K033946 · Msq (M2) , Ltd. · FTC · Apr 5, 2004 · General, Plastic Surgery
Device Facts
Record ID
K033946
Device Name
LOVELY SYSTEM
Applicant
Msq (M2) , Ltd.
Product Code
FTC · General, Plastic Surgery
Decision Date
Apr 5, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Lovely System models (Aria and Harmony) are intended for use in aesthetic and cosmetic applications requiring selective photothermolysis (photocoagulation or coagulation) and hemostasis of soft tissue in the medical specialties of plastic surgery and dermatology as follows: Lovely I (Aria) & Lovely II (Harmony) Models: The Advanced Fluorescence Technology (AFT) with 420-950 nm wavelengths is indicated for: - The treatment of moderate inflammatory acne vulgaris. - The treatment of benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, ephelides (freckles). - The treatment of cutaneous lesions including warts, scars and striae. - The treatment of benign cutaneous vascular lesions including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, erytherna of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and voenous, malformations. - The removal of unwanted hair and to effect stable long-term or permanent hair reduction. Lovely II (Harmony) Model: The Nd: YAG lasers (1064nm) are indicated for: - The removal of black, blue or green tattoos. - Coagulation and hemostasis of vascular lesions and soft tissue, including treatment and clearance of superficial and deep telangiectasias) and reticular veing (0.1-4.0 mm diameter) of the leg. - The non-ablative treatment of facial wrinkles. Lovely II (Harmony) Model: - The UVB Light source (300-380nm) is indicated for: - The treatment of leukoderma, including vitiligo (acquired leukoderma). - The treatment of psoriasis, atopic dermatitis (eczema), and seborrheic dermatitis. The Lovely System models are also indicated for use on all skin types (Fitzpatrick I-VI), including tanned skin.
Device Story
Multi-modality aesthetic platform (Lovely I/Aria, Lovely II/Harmony) utilizing AFT (420-950nm), Nd:YAG laser (1064nm), and UVB (300-380nm) light sources. System comprises console with control electronics/software, water-cooling system, and seven interchangeable handpieces. Physician-operated in clinical settings. Device delivers light/laser energy to target tissues via selective photothermolysis, coagulation, or phototherapy. Console manages timing and dosing parameters via LCD/soft-key interface. Output affects tissue by targeting specific chromophores (melanin, hemoglobin, water) for lesion clearance, hair reduction, or skin rejuvenation. Benefits include non-invasive treatment of dermatological conditions and cosmetic improvements.
Clinical Evidence
No clinical data provided. Equivalence demonstrated through bench testing, compliance with safety standards (IEC 60601-1, IEC 60601-1-2, IEC-60825-1, IEC 601-2-22), and comparison of technical specifications to predicate devices.
Technological Characteristics
Platform supports AFT, UVB, and Nd:YAG laser sources. Wavelengths: 300-1064nm. Includes water-cooling system, LCD control panel, and seven interchangeable handpieces. Complies with 21 CFR 1040.10/1040.11 (Class IV laser), IEC 60601-1, IEC 60601-1-2, IEC-60825-1, and IEC 601-2-22. Software-controlled timing and dosing.
Indications for Use
Indicated for patients of all skin types (Fitzpatrick I-VI), including tanned skin, requiring aesthetic/cosmetic treatment of acne, pigmented/vascular lesions, unwanted hair, tattoos, wrinkles, psoriasis, vitiligo, and dermatitis.
Regulatory Classification
Identification
An ultraviolet lamp for dermatologic disorders is a device (including a fixture) intended to provide ultraviolet radiation of the body to photoactivate a drug in the treatment of a dermatologic disorder if the labeling of the drug intended for use with the device bears adequate directions for the device's use with that drug.
