XPECT CRYPTOSPORIDIUM LATERAL FLOW ASSAY, MODEL 2451020

K031965 · Remel, Inc. · MHJ · Nov 14, 2003 · Microbiology

Device Facts

Record IDK031965
Device NameXPECT CRYPTOSPORIDIUM LATERAL FLOW ASSAY, MODEL 2451020
ApplicantRemel, Inc.
Product CodeMHJ · Microbiology
Decision DateNov 14, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3220
Device ClassClass 2

Indications for Use

Xpect™ Cryptosporidium kit is an in vitro qualitative immunoassay for the detection of Cryptosporidium antigens in preserved and unpreserved fecal specimens. This test is intended as an aid in the laboratory diagnosis of suspected Cryptosporidium infections.

Device Story

Xpect™ Cryptosporidium Lateral Flow Assay is a qualitative immunochromatographic test for detecting Cryptosporidium antigens in human fecal specimens. The device consists of a Mylar-backed nitrocellulose membrane striped with rabbit anti-Cryptosporidium and goat anti-mouse IgG. The user mixes the fecal specimen with a buffered diluent and a conjugate containing colored polystyrene microparticles linked to murine monoclonal anti-Cryptosporidium antibodies. This mixture is dispensed into the device sample well and wicks across the membrane. If Cryptosporidium antigen is present, it forms an immune complex with the conjugate and is captured at the test line, producing a visible red/pink line. A control line confirms proper test function. The assay is performed in a laboratory setting by trained personnel. Results provide qualitative information to assist clinicians in diagnosing Cryptosporidium infections, potentially facilitating timely patient management.

Clinical Evidence

Clinical performance was evaluated across six laboratories using 578 routine fecal specimens compared to microscopy. The study reported a sensitivity of 96.4% (95% CI: 91.2-98.6%) and a specificity of 98.3% (95% CI: 96.6-99.1%). Reproducibility was confirmed across seven sites using blinded samples. Analytical sensitivity (limit of detection) was 6 organisms per 0.1 ml. Cross-reactivity testing showed no interference from a wide range of common enteric pathogens, though Imodium® A-D was found to interfere with low-level antigen detection.

Technological Characteristics

Qualitative immunochromatographic lateral flow assay. Components: Mylar-backed nitrocellulose membrane, colored microparticles conjugated with monoclonal anti-Cryptosporidium antibodies, and rabbit anti-Cryptosporidium capture antibodies. Specimen type: human stool (preserved or unpreserved). Manual test format; no external energy source required. Visual readout.

Indications for Use

Indicated for the qualitative detection of Cryptosporidium antigens in preserved and unpreserved human fecal specimens to aid in the laboratory diagnosis of suspected Cryptosporidium infections.

Regulatory Classification

Identification

Entamoeba histolytica serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Entamoeba histolytica in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identify Entamoeba histolytica directly from clinical specimens. The identification aids in the diagnosis of amebiasis caused by the microscopic protozoan parasite Entamoeba histolytica and provides epidemiological information on diseases caused by this parasite. The parasite may invade the skin, liver, intestines, lungs, and diaphragm, causing disease conditions such as indolent ulcers, an amebic hepatitis, amebic dysentery, and pulmonary lesions.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.

Predicate Devices

Submission Summary (Full Text)

