LUMENIS FAMILY OF IPL AND COMBINATION IPL/ND:YAG SYSTEMS
K030527 · Lumenis · GEX · May 20, 2003 · General, Plastic Surgery
Device Facts
Record ID
K030527
Device Name
LUMENIS FAMILY OF IPL AND COMBINATION IPL/ND:YAG SYSTEMS
Applicant
Lumenis
Product Code
GEX · General, Plastic Surgery
Decision Date
May 20, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Lumenis Family of Intense Pulsed-Light (IPL) and IPL/Nd:YAG laser systems (and the delivery accessories that are used with them to deliver light and/or laser energy) are intended for use in surgical, aesthetic and cosmetic applications requiring selective photothermolysis (photocoagulation or coagulation) and hemostasis of soft tissue in the medical specialties of general and plastic surgery, and dermatology.
Device Story
System consists of console (electronics/software), control panel, and hand-pieces with integrated skin cooling. Uses Intense Pulsed Light (515-1200 nm) and Nd:YAG laser (1064 nm) to deliver energy for selective photothermolysis and hemostasis. Operated by clinicians in general surgery, plastic surgery, or dermatology settings. Real Time Chiller cools epidermis before, during, and after treatment to mitigate thermal injury and patient discomfort. Output allows clinicians to perform aesthetic/surgical procedures; clinical benefit includes effective treatment of vascular/pigmented lesions and hair reduction with reduced side effects.
Clinical Evidence
Clinical data provided to demonstrate safety and effectiveness for indicated uses in general surgery, plastic surgery, and dermatology. No specific study metrics (e.g., sensitivity, AUC) provided in summary.
Technological Characteristics
System includes console, control panel, and hand-pieces with integrated cooling. Energy sources: Intense Pulsed Light (515-1200 nm) and Nd:YAG laser (1064 nm). Features include software-controlled electronics for energy delivery and epidermal cooling. Connectivity and specific software architecture not detailed.
Indications for Use
Indicated for patients requiring treatment of tattoos, benign pigmented epidermal lesions (dyschromia, hyperpigmentation, melasma, ephelides), cutaneous lesions (warts, scars, striae), benign cutaneous vascular lesions (port wine stains, hemangiomas, telangiectasias, rosacea, angiomas, poikiloderma of Civatte, leg veins, venous malformations), and hair removal/reduction (skin types I-V). Nd:YAG component indicated for coagulation/hemostasis of vascular lesions and soft tissue, including leg telangiectasias and reticular veins (0.1-4.0 mm). Real Time Chiller indicated for epidermal cooling to reduce pain/discomfort, minimize thermal injury, and allow higher fluences.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
K0305627
# MAY 2 0 2003
#### Appendix 2 VIII.
## A. 510(k) Summary
## Table 1 - Administrative Information
1/2
| Category: | Comments |
|--------------------------------------------|-----------------------------------------------------------------------------------------|
| Sponsor: | Lumenis, Inc.<br>2400 Condensa Street<br>Santa Clara, CA 95051 |
| Correspondent: | Andrea L. Ruth, RAC<br>Senior Associate II, Regulatory Affairs |
| Contact Numbers: | 408.764.3235 |
| Device Common Name | Laser Powered Surgical Instrument (and Accessories) |
| Device Proprietary Name | Lumenis Family of Intense Pulsed-Light (IPL) and IPL/Nd:YAG<br>Laser Systems |
| Device Classification Name | Laser Surgical Instrument for Use in General and planstic<br>Surgery and in Dermatology |
| Device Classification | 21 CFR § 878.4810 |
| Predicate Devices | IPL Quantum Family; VascuLight Family |
| Predicate Device<br>Manufacturer(s) | Lumenis |
| Predicate Device Reference(s) | K020839 |
| Predicate Device Classification<br>Name(s) | Laser Surgical Instrument for Use in General and planstic<br>Surgery and in Dermatology |
| Predicate Device<br>Classification(s) | 21 CFR §878.4810 |
Date Summary Was Prepared: February 15, 2003.
Description of the Device: Lumenis Family of Intense Pulsed-Light (IPL) and IPL/Nd:YAG laser systems are comprised of the following main components:
- A light/laser system console (including software and control electronics); *
- * A control and display panel; and
- * One or more attached hand-piece(s), which may have integrated skin cooling components.
Intended Use: The Lumenis Family of Intense Pulsed-Light (IPL) and IPL/Nd:YAG laser systems (and the delivery accessories that are used with them to deliver light and/or laser energy) are intended for use in surgical, aesthetic and cosmetic applications requiring selective photothermolysis (photocoagulation or coagulation) and hemostasis of soft tissue in the medical specialties of general and plastic surgery, and dermatology.
