SECTRA ANGIOGRAPHY PACKAGE AND SECTRA CARDIOLOGY PACKAGE
K034059 · Sectra AB · LLZ · Mar 9, 2004 · Radiology
Device Facts
Record ID
K034059
Device Name
SECTRA ANGIOGRAPHY PACKAGE AND SECTRA CARDIOLOGY PACKAGE
Applicant
Sectra AB
Product Code
LLZ · Radiology
Decision Date
Mar 9, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Sectra Angiography and Cardiology Package provides quantitative and reproducible information regarding the calculated dimensions of arterial segments, and quantitative information regarding the calculated dimensions and calculated performance characteristics of the left vertical of the heart, imaged during angiographic x-ray procedures typically performed in cath labs. This information is suitable for use in the following applications: 1. Scientific and research studies, assessing the angiographic condition of patients and the result of device and drug therapeutics, 2. Review and analysis of patient angiographic imaging records, providing additional information to physicians and administrators. Typical users of the device are trained professionals, e.g. physicians, radiologists, cardiologists and scientists.
Device Story
Sectra Angiography and Cardiology Package is a software-based image processing system for cath lab environments. It takes angiographic x-ray images as input to perform quantitative analysis of arteriograms (QCA/QVA) and left ventricular angiographic images (LVA). The device calculates dimensions of arterial segments and performance characteristics of the left ventricle. Operated by physicians, radiologists, and cardiologists on a Sectra IDS5 Radiology Workstation, the system provides quantitative data to support diagnostic processes, research studies, and clinical review. Output is interpreted by the clinician to assess patient condition and therapeutic outcomes. The device does not contact the patient or control life-sustaining equipment; human intervention is required for all clinical decisions.
Clinical Evidence
No clinical data provided. Bench testing only.
Technological Characteristics
Software-based image processing system running on Windows 2000/XP PCs with Sectra IDS5 Radiology Workstation. Complies with ACR/NEMA DICOM 3.0 standards. Developed under ISO 9001:2000. Password-protected access control.
Indications for Use
Indicated for use by trained professionals (physicians, radiologists, cardiologists, scientists) to provide quantitative analysis of arterial segments and left ventricular performance characteristics from angiographic x-ray images. Used for scientific/research studies and clinical review of patient imaging records.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
QCA-View (Cardiovascular Measurement System - View) (K993761)
QCA-CMS (Quantitative Analysis of Coronary Angiograms) (K993763)
QLV-CMS (Quantitative Analysis of Left Ventricular Angiograms) (K993765)
{0}------------------------------------------------
MAR - 9 2004
# 510(k) Summary of Safety & Effectiveness
(as required by 21 CFR 807.92c)
Date Prepared:
Dec 18, 2003
# Submitter's Information:
Sectra Imtec AB Teknikringen 20 SE-583 30 Linköping Sweden Phone: +1 46 13 23 52 00 Fax: +1 46 13 21 21 85
#### Trade Name, Common Name, Classification:
(03,4054
| Trade name: | Sectra Angiography and Cardiology Package |
|----------------------|---------------------------------------------------|
| Common Names: | Picture Archiving and Communications System |
| Classification Name: | Image Processing System (LLZ) (21 CFR § 892.2050) |
### Predicate Devices:
| Applicant: | Medical Imaging Systems (MEDIS) |
|----------------|---------------------------------------------------------------|
| 510(k) Number: | K993761 |
| Device: | QCA-View (Cardiovascular Measurement System - View) |
| Applicant: | Medical Imaging Systems |
| 510(k) Number: | K993763 |
| Device: | QCA-CMS (Quantitative Analysis of Coronary Angiograms) |
| Applicant: | Medical Imaging Systems (MEDIS) |
| 510(k) Number: | K993765 |
| Device: | QLV-CMS (Quantitative Analysis of Left Ventricular Angiograms |
| Applicant: | Medical Imaging Systems (MEDIS) |
| 510(k) Number: | K023970 |
| Device | OVA-CMS (OVA-CMS Analytical software package) |
#### Device Description:
Sectra Angiography and Cardiology Package is intended to support radiologists and cardiologists in the diagnostic process to quantify findings. The device has functionality for quantitative analysis of the arteriograms (QCA and QVA) and quantitative analysis of the left ventricular angiographic images (LVA). Measurements can be applied on angiographic images.
{1}------------------------------------------------
### Indications for Use:
The Sectra Angiography and Cardiology Package provides quantitative and reproducible information regarding the calculated dimensions of arterial segments, and quantitative information regarding the calculated dimensions and calculated performance characteristics of the left vertical of the heart, imaged during angiographic x-ray procedures typically performed in cath labs.
This information is suitable for use in the following applications:
1. Scientific and research studies, assessing the angiographic condition of patients and the result of device and drug therapeutics,
2. Review and analysis of patient angiographic imaging records, providing additional information to physicians and administrators.
Typical users of the device are trained professionals, e.g. physicians, radiologists, cardiologists and scientists.
#### Technological Characteristics:
The Sectra Angiography and Cardiology Package will run on the Windows 2000, and Windows XP operating systems for PCs (as a minimum and depending upon system configuration) featuring a Sectra IDS5 Radiology Workstation.
#### Performance Data:
The subject device is developed according to ISO 9001:2000 and complies with ACR/NEMA Digital Imaging Communications in Medicine version 3.0.
#### Conclusion:
Similar to the predicate devices, the Sectra Angiography and Cardiology Package does not contact the patient, nor does it control any life sustaining devices. Images and information being reviewed, processed, relayed, and or transmitted are interpreted by a trained professional, e.g. radiologist, cardiologist, providing ample opportunity for competent human intervention. Device failures, which might result in partial or failed transmissions, images, or data, may be recovered from storage or re-transmission after correcting the problem(s). Passwords are required for operation and to protect against unauthorized use.
Based on the information supplied in this 510(k), we conclude that the subject device is safe, effective, and substantially equivalent to the predicate devices.
Ri AL
Peter Andersson Regulatory Affairs Officer Sectra Imtec AB Teknikringen 20 SE-58330 Linköping Sweden
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a human figure with outstretched arms.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR - 9 2004
Sectra Imtec AB % Mr. Carl Alletto United States Agent OTech, Inc. 1100 Lakeview Blvd. DENTON TX 76208
Trade/Device Name: Sectra Angiography and Cardiology Package Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ
Dated: December 22, 2003 Received: December 31, 2003
Dear Mr. Alletto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
Re: K034059
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. brydon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
:
510(k) Number: Ko3 4059
Device Name: Sectra Angiography and Cardiology Package
Indications For Use:
.
The Sectra Angiography and Cardiology Package provides quantitative and reproducible information regarding the calculated dimensions of arterial segments, and quantitative information regarding the calculated dimensions and calculated performance characteristics of the left vertical of the heart, imaged during angiographic x-ray procedures typically performed in cath labs.
This information is suitable for use in the following applications: 1. Scientific and research studies, assessing the angiographic condition of patients and the result of device and drug therapeutics, 2. Review and analysis of patient angiographic imaging records, providing additional information to physicians and administrators.
Typical users of the device are trained professionals, e.g. physicians, radiologists, cardiologists and scientists.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | |
|----------------------|--|
| (Per 21 CFR 801.109) | |
OR
| Over-The-Counter Use | |
|----------------------|--|
|----------------------|--|
(Optional Format 1-2-96)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
| 510(k) Number | K034059 |
|---------------|---------|
|---------------|---------|
Nancy Brogdon
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.