CALCIGEN -NAP BONE VOID FILLER

K033611 · Biomet, Inc. · MQV · Apr 27, 2004 · Orthopedic

Device Facts

Record IDK033611
Device NameCALCIGEN -NAP BONE VOID FILLER
ApplicantBiomet, Inc.
Product CodeMQV · Orthopedic
Decision DateApr 27, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3045
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Calcigen™-NaP Bone Void Filler is intended for filling bony voids of the skeletal system (i.e., the extremities, spine and pelvis). These defects may be surgically created osseous defects or osseous defects created from disease or traumatic injury to the bone. The device is intended for bone voids or gaps that are not intrinsic to the structural integrity of the bony structure.

Device Story

Calcigen™-NaP Bone Void Filler is a synthetic, self-setting bone graft substitute. Provided as powder and setting solution; mixed intra-operatively to form a moldable paste. Applied to bone defects via spatula or syringe; hardens in-situ to prevent migration. Resorbed and replaced by bone during healing. Used by surgeons in clinical settings to fill non-structural osseous defects in extremities, spine, or pelvis.

Clinical Evidence

No clinical data provided.

Technological Characteristics

Synthetic calcium sodium phosphate bone void filler. Provided as a two-component system (powder and setting solution) for intra-operative mixing. Self-setting paste that hardens in-situ. Resorbable material.

Indications for Use

Indicated for filling non-structural bony voids in the extremities, spine, and pelvis caused by surgery, disease, or trauma.

Regulatory Classification

Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K033611 APR 2 7 2004 Image /page/0/Picture/2 description: The image shows the word "BIOMET" in a bold, sans-serif font. The letters are all capitalized and outlined in black. The word appears to be a logo or brand name. ### 510(k) Summary | Applicant/Sponsor: | Biomet Manufacturing Corp. | |-------------------------------------------|---------------------------------------------------| | Contact Person: | Mary L. Verstynen<br>Director of Clinical Affairs | | Proprietary Name: | Calcigen™-NaP Bone Void Filler | | Common Name: | Calcium Sodium Phosphate Bone Void Filler | | Classification Name:<br>(21 CFR 888.3045) | Resorbable calcium salt bone void filler device | # Legally Marketed Devices To Which Substantial Equivalence Is Claimed: - · EBI OsteoStim™ Granules Resorbable Bone Graft Substitute (EBI L.P .; K011386) - · α-BSM® Bone Substitute Material (DePuy Orthopaedics, Inc; K011048) - · Pro Osteon® 500R Resorbable Bone Graft Substitute (Interpore Cross; K990131. K980817) - · Norian® SRS® Bone Void Filler (Synthes USA; K011897) - · 3i® Calcium Sodium Phosphate Bone Cement (Biomet, Inc; K003493) #### Device Description: Calcigen™-NaP Bone Void Filler is a synthetic, self-setting paste for use as a bone graft substitute. The product is provided in pre-measured quantities of powder and setting solution. The two components are mixed together intra-operatively to form a thick, moldable paste that is then applied to the bone void or defect. The paste can be shaped to completely fill the bone defect and hardens in-situ to prevent migration out the defect area. It is designed to be applied/delivered to the defect either using a spatula or equivalent instrument or injected thru a syringe into open defects. The material is resorbed and replaced by bone during the healing process. #### Intended Use: Calcigen™-NaP Bone Void Filler is intended for filling bony voids of the skeletal system (i.e., the extremities, spine and pelvis). These defects may be surgically created osseous defects or osseous defects created from disease or traumatic injury to the bone. The device is intended for bone voids or gaps that are not intrinsic to the structural integrity of the bony structure. > MAILING ADDRESS Box 587 Box 587 - 3 - 3 - 3 - 3 - - - - - - - - - - - - - - - - - - - - - - - Warsaw, IN 46582 > > . SHIPPING ADDRESS で ()FFICE 574.267.6639 FAX 574.267.8137 E-MAIL biomet@biomet@biomet.com {1}------------------------------------------------ ## Summary of Technologies: Calcigen™-NaP Bone Void Filler is shown by analysis of available information to be identical in intended use and equivalent in materials and form to currently marketed similar products. The product is identical in chemistry and formulation to 3i® Calcium Sodium Phosphate Bone Cement. #### Non-Clinical Testing: The information presented demonstrates that the Calcigen™-NaP Bone Void Filler is substantially equivalent to currently marketed predicate devices. The differences between Calcigen™-NaP Bone Void Filler and the predicates are not significant and have already been shown not to present additional safety or efficacy issues. #### Clinical Testing: None is provided as the basis of substantial equivalence. {2}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR 2 7 2004 Ms. Mary L. Verstynen Director of Clinical and Regulatory Affairs Biomet Manufacturing Corporation 56 East Bell Drive P.O. Box 587 Warsaw, Indiana 46581-0587 Re: K033611 Trade/Device Name: Calcigen™-NaP Bone Void Filler Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcıum salt bone void filler Regulatory Class: II Product Code: MQV Dated: January 28, 2004 Received: January 30, 2004 Dear Ms. Verstynen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Mcdical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (sec above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please he advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807), labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ #### Page 2 - Ms. Mary L. Verstynen This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours. Miriam C. Provost Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosurc {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): K033611 Device Name: Calcigen™-NaP Bone Void Filler Incications For Use: Calcigen™-NaP Bone Void Filler is intended for filling bony voids of the skeletal system (i.e., the extremities, spine and pelvis). These defects may be surgically created osseous defects or osseous defects created from disease or traumatic injury to the bone. The device is intended for bone volds or gaps that are not intrinsic to the structural integrity of the bony structure. Prescription Use ﻟﺮ (Part 2 1 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) uriam C. Provost (Division Sign-Off) Division of General, Restorative, and Neurological Devices Page 1 of 1 **510(k) Number** K033611
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...