Mendec Spine is indicated for the treatment of pathological fractures of the vertebrae using a vertebroplasty or kyphoplasty procedure. Painful vertebral compression fractures may result from osteoporesis, benign lesions (hemangioma), and malignant lesions (metastatic cancers, myeloma).
Device Story
Mendec Spine is a two-component acrylic bone cement (PMMA-based) used in vertebroplasty or kyphoplasty procedures. The system consists of a powder component (PMMA-styrene copolymer, barium sulphate, benzoyl peroxide) and a liquid component (methylmethacrylate monomer, hydroquinone, N,N-dimethyl-p-toluidine). Components are mixed by the clinician prior to use to initiate polymerization. The resulting cement is injected into the vertebral body to stabilize pathological fractures. The device provides structural support to the fractured vertebra, potentially reducing pain and restoring vertebral integrity. It is intended for use by physicians in a clinical or surgical setting.
Clinical Evidence
No clinical data provided. Substantial equivalence was established through bench testing, including mechanical performance testing per ISO 5833:2002 and biocompatibility testing per ISO 10993 and ASTM F 451-99.
Technological Characteristics
Two-component PMMA-based acrylic bone cement. Powder: PMMA-styrene copolymer, barium sulphate (radiopacifier), benzoyl peroxide (initiator). Liquid: Methylmethacrylate monomer, hydroquinone (stabilizer), N,N-dimethyl-p-toluidine (promoter). Complies with ASTM F 451-99 and ISO 5833:2002. Sterilization: Solid components via ethylene oxide (EO) gas; liquid components via filtration. Single-use.
Indications for Use
Indicated for patients with painful vertebral compression fractures resulting from osteoporosis, benign lesions (hemangioma), or malignant lesions (metastatic cancers, myeloma) requiring vertebroplasty or kyphoplasty.
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
Stryker Spineplex Radiopaque Bone Cement (K032945)
Submission Summary (Full Text)
{0}------------------------------------------------
# K042415
JUN 9 – 2005
ﺎ
. : Mendec Spine
Company confidential
」
r
## Traditional 510(k)
# Summary of Safety and Effectiveness
| Official Correspondent: | Christine Brauer, Ph.D.<br>Regulatory Affairs Consultant<br>One Democracy Plaza<br>6701 Democracy Boulevard, Suite 700<br>Bethesda, MD 20817<br>tel: 301-545-1990<br>fax: 301-545-1992<br>E-MAIL: chrisbrauer@earthlink.net | | |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Trade Name: | Mendec Spine | | |
| Common Name: | Bone Cement | | |
| Classification Name: | Bone Cement, 888.3027 | | |
| Device code: | NDN, LOD | | |
| Manufacturer/Submitter: | TECRES S.p.A FDA Owner number: 9033624<br>Via Andrea Doria 10,<br>37066 Sommacampagna, Verona, Italia<br>Phone: +39 045 9217311<br>Fax: +39 045 9217330 | | |
| Contact person: | Massimo Grazioli, General Manager | | |
| Predicate device: | K033801: Kyphx HV-R Bone Cement (Kyphon Inc.).<br>K041584: Kyphx HV-R Bone Cement (Kyphon Inc.).<br>K032945: Stryker Spineplex Radiopaque Bone Cement (Stryker) | | |
| Indications for use: | Mendec Spine is indicated for the treatment of pathological<br>fractures of the vertebral body using a vertebroplasty or<br>kyphoplasty procedure. Painful vertebral compression fractures<br>may result from osteoporosis, bening lesions (hemangioma), and<br>malignant lesions (metastatic cancers, mylema) | | |
| Device Description: | Like the predicate devices. Mendec Spine acrylic resin is provided<br>as a two-component system. The powder component consists of a<br>PMMA-styrene copolymer with barium sulphate as a radiopacifier<br>and benzoyl peroxide as an initiator. The liquid component consists<br>of methylmethacrylate monomer, with the addition of<br>hydroquinone as a stabiliser and N.N-dimethyl-p-toluidine as<br>promoter. The powder and liquid components are mixed prior to<br>Section 4<br>Page 1/2 | | |
{1}------------------------------------------------
K042415
#### Company confidential
#### Traditional 510(k)
Mendec Spine
## Summary of Safety and Effectiveness
Table 4-1 compares the chemical composition of Mendec use. Spine to the predicate device.
