DEPUY PINNACLE ACETABULAR CUP SYSTEM, ESL MARATHON POLYETHYLENE LINERS, AND DEPUY ULTIMA UNIPOLAR FEMORAL HEADS
K033273 · DePuy Orthopaedics, Inc. · LPH · May 6, 2004 · Orthopedic
Device Facts
Record ID
K033273
Device Name
DEPUY PINNACLE ACETABULAR CUP SYSTEM, ESL MARATHON POLYETHYLENE LINERS, AND DEPUY ULTIMA UNIPOLAR FEMORAL HEADS
Applicant
DePuy Orthopaedics, Inc.
Product Code
LPH · Orthopedic
Decision Date
May 6, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The subject ESL Marathon" Polyethylene Liners are intended to be used with the DePuy Pinnacle® metal acetabular shells, modular femoral heads, unipolar femoral heads, and selfcentering heads to resurface the acetabular socket in cementless total hip arthroplasty. The subject Ultima" Unipolar and the Sclf-Centering" Hip Prosthesis, originally cleared for hemi-arthroplasty procedures, are also intended for use in total hip replacement when used in conjunction with a metal backed UHMWPE bearing surface having an inside diameter corresponding to the outside diameter of the metallic head/cup that is utilized.
Device Story
Modular hip replacement system components; includes Pinnacle® ESL Marathon™ UHMWPE liners, Ultima™ unipolar femoral heads, and Self-Centering™ hip prostheses. Liners lock into titanium alloy (Ti-6Al-4V) acetabular shells; femoral heads (CoCrMo alloy) articulate with liners. Used in total hip arthroplasty to replace damaged hip joint articulation; intended to increase patient mobility and reduce pain. Operated by orthopedic surgeons in clinical/OR settings. System provides bearing surfaces for cementless application; ESL liners feature lateralized neutral or face-changing orientations with Charnley-style bores for stability. Clinical benefit derived from restoring joint function and reducing pain in patients with degenerative or traumatic hip conditions.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on design, material, and manufacturing similarities to legally marketed predicate devices.
Indicated for total hip replacement in patients with sufficient bone stock to support components. Conditions include: osteoarthritis, traumatic/rheumatoid arthritis, congenital hip dysplasia, avascular necrosis, acute traumatic fracture of femoral head/neck, failed previous hip surgery (reconstruction, fixation, arthrodesis, hemiarthroplasty, surface replacement), and ankylosis.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
MAY - 6 2004 510(k) SUMMARY NAME OF FIRM: DePuy Orthopaedics, Inc. 700 Orthopaedic Drive Warsaw, IN 46580 510(K) CONTACT: Natalie S. Heck Manager, Regulatory Affairs DePuy Orthopaedics, Inc. PO Box 988 700 Orthopaedic Drive Warsaw, IN 46581-0988 DePuy Pinnacle® Acetabular Cup System
ESL Marathon™ Polyethylene Liners TRADE NAME: DePuy Ultima™ Unipolar Femoral Heads DePuy Self-Centering™ Hip Prosthesis COMMON NAME: Acetabular Cup Liner Femoral Head CLASSIFICATION: Class II, per 21 CFR, 888.3358 Hip joint metal/polymer/ metal, semi-constrained, porous-coated, uncemented prosthesis DEVICE PRODUCT CODE: 87 LPH SUBSTANTIALLY DePuy Pinnacle® Acetabular Cup System EQUIVALENT DEVICES: Marathon™ Cross-Linked Polyethylene Liners Biomet Tri-polar System DePuy Ultima" Unipolar Femoral Heads Depuy Self-Centering™ Hip Prosthesis
Ko33275
#### DEVICE DESCRIPTION:
The Pinnacle® Acetabular System is part of a modular system designed to replace the natural articular surface of the hip joint in total hip replacement. The acetabular component is provided as two separate units, a porous coated hemispherical outer shell manufactured from titanium alloy (Ti-6Al-4V) and a liner manufactured from cross-linked ultra high molecular weight polyethylene (UHMWPE), which locks into the outer shell. The liner component articulates with a femoral head of an appropriate diameter.
The subject Pinnacle® Enhanced Stability (ESL) Marathon™ liners are cross-linked UHMWPE acetabular cup liners that are available in a lateralized neutral, or lateralized face-changing orientation. The liners have inner diameters (ID) intended for use with modular, unipolar, or self-centering (bipolar) femoral heads within the 28mm-48mm size range. The outer diameters
d
{1}------------------------------------------------
geometrically the same as other Pinnacle Acetabular Cup Liners, in a 44mm-76mm size ra offering. There is an addition of a Charnley-style bore on sizes 36mm-48mm ID to increase stability.
