K041660 · Trimedyne, Inc. · GEX · Sep 8, 2004 · General, Plastic Surgery
Device Facts
Record ID
K041660
Device Name
TRIMEDYNE HOLMIUM LASER SYSTEMS
Applicant
Trimedyne, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 8, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Trimedyne Holmium:YAG Laser System is indicated for superficial incision. excision, resection, ablation, coagulation, hemostasis, and vaporization, with or without an endoscope, in multispecialty applications. When used in conjunction with the Trimedyne Side Firing Needle with Vent Sheath family of optical fiber delivery devices, these laser systems may be used for interstitial incision, excision, resection, ablation, coagulation, hemostasis, and vaporization in multispecialty applications.
Device Story
Trimedyne Holmium Laser System is a Class IV, pulsed, solid-state Holmium:YAG laser; delivers infrared energy at 2.1 µm wavelength with 350 µs pulsewidth. System includes models 1210, 1210-VHP, 1230-30, and 1500-A. Operated by clinicians in clinical/surgical settings; uses menu-driven controls to adjust pulse repetition rate, output energy, and duration. Laser energy delivered via optical fibers (including Side Firing Needle with Vent Sheath for interstitial use) to target tissues for surgical procedures. Output allows for precise tissue interaction (incision, ablation, hemostasis). Benefits include minimally invasive surgical access and effective tissue management across multiple specialties.
Clinical Evidence
No clinical data or non-clinical testing data were submitted in this Premarket Notification.
Technological Characteristics
Holmium:YAG laser; 2.1 µm wavelength; 350 µs pulsewidth; maximum output 100 watts. Class IV laser-powered instrument. Menu-driven control interface for pulse repetition, energy, and duration. Standalone system. Sterilization requirements for delivery fibers apply.
Indications for Use
Indicated for superficial and interstitial incision, excision, resection, ablation, coagulation, hemostasis, and vaporization in dermatology, plastic surgery, gastroenterology, general surgery, urology, gynecology, lithotripsy, orthopedics, ENT, and percutaneous disc decompression. Patient populations include those with soft, mucosal, fatty, cartilaginous, or bony tissue conditions, including tumors, lesions, calculi, BPH, and herniated discs. Contraindications not explicitly listed, though specific clinical criteria (e.g., 12 weeks conservative therapy) apply to disc decompression.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "Trimedyne" in a bold, sans-serif font, with a registered trademark symbol to the right of the word. Above the word "Trimedyne" is the text "k04 1660 1/2" in a handwritten font. Below the word "Trimedyne" is a graphic of three arrows stacked on top of each other.
## SEP - 8 2004
## Trimedyne® Holmium Laser System
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| Submitter Information: | Trimedyne, Inc.<br>15091 Bake Parkway<br>Irvine, CA 92618<br>949-951-3800 |
|------------------------|---------------------------------------------------------------------------|
| Contact Person: | Glenn Yeik<br>President and COO |
| Summary Date: | 20 August 2004 |
#### 11. Device Name
Proprietary: Trimedyne Holmium Laser System, including:
- OmniPulse™ Holmium Laser System (Model 1210) ●
- OmniPulse MAX™ Holmium Laser System (Model 1210-VHP) .
- OmniPulse Jr. TM Holmium Laser System (Model 1230-30) .
- Model 1500-A Holmium Laser System .
Holmium: Yttrium Aluminum Garnet (Holmium: YAG) Laser Common:
Classification: Laser-Powered Instrument
### !!!. Predicate Device
Trimedyne Holmium Laser System
### IV. Device Description
The Trimedyne Holmium Laser System is a medical grade, Class IV, pulsed, solid state Holmium: YAG laser system designed to deliver pulsed infrared laser energy with a wavelength of 2.1 um and 350 microseconds pulsewidth. Menu-driven control options allow the users to select pulse repetition rate, output energy, and lasing duration.
### V. Intended Use
The Trimedyne Holmium:YAG Laser System is indicated for superficial incision. excision, resection, ablation, coagulation, hemostasis, and vaporization, with or without an endoscope, in multispecialty applications.
