FIXION INTERLOCKING PROXIMAL FEMORAL NAILING SYSTEM
K010988 · Disc-O-Tech Medical Technologies, Ltd. · HSB · Jun 19, 2001 · Orthopedic
Device Facts
Record ID
K010988
Device Name
FIXION INTERLOCKING PROXIMAL FEMORAL NAILING SYSTEM
Applicant
Disc-O-Tech Medical Technologies, Ltd.
Product Code
HSB · Orthopedic
Decision Date
Jun 19, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Fixion Interlocking Proximal Femoral Nailing System ("Fixion PF") is intended for use in fixation of fractures in the femur. The Fixion PF is indicated for use in fractures in the femur shaft, proximal femoral fractures, and combinations of these fractures. Proximal femoral fractures include stable and unstable pertrochanteric, intertrochanteric and subtrochanteric (with and without break-off of the minor trochanter), high subtrochanteric fractures and combinations of these fractures. The long Fixion PF may also be used in mid shaft fractures, 5cm below the surgical neck to 5cm proximal to the distal end of the medullar canal. The Fixion PF is also indicated for use in osteotomy, nonunions and malunions, bone reconstruction following tumor resection, grafting and pathological fractures, revision procedures.
Device Story
Fixion PF is an expandable, non-slotted stainless steel intramedullary nail system used for femoral fracture fixation. System includes nail implant, Femoral Neck Peg, Hip Pin, and manual plastic inflation pump. Surgeon inserts nail into medullary canal; uses pump to inflate nail and pegs with sterile saline; expansion secures reinforcement bars against canal wall. Procedure performed in OR under X-ray guidance. Inflation mechanism provides mechanical fixation equivalent to traditional interlocking nails. Benefits include stable fracture fixation via controlled expansion.
Clinical Evidence
Bench testing only. Mechanical performance validated via 4-point bending (ASTM F1264) and Hip Peg testing (ASTM F384-99). No clinical data presented.
Technological Characteristics
316L Stainless Steel (ASTM F138). Expandable, non-slotted cylindrical tube design with reinforcement bars. Mechanical expansion via saline inflation pump. Cannulated design. Sterilization method not specified.
Indications for Use
Indicated for patients requiring fixation of femoral fractures, including shaft, proximal (pertrochanteric, intertrochanteric, subtrochanteric), and combined fractures. Also indicated for osteotomy, nonunions, malunions, bone reconstruction post-tumor resection, grafting, pathological fractures, and revision procedures.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
{0}------------------------------------------------
K010988
#### JUN 1 9 2001 510(K) Summary
# Disc-O-Tech Medical Technologies Ltd. Fixion Interlocking Proximal Femoral Nailing System
#### Company Name
Disc-O-Tech Medical Technologies, Ltd. 3 Hasadnaot St., Herzelia Israel, 46728
### Submitter's Name and Contact Person
Jonathan S. Kahan, Esq. 555 Thirteenth Street, NW Washington, DC 20004 Tel: 202-637-5794 Fax: 202-637-5910
### Date Prepared
March, 2001
#### Trade/Proprietary Name
Fixion The Interlocking Proximal Femoral Nailing System
#### Classification Name
Intramedullary Fixation Rod 21 C.F.R. § 888.3020 Class II
#### Predicate Devices
- 1. Fixion Interlocking Nailing System (K002783) by Disc-O-Tech.
- 2. Proximal Femoral Nail (K970097) by Synthes.
#### Performance Standards
The following standards were used:
- 1. The Interlocking Nail is manufactured from 316L Stainless Steel, which meets the requirements of ASTM F138 - Standard Specification for Stainless steel Bar and Wire for Surgical Implants.
- 2. The 4 point bending mechanical testing was performed according to ASTM F1264 - Standard for Mechanical Performance Considerations for Intramedullary Fixation Devices.
- 3. The Hip Peg testing was performed according to ASTM F384-99 Standard Specification for Metallic Angled Orthopedic Fracture Fixation Devices.
#### Intended Use
The Fixion Interlocking Proximal Femoral Nailing System ("Fixion PF") is intended for use in fixation of fractures in the femur. It is indicated for use in fractures in the femur
{1}------------------------------------------------
shaft and proximal fractures and combinations of these fractures. Proximal femoral fractures include stable and unstable pertrochanteric and subtrochanteric (with and without break-off of the minor trochanter), high subtrochanteric fractures and combinations of these fractures. Shaft fractures include those that are 5cm below the surgical neck to 5cm proximal to the distal end of the medullar canal.
The Fixion PF is also indicated for use in osteotomy, nonunions and malunions, bone reconstruction following tumor resection, grafting and pathological fractures, and revision procedures.
