PRO OSTEON IMPLANT 500R RESORBABLE BONE GRAFT SUBSTITUTE MODEL NUMBERS 5RG05, 5RG10, 5RG15, 5G15L, 5RG20, 5RG30, 5RG30L
Applicant
Interpore Intl.
Product Code
MQV · Orthopedic
Decision Date
Sep 25, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Pro Osteon® 500R Resorbable Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. Pro Osteon 500R is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities, spine and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.
Device Story
Pro Osteon 500R is an osteoconductive, porous, resorbable bone void filler; structure mimics human cancellous bone. Composed of trabecular calcium carbonate coated with thin calcium phosphate layer. Supplied sterile in various shapes/sizes. Used by surgeons familiar with bone grafting; packed into non-structural bony voids/gaps in extremities, spine, or pelvis. Implant provides scaffold for tissue infiltration and bone ingrowth; calcium phosphate layer resorbs slowly, followed by faster resorption of calcium carbonate substructure (within 6 months). Requires rigid fixation techniques for defect stabilization; does not provide mechanical support. Radiopaque; may mask pathology or complicate radiographic assessment of bone ingrowth. Benefits patient by facilitating natural bone healing and replacement of void space.
Clinical Evidence
No clinical data. Evidence based on animal testing (rabbit tibial defect model) demonstrating bone formation and resorption rates, dissolution studies comparing resorption to non-resorbable substitutes, and mechanical testing confirming compressive characteristics equivalent to trabecular bone. Testing confirmed non-pyrogenic and non-toxic status.
Technological Characteristics
Porous, trabecular calcium carbonate implant with 2-10 micron calcium phosphate surface coating. Median pore diameter 435 microns. Resorbable. Supplied sterile. Mechanical properties equivalent to trabecular bone. No software or electronic components.
Indications for Use
Indicated for filling bony voids or gaps in the skeletal system (extremities, spine, pelvis) not intrinsic to structural stability. Suitable for surgically created or traumatic osseous defects. Contraindicated for growth plate fractures, segmental defects, sites subject to excessive impact/stress, significant vascular impairment, metabolic/systemic bone/healing disorders, unstable fractures, sites requiring soft tissue coverage not planned, or infected sites.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
Predicate Devices
Wright Medical Osteoset™ Bone Void Filler (K960978 and K963562)
Submission Summary (Full Text)
{0}------------------------------------------------
SEP 2 5 1998
510(k) Premarket Notification Pro Osteon 500R Resorbable Bone Void Filler
Interpore International - Confidential -K980817
SEP 2 5 1998
# APPENDIX 11
# 510(k) SUMMARY
# FOR
# PRO OSTEON® 500R RESORBABLE BONE VOID FILLER
{1}------------------------------------------------
# 510(k) SUMMARY
#### SUBMITTED BY
David P. Balding Director, Quality Assurance and Regulatory Affairs INTERPORE International 181 Technology Drive Irvine, California 92618 (714) 453-3200
March 2, 1998 Date Submitted:
# CLASSIFICATION, COMMON OR USUAL NAME, DEVICE NAME
| Classification Name: | Bone Void Filler |
|-------------------------|----------------------------------------------|
| Common/Usual Name: | Bone Void Filler |
| Product Classification: | Unclassified |
| Proprietary Name: | Pro Osteon® 500R Resorbable Bone Void Filler |
### PREDICATE DEVICE
Wright Medical Osteoset™ Bone Void Filler (aka Wright Plaster of Paris Pellets) [reference 510(k)'s K960978 and K963562]
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92.
### DEVICE DESCRIPTION AND MATERIALS OF CONSTRUCTION
Pro Osteon 500R Resorbable Bone Void Filler is an osteoconductive porous implant similar in structure to human cancellous bone. It is supplied sterile in various shapes and sizes.
Pro Osteon 500R has a trabecular structure which resembles the multidirectional interconnected porosity of cancellous bone. It has a median pore diameter of 435 microns. The implant is trabecular calcium carbonate covered by a very thin outer layer of calcium phosphate representing approximately 13% of the total implant structure. The calcium phosphate is located on the outer surface of the porosity throughout the entire structure of the implant. It is approximately 2 to 10 microns in thickness and overlays
{2}------------------------------------------------
the calcium carbonate substructure. Once implanted, the calcium phosphate outer laver will slowly resorb, delaving exposure of the underlying and faster resorbing calcium carbonate. Calcium carbonate resorption will normally occur within six months or less.
