Datascope's 7.5Fr. Intra-aortic Balloon Catheters and Accessories are used for intra-aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation during diastole and deflation during systole increases blood supply to the heart muscle and decreases the work of the left ventricle. This is the same intended use as previously cleared for all other Datascope Intra-Aortic Balloon Catheters. The intra-aortic balloon is placed in the descending aorta just below the subclavian artery and is intended to improve cardiovascular functioning during the following situations: Refractory unstable angina . Impending infarction . . Acute MI Refractory ventricular failure . Complications of acute MI (i.e. Acute MR or VSD, or papillary . muscle rupture) Cardiogenic shock . Support for diagnostic, percutaneous revascularization, and . interventional procedures. Ischemia related intractable ventricular arrhythmias . Septic shock . Intraoperative pulsatile flow generation . . Weaning from bypass Cardiac support for non-cardiac surgery . Prophylactic support in preparation for cardiac surgery . Post surgical myocardial dysfunction/low cardiac output syndrome . Myocardial contusion Mechanical bridge to other assist devices . Cardiac support following correction of anatomical defects .
Device Story
Intra-aortic balloon catheter (IAB) for counterpulsation therapy; placed in descending aorta. Device inflates during diastole and deflates during systole to increase myocardial blood supply and reduce left ventricular workload. Used in clinical settings (OR/ICU) by physicians. Modification of existing 8Fr. IAB to 7.5Fr. profile; includes modified reinforced catheter introducer set with color-coded hubs and lock-fit dilator/sheath interface. Provides hemodynamic support for patients with cardiac dysfunction or undergoing cardiac procedures.
Clinical Evidence
No clinical evaluation performed. Substantial equivalence supported by in-vitro bench testing demonstrating comparable functionality and performance to predicate devices.
Technological Characteristics
Co-extruded polyurethane blend catheter; 7.5Fr. diameter; .027" polyimide inner lumen. Includes reinforced catheter introducer set with color-coded hubs (orange for 7Fr.) and lock-fit dilator/sheath interface. Materials are previously FDA-cleared. Mechanical device; no energy source or software.
Indications for Use
Indicated for patients requiring intra-aortic balloon counterpulsation for refractory unstable angina, impending/acute MI, refractory ventricular failure, complications of acute MI (MR, VSD, papillary muscle rupture), cardiogenic shock, septic shock, ischemia-related intractable ventricular arrhythmias, myocardial contusion, or cardiac support during/post-surgery, diagnostic/interventional procedures, or as a mechanical bridge to other assist devices.
Regulatory Classification
Identification
An intra-aortic balloon and control system is a prescription device that consists of an inflatable balloon, which is placed in the aorta to improve cardiovascular functioning during certain life-threatening emergencies, and a control system for regulating the inflation and deflation of the balloon. The control system, which monitors and is synchronized with the electrocardiogram, provides a means for setting the inflation and deflation of the balloon with the cardiac cycle.
Special Controls
*Classification.* (1) Class II (special controls) when the device is indicated for acute coronary syndrome, cardiac and non-cardiac surgery, or complications of heart failure. The special controls for this device are:(i) Appropriate analysis and non-clinical testing must be conducted to validate electromagnetic compatibility and electrical safety of the device;
(ii) Software verification, validation, and hazard analysis must be performed;
(iii) The device must be demonstrated to be biocompatible;
(iv) Sterility and shelf-life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components;
(v) Non-clinical performance evaluation of the device must demonstrate mechanical integrity, durability, and reliability to support its intended purpose; and
(vi) Labeling must include a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(2) Class III (premarket approval) when the device is indicated for septic shock and pulsatile flow generation.
(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before March 31, 2014, for any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976. Any other intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
Datascope's Reinforced Catheter Introducer Set (K002365)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "Datascope" in a bold, serif font. Above the word is a curved line that starts thick on the left and thins out as it curves to the right. The word is the main focus of the image and is centered.
innovation is the best medicine
JUN - 7 2004
K041281
Attachment D
Datascope Corp. Cardiac Assist Division 15 Law Drive Fairfield. NJ 07004 Tel: 973.244.6100 Fax: 973.244.6279
### 510(k) SUMMARY
### FOR
## DATASCOPE'S 7.5Fr. IAB and ACCESSORIES
(Prepared in accordance with 21 CFR Part 807.92)
#### GENERAL INFORMATION A.
| Submitter: | Datascope Corp. |
|-----------------|--------------------------------------|
| | Cardiac Assist Division |
| Address: | 15 Law Drive |
| | Fairfield, NJ 07004 |
| Contact Person: | JoAnn Taylor |
| | Global Regulatory Affairs Specialist |
| Phone: | 973/244-6123 |
| Fax: | 973/244-6243 |
| Date: | May 12, 2004 |
#### DEVICE INFORMATION B.
| Generic Name: | Intra-Aortic Balloon Catheter (IAB) |
|----------------------|------------------------------------------------------------------------------|
| Trade Name: | Datascope's 7.5Fr. IAB and Accessories |
| Classification Name: | Intra-Aortic Balloon Catheters (IABs) are classified<br>under 21CFR 870.3535 |
#### PREDICATE DEVICE INFORMATION C.
Datascope's 7.5Fr. IAB and Accessories are substantially equivalent to the following marketed devices:
| K031569 - Datascope's Blow-Molded 8Fr. IAB, S/E 6/10/03 |
|-----------------------------------------------------------------------|
| K013326 - Datascope's Fidelity 8Fr. IAB, S/E 11/02/01 |
| K002365 - Datascope's Reinforced Catheter Introducer Set, S/E 8/25/00 |
{1}------------------------------------------------
#### DEVICE DESCRIPTION/INTENDED USE D.
Datascope's 7.5Fr. Intra-aortic Balloon Catheters and Accessories are used for intra-aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation during diastole and deflation during systole increases blood supply to the heart muscle and decreases the work of the left ventricle. This is the same intended use as previously cleared for all other Datascope Intra-Aortic Balloon Catheters.
