MODIFICATION TO BARD AQUAGUIDE URETERAL CONDUIT

K033778 · C.R. Bard, Inc. · FED · Dec 30, 2003 · Gastroenterology, Urology

Device Facts

Record IDK033778
Device NameMODIFICATION TO BARD AQUAGUIDE URETERAL CONDUIT
ApplicantC.R. Bard, Inc.
Product CodeFED · Gastroenterology, Urology
Decision DateDec 30, 2003
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The Bard® AquaGuide™ Ureteral Access Sheath is indicated for use in endoscopic urology procedures where ureteral dilation and continued ureteral access is desired for injection and aspiration of fluids and insertion and removal of endoscopes and related instruments.

Device Story

Bard® AquaGuide™ Ureteral Access Sheath is a two-component ureteral dilation system; consists of hydrophilic-coated introducer and dual-lumen sheath with multi-function access hub. Used in endoscopic urology procedures; operated by urologists. Provides continuous ureteral access; allows for injection/aspiration of fluids and passage of endoscopes/instruments. Dual-lumen design reduces potential for hypervolemia. Placed over 0.038" guidewire. Benefits include maintained access and fluid management during procedures.

Clinical Evidence

Bench testing only. Device constructed of biocompatible materials; design features (additional lumen) intended to reduce potential for hypervolemia.

Technological Characteristics

Two-component system; hydrophilic-coated introducer and dual-lumen sheath. Multi-function access hub. Compatible with 0.038" guidewire. Sizes: Introducer 10-13, Sheath 12-16; Lengths: 25-55cm. Biocompatible materials.

Indications for Use

Indicated for patients undergoing endoscopic urology procedures requiring ureteral dilation and access for fluid management and instrument passage.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ DEC 3 0 2003 Image /page/0/Picture/1 description: The image shows the text "C. R. Bard, Inc." on the first line, followed by the address "8195 Industrial Blvd." on the second line. The third line contains the city, state, and zip code: "Covington, GA 30014". The text is in a simple, sans-serif font and is left-aligned. Image /page/0/Picture/2 description: The image shows a handwritten text that appears to be a combination of alphanumeric characters and words. The text includes 'K033778' on the top line, followed by 'page 1 of 2' on the second line. The handwriting is cursive and slightly stylized, with some characters more legible than others. Image /page/0/Picture/3 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are all capitalized and connected, giving the word a solid, block-like appearance. The overall impression is one of strength and stability. ## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990. A. Submitter's Information: | Submitter's Name:<br>Address: | C. R. Bard, Inc., Urological Division<br>8195 Industrial Blvd.<br>Covington, Georgia 30014 | |-------------------------------|--------------------------------------------------------------------------------------------| | Contact Person: | Frances E. Harrison, RAC | | Contact Person's Phone: | (770) 784-6257 | | Contact Person's Fax: | (770) 784-6419 | | Date of Preparation: | November 26, 2003 | - B. Device Name: | Trade Name: | Bard® AquaGuide™ Ureteral Access Sheath | |----------------------|-----------------------------------------| | Common / Usual Name: | Ureteral Conduit | | Classification Name: | Endoscope and Accessories | - C. Predicate Device Name: Bard® AquaGuide™ Ureteral Conduit Trade Name: Same as above - D. Device Description: The Bard® AquaGuide™ Ureteral Access Sheath is a twocomponent ureteral dilation system which contains multiple lumens for injection and aspiration of fluids as well as passage of endoscopes and related instruments. The auptram consists of a hydrophilic-coated introducer with female luer connector and a hydrophilic-coated dual-lumen sheath with multi-function access hub. The system is available in Introducer sizes 10-13, Sheath sizes 12-16 and lengths: 25 - 55cm. - E. Intended Use: The Bard® AquaGuide™ Ureteral Access Sheath is indicated for use in indoscopic urology procedures where ureteral dilation and continued ureteral access is desired for injection and aspiration of fluids and insertion and removal of endoscopes and related instruments. {1}------------------------------------------------ K033778 Page 2 of 2 - F. Technological Characteristics Summary: The Bard® AquaGuide™ Ureteral Access r connological onaracted online coating and is constructed of materials that soften at Oncan has a labholoud hydrophilis comfort. The device can be placed over a 0.038" buidewire, allows for passage of endoscopes and related instruments, allows for injection and aspiration of fluids and provides for continuous flow. - G. Performance Data Summary: The Bard® AquaGuide™ Ureteral Access Sheath is r onstructed of biocompatible materials. The additional lumen reduces the potential for hypervolemia (excess fluid build-up). {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure with three flowing lines representing the body and arms. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC 3 0 2003 Frances E. Harrison, RAC Director, Regulatory Affairs Bard Urological Division C. R. Bard, Inc. 8195 Industrial Blvd. COVINGTON GA 30014 Re: K033778 Trade/Device Name: Bard® AquaGuide™ Ureteral Access Sheath Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: 78 FED Dated: November 26, 2003 Received: December 5, 2003 Dear Ms. Harrison: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreated.1976, the enactment date of the Medical Device Amendments, or to connices that have been reclassified in accordance with the provisions of the Federal Food, Drug, de nees that have been rout do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, thereferer, missus of the Act include requirements for annual registration, listing of Echolar volurole provinstvring practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean r that FDA has made a determination that your device complies with other requirements of the Act that x 1971 here interes and regulations administered by other Federal agencies. You must or any I varrate the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set (21 CI R Part 007), acoming (2) regulation (21 CFR Part 820); and if applicable, the electronic form in the quality of the provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to ocgin harketing of substantial equivalence of your device to a legally prematication. The PDF mailing to and group device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), plcase If you desire specific advice for your dones an one of the following numbers, based on the regulation number at the top of the letter: | 8xx. 1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, 101 questions of the premier A. . Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding Other of Compliance at (301) 37 1 105 Part 807.97) you may obtain. Other general by reletence to premarked nombation (er the Act may be obtained from the Division of Strall information on your responsionalism and Consumer Assistance at its toll-free number (800) 638-2041 or Manufacturers, International and Solless http://www.fda.gov/cdrf/dsma/dsmamain.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## INDICATIONS FOR USE STATEMENT 033778 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Bard® AquaGuide™ Ureteral Access Sheath Device Name: ________________________________________________________________________________________________________________________________________________________________ Indications for Use: The Bard® AquaGuide™ Ureteral Access Sheath is indicated for use in endoscopic urology procedures where ureteral dilation and continued ureteral endoscopic arology procession and aspiration of fluids and insertion and removal of endoscopes and related instruments. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE) -Prescription Use _ (Per 21 CFR 801.109) Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ OR (Optional Format 1/2/96) Nancy C. Brogdon (Division Sign Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...