### System Description
The Fixion Interlocking Proximal Femoral Nailing System is a single use system that consists of the following components:
1. The Nail implant is an expandable non-slotted stainless steel cylindrical tube, with a cap protected, female threaded proximal end with holes for Femoral Neck Peg and Hip Pin, to fix fractures in the proximal femur.
2. The Inflation device is a single-use manual plastic pump that is filled with sterile inflation liquid an used to inflate the nail implant and Femoral Neck Peg.
Once the nail and the Femoral Neck Peg and hip pin are positioned within the medullary canal and femur neck, rotation of the "pump" handle allows for screw and nail diameter increase to their intended diameter under X-ray and controlled pressure.
#### Substantial Equivalence
The Fixion Interlocking Proximal Femoral Nailing System Nail has substantially equivalent intended use and indications as the Fixion™ Interlocking Nail and the Synthes Proximal Femoral Nail, i.e., fixation of fractures of the femur.
The performance characteristics of the Fixion Proximl Femoral Nail have been tested and found to meet the specifications through a series of bench tests.
The Fixion Interlocking Proximal Femoral Nail, like the Fixion TM Interlocking Nail, is made of 316L Stainless Steel and has cannulated design. The cross section of the Fixion Interlocking Proximal Femoral and the Fixion Interlocking nails is circular with reinforcement bars.
Fixation of the Fixion Interlocking Proximal Femoral and Interlocking nails is achieved by inflation and results in the attachment of the 4 reinforcement bars to the medullary canal wall. The addition of Femoral Neck Peg and Hip Pin provides equivalent fixation to the end of the nail, in a manner that is substantially equivalent to that of the Synthes Proximal Femoral Nail. The inflation of the Fixion Interlocking Proximal Femoral nail with saline, which is a non-compressible biocompatible fluid, is identical to the cleared Fixion Interlocking nail and does not raise any new safety and efficacy issues.
{2}------------------------------------------------
# Premarket Notification Truthful and Accurate Statement* (As required by 21 CFR 807.87(j))
I certify that, in my capacity as the General Manager of Disc-O-Tech Medical Technologies, Ltd., I believe to the best of my knowledge, that all data and information submitted in the premarket notification for the Fixion Interlocking Nailing System are truthful and accurate and that no material fact has been omitted.
Art. K
```
Signature
Elad Magal - General Manager Typed Name and Title
Disc-O-Tech Medical Technologies, Ltd. Company
03 / 28 / 01
Date
Premarket Notification 510(k) Number
* Must be signed by a responsible person of the firm required to submit the premarket notification (e.g., not a consultant for the 510(k) submitter)
Page 31 of 39
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, composed of three curved lines that suggest a head, body, and legs. The figure is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
# JUN 1 9 2001
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Disc-O-Tech Medical Technologies, LTD. c/o Mr. Johnathan S. Kahan, Esq. 555 Thirteenth Street, NW Washington, DC 20004
Re: K010988
Trade Name: Fixion Interlocking Proximal Femoral Nailing System Regulation Number: 888.3020 Regulatory Class: II Product Codes: HSB Dated: March 28, 2001 Received: April 2, 2001
Dear Mr. Kahan:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general control provisions of the Act. The general control provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{4}------------------------------------------------
# Page 2 - Mr. Johnathan S. Kahan, Esq.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Psmt dweelllwa kon
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indication for Use
510(K) Number (if known): ____________________________________________________________________________________________________________________________________________________
Fixion Interlocking Proximal Femoral Nailing System Device Name:
Indication for Use:
The Fixion Interlocking Proximal Femoral Nailing System ("Fixion PF") is intended for use in fixation of fractures in the femur. The Fixion PF is indicated for use in fractures in the femur shaft, proximal femoral fractures, and combinations of these fractures. Proximal femoral fractures include stable and unstable pertrochanteric, intertrochanteric and subtrochanteric (with and without break-off of the minor trochanter), high subtrochanteric fractures and combinations of these fractures. The long Fixion PF may also be used in mid shaft fractures, 5cm below the surgical neck to 5cm proximal to the distal end of the medullar canal.
The Fixion PF is also indicated for use in osteotomy, nonunions and malunions, bone reconstruction following tumor resection, grafting and pathological fractures, revision procedures.
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use<br>(per 21 CFR 801.109) | X |
|------------------------------------------|--------------------------------------------------------------------------------------------|
| OR Counter Use | |
| | <div style="text-decoration: overline;">Dsmita elllrus An chu</div><br>(Division Sign-Off) |
| Page 32 of 39 | Division of General, Restorative<br>and Neurological Devices |
| | 510(k) Number K010988 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.