HOMECHOICE APD SET WITH LINEO CONNECTOR, MODELS 5C4469Q, 5C4468Q, 5C4531Q, 5C8302Q, 5C4599Q
Applicant
Baxter Healthcare Corp
Product Code
KDJ · Gastroenterology, Urology
Decision Date
Nov 20, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5630
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The HomeChoice APD Set with Lineo Connector is intended for use with the HomeChoice Automated Personal Cycler. The HomeChoice Integrated APD Set with Lineo Connector is intended for delivery of peritoneal dialysis therapy and is compatible with the HomeChoice Automated PD System only, and is to be used with the MiniCap Extended Life PD Transfer Set, Easy-Lock Extended Life PD Transfer Set, or Easy-Lock Transfer Set. The HomeChoice Low Recirculation Volume APD Set with Lineo Connector is intended for delivery of peritoneal dialysis therapy in pediatric and adult renal failure patients requiring fill volumes of 60 - 1000 ml, using only the HomeChoice Automated Personal Cycler with Low Fill Mode Drain Logic. The Lineo OptiCap is for protection of the Lineo Connector (patient line) of APD Disposable Sets with Lineo Connector that are intended for delivery of peritoneal dialysis therapy.
Device Story
Disposable tubing administration set for automated peritoneal dialysis (APD). Connects patient and dialysis solutions to HomeChoice automated cycler. Incorporates disinfectant feature into patient connector via Lineo Connector/OptiCap. Used in clinical or home settings by patients or caregivers. Facilitates fluid delivery for renal failure therapy. Replaces standard connectors with integrated disinfectant protection to reduce contamination risk.
Clinical Evidence
No clinical data. Bench testing performed including biological evaluation per ISO 10993-1 and functional/physical testing prior to release. Sterilization validated to SAL of 10^-6.
Technological Characteristics
Disposable tubing administration set. Materials meet ISO 10993-1 biological requirements. Features integrated disinfectant patient connector. Sterilized to SAL ≥ 10^-6.
Indications for Use
Indicated for pediatric and adult patients with renal failure requiring peritoneal dialysis therapy. Compatible with HomeChoice Automated PD Systems. Specific sets indicated for fill volumes 60-1000 ml.
Regulatory Classification
Identification
(1) A peritoneal dialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions, and that consists of a peritoneal access device, an administration set for peritoneal dialysis, a source of dialysate, and, in some cases, a water purification mechanism. After the dialysate is instilled into the patient's peritoneal cavity, it is allowed to dwell there so that undesirable substances from the patient's blood pass through the lining membrane of the peritoneal cavity into this dialysate. These substances are then removed when the dialysate is drained from the patient. The peritoneal dialysis system may regulate and monitor the dialysate temperature, volume, and delivery rate together with the time course of each cycle of filling, dwell time, and draining of the peritoneal cavity or manual controls may be used. This generic device includes the semiautomatic and the automatic peritoneal delivery system.(2) The peritoneal access device is a flexible tube that is implanted through the abdominal wall into the peritoneal cavity and that may have attached cuffs to provide anchoring and a skin seal. The device is either a single use peritioneal catheter, intended to remain in the peritoneal cavity for less than 30 days, or a long term peritoneal catheter. Accessories include stylets and trocars to aid in the insertion of the catheter and an obturator to maintain the patency of the surgical fistula in the abdominal wall between treatments. (3) The disposable administration set for peritoneal dialysis consists of tubing, an optional reservoir bag, and appropriate connectors. It may include a peritoneal dialysate filter to trap and remove contaminating particles. (4) The source of dialysate may be sterile prepackaged dialysate (for semiautomatic peritoneal dialysate delivery systems or “cycler systems”) or dialysate prepared from dialysate concentrate and sterile purified water (for automatic peritoneal dialysate delivery systems or “reverse osmosis” systems). Prepackaged dialysate intended for use with either of the peritoneal dialysate delivery systems is regulated by FDA as a drug.
Special Controls
*Classification.* Class II (special controls). The following accessories are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9: A catheter finger grip that is non-patient contacting and intended for single use with a peritoneal catheter; a continuous ambulatory peritoneal dialysis (CAPD) belt; and a catheter stand that does not include weigh scales.
{0}------------------------------------------------
K031676
Page 1/2
NOV 2 0 2003
## 510(K) SUMMARY
| Submitter's Name: | David E. Curtin, RAC |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 1620 Waukegan Rd. MPGR-A2E |
| Phone: | (847) 473-6079 |
| Fax: | (847) 473-6952 |
| Contact: | David E. Curtin |
| Date Prepared: | May 30, 2003 |
| Trade Name: | HomeChoice APD Set with Lineo Connector |
| Common Name: | Sets, Administration, for Peritoneal Dialysis, Disposable |
| Classification Name: | Peritoneal Dialysis System and Accessories 21 CFR 876.5630 |
| Equivalent Predicate: | HomeChoice APD Set (Baxter Healthcare Corp., K923065),<br>HomeChoice Low Recirculation Volume APD Set (Baxter Healthcare Corp., K012988), High Dose Disconnect Cap and<br>MiniCap with Povidone Iodine (Baxter Healthcare Corp., K972579) and Disconnect Caps with Povidone Iodine (Baxter Healthcare Corp., K895631) |
| Device Description: | HomeChoice APD Set with Lineo Connector is a disposable<br>tubing administration set used in automated peritoneal dialysis<br>therapies. It provides for connection of the patient and the<br>peritoneal dialysis solutions to the automated cycler. |
| Intended Use: | The HomeChoice APD Set with Lineo Connector is intended<br>for use with the HomeChoice Automated Personal Cycler |
:
{1}------------------------------------------------
.