When Pro Osteon 500R is placed in direct contact with viable bone, the reticulated spaces in the implant are infiltrated with tissue. Bone formation occurs in apposition to the Pro Osteon 500R surface and within the interstices of the implant skeleton. As the implant resorbs, bone and soft tissue grow into the space previously occupied by the implant.
# INDICATIONS FOR USE
Pro Osteon 500R Resorbable Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. Pro Osteon 500R is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities, spine and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.
# COMPARISON TO THE PREDICATE DEVICE
Pro Osteon 500R is technologically very similar to the predicate device with the following exceptions:
- Pro Osteon 500R is a composite material of calcium carbonate and 1. calcium phosphate. The predicate device is composed of calcium sulfate.
- Pro Osteon 500R has a resorption rate of 6 months or less versus 2. Osteoset's 30 to 60 days.
- Pro Osteon 500R provides an interconnected, porous scaffold which 3. supports new bone ingrowth, while the predicate device is non-porous and is quickly resorbed in vivo.
{3}------------------------------------------------
# CONTRAINDICATIONS, WARNINGS, PRECAUTIONS AND COMPLI-CATIONS
- 1. Contraindications
Pro Osteon 500R is contraindicated for fractures of the growth plate; for segmental defects; for indications which may be subjected to excessive impact or stresses; when there is significant vascular impairment proximal to the graft site; when there are metabolic or systemic disorders that affect bone or wound healing; when stabilization of the fracture is not possible; in cases where intraoperative soft tissue coverage is not planned or possible, and in infected sites.
- 2. Warnings and Precautions
Pro Osteon 500R does not possess sufficient mechanical strength to support reduction of a defect site prior to soft and hard tissue ingrowth. Rigid fixation techniques are recommended as needed to assure rigid stabilization of the defect in all planes. Pro Osteon 500R Resorbable Bone Void Filler is intended for use by surgeons familiar with bone grafting and rigid fixation techniques. Complete postoperative wound closure is essential.
Pro Osteon 500R is radiopaque until resorbed. Radiopacity may mask underlying pathological conditions. Radiopacity may also make it difficult to radiographically assess the ingrowth of new bone.
- 3. Complications
Potential complications using this device are the same as those encountered in autogenous bone grafting procedures and include the following: superficial wound infection, deep wound infection, deep wound infection with osteomyelitis, nonunion, wound dehiscence, delayed union, malunion, loss of reduction, refracture, cyst recurrence, hematoma, and cellulitis.
### TESTING SUMMARY
Animal testing demonstrates bone formation rate and resorption of Pro Osteon 500R is substantially equivalent to the predicate device for the repair of tibial defects in rabbits. Dissolution studies demonstrate that Pro Osteon 500R has dissolution rates which are similar to the predicate device when
{4}------------------------------------------------
compared to essentially non-resorbable bone graft substitute materials. Mechanical testing demonstrates that Pro Osteon 500R has compressive characteristics substantially equivalent to the predicate device and to trabecular bone. Testing indicates the product to be non-pyrogenic and nontoxic. -----------------------------------------------------------------------------------------------------------------------------------------------------------------------
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
The product design, material of construction, and function as a bone void filler are substantially equivalent to the FDA cleared and legally marketed predicate device.
- Pro Osteon 500R and the predicate device are both calcium salts. Pro 1. Osteon 500R is a composite of calcium carbonate and calcium phosphate. Osteoset is composed of calcium sulfate and stearic acid.
- 2. Pro Osteon 500R and the predicate device are resorbed following implantation.
- 3. Pro Osteon 500R and the predicate device provide an environment for new bone ingrowth.
{5}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. David P. Balding ·Interpore International 181 Technology Drive Irvine, California 92618-2402
SEP 2 5 1998 Re: K980817 Pro Osteon® Resorbable Bone Void Filler Requlatory Class: Unclassified Product Code: MQV June 30, 1998 Dated: Received: July 1, 1998
Dear Mr. Balding:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. Please note: concerning your device in the Federal Register. this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
#### Page 2 - Mr. David P. Balding
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
510(k) Number (if known): K980817
Device Name: Pro Osteon® 500R Resorbable Bone Void Filler
#### Indications for Use:
Pro Osteon® 500R Resorbable Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. Pro Osteon 500R is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities, spine and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
(Division Sign-Off)
Division of General Restorative Devices
510(k) Number K980817
Prescription Use
(PER 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.