{0} # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: K031965 B. Analyte: Cryptosporidium antigens C. Type of Test: Chromatographic immunoassay D. Applicant: Remel Inc. E. Proprietary and Established Names: Xpect™ Cryptosporidium Lateral Flow Assay F. Regulatory Information: 1. Regulation section: 21 CFR Part 866.3220 Entamoeba histolytica serological reagents 2. Classification: Class II 3. Product Code: MHJ – Cryptosporidium spp. 4. Panel: 83 (Microbiology) G. Intended Use: 1. Intended use(s): Xpect™ Cryptosporidium kit is an in vitro qualitative immunoassay for the detection of Cryptosporidium antigens in preserved and unpreserved fecal specimens. This test is intended as an aid in the laboratory diagnosis of suspected Cryptosporidium infections. 2. Indication(s) for use: Xpect™ Cryptosporidium kit is an in vitro qualitative immunoassay for the detection of Cryptosporidium antigens in preserved and unpreserved fecal specimens. This test is intended as an aid in the laboratory diagnosis of suspected Cryptosporidium infections. 3. Special condition for use statement(s): Not applicable 4. Special instrument Requirements: Not applicable H. Device Description: The kit contains 20 test devices consisting of a membrane striped with rabbit anti-Cryptosporidium, and goat anti-mouse IgG. A mixture of specimen, buffered diluent, and conjugate containing colored micro-particles linked to murine monoclonal antibody specific for Cryptosporidium, is dispensed into the sample well of the device. The test sample wicks along a membrane containing capture antibody stripes. The Cryptosporidium immune complex, if present, reacts with {1} Page 2 of 5 anti-Cryptosporidium antibody at the test line. Conjugate not bound at the test line is later captured at the control line containing anti-mouse antibody. A red line of any intensity will appear at the Cryptosporidium test position if Cryptosporidium antigen is present. A line in the Control position indicates that the test is working properly. The device also contains a procedure card; disposable transfer pipettes, dilution tubes, and instructions for use. I. Substantial Equivalence Information: 1. Predicate device name(s): BD ColorPAC Giardia/Crypto Rapid Assay 2. Predicate K number(s): K 983399 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | Detection of Cryptosporidium antigens in fecal specimens | Detection of Giardia and Cryptosporidium antigens in fecal specimens | | Technology | Qualitative immunochromatographic assay | Qualitative immunochromatographic assay | | Antibodies:conjugate | Monoclonal anti-Cryptosporidium | Rabbit anti-Giardia, monoclonal anti-Giardia and Cryptosporidium | | Specimen type | Human stool preserved in 10% formalin, SAF, MIF or Cary Blair | Human stool preserved in 10% formalin, SAF, MIF, or Cary Blair | | Differences | | | | Item | Device | Predicate | | Capture antibodies or molecules | Rabbit anti-Cryptosporidium, goat anti-mouse IgG | Mouse anti-Cryptosporidium, goat anti-mouse IgG, Avidin derivative | | Material: membrane | Mylar-backed nitrocellulose | Nitrocellulose | | Material: conjugate | Anti-mouse and anti-Cryptosporidium colored polystyrene microparticles diluted in buffer | Colloidal dye labeled monoclonal antibodies to Giardia and Cryptosporidium | | Sample volume | 100μl | 50μl | J. Standard/Guidance Document Referenced (if applicable): Not applicable {2} Page 3 of 5 K. Test Principle: The Xpect™ Cryptosporidium Lateral Flow Assay is a chromatographic immunoassay that detects the presence of Cryptosporidium antigens. The test utilizes sample wicking to capture Cryptosporidium antigens on discrete test lines containing antigen-specific antibodies for each organism. A specimen is added to a dilution tube containing a buffered solution. A conjugate containing colored micro-particles linked to monoclonal antibodies specific for Cryptosporidium is added. The mixture is dispensed into the sample well of the device and wicks across a membrane containing capture antibody strips. The Cryptosporidium immune complexes, if present, react with anti-Cryptosporidium antibody at the test line. Conjugates not bound at the test lines are later captured at the control line containing anti-mouse antibody. A pink line will appear at the Cryptosporidium test position if Cryptosporidium antigen is present. A line in the Control position indicates that the test is working properly. L. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Reproducibility testing was conducted at seven sites, including one in-house site, on three separate days with ten blinded samples of varying activity. All samples tested for Cryptosporidium produced the expected result. b. Linearity/assay reportable range: Not applicable c. Traceability (controls, calibrators, or method): Not applicable d. Detection limit: For Cryptosporidium, detection limit was 6 organisms per 0.1 ml. of specimen e. Analytical specificity: Cross-reactivity: No cross-reactivity was observed using samples containing the following organisms: Ascaris lumbricoides, Blastocystis hominis, Campylobacter coli, Campylobacter jejuni, Candida albicans, Chilomastix mesnili, Cyclospora cayetanensis, Dientamoeba fragilis, Endolimax nana, Entamoeba coli, Entamoeba hartmanni, Entamoeba histolytica, Enterobius vermicularis, Escherichia coli, hookworm, Hymenolepis nana, Iodamoeba butschlii, Isospora sp., Microsporidia, Rotavirus, Salmonella choleraesuis subsp. choleraesuis serotype Typhimurium, Shigella dysenteriae, Shigella flexneri, Shigella sonnei, Strongyloides stercoralis, Taenia sp., and Trichuris trichiura. Cross-reactivity to Astrovirus and Caliciviruses has not been established. Interfering Substances: Prior to testing, positive and negative samples were spiked (20% v/v) with blood, mucin, fecal fat or the following over-the-counter anti-diarrheal products: Pepto-Bismol®, Imodium® A-D, and Kaopectate® (active ingredients: bismuth {3} Page 4 of 5 subsalicylate, loperamide HCl, and attapugite respectively). Testing indicated that none of these substances interfered with the expected result except for Imodium® A-D. Imodium® A-D at 20% (v/v) in stool interfered with the detection of low levels of Cryptosporidium antigen. f. Assay cut-off: The assay can detect 6 Cryptosporidium organisms per 0.1 ml of specimen. Clinically relevant detection limits were determined using true clinical specimens diluted to an end point titration with the Xpect Cryptosporidium test. Sequential serial dilutions were tested until an endpoint dilution was reached. The endpoint dilution was defined as one dilution above where the sample became negative. The quantity of organisms detected at the titration endpoint in each specimen was calculated from the numbers seen microscopically in a 10 µl sample using DFA. 2. Comparison studies: a. Method comparison with predicate device: Percent Agreement: Xpect™ & Predicate Device vs. Microscopy: The Xpect™ Cryptosporidium and a commercially available lateral flow test (the "Predicate Device") were compared side-by-side to microscopy. The Percent Agreement of each assay versus microscopy was as follows: | Cryptosporidium | Microscopy | | | | | --- | --- | --- | --- | --- | | | | + | | - | | Xpect™ | + | 28 | 6 | | | | - | 2 | 110 | Agreement | | Total | | 30 | 116 | 94.5% (138/146) | | Predicate | + | 29 | 21 | | | --- | --- | --- | --- | --- | | | - | 1 | 95 | Agreement | | Total | | 30 | 116 | 84.9% (124/146) | b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical sensitivity: Sensitivity and Specificity Compared to Microscopy: The performance of the Xpect™ Cryptosporidium was evaluated at six geographically diverse laboratories. The overall sensitivity and specificity of the test was compared to microscopy using 578 specimens. Samples tested were {4} Page 5 of 5 samples routinely submitted for microscopic exam. Performance relative to patients’ clinical status has not been established. Data is listed below. | Cryptosporidium | Microscopy | | | | --- | --- | --- | --- | | | | + | - | | Xpect™ | + | 108 | 8 | | | - | 4 | 458 | | | Total | 112 | 466 | Sensitivity: 96.4% (108/112); 95% CI = 91.2-98.6% Specificity: 98.3% (458/466); 95% CI = 96.6-99.1% b. Clinical specificity: Refer to (a) above c. Other clinical supportive data (when a and b are not applicable): Not applicable 4. Clinical cut-off: See assay cut off above 5. Expected values/Reference range: Expected values were established from literature. Prevalence rates for Cryptosporidium range from 1-3% in North America and Europe to 7-9% in less developed countries. Rates tend to be higher in children under 2 years and in persons with weakened immune systems. M. Conclusion: In clinical settings, the Xpect™ Cryptosporidium Lateral Flow Assay is substantially equivalent in performance to the predicate device and to microscopic examination for the identification of Cryptosporidium in fecal specimens.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...