MODIFICATION TO EN-BLOC BIOPSY SYSTEM

K033427 · Neothermia Corp. · KNW · Nov 10, 2003 · Gastroenterology, Urology

Device Facts

Record IDK033427
Device NameMODIFICATION TO EN-BLOC BIOPSY SYSTEM
ApplicantNeothermia Corp.
Product CodeKNW · Gastroenterology, Urology
Decision DateNov 10, 2003
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.1075
Device ClassClass 2

Indications for Use

The en-bloc Biopsy System is intended for diagnostic sampling of breast tissue during a breast biopsy procedure.

Device Story

The en-bloc Biopsy System is a percutaneous, automated, vacuum-assisted electrosurgical device used for breast tissue biopsy. It consists of a hand-held biopsy handle, a single-use biopsy probe, and a control unit. The probe features two sets of active electrodes (precursor and cutting/capture) at its distal end, encased in a stainless steel cannula with an outer plastic sleeve. An annular gap between the sleeve and cannula provides vacuum-assisted removal of gaseous electrosurgical byproducts and fluids, while also maintaining the cutting arc during tissue capture. The device is operated by a clinician in a clinical setting. The system automates the electrosurgical cutting and simultaneous capture of an incised tissue volume, facilitating diagnostic sampling.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Percutaneous, vacuum-assisted electrosurgical device. Components: hand-held handle, single-use probe with dual active electrodes (precursor and cutting/capture), stainless steel cannula, outer plastic sleeve. Energy source: high-frequency electrosurgical generator. Connectivity: integral cable to control unit. Vacuum system for byproduct removal and cutting arc maintenance.

Indications for Use

Indicated for diagnostic sampling of breast tissue during a breast biopsy procedure.

Regulatory Classification

Identification

A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Ko 33 427 NOV 1 0 2003 #### 510(k) SUMMARY # Neothermia Corporation's en-bloc Biopsy System™ # Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared Sherrie Coval-Goldsmith VP. RA/QA Neothermia Corporation One Apple Hill, Suite 316 Natick, Massachusetts 01760 Phone: (508) 655-7820 Facsimile: (508) 655-7822 Date Prepared: October 21, 2003 # Name of Device and Name/Address of Sponsor | Common or Usual Name: | Electrosurgical Generator | |----------------------------|----------------------------------------------------------------------------------------------------------------------------------| | Trade or Proprietary Name: | en-bloc Biopsy System™ | | Classification Name: | Electrosurgical Cutting & Coagulation Device &<br>Accessories (21 C.F.R. § 878.4400)<br>Biopsy Instrument (21 C.F.R. § 876.1075) | Neothermia Corporation One Apple Hill, Suite 316 Natick, Massachusetts 01760 Phone: (508) 655-7820 Facsimile: (508) 655-6239 #### Predicate Devices Neothermia Corp.'s en-bloc Biopsy System™ #### Intended Use The en-bloc Biopsy System is intended for diagnostic sampling of breast tissue during a breast biopsy procedure. ### Technological Characteristics The en-bloc is a percutaneous high frequency, automated, vacuum-assisted electrosurgical device used to remove tissue by automated electrosurgical cutting and {1}------------------------------------------------ simultaneous capture of an incised tissue volume. The Neothermia en-bloc ™ consists of a hand-held biopsy handle, upon which the single-use en-bloc Biopsy Probe is attached, with an integral cable to connect the handle to the control unit. The Probe™ contains two sets of active electrodes at its distal end - a precursor electrode and cutting/capture electrodes. The shaft of the Probe™ is encased in a stainless steel cannula. An outer plastic sleeve surrounds this stainless steel cannula and an annular gap between the sleeve and the cannula provides a conduit for vacuum-assisted removal of the gaseous products of electrosurgical cutting and any liquids (e.g., blood) that may accumulate at the distal end of the Probe during the biopsy procedure. The vacuum also helps maintain the required cutting arc during automated tissue capture. # Substantial Equivalence The modified cradle has the same intended use, principles of operation, and technological characteristics as the previously cleared predicate devices. The modified device and its predicate devices are both electrosurgical devices used to biopsy breast tissue. The change does not raise new questions of safety or efficacy. The modified cradle is substantially equivalent to Neothermia's cleared en-bloc cradles. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 NOV 1 0 2003 Ms. Sherrie Coval-Goldsmith Vice President, Regulatory Affairs Neothermia Corporation One Apple Hill, Suite 316 Natick, Massachusetts 01760 Re: K033427 Trade/Device Name: en-bloc Biopsy System™ Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology urology biopsy instrument Regulatory Class: II Product Code: KNW Dated: October 27, 2003 Received: October 31, 2003 Dear Ms. Coval-Goldsmith: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ #### Page 2 - Ms. Sherrie Coval-Goldsmith This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Miriam C. Provost Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Number (if known): ___________________________________________________________________________________________________________________________________________________ Device Name: en-bloc Biopsy System™ Indications for Use: The en-bloc Biopsy System is intended for diagnostic sampling of breast tissue during a breast biopsy procedure. (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) CDRH, Office of Device Evaluation (ODE) . ____________________________________________________________________________________________________________________________________________________________________________ Prescription Use _____________________________________________________________________________________________________________________________________________________________ OR Over-The-Counter Use_ (Per 21 C.F.R. 801.109) (Optional Format 1-2-96) Miriam C. Provost (Division Sign-Off) Division of General, Restorative and Neurological Devices 510(k) Number K033427
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...