MODIFICATION TO FIXION INTRAMEDULLARY NAILING SYSTEM (FIXION IM)

K032588 · Disco-O-Tech Medical Technologies, Ltd. · HSB · Sep 12, 2003 · Orthopedic

Device Facts

Record IDK032588
Device NameMODIFICATION TO FIXION INTRAMEDULLARY NAILING SYSTEM (FIXION IM)
ApplicantDisco-O-Tech Medical Technologies, Ltd.
Product CodeHSB · Orthopedic
Decision DateSep 12, 2003
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3020
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Femur: - fractures in the femur shaft, proximal femoral fractures, and combinations of these fractures: - fractures including stable and unstable pertrochanteric, proximal femoral intertrochanteric, and subtrochanteric (with and without break-off of the minor trochanter), high subtrochanteric fractures and combinations of these fractures: - use in osteotomy, nonunions and malunions, bone reconstruction following tumor resection, grafting and pathological fractures, and revision procedures in the femur: - mid shaft fractures in the femur (5 cm below the surgical neck to 5 cm proximal to the distal end of the medullar canal); - comminuted shaft fractures: - fixation of short distal or proximal fragments Tibia: - diaphyseal shaft fractures in the tibia; - comminuted shaft fractures; - fixation of short distal or proximal fragments Humerus: - diaphyseal shaft fractures in the humerus; - comminuted shaft fractures; - fixation of short distal or proximal fragments

Device Story

Fixion Intramedullary Nailing System consists of stainless steel nails and hip pegs for orthopedic fracture fixation. Device functions as internal fixation rod; inserted into medullary canal to stabilize bone fragments. Used by orthopedic surgeons in clinical/surgical settings. System design utilizes mechanical stabilization of bone segments; provides structural support for healing. Benefits include stabilization of complex fractures, including comminuted and pathological types, and support for bone reconstruction procedures.

Clinical Evidence

Bench testing only.

Technological Characteristics

Stainless steel (ASTM F 138/ISO 5832-1 and ASTM F 1314). Intramedullary fixation rod design. Designed to meet ASTM F 565 standards for care and handling of orthopedic implants.

Indications for Use

Indicated for patients requiring fixation of femur, tibia, or humerus fractures, including shaft, proximal, pertrochanteric, intertrochanteric, subtrochanteric, comminuted, and pathological fractures, as well as osteotomy, nonunions, malunions, bone reconstruction, and revision procedures.

