The deflectable catheter system is indicated to provide a pathway through which diagnostic and therapeutic transvenous devices are introduced within the chambers and coronary vasculature of the heart, and for introducing balloon catheters into the coronary sinus or leads into vessels of the left heart via the coronary sinus.
Device Story
System provides percutaneous access pathway for cardiac procedures; includes needle, syringe, guide wire, adjustable hemostasis valve, deflectable catheter, dilator, and slitters. Used in clinical settings by physicians to facilitate introduction of diagnostic/therapeutic devices into heart chambers, coronary vasculature, or coronary sinus. Deflectable mechanism allows navigation within vasculature. Single-use, sterile device. Output is physical access pathway; enables delivery of secondary devices (leads/balloons) to target cardiac sites. Benefits include facilitated navigation and reduced blood loss during implantation.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included environmental conditioning, mechanical integrity, compatibility, and package/shelf-life validation. All results met specified requirements. Biocompatibility testing performed on all blood-contacting materials per ISO 10993-1.
Technological Characteristics
Percutaneous catheter system; includes deflectable catheter, dilator, hemostasis valve, guide wire, and slitters. Materials are biocompatible per ISO 10993-1. Sterilization via 100% Ethylene Oxide (EtO). Mechanical design facilitates navigation in coronary vasculature.
Indications for Use
Indicated for patients requiring transvenous access to heart chambers, coronary vasculature, or coronary sinus for the introduction of diagnostic/therapeutic devices, balloon catheters, or left-heart leads.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Model 6216 Medtronic Attain LDS Left-heart delivery system (K012130)
Model 6218 Medtronic Attain Access Left-heart delivery system (K012083)
MedAmicus, Inc. Percutaneous Venous Introducer (K965167)
Submission Summary (Full Text)
{0}------------------------------------------------
510(k) Premarket Notification Attachment C
Model 10600 Deflectable Catheter System
K
## 510(k) Summary of Substantial Equivalence
| Date Prepared: | October 19, 2001 | |
|---------------------------|---------------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| Submitter: | Medtronic, Inc.<br>7000 Central Ave. N.E.<br>Minneapolis, MN 55432 | |
| Contact: | Kristy K. Mollner, RAC<br>Product Regulation Manager<br>Telephone:<br>Facsimile:<br>E-mail: | (763) 514-4189<br>(763) 514-6424<br>kristina.mollner@medtronic.com |
| Trade / Proprietary Name: | Medtronic Model 10600 Deflectable Catheter System | |
| Common Name: | Catheter, Percutaneous | |
| Device Classification: | Class II, 21 CFR 870.1250 | |
| Product Code: | 74 DQY | |
#### Device Description
The Medtronic Model 10600 Deflectable Catheter System features a percutaneous needle and syringe to access the venous insertion site, a guide wire to access the vein, an adjustable hemostasis valve to reduce blood loss during the implant procedure, a deflectable catheter to introduce a transvenous device, a deflectable catheter dilator to facilitate deflectable catheter passage, and slitters to remove the deflectable catheter.
The Model 10600 Deflectable Catheter System is provided STERILE and is intended for single use only.
The Model 10600 Deflectable Catheter System combines devices that are either cleared for market distribution via 510(k) or are exempt from Premarket notification because of Class I designation.
#### Indications for Use
The Medtronic Model 10600 Deflectable Catheter System is indicated to provide a pathway through which diagnostic and therapeutic devices are introduced within the chambers and coronary vasculature of the heart, and for introducing balloon catheters into the coronary sinus or leads into vessels of the left heart via the coronary sinus.