The Ultima" Unipolar femoral head is provided in a size range of 38mm to 63mm OD, in Imm increments. The subject heads have a tapered bore which can receive a variety of adapter sleeves, originally cleared in K965156 (01-24-07) which are machined from CoCrMo alloy in wrought bar. The femoral head is also used in total hip arthroplasty when used in conjunction with a UHMWPE bearing surface having an inside diameter corresponding to the outside diameter of the modular femoral head that is utilized.
The Self-Centering " Hip Prosthesis is a component consisting of a metallic cup and an UHMWPE insert and plastic retaining ring. It is used with a DePuy femoral hip stem and modular metal ball with a head diameter corresponding to the inside diameter of the Self-Centering " Hip polyethylene insert (22mm - 28mm), to replace the femoral head and neck. The Self-Centering Hips are provided in a size range of 39mm through 57mm OD, in 2 mm increments. The bearing inserts are offered in two inner diameters, to accept 22mm and 28mm femoral components. The Self-Centering Hip is being submitted as part of a Total Hip Prosthesis when used in conjunction with a UHMWPE bearing surface having an inside diameter corresponding to the outside diameter of the metallic cup that is utilized.
## INDICATIONS AND INTENDED USE:
#### Intended Use:
The subject ESL Marathon" Polyethylene Liners are intended to be used with the DePuy Pinnacle® metal acetabular shells, modular femoral heads, unipolar femoral heads, and selfcentering heads to resurface the acetabular socket in cementless total hip arthroplasty.
The subject Ultima" " Unipolar and the Sclf-Centering" Hip Prosthesis, originally cleared for hemi-arthroplasty procedures, are also intended for use in total hip replacement when used in conjunction with a metal backed UHMWPE bearing surface having an inside diameter corresponding to the outside diameter of the metallic head/cup that is utilized.
#### Indications:
The Pinnacle" Enhanced Stability Liner is indicated for use in total hip replacement procedures. Total hip arthroplasty is intended to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. Total hip replacement is indicated in the following conditions:
- 1. A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, theumatoid arthritis, or congenital hip dysplasia.
- 2. Avascular necrosis of the femoral head.
- 3. Acute traumatic fracture of the femoral head or neck.
- 4. Failed previous hip surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.
- 5. Certain cases of ankylosis.
The Pinnacle" ESL is indicated for use with the Pinnacle" Acetabular Cup in cementless application.
Image /page/1/Picture/17 description: The image shows a black and white abstract design. The design is roughly circular in shape, with an irregular border. The interior of the circle is filled with a pattern of small, irregular shapes, creating a textured appearance.
{2}------------------------------------------------
Indications for the Ultima™ Unipolar heads and the Self-Centering™ Hip Prostheses
previously cleared for Hemi-Hip Arthroplasty remain unchanged. Additions to the indications
for use include:
Self-Centering™ Hip Prostheses and unipolar femoral heads are also intended to be used for total hip arthroplasty to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components, when used in conjunction with a UHMWPE bearing surface having an inside diameter corresponding to outside diameter of the metallic cup that is utilized. Use in total hip replacement is indicated in the following additional conditions:
1. A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, or congenital hip dysplasia
2. Failed previous hip surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty ,or total hip replacement.
### BASIS OF SUBSTANTIAL EQUIVALENCE:
The DePuy Pinnacle® ESL Marathon Polyethylene Liners (lateralized neutral and lateralized face-changing), Ultima™ Unipolar and Self-Centering devices described in this submission are substantially equivalent to the predicate devices based on similarities in design, intended use, material and manufacturing methods.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract image of an eagle with three stylized lines representing its wings.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY - 6 2004
Ms. Natalie Heck Manager, Regulatory Affairs Depuy Orthopaedics, Inc. P.O. Box 988 700 Orthopaedic Drive Warsaw, Indiana 46581-0988
Received: February 11, 2004
Re: K033273
Trade/Device Name: Pinnacle® Acetabular System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: II Product Code: LPH Dated: February 10, 2004
Dear Ms. Heck:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Ms. Natalie Heck
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours,
Sincerely yours,
Mark A. Millican
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
pge 171
#### 510(k) Premarket Notification
# 510(k) Number (if known) _
Device Name: Pinnacle® Acetabular System
## Indications for Use:
The Pinnacle® Enhanced Stability Liner (ESL) is indicated for use in total hip replacement procedures. Total hip arthroplasty is intended to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. Total hip replacement is indicated in the following conditions:
- 1. A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, or congenital hip dysplasia.
- 2. Avascular necrosis of the femoral head.
- 3. Acute traumatic fracture of the femoral head or neck.
- 4. Failed previous hip surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.
- 5. Certain cases of ankylosis.
The Pinnacle® ESL is indicated for use with the Pinnacle® Acetabular Cup in cementless application.
(cont.)
Concurrence of CDRH, Office of Device Evaluation:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.