When used in conjunction with the Trimedyne Side Firing Needle with Vent Sheath family of optical fiber delivery devices, these laser systems may be used for interstitial incision, excision, resection, ablation, coagulation, hemostasis, and vaporization in multispecialty applications.
{1}------------------------------------------------
# 2/2
### Technological Characteristics VI.
The laser system is a Holmium:YAG laser that emits light at a wavelength of 2.1 µm (near infrared) and a pulsewidth of 350 microseconds. The laser has the capability of attaining a maximum output of 100 watts of power.
#### VII. Nonclinical Data
No non-clinical data were submitted in this Premarket Notification.
### VIII. Clinical Data
No clinical tests were submitted in this Premarket Notification.
### IX. Conclusion
The Trimedyne Holmium:YAG Laser System is substantially equivalent to the predicate device described in this Premarket Notification. Therefore, upon clearance of this submission, the Trimedyne Holmium: YAG Laser System will be marketed with the proposed modified/clarified indications.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of a stylized eagle with three lines representing its wings and head.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP = 8 2004
Trimedyne, Inc. c/o Ms. Laura Danielson 510(k) Program Manager TÜV Product Service 1775 Old Highway 8 NW, Suite 104 New Brighton, Minnesota 55112-1891
Re: K041660 Trade/Device Name: Trimedyne Holmium Laser System, including: OmniPulse™ Holmium Laser System (Model 1210) OmniPulse MAX™ Holmium Laser System (Model 1210-VHP) OmniPulse Jr.TM Holmium Laser System (Model 1230-30) Model 1500-A Holmium Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: August 23, 2004 Received: August 24, 2004
Dear Ms. Danielson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r loase be act noce a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or uny I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) rms lotet notification. The FDA finding of substantial equivalence of your device to a legally promated predicated with a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please if you desire of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other Ectorers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark A. Milkman
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
K041660 510(k) Number:
Device Name: Trimedyne Holmium Laser System, including:
- OmniPulse™ Holmium Laser System, Model 1210 .
- OmniPulse MAX™ Holmium Laser System, Model 1210-VHP .
- OmniPulse Jr.™ Holmium Laser System, Model 1230-30 .
- Model 1500-A Holmium Laser System .
## Indications for Use:
Superficial incision, excision, resection, ablation, coagulation, hemostasis, and vaporization, with or without an endoscope, in the following indications:
- Dermatology and Plastic Surgery of soft, mucosal, fatty, and cartilaginous tissues, in l . therapeutic plastic, therapeutic dermatological and aesthetic surgical procedures, including:
- scars .
- tattoo removal ﮩ
- vascular lesions (including port wine stains, ◆ hemangioma, telangiectasia [facial, leg], and rosacea)
- COLIS
- papillomas
- basal cell carcinomas
- lesions of skin and subcutaneous tissue
- plantar warts
- periungual and subungual warts
- debridement of decubitus ulcer
- skin tag vaporization
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
ice of Device Eyaluation (ODE) Concurrence of CDRH, C
Mark A. Milleson
Division of General, Restorative, and Neurological Devices
510/k) Nu
K041660
Page 1 of 4
{5}------------------------------------------------
### Indications - continued
#### 2. Gastroenterological/Gastrointestinal Surgery, including:
- cholecystectomy +
- lysis of adhesions �
- appendectomy +
- biopsy +
- pylorostenotomy +
- benign and malignant lesions ◆
- rectal polyps of sigmoid colon ◆
- gall bladder calculi �
- biliary/bile duct calculi
- benign and malignant neoplasm
- polyps .
- colitis +
- ulcers
- angiodysplasia
- hemorrhoids
- varices
- esophagitis esophageal ulcer
- Mallory-Weiss tear
- gastric ulcer
- duodenal ulcer
- non-bleeding ulcer
- gastric erosions
- colorectal cancer
- gastritis
- bleeding tumors
-
- pancreatitis
- vascular malformations
- telangiectasias
- telangiectasias of the Osler-Weber-Rendu disease
### 3. General Surgery of soft tissue, including:
- skin incision +
- tissue dissection �
- excision of external tumors and lesions +
- complete or partial resection of internal organs +
- tumors and lesions .
