ATTAIN 6226DEF DEFLECTABLE CATHETER DELIVERY SYSTEM
Applicant
Medtronic Vascular
Product Code
DQY · Cardiovascular
Decision Date
Oct 21, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The deflectable catheter delivery system is indicated to provide a pathway through which diagnostic and therapeutic transvenous devices are introduced within the chambers and coronary vasculature of the heart, and for introducing balloon catheters into the coronary sinus or leads into vessels of the left heart via the coronary sinus.
Device Story
Attain 6226DEF is a percutaneous delivery system for transvenous diagnostic/therapeutic devices. System components include needle, syringe, guide wire, valve, deflectable catheter, dilator, and slitter. Used in clinical settings by physicians to access heart chambers and coronary vasculature. Physician inserts needle/guide wire to establish venous access; deflectable catheter and dilator are introduced to navigate to target site; valve minimizes blood loss. Catheter tip deflection allows precise positioning for lead or balloon catheter placement. Once device is deployed, slitter facilitates catheter removal. System modifications from predicate 10600 include reduced catheter/dilator length, softer catheter tip, altered articulation shape, updated valve, and universal slitter. Benefits include improved navigation and access to left-heart vessels via coronary sinus.
Clinical Evidence
Bench testing only. In vitro performance testing and biocompatibility/sterilization validation were conducted. All materials comply with ISO 10993. No clinical data presented.
Indicated for patients requiring transvenous access to heart chambers, coronary vasculature, or coronary sinus for diagnostic/therapeutic device placement, including balloon catheters or left-heart leads.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Medtronic 10600 Deflectable catheter system (K013517)
Medtronic 6228SLT Universal Slitter (Class I exempt)
Medtronic Attain Access 6218 Left-Heart Delivery System (K012083)
Submission Summary (Full Text)
{0}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the department's logo, which features three stylized human profiles facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the perimeter of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
2003 NOV 3
Medtronic, Inc. c/o Ms. Kristina Mollner Sr. Regulatory Affairs Specialist Cardiac Rhythm Management 7000 Central Avenue, NE Minneapolis, MN 55432
Re: K032312
Trade Name: ATTAIN™ 6226DEF Deflectable Catheter Delivery System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II (two) Product Code: DOY Dated: September 19, 2003 Received: September 22, 2003
Dear Ms. Mollner:
This letter corrects our substantially equivalent letter of October 21, 2003 regarding the designation as class III (three) in the letter. The correct class for the device is class II (two).
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the [kit/tray] have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.
{1}------------------------------------------------
#### Page 2 - Ms. Kristina Mollner
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration (21 CFR Part 807); listing (21 CFR Part 807), labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on the labeling regulation, please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
O. Malla
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
OCT 2 1 2003
# 510(K) Summary of Substantial Equivalence
| Date prepared | 25 July, 2003 |
|---------------------------|-------------------------------------------------------------------------------------------------------------|
| Submitter: | Medtronic Ireland<br>Parkmore Business Park West<br>Galway<br>Ireland |
| Submission Correspondent: | Kristina Mollner<br>Sr. Regulatory Affairs Specialist<br>7000 Central Avenue, N.E.<br>Minneapolis, MN 55432 |
| Telephone: | (763) 514-3914 |
| Fax: | (763) 514-6424 |
| E-Mail: | kristina.mollner@medtronic.com |
| Proprietary Name: | Attain™ 6226DEF Deflectable Catheter Delivery<br>System |
| Common Name: | Catheter, Percutaneous |
| Device Classification: | Class II, 21 CFR § 870.1250 |
| Product Code: | 74 DQY |
# Performance Standard
Performance standards do not currently exist for these devices. None established under Section 514.
{3}------------------------------------------------
# Device Description
The 6226DEF system features a percutaneous needle and syringe to access the venous insertion site, a guide wire to access the vein, a valve to reduce blood loss during the implant procedure, a deflectable catheter to introduce a transvenous device, a deflectable catheter dilator to facilitate deflectable catheter passage, and a slitter to remove the deflectable catheter.
The Attain 6226DEF Deflectable Catheter Delivery System combines devices that are either pre-amendment devices, devices that are cleared for market distribution via 510(k), or are exempt from premarket notification because of Class I designation.
The key changes being made to the system versus the predicate 10600 device are a decrease in the length of the deflectable catheter, a reduction in the stiffness of the deflectable catheter tip, and a change in the articulated shape of the catheter. The dilator length is also decreased in line with the catheter. In addition the slitters are replaced by a Universal Slitter and the adjustable hemostasis valve is replaced by an introducer valve. The entire system is packaged in a revised packaging configuration.