Regulatory Classification

Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ SEP 1 2 2003 Disc-O-Tech Medical Technologies, Ltd. Fixion Intramedullary Nailing Systems 32588 page 1 of 3 # 510(K) Summary # Disc-O-Tech Medical Technologies, Ltd. # Fixion® Intramedullary Nailing System; Fixion® Interlocking Intramedullary Nailing System; Fixion Interlocking Proximal Femoral Intramedullary Nailing System ### Company Name Disc-O-Tech Medical Technologies, Ltd. 3 Hasadnaot St., Herzliya Israel, 46728 ### Submitter's Name and Contact Person Yael Rubin Disc-O-Tech Medical Technologies, Ltd. 3 Hasadnaot St., Herzliya Israel, 46728 Tel: +972 9 9511511 Fax: +972 9 9548939 # Date Prepared August 2003 # Trade/Proprietary Name Fixion® Intramedullary Nailing System; Fixion® Interlocking Intramedullary Nailing System; Fixion® Interlocking Proximal Femoral Intramedullary Nailing System; # Classification Name Intramedullary Fixation Rod 21 CFR § 888.3020 Class II ### Predicate Devices - Fixion Intramedullary Nailing System (K990717, K003215, K003215, K010901, l . K021324, K023437) by Disc-O-Tech. - Fixion Interlocking Intramedullary Nailing System (K002783, K013449, K023437) 2. {1}------------------------------------------------ by Disc-O-Tech. - Fixion Interlocking Proximal Femoral Intramedullary Nailing System (K010988, 3. K012967, K023437) by Disc-O-Tech. - Unreamed Tibial Nail (K914453), by Synthes. 4. - Universal Femoral Nail (K914371), by Synthes. న్. ### Performance Standards The following standards were used: - The Fixion Intramedullary Nailing Systems Nails and Hip Peg are manufactured 1. from Stainless Steel, which meets the requirements of ASTM F 138 (UNS31673)/ ISO 5832-1, and from Stainless Steel, which meets the requirements of ASTM F 1314 (UNS S20910). - The Fixion Intramedullary Nailing Systems are designed to meet the requirements of 2. ASTM F 565 - Standard Practice for Care and Handling of Orthopedic Implants and Instruments. ### Intended Use Femur: - V fractures in the femur shaft, proximal femoral fractures, and combinations of these fractures: - イ fractures including stable and unstable pertrochanteric, proximal femoral intertrochanteric, and subtrochanteric (with and without break-off of the minor trochanter), high subtrochanteric fractures and combinations of these fractures: - V use in osteotomy, nonunions and malunions, bone reconstruction following tumor resection, grafting and pathological fractures, and revision procedures in the femur: - イ mid shaft fractures in the femur (5 cm below the surgical neck to 5 cm proximal to the distal end of the medullar canal); - く comminuted shaft fractures: - イ fixation of short distal or proximal fragments #### Tibia: - > diaphyseal shaft fractures in the tibia; - V comminuted shaft fractures; - > fixation of short distal or proximal fragments {2}------------------------------------------------ #### Humerus: - diaphyseal shaft fractures in the humerus; V - イ comminuted shaft fractures; - イ fixation of short distal or proximal fragments #### Substantial Equivalence The modified Fixion Intramedullary Nailing Systems have the following similarities to those that previously received 510(k) concurrence: - Have the same intended use V - イ Have the same operating principles - イ Incorporate the same design principles - V Incorporate either the same or similar materials - イ Have the same packaging and undergo the same sterilization processes, using the same materials and processes. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image is a black and white circular logo for the Department of Health & Human Services - USA. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the top half of the circle. The bottom half of the circle contains a stylized image of what appears to be an abstract symbol, possibly representing a human figure or a stylized bird. SEP 1 2 2003 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Yael Rubin Director of Regulatory Affairs Disc-O-Tech Medical Technologies, Ltd. 3 Hasadnaot St. Herzliya, Israel 46728 Re: K032588 Trade/Device Names: Fixion® Intramedullary Nailing System, Fixion® Interlocking Intramedullary Nailing System, Fixion® Interlocking Proximal Femoral Inramedullary Nailing System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: II Product Code: HSB Dated: August 20, 2003 Received: August 22, 2003 Dear Mr. Rubin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2 - Mr. Yael Rubin This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours. Muriam C. Provost Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indication for Use 510(K) Number (if known): _ 1030588 Device Name: Fixion® Intramedullary Nailing System Fixion® Interlocking Intramedullary Nailing System Fixion® Interlocking Proximal Femoral Intramedullary Nailing System #### Indication for Use: These devices are indicated for the following: Femur: - fractures in the femur shaft, proximal femoral fractures, and combinations of these । fractures; - proximal femoral fractures including stable and unstable pertrochanteric, intertrochanteric, and subtrochanteric (with and without break-off of the minor trochanter), high subtrochanteric fractures and combinations of these fractures; - use in osteotomy, nonunions and malunions, bone reconstruction following tumor resection, grafting and pathological fractures, and revision procedures in the femur; - mid shaft fractures in the femur (5 cm below the surgical neck to 5 cm proximal to the distal end of the medullar canal); - comminuted shaft fractures; - - fixation of short distal or proximal fragments - #### Tibia: - diaphyseal shaft fractures in the tibia; - - comminuted shaft fractures; - - fixation of short distal or proximal fragments - #### Humerus: - diaphyseal shaft fractures in the humerus; - comminuted shaft fractures: ー - fixation of short distal or proximal fragments - ### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use | X | OR | Over the Counter Use | | |----------------------|---|----|----------------------|--| | (per 21 CFR 801.109) | | | | | Miriam C. Provost (Division Sign-Off) Division of General, Restorative and Neurological Devices | Page 22 of 27 | 510(k) Number | K032588 | |---------------|---------------|---------| |---------------|---------------|---------|
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...