#### Table 4-1 Chemical composition of Mendec Spine and Kyphx HV-R
| Chemical composition | KyphX HV-R<br>#K033801<br>#K041584 | Spineplex<br>#K032945 | Mendec Spine |
|------------------------------|------------------------------------|-----------------------|--------------|
| Powder | 20 g | 20 g | 20 g |
| Polymethymethacrylate | $68,0% w/w$ | $68,50% w/w$ | $67,50% w/w$ |
| Barium sulphate | $30,0% w/w$ | $30,0% w/w$ | $30,0% w/w$ |
| Benzoyl peroxide | $2,0% w/w$ | $1,5% w/w$ | $2.5% w/w$ |
| Liquid | 9 g | 9.96 ml (9.05 g) | 9.4 g |
| Methylmethacrylate (monomer) | $99,1% w/w$ | $97,4% w/w$ | $99.1% w/w$ |
| N.N-dimethyl-p-toluidine | $0,90% w/w$ | $2,60% w/w$ | $0,90% w/w$ |
| Hydroquinone | 75 ppm | 75 ppm | 75 ppm |
Mechanical tests: Mendec Spine acrylic resin was tested in direct comparison to the predicate device and verified substantially equivalent, as defined by ISO 5833:2002, "Implants for Sugery - Acrylic resin cements" Non-clinical test results: Performance testing demonstrated that Mendec Spine is substantially equivalent to the Kyphx HV-R with regard to functional characteristics Clinical results; Clinical information demonstrates that the intended use is substantially equivalent to the predicate indication for the treatment of pathological vertebral body fractures and does not adversely impact safety and effectiveness. Biocompatibility:
The materials used in Mendee Spine complies with ASTM F 451-99 point 10 and ISO 10993
Sterilization: The Mendee Spine acrylic resin is sterile and non-pyrogenic. The solid components are sterilised with ethylene oxide (EO) gas to a Sterility Assurance Level (SAL) of 10th. The liquid components are sterilized with filtration methods to a SAL of 102. The outer packaging containing the liquid component is sterifised with ethylene oxide (EO) gas. The acrylic resin is intended for single use only.
> Section 4 Page 2/3
{2}------------------------------------------------
K042415
Mendec Spine
Company confidential
LONGLAND REPRENTWERSON ..... ... ... ... ... .. .. .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
## Traditional 510(k)
INSUL INCLUSION IN MARKET PURGE
BARRE BL. GELLERS AND AND AND BELLER
www.wanawar.ua
consistences ... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
## Summary of Safety and Effectiveness
Substantial Equivalence: The chemical constituents in Mendec Spine acrylic resin are substantially equivalent to those in the predicates Kyphx HR-V (#K033801; #K041584) and in the predicate Spineplex (#K032945).
> Section 4 Page 3/3
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 9 - 2005
Tecres Spa C/o Christine L. Brauer, Ph.D Regulatory Affairs Consultant Brauer Device Consultants, LLC One Democracy Plaza 6701 Democracy Boulevard, Suite 700 Bethesda, Maryland 20817
Re: K042415
Trade/Device Name: Mendec Spine Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: II Product Code: NDN Dated: May 18, 2005 Received: May 18, 2005
Dear Dr. Brauer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as sct
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 -Christine L. Brauer, Ph.D
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120 . Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
H.A. Chuds
2. Miriam Provost, Ph.D.
Miriam Prov st. Ph.D Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): K042415
Device Name: Mendec Spine
Indications For Use:
Mendec Spine is indicated for the treatment of pathological fractures of the vertebrae using a vertebroplasty or kyphoplasty procedure. Painful vertebral compression fractures may result from osteoporesis, benign lesions (hemangioma), and malignant lesions (metastatic cancers, myeloma).
Prescription Use ﺮ ﺍﻟﻤﺮﺍﺟﻊ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
H. Curtis
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
(Division Sign-Off Division of General, Restorative, and Neurological Devices
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.