K991485 · Smith & Nephew, Inc. · LPH · Jul 12, 1999 · Orthopedic
Device Facts
Record ID
K991485
Device Name
SYNERGY POROUS SIZE 8 HIP STEM
Applicant
Smith & Nephew, Inc.
Product Code
LPH · Orthopedic
Decision Date
Jul 12, 1999
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Total hip components are indicated for uncemented use only in individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJ) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; old, remote osteomyclitis with an extended drainage-free period; nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; fernoral osteotomy, or Girdlestonc resection; fracture dislocation of the hip; and correction of deformity. The Synergy Porous Size 8 Hip Stem is designed for single use only.
Device Story
Synergy Porous Size 8 Hip Stem is a metallic femoral component for total hip arthroplasty. Device is implanted by orthopedic surgeons in clinical settings to replace damaged hip joints. It functions as a structural prosthesis, designed for uncemented fixation. It is compatible with modular cobalt chrome or ceramic femoral heads via a 12/14 taper. The device provides mechanical stability and restores joint function for patients with degenerative or traumatic hip conditions. It is intended for single use only.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and design similarities to legally marketed predicate devices.
Technological Characteristics
Material: Titanium alloy (Ti-6Al-4V, ASTM F1472). Design: Porous tapered hip stem. Compatibility: Modular cobalt chrome or ceramic femoral heads with 12/14 taper. Fixation: Uncemented. Single-use only.
Indications for Use
Indicated for individuals undergoing primary and revision hip surgery for trauma, inflammatory joint disease (e.g., rheumatoid arthritis), noninflammatory degenerative joint disease (e.g., osteoarthritis), avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, pelvic fracture, diastrophic variant, remote osteomyelitis, nonunion, femoral neck/trochanteric fractures, femoral osteotomy, Girdlestone resection, fracture dislocation, or deformity correction.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Predicate Devices
Global Taper "Tapered" Hip Stem (Smith & Nephew)
Revision (Echelon) Hip Stem (Smith & Nephew)
Versys® Hip Stem (Zimmer)
Perfecta IMCTM Hip Stem (Wright Medical)
AML® Hip Stem (DePuy)
Submission Summary (Full Text)
{0}------------------------------------------------
K991485
## 510(k) Summary
Submitter's name: Submitter's address: Submitter's telephone number: Contact person: Date summary prepared: Trade or proprietary device name: Common or usual name:
Classification name:
Smith & Nephew, Inc. 1450 Brooks Road, Memphis, TN 38116 901-399-6487 David Henley April 23, 1999 Synergy Porous Size 8 Hip Stem Syncrgy Forous ont - Porous Tapered Hip Stem
21 CFR 888.3358 hip joint metal/polymer/metal semi-21 CFR 866.5578 mp John mean posted uncemented prosthesis-Class II 87LPH
## Substantially Equivalent Legally Marketed Devices
- Global Taper "Tapered" Hip Stem Smith & Nephew .
- Revision (Echelon) Hip Stem Smith & Nephew . �
- Versys® Hip Stem Zimmer .
- Perfecta IMCTM Hip Stem Wright Medical �
- AML® Hip Stem DePuy ●
#### Device Description
Device Description
The Synergy Porous Size 8 Hip Stem is manufactured from titanium material (Ti-6AI-4V, ASTM F1472) The Synergy Porous Size o Hip Stein is nambidered from in them.
and is designed for use with existing cobalt chrome or ceramic modular femoral heads with a 12/14 taper.
## Device Intended Use
Device Intended Use I of a mp components are indicated ancements associations hips damaged as a result of trauma,
surgery where other treatments or devices have failed in rehabilitation inint di surgery where other treatments of devices have in rendemannatory degeneration in disease inflammatory Joint disease such as medification in the recosis, traumatic arthoritos (NIDD) of any of Its composite uraginses such as estocal director of tremote of the province of the provimal silpped capital epiphysis, fused inf, fracture of the period tractures of the proximal
an extended drainage-free period; nonunion, femoral neck fractures fractores of the pro all extended dramage-free period, nonanageable using other techniques; femoral osteotomy, or femal with nead in Forrement dislocation of the hip; and correction of deformity.
The Synergy Porous Size 8 Hip Stem is designed for single use only.
## Technological characteristics:
The Synergy Porous Size 8 Hip Stem is similar to the legally marketed devices listed above. All of these I he Synergy I Grous Gree of Thp Seem is Smith, as with of the Synergy Porous Size 8 Hip Stem and have the same technological characteristics.
#### Performance characteristics:
x croomance that the Synergy Porous Size 8 Hip Stem is substantially equivalent to legally marketed devices.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a sans-serif font. The text is arranged to follow the curve of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 1 2 1999
Mr. David Henley Clinical/Regulatory Affairs Specialist Smith & Nephew, Inc. 1450 Brooks Rd. Memphis, Tennessee 38116
Re: K991485
Trade Name: Synergy porous Size 8 Hip Stem Regulatory Class: II Product Code: LPH and LZO Dated: April 27, 1999 Received: April 28, 1999
Dear Mr. Henley:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commercc prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
Page 2 -- Mr. David Henley
This letter will allow you to begin marketing your device as described in your 510(k) promated I his letter will anow you to begin marketing your device to a legally marketed nothication. The PDA inding of succemar vqur device and thus, permits your device to procced to the market.
If you desire specific advice for your device on our labeling regulation (21 CFF Part 801 and II you desire specific advice for your dones), please contact the Office of Compliance at additionally 609.10 for in vitty diagnostions on the promotion and advertising of your device, (301) 594-4037. Addinonally, for questions on assisted on associated the regulation prease contact the Office or Comphanee at (301) of the other (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Information on your responsionnes ander the rice may of -2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications Statement Synergy Porous Size 8 Hip Stem
Total hip components are indicated for uncemented use only in individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJ) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; old, remote osteomyclitis with an extended drainage-free period; nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; fernoral osteotomy, or Girdlestonc resection; fracture dislocation of the hip; and correction of deformity.
signature
Orvision Sid vision of General Re 510(k) Number
Prescription Use
(Per 21 CFR 801.109)
Syncrgy Porous Sz. 8 Hip Stem CONFIDENTIAL
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.