URINE COCAINE METABOLITE SCREEN FLEX REAGENT CARTRIDGE, CATALOG# DF92A
Device Facts
| Record ID | K033713 |
|---|---|
| Device Name | URINE COCAINE METABOLITE SCREEN FLEX REAGENT CARTRIDGE, CATALOG# DF92A |
| Applicant | Dade Behring, Inc. |
| Product Code | DIO · Clinical Toxicology |
| Decision Date | Feb 9, 2004 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.3250 |
| Device Class | Class 2 |
Indications for Use
The COC Flex® reagent cartridge used on the Dimension® clinical chemistry system provides reagents for an in vitro diagnostic test intended for the qualitative and semi-quantitative determination of benzoylecgonine (cocaine metabolite) in human urine using a cutoff of 150 or 300 ng/mL. Measurements obtained with the COC method are used in the diagnosis and treatment of cocainc use or overdose. The COC method provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.
Device Story
Single-use, 8-well plastic cartridge containing polyclonal antibody, substrate, and enzyme-labeled benzoylecgonine. Used on Dimension clinical chemistry analyzers; operated by qualified laboratory personnel. Input: human urine sample. Principle: homogeneous enzyme immunoassay; drug in sample competes with enzyme-labeled drug for antibody binding sites. Unbound conjugate catalyzes G6PDH oxidation, reducing NAD+ to NADH. Analyzer measures rate of absorbance increase at 340nm. Output: qualitative or semi-quantitative benzoylecgonine concentration. Results are preliminary; require confirmation via GC/MS. Semi-quantitative output assists in sample dilution for confirmatory testing and quality control. Not for point-of-care use.
Clinical Evidence
Bench testing only. Precision evaluated per NCCLS EP5-A over 20 days. Method comparison performed on 125 urine samples against predicate and GC/MS. Qualitative agreement with GC/MS: 99% positive, 91% negative. Semi-quantitative agreement with GC/MS: 96% positive, 91% negative. Analytical sensitivity (LOD) is 35 ng/mL. Cross-reactivity and interference testing performed with various drugs and endogenous compounds.
Technological Characteristics
Plastic 8-well cartridge format. Homogeneous enzyme immunoassay (G6PDH-based). Energy source: automated clinical chemistry analyzer (optical measurement at 340nm). Connectivity: integrated with Dimension analyzer family. Sterilization: not specified.
Indications for Use
Indicated for the qualitative and semi-quantitative measurement of benzoylecgonine in human urine to aid in the diagnosis and treatment of cocaine use or overdose. Intended for use by qualified laboratory personnel.
Regulatory Classification
Identification
A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.
Special Controls
*Classification.* Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
- Syva’s Emit II Plus Cocaine Metabolite Assay for the 30-R Biochemical System (K031512)