DATASCOPE CA 40 8FR. IABS, MODEL 0684-00-04XX

K031569 · Datascope Corp. · DSP · Jun 10, 2003 · Cardiovascular

Device Facts

Record IDK031569
Device NameDATASCOPE CA 40 8FR. IABS, MODEL 0684-00-04XX
ApplicantDatascope Corp.
Product CodeDSP · Cardiovascular
Decision DateJun 10, 2003
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.3535
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Datascope's CA 40 8Fr. Intra-Aortic Balloon Catheters have the following indications for use: Refractory unstable angina Impending infarction Acute MI Refractory ventricular failure Complications of acute MI (i.e. Acute MR or VSD, or papillary muscle rupture) Cardiogenic shock Support for diagnostic, percutaneous revascularization, and interventional procedures. Ischemia related intractable ventricular arrhythmias Septic shock Intraoperative pulsatile flow generation Weaning from bypass Cardiac support for non-cardiac surgery Prophylactic support in preparation for cardiac surgery Post surgical myocardial dysfunction/low cardiac output syndrome Myocardial contusion Mechanical bridge to other assist devices Cardiac support following correction of anatomical defects

Device Story

Intra-aortic balloon catheter (IAB) placed in descending aorta below subclavian artery; provides mechanical circulatory support. Device consists of balloon membrane and tapered catheter tip. Used in clinical settings (OR, ICU) by trained medical professionals. Functions by inflating/deflating to improve cardiovascular performance; provides pulsatile flow, cardiac support, and hemodynamic stabilization. Benefits include improved perfusion and reduced myocardial workload in acute cardiac failure or surgical contexts.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing only.

Technological Characteristics

8Fr. intra-aortic balloon catheter. Features modified balloon membrane material and tapered catheter tip. Tip material is previously FDA-cleared. Mechanical device; no software or energy source integral to the catheter itself.

Indications for Use

Indicated for patients requiring hemodynamic support due to refractory unstable angina, impending/acute MI, cardiogenic/septic shock, ventricular failure, post-surgical myocardial dysfunction, or complications of acute MI (MR, VSD, rupture). Also indicated for prophylactic support during cardiac/non-cardiac surgery, weaning from bypass, and as a mechanical bridge to other assist devices.

Regulatory Classification

Identification

An intra-aortic balloon and control system is a prescription device that consists of an inflatable balloon, which is placed in the aorta to improve cardiovascular functioning during certain life-threatening emergencies, and a control system for regulating the inflation and deflation of the balloon. The control system, which monitors and is synchronized with the electrocardiogram, provides a means for setting the inflation and deflation of the balloon with the cardiac cycle.