Predicate Devices
Lumenis Family of Intense Pulse-Light (IPL) and IPL/ND:YAG Laser Systems (K030342, K030527, K024093)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of a human figure with three wavy lines extending from the figure's head.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
#### APR 2 8 2004
Ms. Anne Worden Regulatory Consultant MSq(M2) Ltd. c/o A.Worden Consulting 3637 Bernal Avenue Pleasanton, California 94566
Re: K033946
Trade/Device Name: Lovely System Models: Lovely I (Aria) & Lovely II (Harmony) Regulation Number: 21 CFR 878.4810 and 21 CFR 878.4630 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in . dermatology and Ultraviolet lamp for dermatologic disorders Regulatory Class: II Product Code: GEX and FTC Dated: December 16, 2003 Received: January 9, 2004
Dear Ms. Worden:
This letter corrects our substantially equivalent letter of April 5, 2004 regarding the Indications of Use.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
#### Page 2 - Ms. Anne Worden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Purist
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure New Indications for Use
{2}------------------------------------------------
#### INDICATIONS FOR USE
#### 510(k) Number (if known):
#### Device Name: Lovely System Models: Lovely I (Aria) & Lovely II (Harmony)
#### Indications for Use:
The Lovely System models (Aria and Harmony) are intended for use in aesthetic and cosmetic applications requiring selective photothermolysis (photocoagulation or coagulation) and hemostasis of soft tissue in the medical specialties of plastic surgery and dermatology as follows:
#### Lovely I (Aria) & Lovely II (Harmony) Models:
The Advanced Fluorescence Technology (AFT) with 420-950 nm wavelengths is indicated for:
- The treatment of moderate inflammatory acne vulgaris.
- The treatment of benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, ephelides (freckles).
- . The treatment of cutaneous lesions including warts, scars and striae.
- The treatment of benign cutaneous vascular lesions including port wine stains, . hemangiomas, facial, truncal and leg telangiectasias, rosacea, erytherna of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and voenous, malformations.
- The removal of unwanted hair and to effect stable long-term or permanent hair reduction. ●
#### Lovely II (Harmony) Model:
The Nd: YAG lasers (1064nm) are indicated for:
- The removal of black, blue or green tattoos. .
- Coagulation and hemostasis of vascular lesions and soft tissue, including treatment and clearance of superficial and deep telangiectasias) and reticular veing (0.1-4.0 mm diameter) of the leg.
- The non-ablative treatment of facial wrinkles. .
#### Lovely II (Harmony) Model:
- The UVB Light source (300-380nm) is indicated for:
- The treatment of leukoderma, including vitiligo (acquired leukoderma). .
- The treatment of psoriasis, atopic dermatitis (eczema), and seborrheic dermatitis.
The Lovely System models are also indicated for use on all skin types (Fitzpatrick I-VI), including tanned skin.
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CBRE, office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Restorative.
# and Neurological Devices
| Prescription Use<br>(Per 21 CFR 801.109) | |
|------------------------------------------|-----------------------------|
| 510(k) Number | K033946 |
| | OR |
| | Over the Counter Use |
| | 10-8 |
| | Page Revised March 30, 2004 |
{3}------------------------------------------------
APR - 5 2004
Page 1 of 4
### 510(K) SUMMARY
## LOVELY SYSTEM 510(k) Number K_033946
| Applicant's Name: | MSq(M²) Ltd.<br>7 Haeshel St. P.O.B 3021<br>Caesarea Industrial Park 38900 Israel<br>Tel: 04-6275357<br>Fax: 04-6275368<br>e-mail: zivk@MSq.co.il |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Arava Hacohen<br>Push-med Ltd.<br>117 Ahuzah St.<br>Ra'anana 43373, Israel<br>Tel: 972-9-7718130<br>Fax: 972-9-7718131<br>e-mail: arava@push-med.com |
| Date Prepared: | December 2003 |
| Trade Name: | Lovely System |
| Classification Name: | Laser Instrument, Surgical, Powered |
| Classification: | FDA has classified laser device as a class II device (product<br>code GEX) and it is reviewed by the General & Plastic Surgery<br>Panel. |
| Predicate Devices: | The Lovely System is substantially equivalent in terms of<br>technological characteristics, performance, intended use,<br>indications for use and operator interface to the following<br>predicate devices:<br>Lumenis Family of Intense Pulse-Light (IPL) and<br>IPL/ND:YAG Laser Systems ("Lumenis Combined IPL")<br>cleared under K030342, K030527and K024093. Estelux™ (Palomar Medical Technologies) cleared under<br>K020453. Altus Medical Family of Altus Medical CoolGlide Aesthetic<br>Lasers with Optional Pulse Light Handpiece ("CoolGlide")<br>cleared under K023954. ClearLight Phototherapy Device, Model CL420<br>( |
{4}------------------------------------------------
Page 2 of 4
· Medlite Q-Switched Laser ("Medlite") (Continuum Electro-Optics, Inc.) cleared under K970808, K973719, K983054, K011677, K014234 and K022709.
