using DICOM or similar interface standards. The device assists Cardiology surgeons when doing preoperative planning and post-operative follow-up. Typical users of this system are trained professionals, for example orthopedic surgeons, physicians, and radiologists.
Device Story
iPACS CARDIO™ is a software add-on for the iPACS PRISMI™ system; functions as a Picture Archiving and Communication System (PACS) for digital images and Cine objects. Input consists of medical images from various modalities via DICOM or similar standards. Software provides tools for image manipulation, display, and distribution; assists cardiology surgeons in preoperative planning and post-operative follow-up. Operated by trained professionals (physicians, radiologists, surgeons) in clinical environments. Output is displayed on workstations; interpreted by clinicians to inform surgical planning and patient follow-up. Device does not contact patients or control life-sustaining equipment; requires human intervention for interpretation.
Clinical Evidence
No clinical data. Bench testing only; hazard analysis performed.
Technological Characteristics
Software-based PACS; runs on Windows 2000 or Windows XP. Interfaces with DICOM-compliant modalities and storage/printing devices. No patient contact; no life-sustaining control. Level of concern: Minor.
Indications for Use
Indicated for trained professionals including cardiology surgeons, orthopedic surgeons, physicians, and radiologists for the manipulation, display, and distribution of medical images for preoperative planning and post-operative follow-up.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Sectra Angiography Package and Sectra Cardiology Package (K034059)
{0}------------------------------------------------
K042781
Image /page/0/Picture/1 description: The image contains the words "RealTime" and "Image" stacked on top of each other. The word "RealTime" is in a larger font than the word "Image". There is also a small square image in the upper left corner of the image. The image is blurry and difficult to make out.
# OCT 1 9 2004
510(k) Summary of Safety and Effectiveness
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.
Date Prepared: September 3, 2004
Submitter's Information: 21 CFR 807.92(a)(1) Mr. Zvi Eintracht, CEO Real Time Image Inc. 1111 Bayhill Dr, San Bruno, CA94066, suite 290 Tel: 650.616.4671 Fax: 650.616.4680
Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2) iPACS CARDIO™ Trade Name: Picture Archiving Communications System Common Name: 892.2050 Device Classification: System, Image Processing Name:
Predicate Device: 21 CFR 807. 92(a)(3)
| Device Classification Name | system, image processing, radiological |
|---------------------------------------------|-------------------------------------------------------------|
| 510(k) Number | K034059 |
| Regulation Number | 892.2050 Class II |
| Device Name | Sectra Angiography Package and Sectra<br>Cardiology Package |
| Applicant | Sectra-imtec ab |
| Product Code | LLZ |
| Decision Date | 03/09/2004 |
| Decision | Substantially equivalent (SE) |
| Classification Advisory Committee Radiology | |
| Type | Traditional |
#### Device Description: 21 CFR 807 92(a)(4)
iPACS CARDIO™ is a software product that is added to the iPACS PRISMI™ device (K030751) and displays and manages various digital images and Cine objects in a Picture Archive and Communication System (PACS) environment and is intended to assist cardiology surgeons when doing preoperative planning and post-operative follow-up. iPACS CARDIO™ also includes various tools and can access additional tools within iPACS Prism.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the words "RealTime Image" in a sans-serif font. The word "RealTime" is in a larger font size than the word "Image". There is a square graphic in the upper left corner of the image. The graphic is a square with a smaller square to the left of it.
### Indications for Use: 21 CFR 807 92(a)(5)
iPACS CARDIO™is intended for the manipulation, display, and distribution of medical images. It can show images from different modalities and interfaces to various image storage and printing devices using DICOM or similar interface standards.
The device assists Cardiology surgeons when doing preoperative planning and post-operative follow-up.
Typical users of this system are trained professionals, for example orthopedic surgeons, physicians, and radiologists.
Technological Characteristics: 21 CFR 807 92(a)(6)
iPACS CARDIO™ runs together with iPACS Prism on Windows 2000 or Windows XP operating systems, (depending upon system configuration). The device does not contact the patient, nor does it control any life sustaining devices. A physician, providing ample opportunity for competent human intervention interprets images and information being displayed and printed.
#### Conclusion: 21 CFR 807 92(b)(1)
The 510(k) Pre-Market Notification for iPACS CARDIO™ contains adequate information and data to enable FDA - CDRH to determine substantial equivalence to the predicate device.
iPACS CARDIO™ has been and will be manufactured in accordance with the voluntary standards listed in the enclosed voluntary standard survey. The submission contains the results of a hazard analysis and the "Level of Concern for potential hazards has been classified as "Minor".
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# OCT 1 9 2004
Real Time Image, Inc. 1% Mr. N. E. Devine Responsible Third Party Official Entela, Inc. 3033 Madison Ave., SE GRAND RAPIDS MI 49548
Re: K042781
Trade/Device Name: iPACS CARDIO™ Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications systems
Regulatory Class: II Product Code: 90 LLZ Dated: October 6, 2004 Received: October 6, 2004
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced we have reviewed your bother ovece is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to the cherosate) to regally mant date of the Medical Device Amendments, or to devices that have been May 20, 1770, the chavannith the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the do not require approval or a provisions of the Act. The general controls provisions of the Act do floor to any go annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket a your al), it may be subject to such additional controls. Existing major regulations affecting your Apployally, it the 90 babyer of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that I rease of actives made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the r occares statues ats including, but not limited to registration and listing (21 CFR Part 807); labeling (2) CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) (21 Cric Fart 801); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I ms letter will anow you was wegin many of substantial equivalence of your device to a legally premaired predicated on " rivesults in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specifice at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the 001:57).
Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmadsmamain.html
Sincerely yours,
Nancy C. brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo with a black square in the upper left corner. A smaller black square is located to the left of the text. The text below the squares reads "RealTime Image" with the words stacked on top of each other.
# Indications for Use
510(k) Number (if known):
Device Name: iPACS CARDIO™
Indications For Use:
iPACS CARDIO™is intended for the manipulation, displaying, and distribution of medical images. It can show images from different modalities and interfaces to various image storage and printing devices using DICOM or similar interface standards.
The device assists Cardiology surgeons when doing preoperative planning and post-operative follow-up.
Typical users of this system are trained professionals, for example orthopedic surgeons, physicians, and radiologists.
Prescription Use
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C. Swagdon
(Division Sign-Off) ()
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K042781
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.