K022913 · Nxstage Medical, Inc. · KPO · Oct 21, 2002 · Gastroenterology, Urology
Device Facts
Record ID
K022913
Device Name
NXSTAGE PREMIXED DIALYSATE
Applicant
Nxstage Medical, Inc.
Product Code
KPO · Gastroenterology, Urology
Decision Date
Oct 21, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5820
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
NxStage Premixed Dialysate is indicated for use with renal replacement therapy systems that utilize sterile premixed dialysate.
Device Story
NxStage Premixed Dialysate consists of sterile, non-pyrogenic solutions provided in single-use flexible PVC bags (1000-5000ml). Solutions are used with renal replacement therapy systems requiring sterile premixed dialysate. Physicians prescribe specific electrolyte compositions from the available family of solutions to meet individual patient needs. The device functions as a fluid medium for solute clearance during renal replacement therapy. It is intended for clinical use by healthcare professionals in settings where renal replacement therapy is performed.
Clinical Evidence
Bench testing only. Validation was performed on a batch produced under controlled manufacturing conditions representative of commercial production. Assays confirmed the manufacturer's capability to produce solutions meeting all end-product specifications.
Technological Characteristics
Sterile, non-pyrogenic dialysate solutions. Packaging: single-use flexible PVC bags (1000-5000ml). Chemical composition varies by formulation to meet specific electrolyte requirements. Manufactured by B. Braun Schiwa, Germany.
Indications for Use
Indicated for use with renal replacement therapy systems utilizing sterile premixed dialysate in patients requiring renal replacement therapy.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
{0}------------------------------------------------
K022913
Page 1 of 3
# NxStage Medical, Inc. Premixed Dialysate 510(k) Premarket Notification
OCT 21 2002
# Section VI: Summary of Safety and Effectiveness
This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of SMDA 1990.
#### A. Submitter's Information:
| Name: | NxStage Medical, Inc. |
|----------------------|---------------------------------------------------------|
| Address: | 439 South Union Street, Suite 501<br>Lawrence, MA 01843 |
| Phone: | (978) 687-4700 |
| Fax: | (978) 687-4800 |
| Contact Person: | Norma LeMay<br>Senior Regulatory Affairs Specialist |
| Date of Preparation: | August 23, 2002 |
#### B. Device Name:
| Trade Name: | NxStage Premixed Dialysate |
|----------------------|-----------------------------------------------------------------------|
| Common/Usual Name: | Premixed Dialysate |
| Classification Name: | Hemodialysis system and accessories<br>(21CFR 876.5820) |
| Product Code: | 78 KP0 - Dialysate Concentrate for Hemodialysis<br>(Liquid or Powder) |
### C. Predicate Device Name:
The predicate devices for the NxStage Premixed Dialysate are:
- NxStage Dialysate Concentrate (K013655, 02/04/02) .
- Baxter Premixed Dialysate for Hemodiafiltration (K910270, 4/18/1991) .
- . PrismaSate Dialysis Solutions (K013448, 01/15/02)
#### D. Device Description/Indications for Use:
The NxStage Premixed Dialysate Solutions are a family of premixed dialysate solutions which are sterile, non-pvrogenic solutions to be provided in single use flexible PVC bags varying in sizes from 1000 to 5000ml. The premixed dialysate solutions are intended for use with renal replacement therapy systems that utilize sterile premixed dialysate. A family of dialysate solutions will allow the physician to prescribe different electrolyte compositions that meet the specific needs of individual patients.
The NxStage Premixed Dialysate Solutions are manufactured by B. Braun Schiwa located in Glandorf, Germany.
{1}------------------------------------------------
K022913
Page 2 of 3
## NxStage Medical, Inc. Premixed Dialysate 510(k) Premarket Notification
# Section VI: Summary of Safety and Effectiveness
#### Indications for Use
NxStage Premixed Dialysate is indicated for use with renal replacement therapy systems that utilize sterile premixed dialysate.
#### E. Substantial Equivalence:
510(k) Substantial Equivalence Decision Making Process
#### 1. Does the new device have the same Indication Statement?
YES - The indication for use for the NxStage Premixed Dialysate Solutions is equivalent to that of the predicate devices.
