K034066 · B.Braun Medical, Inc. · KPO · Mar 3, 2004 · Gastroenterology, Urology
Device Facts
Record ID
K034066
Device Name
B.BRAUN PREMIXED DIALYSATE
Applicant
B.Braun Medical, Inc.
Product Code
KPO · Gastroenterology, Urology
Decision Date
Mar 3, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5820
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
B. Braun Premixed Dialysate is indicated for use with renal replacement therapy systems that utilize sterile premixed dialysate.
Device Story
B. Braun Premixed Dialysate consists of sterile, non-pyrogenic electrolyte solutions provided in single-use flexible PVC bags (1000 mL to 5000 mL). These solutions are used with renal replacement therapy systems that require sterile premixed dialysate. Physicians prescribe specific electrolyte compositions from the available family of solutions to meet individual patient needs. The device functions as the dialysate fluid in renal replacement procedures, facilitating the removal of waste products and fluid balance management. It is intended for clinical use by healthcare professionals in settings where renal replacement therapy is performed.
Clinical Evidence
No clinical data provided; substantial equivalence is based on chemical composition, manufacturing, and sterilization process comparisons to legally marketed predicate devices.
Technological Characteristics
Sterile, non-pyrogenic electrolyte solutions in flexible PVC bags (1000-5000 mL). Designed for use with renal replacement therapy systems. Sterilization process is consistent with predicate devices.
Indications for Use
Indicated for use with renal replacement therapy systems requiring sterile premixed dialysate. Patient population includes individuals requiring renal replacement therapy; specific age or gender not restricted.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
Baxter Premixed Dialysate for Hemodiafiltration (K910270)
Submission Summary (Full Text)
{0}------------------------------------------------
KO34066
## mar - 3 2004
## 510(k) Summary 8.0
| SUBMITTER: | B. Braun Medical Inc.<br>901 Marcon Boulevard<br>Allentown, PA 18109-9341<br>(610) 266-0500, ext. 2280<br>Contact: Patricia D. Wilson, Regulatory Affairs Specialist |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| DEVICE NAME: | B. Braun Premixed Dialysate |
| COMMON OR USUAL<br>NAME: | Premixed Dialysate |
| DEVICE<br>CLASSIFICATION: | Class II, 21 CFR § 876.5820, Hemodialysis System and<br>Accessories (Product Code KPO) |
| PREDICATE DEVICE: | NxStage Medical, Inc. - NxStage Premixed Dialysate<br>(K022913)<br>Gambro Renal Products – Gambro PrismaSate Dialysis<br>Solutions for Continuous Renal Replacement Therapy<br>(K013448)<br>Baxter Healthcare Corporation -Baxter Premixed Dialysate for<br>Hemodiafiltration (K910270) |
| DESCRIPTION: | The B. Braun Premixed Dialysate Solutions are a family of<br>premixed dialysate solutions which are sterile, non-pyrogenic<br>solutions to be provided in single use flexible PVC bags<br>varying in sizes from 1000 mL to 5000 mL. The premixed<br>dialysate solutions are intended for use with renal replacement<br>therapy systems that utilize sterile premixed dialysate. A<br>family of dialysate solutions will allow the physician to<br>prescribe different electrolyte compositions that meet the<br>specific needs of individual patients. |
| INTENDED USE: | B. Braun Premixed Dialysate is indicated for use with renal<br>replacement therapy systems that utilize sterile premixed<br>dialysate. |
| SUBSTANTIAL<br>EQUIVALENCE: | B. Braun believes that, within the meaning of the Medical<br>Device Amendments of 1976, the B. Braun Premixed<br>Dialysate addressed in this 510(k) premarket notification<br>is substantially equivalent to the following medical devices<br>in commercial distribution: |
{1}------------------------------------------------
K034066 page 2 of 2
## SUBSTANTIAL EQUIVALENCE (continued):
- NxStage Premixed Dialysate (K022913, cleared 10/21/02) .
- PrismaSate Dialysis Solutions (K013448, cleared 01/15/02) .
- Baxter Premixed Dialysate for Hemodiafiltration . (K910270, cleared 04/18/91)
The B. Braun Premixed Dialysate includes the same chemical composition range, and has the same manufacturing, packaging, and sterilization process as the NxStage Premixed Dialysate (K022913). The B. Braun Premixed Dialysate is also similar to the Gambro PrismaSate Solutions (K013448), and Baxter Premixed Dialysate (K910270), with regard to composition, sterility, and packaging.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with three stripes extending from its back.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR - 3 2004
Ms. Patricia Wilson Regulatory Affairs Specialist B. Braun Medical, Inc. 901 Marcon Boulevard ALLENTOWN PA 18109
Re: K034066
Trade/Device Name: B. Braun Premixed Dialysate Regulation Number: 21 CFR §876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Codc: 78 KPO Dated: Deccmber 30, 2003 Received: December 31, 2003
Dear Ms. Wilson:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roviewed your betermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreased 976, the enactment date of the Medical Device Amendments, or to conniner of the 11th 11th 20, 11:10 accordance with the provisions of the Federal Food, Drug, de neces that have been require approval of a premarket approval application (PMA). and Cosmetion (110) and the device, subject to the general controls provisions of the Act. The r ou may, diereler, mains of the Act include requirements for annual registration, listing of general controls provisions of nactice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it your device is elamititional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean i Rase be actived that i Dr Princem that your device complies with other requirements of the Act that 1177 Hass Interes and regulations administered by other Federal agencies. You must of any I edular Battatoo and one of including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set (21 CFR Part 801), as only (QS) regulation (21 CFR Part 820); and if applicable, the electronic rently in and ation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
l'age 2
This letter will allow you to begin marketing your device as described in your Section 510(k)
the first and the first of the said the similars of your device to a locally This letter will allow you to begill marketing your avones of your device of your device to a legally premarket notification. The PDA miding of basification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
r the regulation (2010-11-24, 11-4 in increases on the regulation It you desire specific advice tor your device on on the following numbers, based on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, for questions on the promotion on the regulation entitled, "Misoranding
Office of Compliance at (301) 594-4639. Also, possessed on be includes and and Office of Compliance at (501) 594-1057. This, production of Super the Division of Sma by reference to premarket nouncation (27 or the Act may be obtained from the Division of Small information on your responsionities under the researce at its toll-fried mumber (800) 638-2041 or
Manufacturers, International and Consumer Assistance at its toll-fremation h Manufacturers, International and Collsamler I assistement of the Courcedridsmaldsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use Statement 2.0
Page _
| 510(k) Number (if known): | K034066 |
|---------------------------|---------|
|---------------------------|---------|
Device Name:
B. Braun Premixed Dialysate
Indications For Use:
B. Braun Premixed Dialysate is indicated for use with renal replacement therapy systems that utilize sterile premixed dialysate.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use **__**
Nancy C. Brogdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number. K034066
000005
510(k) Premarket Notification B. Braun Premixed Dialysate
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.