| Summary of the<br>Technological<br>Characteristics<br>Compared to the | The general design and material of the HomeChoice APD Set<br>with Lineo Connector is to the HomeChoice APD Set<br>(K923065 and K012988) and High Dose Disconnect<br>Cap/MiniCap (K972579 and K895631) in that it incorporates<br>the disinfectant feature of the Disconnect Cap/MiniCap into the<br>patient connector of the set. The technological characteristics<br>of the HomeChoice APD Set with Lineo Connector do not<br>raise any new types of safety and effectiveness issues, when<br>compared to the predicate product. |
|-----------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate Device: | HomeChoice APD Set (Baxter Healthcare Corp), HomeChoice<br>Low Recirculation Volume APD Set (Baxter Healthcare<br>Corp.), High Dose Disconnect Cap and MiniCap with<br>Povidone Iodine (Baxter Healthcare Corp.) and Disconnect<br>Caps with Povidone Iodine (Baxter Healthcare Corp) |
| Clinical Data: | N/A |
| Conclusions Drawn | Components of the subject HomeChoice APD Set with Lineo<br>Connector have met the biological requirements of ISO 10993-<br>1: Biological Evaluation of Medical devices - Part: Guidance<br>on selection of tests.<br><br>Sterilization of the HomeChoice APD Set with Lineo<br>Connector is by a method determined and validated to ensure<br>an SAL of ≥ 106. Functional and physical testing is performed<br>prior to product release |
Additional
Information Requested by FDA:
None to date
:
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle or bird with three curved lines representing its body and wings. The bird is facing left. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird.
Public Health Service
NOV 2 0 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
David E. Curtin, RAC Associate Director, Regulatory Affairs Renal Division Baxter Healthcare Corporation 1620 Waukegan Road MCGAW PARK IL 60085
Re: K031676
Trade/Device Name: HomeChoice Automated PD Set with Lineo Connector; #5C4469Q, 5C4468Q HomeChoice Integrated APD Set with Lineo Connector; #5C4531Q HomeChoice Low Recirculation Volume APD Set with Lineo Connector; #5C8302Q and Lineo Opticap; #5C4599Q
Regulation Number: 21 CFR§ 876.5630 Regulation Name: Peritoneal Dialysis System and Accessories Regulatory Class: II Product Code: 78 KDJ Dated: August 29, 2003 Received: September 2, 2003
Dear Mr. Curtin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050,
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Page 2
{4}------------------------------------------------
510(k) Number (if known): K031676
Device Name: HomeChoice Integrated APD Set with Lineo Connector
Indications For Use:
The HomeChoice Integrated APD Set with Lineo Connector is intended for delivery of peritoneal dialysis therapy and is compatible with the HomeChoice Automated PD System only, and is to be used with the MiniCap Extended Life PD Transfer Set, Easy-Lock Extended Life PD Transfer Set, or Easy-Lock Transfer Set.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
Nancy C Brogdon
(Division Sign-Off)
22
Division of Reproduct and Radiological Devices 510(k) Number
G:\510k\Lineo\response!\Indications for Use Statement
{5}------------------------------------------------
510(k) Number (if known):________________________________________________________________________________________________________________
Device Name: HomeChoice Low Recirculation Volume APD Set with Lineo Connector
Indications For Use:
The HomeChoice Low Recirculation Volume APD Set with Lineo Connector is intended for delivery of peritoneal dialysis therapy in pediatric and adult renal failure patients requiring fill volumes of 60 - 1000 ml, using only the HomeChoice Automated Personal Cycler with Low Fill Mode Drain Logic.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
510(k) Number
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
Nancy C Brogdon
(Division Sign-Off)
Division of Reproductive, Abdomi nd Radiological Dev
G:\510k\Lineo\response 1\Indications for Use Statement
23
{6}------------------------------------------------
K031676 510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
Device Name: Lineo OptiCap
Indications For Use:
The Lineo OptiCap is for protection of the Lineo Connector (patient line) of APD Disposable Sets with Lineo Connector that are intended for delivery of peritoneal dialysis therapy.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
1 Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use_
Nancy C Brogdon
Division Sign Off
Reproductive, Abdom and Radiological D 510(k) Number
24
{7}------------------------------------------------
510(k) Number (if known): K031676
Device Name: HomeChoice Automated PD Set with Lineo Connector
Indications For Use:
The HomeChoice Automated PD Set with Lineo Connector is intended for delivery of peritoneal dialysis therapy and is compatible with the HomeChoice Automated PD System only, and is to be used with the MiniCap Extended Life PD Transfer Set, Easy-Lock Extended Life PD Transfer Set, or Easy-Lock Transfer Set.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
Nancy Brogdon
(Division Sign-Off)
vision of Reproductive, Abdo and Radiological De 510(k) Number
G:\510k\Lineo\response 1\Indications for Use Statement
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.