K032083 · Nuvasive, Inc. · ETN · Oct 24, 2003 · Ear, Nose, Throat
Device Facts
Record ID
K032083
Device Name
NUVASIVE NEUROVISION JJB SYSTEM
Applicant
Nuvasive, Inc.
Product Code
ETN · Ear, Nose, Throat
Decision Date
Oct 24, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.1820
Device Class
Class 2
Indications for Use
The NeuroVision JJB System is intended to provide stimulation of spinal nerves during surgery of the spine, and intraoperative electromyographic (EMG) nerve surveillance to assist, by administration of brief electrical stimulus pulses to tissues and nerves at the operative site, and EMG monitoring of muscle groups associated with those nerves, in the following: a) Location and evaluation of spinal nerves, using the System's Screw Test, Detection, and/or Free Run EMG Functions; and b) Location and identification of spinal nerves, using the System's Nerve Retraction Function. The NeuroVision JJB System is designed for use in conjunction with other NuVasive devices to assist in gaining controlled access to, and visualization of, the spinal nerve root, foramina, intervertebral disc, and surrounding tissues of the spine via uniportal or biportal approach, where anatomical restrictions safely permit.
Device Story
NeuroVision JJB System provides intraoperative nerve monitoring during spinal surgery; utilizes electrical stimulus pulses delivered to tissues/nerves via conductive probes; monitors muscle group responses via surface electrodes. System comprises patient module, control unit with embedded computer/touch screen, and various probes/electrodes. Used by surgeons in OR to locate/identify spinal nerves, nerve roots, and surrounding tissues. Provides real-time EMG surveillance; assists in gaining controlled access to spinal structures. Helps reduce incidence of nerve root injury; aids in surgical decision-making regarding nerve proximity and retraction.
Clinical Evidence
No clinical data provided; device relies on established principles of intraoperative neurophysiologic monitoring.
Technological Characteristics
System includes reusable patient module, control unit with embedded computer, and disposable/reusable conductive probes/electrodes. Operates via electrical nerve stimulation and EMG monitoring. Software-based control unit with touch screen interface.
Indications for Use
Indicated for patients undergoing spinal surgery requiring intraoperative nerve surveillance and stimulation. Used to assist in the location, evaluation, and identification of spinal nerves and associated muscle groups during spinal procedures.
Regulatory Classification
Identification
A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.
{0}------------------------------------------------
## VII. 510(k) Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular §807.92, the following summary of information is provided:
#### A. Submitted by:
Laetitia Bernard Manager of Regulatory Affairs and Quality Assurance NuVasive, Incorporated 10065 Old Grove Road San Diego, California 92131 Telephone: (858) 271-7070 Telefacsimile: (858) 271-7101
#### B. Device Name
| Trade or Proprietary Name: | NuVasive NeuroVision JJB System |
|----------------------------|-------------------------------------------|
| Common or Usual Name: | Electromyography (EMG) monitor/stimulator |
| Classification Name: | Surgical nerve stimulator/locator |
| Device Class: | Class II |
| Classification: | §874.1820, §882.1870 |
| Product Code: | 77ETN, 84GWF |
#### C. Predicate Devices
The subject NeuroVision JJB System is substantially equivalent to the NeuroVision JJB System currently manufactured and distributed commercially in the U.S. by NuVasive.
#### D. Device Description
The NVJJB System utilizes conventional neurophysiogic monitoring to reduce the incidence of injury to nerve roots during instrumented spine surgery. In the procedure, stimulus evoked electromyography is used to determine changes in nerves. Corresponding muscle groups are monitored using surface electrodes, while stimulation is used to detect nerve responses.
{1}------------------------------------------------
# NuVasive®, Inc.
The NeuroVision JJB System consists of a reusable Patient Module, a Control Unit comprised of an embedded computer with touch screen controls and an interface card, and an assortment of disposable and reusable conductive probes, electrodes, and electrode leads.
#### E. Intended Use
The NeuroVision JJB System is intended to provide stimulation of spinal nerves during surgery of the spine, and intraoperative electromyographic (EMG) nerve surveillance to assist, by administration of brief electrical stimulus pulses to tissues and nerves at the operative site, and EMG monitoring of muscle groups associated with those nerves, in the following: a) Location and evaluation of spinal nerves, using the System's Screw Test, Detection, and/or Free Run EMG Functions; and b) Location and identification of spinal nerves, using the System's Nerve Retraction Function.
The NeuroVision JJB System is designed for use in conjunction with other NuVasive devices to assist in gaining controlled access to, and visualization of, the spinal nerve root, foramina, intervertebral disc, and surrounding tissues of the spine via uniportal or biportal approach, where anatomical restrictions safely permit.
#### F. Comparison to Predicate Devices
The subject device has indications for use identical to those of its predicate, and employs the same principles of operation. Due to this equivalency, the device raises no new safety or effectiveness issues.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows a logo with a stylized bird inside a circle. The bird is composed of three curved lines that suggest movement or flight. The circle surrounding the bird is not a solid line but appears to be made up of small, indistinct shapes, possibly letters or symbols. The overall design is simple and abstract, focusing on conveying a sense of freedom or progress.
### Public Health Service
OCT 2 4 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Laetitia Bernard Manager of Regulatory Affairs and Quality Assurance Nuvasive , Inc. 10065 Old Grove Road San Diego, California 92131
Re: K032083
Trade/Device Name: NeuroVision JJB System Regulation Number: 21 CFR 874.1820, 890.1375 Regulation Name: Surgical nerve stimulator/locator Diagnostic electromyograph Regulatory Class: II Product Code: ETN, GWP
Dated: September 24, 2003 Received: September 25, 2003
Dear Ms. Bernard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA mav publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Laetitia Bernard
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Sincerely yours,
Mark N Millburn
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### D. Indications for Use Statement
510(k) Number (if known): Ko320 83
Device Name: NeuroVision JJB System
Indications for Use:
The NeuroVision JJB System is intended to provide stimulation of spinal nerves during surgery of the spine, and intraoperative electromyographic (EMG) nerve surveillance to assist, by administration of brief electrical stimulus pulses to tissues and nerves at the operative site, and EMG monitoring of muscle groups associated with those nerves, in the following: a) Location and evaluation of spinal nerves, using the System's Screw Test, Detection, and/or Free Run EMG Functions; and b) Location and identification of spinal nerves, using the System's Nerve Retraction Function.
The Neuro Vision JJB System is designed for use in conjunction with other NuVasive devices to assist in gaining controlled access to, and visualization of, the spinal nerve root, foramina, intervertebral disc, and surrounding tissues of the spine via uniportal or biportal approach, where anatomical restrictions safely permit.
Mark A. Milliken
Jivision Sign-Off Division of General, Restorative and Neurological Devices
KO32093 10(k) Number -
Image /page/4/Picture/11 description: The image shows the text "Concurrence of CDRH, Office of Device Evaluation (ODE)" at the top. Below this text, there is an arrow pointing to the text "Prescription Use (Per 21 CFR 801.109)". The word "OR" is in the middle of the image, and the text "Over-The-Counter Use" is on the right side of the image.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.