# Indications for Use
The deflectable catheter delivery system is indicated to provide a pathway through which diagnostic and therapeutic transvenous devices are introduced within the chambers and coronary vasculature of the heart, and for introducing balloon catheters into the coronary sinus or leads into vessels of the left heart via the coronary sinus.
{4}------------------------------------------------
# Substantially Equivalent Devices
#### Attain 6226DEF Deflectable catheter delivery system predicate devices
The design, technology, features, functions and intended use of the 6226DEF Deflectable Catheter Delivery System are substantially equivalent to the following predicate devices currently in interstate commerce.
| Attain 6226DEF<br>Deflectable<br>Catheter Delivery<br>System | Predicate Device | Predicate Device<br>Manufacturer | Predicate 510(k) |
|--------------------------------------------------------------|------------------------------------------------|-------------------------------------------------------|----------------------------------------------------------|
| Deflectable<br>catheter | Medtronic 10600<br>Deflectable catheter system | Medtronic, Inc.<br>Minneapolis, MN 55432 | K013517 |
| Guide Catheter<br>Dilator | Medtronic 10600<br>Deflectable catheter system | Medtronic, Inc.<br>Minneapolis, MN 55432 | K013517 |
| Guide Wire | Medtronic 10600<br>Deflectable catheter system | Medtronic, Inc.<br>Minneapolis, MN 55432 | K013517 |
| Introducer valve | Cook 6228VAL Introducer<br>Valve | Cook Vascular Incorporated<br>Leechburg, PA 15656 USA | K010128 |
| Universal Slitter | Medtronic 6228SLT<br>Universal Slitter | Medtronic, Inc.<br>Minneapolis, MN 55432 | Class I device, exempt<br>from premarket<br>notification |
| Needle | Medtronic 10600<br>Deflectable catheter system | Medtronic, Inc.<br>Minneapolis, MN 55432 | K013517 |
| Syringe | Medtronic 10600<br>Deflectable catheter system | Medtronic, Inc.<br>Minneapolis, MN 55432 | K013517 |
Additionally, the biocompatibility of the Attain 6226DEF Deflectable Catheter Delivery System is substantially equivalent to the following device.
| Attain 6226DEF<br>Deflectable<br>Catheter Delivery<br>System | Predicate Device | Predicate Device<br>Manufacturer | Predicate 510(k) |
|--------------------------------------------------------------|---------------------------------------------------------------|------------------------------------------|------------------|
| Deflectable<br>catheter | Medtronic Attain Access<br>6218 Left-Heart Delivery<br>System | Medtronic, Inc.<br>Minneapolis, MN 55432 | K012083 |
{5}------------------------------------------------
#### CONFIDENTIAL
May not be reproduced without written permission from Medtronic, Inc.
## Summary of Studies
In vitro testing and biocompatibility/sterilization testing was performed to support substantial equivalence to the predicate device. The results of this testing indicates that the Attain 6226DEF meets all of its design and performance requirements.
## Biocompatibility Information
Materials/components of the Attain 6226DEF kit were evaluated for compliance to ISO10993. All materials were found to be biocompatible and in compliance to ISO10993.
### Sterilization Validation
The Attain 6226DEF Deflectable catheter delivery system is sterilized using a 100% Ethylene Oxide (ETO) sterilization process. Appropriate processes for sterilizing the devices were validated.
## Conclusion (Statement of Equivalence)
Through the data and information presented, as well as similarities to legally marketed devices, Medtronic, Inc. considers the Attain 6226DEF deflectable catheter delivery system to be substantially equivalent to the previously discussed legally marketed predicate devices.
{6}------------------------------------------------
#### CONFIDENTIAL May not be reproduced without written permission from Medtronic, Inc.
# Indications for Use Statement
| 510(k) Number (if known): | N/A |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Attain™ 6226DEF Deflectable catheter delivery system |
| Indications For Use: | The deflectable catheter delivery system is indicated to provide a pathway through which diagnostic and therapeutic transvenous devices are introduced within the chambers and coronary vasculature of the heart, and for introducing balloon catheters into the coronary sinus or leads into vessels of the left heart via the coronary sinus. |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use<br>(Per 21 CFR 801.109) | X |
|------------------------------------------|---|
|------------------------------------------|---|
OR
| (Division Sign-Off) | |
|------------------------------------|--|
| Division of Cardiovascular Devices | |
| Over-The-Counter Use | |
|----------------------|--|
|----------------------|--|
| 510(k) Number | K032312 |
|---------------|---------|
|---------------|---------|
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Special 510(k): Device Modification Premarket Notification Attain™ 6226DEF Deflectable Catheter Delivery System
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.