- tissue ablation
- mastectomy
- hepatectomy
- pancreatectomy
- splenectomy
- thyroidectomy
- parathyroidectomy
- hemiorrhaphy
- tonsillectomy
- lymphadenectomy
- partial nephrectomy
- pilonidal cystectomy
- resection of lipoma
- pelvic adhesiolysis
- debridement of decubitus ulcer
- hemorrhoids
- pilodidal cyst removal and repair
- debridement of statis ulcer
- biopsy
### 4. Genitourinary Surgery/Urology, including:
- superficial urinary bladder tumors
- invasive bladder carcinoma
- urethral strictures
- lesions of the external genitalia
- bladder
- urethral and ureteral tumors
- condylomas
- urethral and penile hemangioma
- bladder neck obstructions
- holmium laser incision, excision, resection, ablation, hemostasis, vaporization, and enucleation in the treatment of benign prostatic hyperplasia (BPH)
N Millican
Restorative
K041660
{6}------------------------------------------------
## Indications - continued
- 5. Gynecological Surgery during open and endoscopic procedures, including:
- condyloma acuminata
### Lithotripsy and Percutaneous Urinary Lithotripsy, including: 6.
- fragmentation of urinary calculi �
- fragmentation of urethral calculi
- fragmentation of kidney calculi
- treatment of distal impacted fragment of steinstrasse when guide wires cannot be passed
- Orthopedic Surgery in pathological soft and cartilaginous tissue in small and large joints, 7. including:
- knee meniscectomy .
- knee synovectomy ◆
- chondromalacia and tears
- loose body debridement +
- lateral retinacular release .
- debridement of the degenerative knee
- plica removal
- ligament and tendon release
- contouring and sculpting of articular surfaces
- debridement of inflamed synovial tissue .
- capsulectomy in the knee .
- chondroplasty in the knee
- chondromalacia ablation
- 8. Otorhinolaryngology (ENT) Surgery in soft, mucosal, cartilaginous and bony tissue, including:
- endosinus surgery �
- functional endoscopic sinus surgery +
- turbinate procedures (e.g., turbinectorny) .
- dacryocystorhinostomy (DCR)
- ethmoidectomy
- polypectomy
- . maxillary antrostomy
- frontal sinusotomy
- sphenoidotomy
Mark N. Wilkerson
510(k) Number
10(k) Number.
{7}------------------------------------------------
### Indications - continued
- Percutaneous Cervical, Lumbar, and Thoracic Disc Decompression / Discectomy: ರು. Superficial incision, excision, resection, ablation, coagulation, hemostasis and vaporization, with or without an endoscope, in the following:
- Percutaneous Lumbar Disc Decompression/Discectorny in soft, cartilaginous, and bony tissue, including: .
- foraminoplasty -
- Percutaneous Cervical Disc Decompression/Discectorny in soft tissue, in patients with:
- uncomplicated ruptured or herniated discs -
- neck pain with radiation down the arm .
- symptoms and signs of sensory loss, tingling, numbness, muscle weakness, and/or decreased deep tendon reflexes
- MRI, CT, myelogram, or discogram findings of disc herniation consistent with patient signs and symptoms
- positive electromyography and/or nerve conduction studies -
- no improvement after 12 weeks of conservative therapy (i.e., physical therapy, traction, bed rest, exercises, and medication)
- Percutaneous Thoracic Disc Decompression/Discectorny in soft tissue, in patients with:
- uncomplicated ruptured or herniated discs ー
- thoracic and intercostal intractable pain -
- paresthesias at levels appropriate to the herniated discs visualized on MRI and CT-myclography -
- MRI, CT, myelogram, or discogram findings of disc herniation consistent with patient signs and symptoms
- no improvement after 12 weeks of conservative therapy (i.e., physical therapy, traction, bed rest, exercises, and medication)
Interstitial incision, excision, resection, ablation, coagulation, hemostasis, and vaporization in multispecialty applications; interstital applications should only be performed using the Trimedyne Side Firing Needle with Vent Sheath family of optical fiber delivery devices.
Mark N. Mulkerson
al. Fortorative. ological Devices
510(k) Number K041660
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.