The intra-aortic balloon is placed in the descending aorta just below the subclavian artery and is intended to improve cardiovascular functioning during the following situations:
- Refractory unstable angina .
- Impending infarction .
- . Acute MI
- Refractory ventricular failure .
- Complications of acute MI (i.e. Acute MR or VSD, or papillary . muscle rupture)
- Cardiogenic shock .
- Support for diagnostic, percutaneous revascularization, and . interventional procedures.
- Ischemia related intractable ventricular arrhythmias .
- Septic shock .
- Intraoperative pulsatile flow generation .
- . Weaning from bypass
- Cardiac support for non-cardiac surgery .
- Prophylactic support in preparation for cardiac surgery .
- Post surgical myocardial dysfunction/low cardiac output syndrome .
- Myocardial contusion �
- Mechanical bridge to other assist devices .
- Cardiac support following correction of anatomical defects .
#### TECHNOLOGICAL CHARACTERISTICS E.
Datascope's 7.5Fr. IABs and Accessories are substantially equivalent to the predicate devices with regard to intended use.
The modification to the Datascope predicate 8Fr. IAB is dimensional only. The dimension of the co-extruded polyurethane blend catheter will be changed from 8Fr. to 7.5Fr. and the polyimide inner lumen dimension will be changed from .030" to .027".
To accommodate the smaller 7.5Fr. IAB catheter, Datascope's predicate 8Fr. Reinforced Catheter Introducer has been modified from 8Fr. to 7.5Fr., and the Stepped Dilator from 5-8Fr. to 4-7.5Fr.
{2}------------------------------------------------
Modifications to the 8Fr. Reinforced Catheter Introducer Set also include a change in the material of the sheath hub and internal tubing, color code the introducer hub and dilator to industry standard (orange for 7Fr.), change the hub cap design from round to square, change dilator/sheath interface from a snap-fit to a lock-fit design, and add an alternate source contract manufacturer. The material changes are all previously FDA cleared materials.
These modifications to the 8Fr. IABs and Accessories have been demonstrated not to affect safety or efficacy of the device.
#### NON-CLINICAL TESTS F.
The results of in-vitro tests conducted demonstrate that the functionality and performance characteristics of the device are comparable to the currently marketed devices.
#### CLINICAL TESTS G.
There has been no clinical evaluation of the new device in the U.S.
#### H. CONCLUSIONS
Based on the information presented in this 510(k) premarket notification, Datascope's 7.5Fr. IAB and Accessories are considered substantially equivalent to Datascope's currently marketed IABs and accessories.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized caduceus, a symbol of medicine and health. The caduceus is composed of three wavy lines that represent the staff of Asclepius, the Greek god of healing.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN - 7 2004
Datascope Corp. c/o Ms. JoAnn Taylor Global Regulatory Affairs Specialist 15 Law Drive Fairfield, NJ 07004
Re: K041281
Datascope's 7.5Fr. IAB and Accessories Regulation Number: 21 CFR 870.3535 Regulation Name: Intra-Aortic Balloon and Control System Regulatory Class: Class III (three) Product Code: DSP Dated: May 12, 2004 Received: May 13, 2004
Dear Ms. Taylor:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roviewed your becement of the device is substantially equivalent (for the indications forched a o re and in sure) to legally marketed predicate devices marketed in interstate for ass stated in the May 28, 1976, the enactment date of the Medical Device Amendments, or to commerce prior to May 20, 1978, in excordance with the provisions of the Federal Food, Drug, de nees that have been require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, morelore, manner in to Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can may be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 – Ms. JoAnn Taylor
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I that FDA has made a determination that your device complies with other requirements of the Act that I DT has intates and regulations administered by other Federal agencies. You must or any I cactar stutures and regulaments, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CFR Part 807); adomig (21 CFR Part 820); and if applicable, the electronic forth in the quart) Bybellio (Sections 531-542 of the Act); 21 CFR 1000-1050. product raction control provisions (seting your device as described in your Section 510(k) This letter will anow you to begin maing of substantial equivalence of your device to a legally prematice notification. The President for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to (2017) 594-1057. Thiss, production as about 97) you may obtain. Other general information on your presponsibilities under the Act may be obtained from the Division of Small Manufacturers, Itsponsional and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Dina R. Vachner
GA Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name:Datascope's 7.5Fr. Intra-Aortic Balloon Catheters and Accessories
Indications For Use:
Datascope's 7.5Fr. Intra-Aortic Balloon Catheters and Accessories have the following indications for use:
- Refractory unstable angina .
- Impending infarction .
- � Acute MI
- Refractory ventricular failure .
- Complications of acute MI (i.e. Acute MR or VSD, or papillary muscle rupture) .
- Cardiogenic shock .
- Support for diagnostic, percutaneous revascularization, and interventional procedures. .
- Ischemia related intractable ventricular arrhythmias ●
- Septic shock .
- Intraoperative pulsatile flow generation .
- Weaning from bypass .
- Cardiac support for non-cardiac surgery .
- Prophylactic support in preparation for cardiac surgery .
- Post surgical myocardial dysfunction/low cardiac output syndrome .
- Myocardial contusion .
- Mechanical bridge to other assist devices .
- Cardiac support following correction of anatomical defects .
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(Please Do NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Duna R. Lochner
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number_k04128
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.