Special Controls

*Classification.* (1) Class II (special controls) when the device is indicated for acute coronary syndrome, cardiac and non-cardiac surgery, or complications of heart failure. The special controls for this device are:(i) Appropriate analysis and non-clinical testing must be conducted to validate electromagnetic compatibility and electrical safety of the device; (ii) Software verification, validation, and hazard analysis must be performed; (iii) The device must be demonstrated to be biocompatible; (iv) Sterility and shelf-life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components; (v) Non-clinical performance evaluation of the device must demonstrate mechanical integrity, durability, and reliability to support its intended purpose; and (vi) Labeling must include a detailed summary of the device- and procedure-related complications pertinent to use of the device. (2) Class III (premarket approval) when the device is indicated for septic shock and pulsatile flow generation. (c) *Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before March 31, 2014, for any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976. Any other intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K031569 ### Attachment D ### SUMMARY OF SAFETY AND EFFECTIVENESS FOR DATASCOPE'S CA 40 8Fr. Intra-Aortic Balloon Catheters Prepared in accordance with 21 CFR Part 807.92) #### GENERAL INFORMATION A. | Submitter: | Datascope Corp.<br>Cardiac Assist Division | |-----------------|------------------------------------------------------| | Address | 15 Law Drive<br>Fairfield, NJ 07004 | | Contact Person: | JoAnn Taylor<br>Global Regulatory Affairs Specialist | #### B. DEVICE INFORMATION | Generic Name: | Intra-Aortic Balloon Catheter (IAB) | |----------------------|------------------------------------------------------------------------------| | Trade Name: | Datascope's 8 Fr. Intra-Aortic Balloon Catheter (IAB) | | Classification Name: | Intra-Aortic Balloon Catheters (IAB) are classified under 21 CFR<br>870.3535 | #### C. PREDICATE DEVICE INFORMATION Datascope's CA 40 Intra-Aortic Balloon Catheters (IAB) are substantially equivalent to the following marketed devices: - K013326, Datascope Fidelity 8Fr. Alt B Intra-Aortic Balloon Catheters (S/E . 11/02/01). - K003598, Datascope Profile 8Fr. Intra-Aortic Balloon Catheters w/Alt B and Gas . Lumen Insert (S/E 12/21/00) {1}------------------------------------------------ #### DEVICE DESCRIPTION/INTENDED USE D. The intra-aortic balloon is placed in the descending aorta just below the subclavian artery and is intended to improve cardiovascular functioning during the following situations: - Refractory unstable angina . - Impending infarction . - Acute MI . - Refractory ventricular failure ● - Complications of acute MI (i.e. Acute MR or VSD, or papillary muscle rupture) . - Cardiogenic shock ● - Support for diagnostic, percutaneous revascularization, and interventional . procedures. - Ischemia related intractable ventricular arrhythmias ● - Septic shock . - Intraoperative pulsatile flow generation ● - Weaning from bypass ● - Cardiac support for non-cardiac surgery ● - Prophylactic support in preparation for cardiac surgery . - Post surgical myocardial dysfunction/low cardiac output syndrome ● - Myocardial contusion ● - Mechanical bridge to other assist devices ● - Cardiac support following correction of anatomical defects . #### TECHNOLOGICAL CHARACTERISTICS E. Datascope's CA 40 8Fr. IAB Catheters are substantially equivalent to the predicate devices with regard to intended use. Modifications to Datascope's predicate 8 Fr. IAB catheters include a material modification of Datascope's currently marketed balloon membrane to enable the balloon to be manufactured using different molding process. The catheter tip material is that of a previously FDA cleared IAB tip material and the catheter's tip dimension has been tapered. These modifications to the 8Fr. IAB catheters have been demonstrated not to affect safety or efficacy of the device. #### F. NON-CLINICAL TESTS The results of in-vitro tests conducted demonstrate that the functionality and performance characteristics of the device are comparable to the currently marketed devices. {2}------------------------------------------------ #### G. CLINICAL TESTS There have been no clinical evaluations of the new device. #### H. CONCLUSIONS Based on the information presented in this 510(k) premarket notification, Datascope's CA 40 8 Fr. IAB Catheters are considered substantially equivalent to Datascope's currently marketed IABs. {3}------------------------------------------------ Public Health Service JUN 1 0 2003 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Datascope Corporation Cardiac Assist Division c/o Ms. JoAnn Taylor Global Regulatory Affairs Specialist 15 Law Drive Fairfield. NJ 07004 Re: K031569 8Fr. Intra-Aortic Balloon Catheter Regulation Number: 21 CFR 870.3535 Regulation Name: Intra-Aortic Balloon Catheter Regulatory Class: Class III (three) Product Code: DSP Dated: May 19, 2003 Received: May 20, 2003 Dear Ms. Taylor: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Ms. JoAnn Taylor Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmaldsmamain.html Sincerely yours, Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Attachment B ## Indications for Use Statement | 510(k) Number | K031569 | |------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name | Datascope's CA 40 8Fr. Intra-Aortic Balloon Catheters | | Indications<br>for Use | Datascope's CA 40 8Fr. Intra-Aortic Balloon Catheters have the following<br>indications for use:<br><ul><li>Refractory unstable angina</li><li>Impending infarction</li><li>Acute MI</li><li>Refractory ventricular failure</li><li>Complications of acute MI (i.e. Acute MR or VSD, or papillary muscle rupture)</li><li>Cardiogenic shock</li><li>Support for diagnostic, percutaneous revascularization, and interventional procedures.</li><li>Ischemia related intractable ventricular arrhythmias</li><li>Septic shock</li><li>Intraoperative pulsatile flow generation</li><li>Weaning from bypass</li><li>Cardiac support for non-cardiac surgery</li><li>Prophylactic support in preparation for cardiac surgery</li><li>Post surgical myocardial dysfunction/low cardiac output syndrome</li><li>Myocardial contusion</li><li>Mechanical bridge to other assist devices</li><li>Cardiac support following correction of anatomical defects</li></ul> | PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CCRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Cardiovascular Devices | Prescription Use<br>(Per 21 CFR 801.109) | X | |------------------------------------------|-------------------| | Over-The-Counter Use | _________________ | | 510(k) Number | K031569 |
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...