This device is indicated for patients with severe knee pain and disability due to: - Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. - Collagen disorders, and/or avascular necrosis of the femoral condyle. - Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy - Moderate valgus, varus, or flexion deformities. - The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery. The device is indicated for use when both cruciate ligaments have been excised. The device is intended for use as part of a cemented or uncemented knee prosthesis.
Device Story
NexGen LPS-Flex Prolong articular surfaces are components of a total knee prosthesis system. Device replaces damaged joint surfaces in patients with severe knee pain/disability. Components are manufactured from highly crosslinked polyethylene. Device is used in orthopedic surgery; implanted by surgeons. Functions as a bearing surface within a semiconstrained knee prosthesis system; provides stability in flexion when both cruciate ligaments are excised. Benefits include restoration of joint function and pain relief. Device is substantially equivalent to existing NexGen components, utilizing similar design and materials with minor dimensional modifications.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Articular surface components made of highly crosslinked polyethylene. Semiconstrained design for patellofemoral/tibial use. Compatible with cemented or uncemented femoral and tibial baseplate components.
Indications for Use
Indicated for patients with severe knee pain and disability due to rheumatoid, osteoarthritis, traumatic, or polyarthritis; collagen disorders; avascular necrosis of femoral condyle; post-traumatic joint loss; or moderate valgus, varus, or flexion deformities. Indicated for salvage of failed surgical attempts or cases requiring stability in flexion. Indicated for use when both cruciate ligaments are excised.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
NexGen Porous, Uncemented Femoral and Tibial Baseplate Components (K031061)
Submission Summary (Full Text)
{0}------------------------------------------------
007 1 3 2004
## Summary of Safety and Effectiveness
| Submitter: | Zimmer, Inc.<br>P.O. Box 708<br>Warsaw, IN 46581-0708 |
|----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Brandon Hipsher<br>Specialist, Corporate Regulatory Affairs<br>Telephone: (574) 371-8083<br>Fax: (574) 372-4605 |
| Date: | October 12, 2004 |
| Trade Name: | NexGen® Complete Knee Solution LPS-Flex<br>Prolong™ Highly Crosslinked Polyethylene<br>Articular Surfaces |
| Common Name: | Total Knee Prosthesis |
| Classification Name<br>and Reference: | Knee joint patellofemorotibial polymer/metal/<br>polymer semi-constrained cemented prosthesis<br>21 CFR § 888.3560 (JWH)<br>Knee joint patellofemorotibial metal/polymer<br>porous-coated uncemented prosthesis<br>21 CFR § 888.3565 (MBH) |
| Predicate Devices: | LPS-Flex Fixed Bearing Articular Surface<br>Components, manufactured by Zimmer, Inc.,<br>K991581, cleared July 30, 1999.<br><br>Prolong Highly Crosslinked Polyethylene Cruciate<br>Retaining (CR) Articular Surface Components,<br>manufactured by Zimmer, Inc., K013991, cleared<br>December 27, 2001.<br><br>NexGen Porous, Uncemented Femoral and Tibial<br>Baseplate Components, manufactured by Zimmer,<br>Inc., K031061, cleared October 9, 2003. |
| Device Description: | The NexGen LPS-Flex Prolong articular surfaces<br>are part of the NexGen system of semiconstrained, |
| Intended Use: | This device is indicated for patients with severe<br>knee pain and disability due to:<br>- Rheumatoid arthritis, osteoarthritis,<br>traumatic arthritis, polyarthritis.<br>- Collagen disorders, and/or avascular<br>necrosis of the femoral condyle.<br>- Post-traumatic loss of joint configuration,<br>particularly when there is patellofemoral<br>erosion, dysfunction or prior patellectomy<br>- Moderate valgus, varus, or flexion<br>deformities.<br>- The salvage of previously failed surgical<br>attempts or for a knee in which satisfactory<br>stability in flexion cannot be obtained at the<br>time of surgery.<br>The device is indicated for use when both cruciate<br>ligaments have been excised.<br>The device is intended for use as part of a cemented<br>or uncemented knee prosthesis. |
| Comparison to Predicate Device: | Except for a change in material and minor<br>dimensional modifications, LPS-Flex Prolong<br>articular surfaces are identical to the predicate<br>device. The modifications do not change the<br>intended use or the fundamental scientific<br>technology. The device is packaged using the same<br>materials and processes. |
| Performance Data (Nonclinical<br>and/or Clinical): | Non-Clinical Performance and Conclusions:<br>Performance testing completed as part of the design<br>assurance process demonstrated that this device is<br>safe and effective and substantially equivalent to the<br>predicate device.<br>Clinical Performance and Conclusions:<br>Clinical data and conclusions were not needed for<br>this device. |
{1}------------------------------------------------
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo is circular in shape, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle, with its wings spread and its head turned to the right.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 3 2004
Mr. Brandon Hipsher Specialist, Corporate Regulatory Affairs Zimmer, Inc. P.O. Box 708 Warsaw, Indiana 46581-0708
Re: K042271
K042271
Trade/Device Name: NexGen® Complete Knee Solution LPS-Flex Prolong™ Highly Crosslinked Polyethylene Articular Surfaces Regulation Number: 21 CFR 888.3560 and 21 CFR 888.3565 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semiconstrained cemented prosthesis and Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis Regulatory Class: II Product Code: JWH and MBH
Dated: September 10, 2004 Received: September 13, 2004
### Dear Mr. Hipsher
We have reviewed your Section 510(k) premarket notification of intent to market the device we have reviewed your bector. 910(s) pe device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 101 use stated in the encreate) to ttg_as) actment date of the Medical Device Amendments, or to commence prices that have been reclassified in accordance with the provisions of the Federal Food, Drug, devices that have been recuire approval of a premarket approval application (PMA). and Cosmetic 710-1710-11-28, subject to the general controls provisions of the Act. The Tourmay, therefore, thanket the act include requirements for annual registration, listing of general controlly provisions practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (800 aborto) als. Existing major regulations affecting your device can may be subject to suen additional begulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I loase oe actived that I be mination that your device complies with other requirements of the Act that I Dri has made a assocd regulations administered by other Federal agencies. You must
{3}------------------------------------------------
## Page 2 - Mr. Brandon Hipsher
comply with all the Act's requirements, including, but not limited to: registration and listing (21 comply with an the rice s requirements, 01); good manufacturing practice requirements as set CITY art 8077, adoling (21 CFR Part 820); and if applicable, the electronic forul in the qualis) byevelsions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) This iciter will anow you to ogen mailing of substantial equivalence of your device to a legally premits to the nevice results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acvice 10. Jour acon 2011) 276-0120. Also, please note the regulation entitled, contact the Office of Computer in (Et notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark A. Millican
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
# 510(k) Number (if known): Ko4227 |
#### Device Name:
NexGen® Complete Knee Solution LPS-Flex Prolong™ Highly Crosslinked Polyethylene Articular Surfaces
#### Indications for Use:
This device is indicated for patients with severe knee pain and disability due to:
- Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. -
- Collagen disorders and/or avascular necrosis of the femoral condyle. ।
- Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy
- Moderate valgus, varus or flexion deformities. -
- The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery.
The device is indicated for use when both cruciate ligaments have been excised. The device is intended for use as part of a cemented or uncemented knee prosthesis.
Prescription Use _X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(Please do not write below this line - Continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
R. Mark A. Millauer
Sion-(){}) Division of General, Restorative, and Neurological Devices
K04227
**510(k) Number**_
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.