#### Proposed NxStage Premixed Dialysate
NxStage Premixed Dialysate is indicated for use with renal replacement therapy systems that utilize sterile premixed dialysate.
#### NxStage Dialysate Concentrate (K013655)
NxStage Dialysate Concentrate, after dilution, is indicated for use with renal replacement therapy systems that utilize sterile premixed dialysate.
#### Baxter Premixed Dialysate for Hemodiafiltration (K910270)
Baxter Premixed Dialysate is indicated for acute dialysis modalities such as continuous arteriovenous hemodiafiltration (CAVHD), and continuous venousvenous hemodiafiltration (CVVHD), when treating acute renal failure patients with hypervolemia and uremia that requires high solute clearance.
#### PrismaSate Dialysis Solution (K013448)
Gambro PrismaSate Solutions are indicated for use as a dialysate in continuous renal replacement therapy.
#### 2. Does the new device have the same technological characteristics (e.g., design, materials, etc.)
Yes - The chemical composition ranges, for the common constituents, of the NxStage Premixed Dialysate are identical to that of the PrismaSate Dialysis Solutions, K013448 and therefore raise no new types of safety or effectiveness issues. As described in this submission, there are no major differences in the technology characteristics as compared to the predicate devices.
#### 3. Are the descriptive characteristics precise enough to ensure equivalence?
No - The chemical composition ranges, for the common constituents, of the NxStage Premixed Dialysate are identical to that of the PrismaSate Dialysis Solutions, K013448 and similar to the Baxter Premixed Dialysate, K910270 and the NxStage Concentrate, K013655. However, variations in product formulations and sterilization methods may impact final characteristics of the product.
{2}------------------------------------------------
## NxStage Medical, Inc. Premixed Dialysate 510(k) Premarket Notification
# Section VI: Summary of Safety and Effectiveness
#### 4. Are performance data available to assess equivalence?
YES - NxStage Medical, Inc. has provided data for a validation batch of the NxStage Premixed Dialysate Solutions produced under controlled manufacturing conditions representative of those to be used to produce the dialysate for commercial distribution. Results of the batch validation assays have demonstrated that the dialysate manufacturer is capable of producing a solution meeting all endproduct specifications.
#### 5. Does performance data demonstrate equivalence?
YES - NxStage Medical, Inc. believes that the information and data provided in this submission clearly describes the NxStage Premixed Dialysate Solutions and demonstrates that the solutions are substantially equivalent to the predicate devices.
#### F. Safety Summary
Both the bag label and Instructions for Use include indications for use, cautions and warnings, as well as the general operating instructions required for proper use of the NxStage Premixed Dialysate Solutions. This information promotes safe and effective use of the device.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle faces left and is positioned within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
Public Health Service
od and Drug Administration 200 Corporate Boulevard Rockville MD 20850
OCT 21 2002
NxStage Medical, Inc. c/o Mr. Donald James Sherratt Intertek Testing Services 70 Codman Hill Road BOXBOROUGH MA 01719
#### Re: K022913
Trade/Device Name: NxStage Premixed Dialysate Regulation Number: 21 CFR 8876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: 78 KPO Dated: October 4, 2002 Received: October 7, 2002
Dear Mr. Sherratt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy Chrogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Page 2 .
{5}------------------------------------------------
Page 1 of 1
## Indications for Use Statement
Page 1 of 1
510(k) Number (if known):
Device Name:
. . . . . . . .
): KOZZ913
NxStage Premixed Dialysate
Indications for Use:
NxStage Premixed Dialysate is indicated for use with renal replacement therapy systems that utilize sterile premixed dialysate.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | <div>✓</div> | OR | Over-The-Counter |
|----------------------|--------------|----|------------------|
| Use | | | |
| (Per 21 CFR 801.109) | | | |
| (Division Sign-Off) | <div> <img alt="signature" src="signature.png"/> </div> |
|---------------------------------------------------------------|---------------------------------------------------------|
| Division of Reproductive, Abdominal, and Radiological Devices | |
| 510(k) Number | K022913 |
0 ປີ 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.