{1}------------------------------------------------
Kozo iji) 2/2
Traditional 510(k) Submission February 14, 2003
Safety and Effectiveness Information: Clinical data was provided to demonstrate that the Lumenis Family of IPL and IPL/Nd:YAG laser (1064 nm) systems are safe and effective for their intended and indicated uses in the medical specialties of general and plastic surgery, and dermatology.
Conclusion: The Lumenis Family of Intense Pulsed-Light (IPL) and IPL / Nd: Y AG laser systems were shown to be substantially equivalent to the predicate devices cleared in K020839. The Lumenis Family of Intense Pulsed-Light (IPL) and IPL/Nd:YAG laser systems has identical intended use, as well as functional and design features as the currently marketed predicate devices. The only change is expansion of labeled indications for use, which fall within the scope of the intended use.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## MAY 2 0 2003
Ms. Andrea L. Ruth, RAC Senior Associate II, Regulatory Affairs Lumenis, Inc. 2400 Condensa Street Santa Clara, California 95051
Re: K030527
Trade/Device Name: Lumenis Family of Intense Pulsed-Light (IPL) and IPL/Nd:YAG Laser Systems Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: February 14, 2003 Received: February 19, 2003
Dear Ms. Ruth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Andrea L. Ruth, RAC
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Sincerely, yours,
Muh M. Milsener
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## 510(k) Number (if Known): K030527
#### Lumenis Family of Intense Pulsed-Light (IPL) and IPL/Nd:YAG Device Name: Laser Systems
Indications For Use:
The Lumenis Family of Intense Pulsed-Light (IPL) and IPL/Nd:YAG laser systems (and the delivery accessories that are used with them to deliver light and/or laser energy) are indicated for use in surgical, aesthetic and cosmetic applications requiring selective photothermolysis (photocoagulation or coagulation) and hemostasis of soft tissue in the medical specialties of general and plastic surgery, and dermatology as follows:
Intense Pulsed Light Energy / Wavelengths (515 - 1200 nm) are indicated for:
- The treatment of tattoos; 关
- The treatment of benign pigmented epidermal lesions including dyschromia, * hyperpigmentation, melasma, ephelides(freckles);
- * The treatment of cutaneous lesions including warts, scars and striae;
- * The treatment of benign cutaneous vascular lesions, including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations;
- The removal of unwanted hair from all skin types, and to effect stable long-term, * or permanent1, hair reduction in skin types I-V.
***Continued on Following Page***
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|--|
|--------------------------------------------------------|--|
| Prescription Use<br>(Per 21 CFR 801.109) | <div style="text-align:center;">✓</div> | OR Over-The-Counter Use<br>(Optional Format 1-2-96) | |
|------------------------------------------|-----------------------------------------|-----------------------------------------------------|--|
|------------------------------------------|-----------------------------------------|-----------------------------------------------------|--|
1 Permanent hair reduction is defined as a long-term stable reduction in the number of hairs regrowing after a treatment regimen.
LUMENIS, INC. MODIFICATION TO IPL AND IPL/ND:YAG COMBINATION PRODUCT FAMILIES CONFIDENTIAL
{5}------------------------------------------------
510(k) Number (if Known): K030527
Device Name:
## Lumenis Family of Intense Pulsed-Light (IPL) and IPL/Nd:YAG Laser Systems
### *** Continued from Previous Page ***
JAII CHILIS INC
Nd:YAG Laser Wavelength (1064 nm) is indicated for the coagulation and hemostasis of vascular lesions and soft tissue, including:
- Treatment and clearance of superficial and deep telangiectasias (venulectasias) * and reticular veins (0.1 - 4.0 mm diameter) of the leg.
- The removal of unwanted hair from all skin types, and to effect stable long-term, * or permanent1, hair reduction in skin types I-V through selective targeting of melanin in hair follicles.
The Real Time Chiller is indicated for use in cooling the epidermis at the treatment site prior to, during and after light or laser treatment in general surgery, plastic surgery and dermatology to:
- Reduce pain during and/or associated with light or laser treatment (via partial * anesthesia from cooling);
- Reduce discomfort during and/or associated with light or laser treatment; *
- Minimize thermal injury, including thermal necrosis, to non-target skin and skin * structures during and/or associated with light or laser treatment, thus reducing complications such as scabbing, scarring, hyperand/or possible hypopigmentation;
- * Allow the use of higher light or laser fluences for light or laser treatments (such as for hair removal and the treatment of vascular or pigmented lesions); and
- * Reduce potential side effects of light or laser treatments (such as for hair removal and the treatment of vascular or pigmented lesions).
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|--------------------------|
| Prescription Use | <img alt="line" src=""/> |
| (Per 21 CFR 801.109) | |
OR
| Over-The-Counter Use | <img alt="line" src=""/> |
|--------------------------|--------------------------|
| (Optional Format 1-2-96) | |
LUMENIS, INC. MODIFICATION TO IPL AND IPL/ND: YAG COMBINATION PRODUCT FAMILIES CONFIDENTIAL
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.