#### Performance Standards:
Diode Laser complies with:U.S. Federal The Lovely Performance Standards 21 CFR 1040.10 and 21 CFR 1040.11 for class IV Laser Products.
In addition, the device complies with the European Medical Directive 93/42/EEC concerning medical devices (Annex II) and with the voluntary standards IEC 60601-1, IEC 60601-1-2, IEC-60825-1 and IEC 601-2-22.
#### Intended Use / Indication for Use:
The Lovely System is intended for use in aesthetic and cosmetic selective photothermolysis requiring applications (photocoagulation or coagulation) and hemostasis of soft tissue in the medical specialties of plastic surgery and dermatology as follows:
The Applied Fluoresce Technology (AFT) with 420-950mm wavelengths is indicated for:
- · The treatment of moderate inflammatory acne vulgaris.
- · The treatment of benign pigmented epidermal lesions hyperpigmentation, melasma, dyschromia. including ephelides (freckles).
- The treatment of cutaneous lesions including warts, scars 트 and striae.
- . The treatment of benign cutaneous vascular lesions including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations.
- The removal of unwanted hair and to effect stable long-마 term or permanent hair reduction.
The Nd: Y AG laser (1064nm) is indicated for:
- The removal of black, blue or green tattoo. ●
- Coagulation and hemostasis of vascular lesions and soft 드 tissue, including treatment and clearance of superficial and deep telangiectasias (venulectasias) and reticular veins (0.1-4.0 mm diameter) of the leg.
- . The non-ablative treatment of facial wrinkles.
The UVB Light source (300-380nm) is indicated for:
- The treatment of leukoderma. l
{5}------------------------------------------------
K033946
Page 3, fy
- The treatment of psoriasis, vitiligo, atopic dermatitis E (eczema), leukoderma, and seborrheic dermatitis
The lovely System is indicated for use on all skin types (Fitzpatrick I-VI), including tanned skin.
Device Description: The Lovely is a platform for AFT, UVB and Nd:YAG Laser Technology for various Aesthetic and Cosmetic applications. The Lovely has an option to operate seven different handpieces that emits light at 300nm to 1064nm for the following applications:
- Hair removal 트
- Skin Rejuvenation 트
- I Vascular and Pigmented lesion
- 트 Moderate Acne
- 트 Deep leg veins
- . Tattoo removal
- 트 Psoriasis and vitiligo
The system includes:
- A light/laser system console (including software and control 트 electronics) that controls timing and dosing parameters. The console includes a water-cooling system used to remove heat from the power supply and the optical bench.
- A control and display panel including LCD and soft-keys. 트
- Seven optional handpiece(s) that includes the light source, 트 electronics and pipes of the water cooling system.
Substantial Equivalence: The Lovely System is substantially equivalent in terms of technological characteristics, performance, intended use. indications for use and operator interface to the following predicate devices:
- Family of of Intense Pulse-Light (IPL) and · Lumenis IPL/ND:YAG Laser Systems ("Lumenis Combined IPL") cleared under K030342, K030527and K024093.
- · Estelux™ (Palomar Medical Technologies) cleared under K020453.
- · Altus Medical Family of Altus Medical CoolGlide Aesthetic Lasers with Optional Pulse Light Handpiece ("CoolGlide") cleared under K023954.
- Phototherapy Device, Model CL420 · ClearLight ("ClearLight") (CureLight Ltd) cleared under K013623 and K030338.
- Belear™ (Lumenis, Inc.) cleared under K020591 and . K021762.
{6}------------------------------------------------
K033946
page 4 of 4
- · Medlite Q-Switched Laser ("Medlite") (Continuum Electro-Optics, Inc.) cleared under K970808, K973719, K983054, K011677, K014234 and K022709.
There are no unique applications, indications, material or specifications presented below. Evidence of equivalence has been demonstrated through:
- · The Lovely intended use and indications for use were previously cleared by FDA for the predicate devices.
- · The technical characteristics of the Lovely are similar to those of the predicate devices.
- · Laser output values of the Lovely are well within previous cleared values of the predicate devices as described.
- · The predicate devices and other previous cleared devices with similar energy output have a proven safety and effectiveness in the treatment of the claimed indications.
- · Safety and performance testing.
Therefore, the Lovely System is substantially equivalent to its predicate devices cited above and raises no new safety and/or effectiveness issues.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.