SMITH & NEPHEW MODULAR HIP

K042127 · Smith & Nephew, Inc. · MEH · Nov 19, 2004 · Orthopedic

Device Facts

Record IDK042127
Device NameSMITH & NEPHEW MODULAR HIP
ApplicantSmith & Nephew, Inc.
Product CodeMEH · Orthopedic
Decision DateNov 19, 2004
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 888.3353
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Total hip components are indicated for uncemented use in individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; old, remote osteomyelitis with an extended drainage-free period; nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity. Smith & Nephew Modular Hip components are for single use only.

Device Story

Smith & Nephew Modular Hip is a prosthetic hip joint system consisting of primary and revision stems used with porous and HA-coated modular sleeves. The system features a 12/14 taper and is designed for use with existing Smith & Nephew femoral heads and acetabular components. The device is intended for surgical implantation by a physician to replace damaged hip joints. It functions as a mechanical support structure to restore hip function in patients with degenerative or traumatic joint conditions. The modular design allows for customization during surgery to address specific patient anatomy and clinical needs. The device is for single use only.

Clinical Evidence

Bench testing only. Mechanical test data indicated the device is equivalent to devices currently used clinically and capable of withstanding expected in vivo loading.

Technological Characteristics

Modular hip stems and sleeves manufactured from Ti-6Al-4V alloy conforming to ASTM F 1472 or ASTM F 136. Features 12/14 taper. Designed for uncemented use. Compatible with existing Smith & Nephew femoral heads and acetabular components.

Indications for Use

Indicated for individuals undergoing primary and revision hip surgery due to trauma, inflammatory joint disease (e.g., rheumatoid arthritis), or noninflammatory degenerative joint disease (e.g., osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, pelvic fracture, diastrophic variant, remote osteomyelitis, nonunion, femoral neck/trochanteric fractures, femoral osteotomy, Girdlestone resection, fracture dislocation, or deformity).

Regulatory Classification

Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K042/27 (fg lof 1) NOV 1 9 2004 Smith & Nephew, Inc. Summary of Safety and Effectiveness Smith & Nephew Modular Hip Date of Summary: November 2, 2004 ## Contact Person and Address Janet Johnson Akil Director, Regulatory Affairs Smith & Nephew, Inc. Orthopaedic Division 1450 Brooks Road Memphis, TN 38116 (901) 399-5153 Name of Device: Smith & Nephew Modular Hip Common Name: Prosthetic hip joint ## Device Description The Smith & Nephew Modular Hip consists of primary and revision stems used with porous and HA coated modular sleeves. The stems feature a 12/14 taper and are used with existing Smith & Nephew femoral heads and acetabular components. Modular hip stems and sleeves are manufactured from Ti-6Al-4V conforming to ASTM F 1472 or ASTM F 136. ## Device Classification 21 CFR 888.3353 Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis - Class II ## Mechanical and Clinical Data A review of the mechanical test data indicated that the Smith & Nephew Modular Hip is equivalent to devices currently used clinically and is capable of withstanding expected in vivo loading without failure. #### Indications for Use Total hip components are indicated for uncemented use in individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; old, remote osteomyelitis with an extended drainage-free period; nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity. Smith & Nephew Modular Hip components are for single use only. #### Substantial Equivalence Information The substantial equivalence of the Smith & Nephew Modular Hip is based on its similarities in indications for use, design features, operational principles, and material composition to the following predicate devices – Joint Medical Products S-ROM Total Hip System (K921111, K934412, K954935, and K961939) and the Smith & Nephew Synergy Hip System (K963509, K991485, and K0029961. {1}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 NOV 1 9 2004 Mr. Jason Sells Regulatory Affairs Specialist Orthopaedics Smith & Nephew, Inc. 1450 Brooks Road Memphis. Tennessee 38116 Re: K042127 Trade Name: Smith & Nephew Modular Hip Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: II Product Code: MEH Dated: August 5, 2004 Received: August 6, 2004 Dear Mr. Sells: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ ## Page 2 - Mr. Jason Sells This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours. R. Merle A. Millhaus Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Indications for Use 16042127 # 510(k) Number (if known): Device Name: Smith & Nephew Modular Hip ### Indications for Use: Total hip components are indicated for uncemented use in individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJ) or any of its composite diagnoses such as osteoarthritis: avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; old, remote osteomyelitis with an extended drainage-free period; nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity. Smith & Nephew Modular Hip components are for single use only. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Division of General, Restorative, and Neurological De **510(k) Number** K042127 Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...