EMIT II PLUS COCAINE METABOLITE ASSAY
Device Facts
| Record ID | K031512 |
|---|---|
| Device Name | EMIT II PLUS COCAINE METABOLITE ASSAY |
| Applicant | Dade Behring, Inc. |
| Product Code | DIO · Clinical Toxicology |
| Decision Date | Jul 23, 2003 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.3250 |
| Device Class | Class 2 |
| Attributes | Real-World Evidence |
Real-World Evidence
| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
|---|---|---|---|---|---|
| K031512 · Jul 23, 2003 | EMIT II PLUS COCAINE METABOLITE ASSAY | Dade Behring, Inc. | Retrospective human urine samples | Retrospective clinical urine samples were used to compare the performance of the EMIT II Plus Cocaine Metabolite Assay against GC/MS reference results to establish substantial equivalence. | Method comparison; Retrospective samples; GC/MS comparison |
Clinical Evidence
| Study Design | Population | Comparator | Key Endpoints |
|---|---|---|---|
| Retrospective method comparison | Human urine samples; Sample Size: 125 (120 native retrospective samples and 5 diluted samples); Number of Sites: 1 (manufacturer's site) | GC/MS | Qualitative agreement at 150 ng/ml and 300 ng/ml cutoffs |
Indications for Use
The Emit® II Plus Cocaine Metabolite Assay is a homogeneous enzyme immunoassay with a 150 ng/mL or 300 ng/mL cutoff (SAMSHA initial test cutoff level). The assay is intended for use in the qualitative and semi-quantitative analyses of benzoylecgonine (cocaine metabolite) in human urine. Emit® II Plus assays are designed for use with a number of chemistry analyzers. The Emit® II Plus Monoclonal Cocaine Metabolite Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.
Device Story
The EMIT II Plus Cocaine Metabolite Assay is a homogeneous enzyme immunoassay used in clinical laboratories on automated chemistry analyzers. It detects benzylecgonine in human urine samples. The assay utilizes antibodies, substrates, and enzyme-labeled benzylecgonine; the presence of the drug in the sample competes with the enzyme-labeled drug for antibody binding sites, modulating enzyme activity. The analyzer measures the resulting signal to provide qualitative or semi-quantitative results. Results are preliminary and require confirmation via gas chromatography/mass spectrometry (GC/MS). The assay assists clinicians in identifying cocaine use; it is intended for professional use in clinical settings.
Clinical Evidence
Method comparison study performed against GC/MS reference method. At 150 ng/mL cutoff, 93% agreement observed. At 300 ng/mL cutoff, 94% agreement observed. Data demonstrates performance consistency with predicate.
Technological Characteristics
Homogeneous enzyme immunoassay; utilizes chemistry analyzers; qualitative and semi-quantitative detection of benzoylecgonine; reagent-based system.
Indications for Use
Indicated for qualitative and semi-quantitative analysis of benzylecgonine (cocaine metabolite) in human urine for prescription use.
Regulatory Classification
Identification
A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.
Special Controls
*Classification.* Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
- Syva EMIT II Plus Cocaine Metabolite Assay (K993988)