{1}------------------------------------------------
510(k) Premarket Notification Attachment C
# Substantially Equivalent Devices
# Medtronic Model 10600 Deflectable Catheter System Predicate Devices
| Medtronic Model 10600<br>Deflectable Catheter<br>System | Predicate Device | Predicate Device<br>Manufacturer | 510(k) |
|------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|-----------------------------------------------------------------|---------|
| Guide Catheter | Model 6216 Medtronic Attain<br>LDS Left-heart delivery system | Medtronic, Inc.<br>Minneapolis, MN 55432 | K012130 |
| | Model 6218 Medtronic Attain<br>Access Left-heart delivery<br>system | Medtronic, Inc.<br>Minneapolis, MN 55432 | K012083 |
| | GCIII (Vector, Vector X)<br>Coronary Guiding Catheter | Medtronic Interventional<br>Vascular (MIV)<br>Danvers, MA 01923 | K950179 |
| | Cardima Naviport Deflectable<br>Guiding Catheter | Cardima<br>Fremont, CA 94538 | K974683 |
| | SafeSheath CSG (Same as<br>SafeSheath MSP) | Thomas Medical<br>Malvern, PA 19355 | K003731 |
| | Marinr Series EP Diagnostic<br>Catheters | Medtronic CardioRhythm<br>Minneapolis, MN 55432 | K931794 |
| Guide Catheter Dilator<br>Adjustable Hemostasis<br>Valve<br>Guide Wire<br>Guide Catheter Slitters<br>Guide Catheter<br>Percutaneous Introducer | Model 6216 Medtronic Attain<br>LDS Left-heart delivery system | Medtronic, Inc.<br>Minneapolis, MN 55432 | K012130 |
| | Model 6218 Medtronic Attain<br>Access Left-heart delivery<br>system | Medtronic, Inc.<br>Minneapolis, MN 55432 | K012083 |
#### Additionally, the sterilization process and biocompatibility of the Medtronic 10600 Deflectable Catheter System is substantially equivalent to the following devices:
| MedAmicus, Inc. Percutaneous Introducer<br>(also known as Solotrak) | MedAmicus, Inc.<br>Plymouth, MN 55447 | K893658 |
|---------------------------------------------------------------------|------------------------------------------|---------|
| MedAmicus, Inc. Percutaneous Venous<br>Introducer | MedAmicus, Inc.<br>Plymouth, MN 55447 | K965167 |
| Medtronic Attain LDS Model 6216 Left-<br>heart delivery system | Medtronic, Inc.<br>Minneapolis, MN 55432 | K012130 |
| Medtronic Attain Access Model 6218<br>Left-heart delivery system | Medtronic, Inc.<br>Minneapolis, MN 55432 | K012083 |
Ironic confiderial
..............................................................................................................................................................................
{2}------------------------------------------------
510(k) Premarket Notification Attachment C
### Summary of Studies
Device integrity testing was conducted to support that the Medtronic Model 10600 Deflectable Catheter System is equivalent to the predicate devices. Testing included: Environmental Conditioning, Mechanical Testing, Compatibility Testing, Package/Shelf Life Testing. All device test results for the Medtronic Model 10600 Deflectable Catheter System met specified requirements.
#### Biocompatibility Information
Biocompatibility testing was conducted on all blood contacting materials. The testing performed on this product was done in accordance with ISO 10993-1 and results demonstrate that all materials were found to be biocompatible.
#### Sterilization Validation
The Medtronic Model 10600 Deflectable Catheter System is sterilized using a validated 100% Ethylene Oxide (EtO) sterilization process.
#### Conclusion
Through the data and information presented, numerous similarities support a determination of substantial equivalence and therefore market clearance of the Medtronic Model 10600 Deflectable Catheter System through this 510(k) Pre-market Notification.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus, with three lines curving upwards and a wavy line at the bottom.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 2 5 2002
Ms. Kristy K. Mollner, RAC Product Regulation Manager Cardiac Rhythm Management Medtronic, Inc. 7000 Central Avenue NE Minneapolis, MN 55432-3576
Re: K013517
Trade Name: Model 10600 Deflectable Catheter System Regulation Number: 21 CFR 870.1250 Regulation Name: Catheter, Percutaneous Regulatory Class: Class II (two) Product Code: DQY Dated: February 1, 2002 Received: February 4, 2002
Dear Ms. Mollner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Ms. Kristy K. Mollner, RAC
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Sincerely, years,
Donna-Bea Tillman, Ph.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Premarket Notification Indications for Use
Model 10600 Deflectable Catheter System K
# INDICATIONS FOR USE
510(k) Number (if known):
N/A K013517
Device Name:
Indications For Use:
Model 10600 Deflectable Catheter System
The deflectable catheter system is indicated to provide a pathway through which diagnostic and therapeutic transvenous devices are introduced within the chambers and coronary vasculature of the heart, and for introducing balloon catheters into the coronary sinus or leads into vessels of the left heart via the coronary sinus.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use_
(Optional Format 1-2-96)
Division of Cardiovascular & Respiratory Devices
510(k) Number K013517
editronic confideral
10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.