HIGH DOSE DISCONNECT CAP AND MINICAP WITH POVIDONE-IODINE, HIGH DOSE DISCONNECT CAP WITH POVIDONE-IODINE
Applicant
Baxter Healthcare Corp
Product Code
KDJ · Gastroenterology, Urology
Decision Date
Oct 1, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5630
Device Class
Class 2
Indications for Use
The High Dose Disconnect Cap and MiniCap with Povidone-Iodine Solution (5C4486) are single use caps that are intended for use with appropriate Peritoneal Dialysis (PD) products. The High Dose Cap is designed to isolate (cap off) the Easy-Lock™ connector (patient line) of the PD disposable set. The MiniCap is designed to isolate (cap off) the Easy-LockTM connector of the patient's transfer set. The High Dose Disconnect Cap with Povidone-Iodine Solution (5C4456) is a single use cap that is intended for use with appropriate Peritoneal Dialysis (PD) disposable products. The High Dose Cap is designed to isolate (cap off) the Easy-Lock™ connector (patient line) of the PD disposable set.
Device Story
Single-use disconnect caps for peritoneal dialysis (PD) therapy; designed to isolate Easy-Lock™ connectors on PD disposable sets and patient transfer sets. Each cap contains a povidone-iodine impregnated polyurethane sponge to disinfect the mating interface between the cap and the connector. Used by patients during PD cycler therapy disconnection. Provides integral seal to mating part; exterior ribs facilitate manual operation. Benefits include maintenance of sterile interface during disconnection. No complex electronics or software.
Clinical Evidence
Bench testing only. Performance testing included microbiological kill efficacy against test organisms, cap/connector seal pressure testing, and ease-of-use torque testing. Results demonstrated desired antimicrobial efficacy, successful seal integrity under pressure, and torque requirements comparable to predicate devices.
Technological Characteristics
Single-use caps; materials include polyurethane sponge impregnated with povidone-iodine. Features exterior ribs for manual handling and internal spacer. Mechanical sealing principle. No energy source, connectivity, or software.
Indications for Use
Indicated for patients undergoing peritoneal dialysis (PD) therapy requiring isolation (capping off) of the Easy-Lock™ connector on PD disposable sets and patient transfer sets.
Regulatory Classification
Identification
(1) A peritoneal dialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions, and that consists of a peritoneal access device, an administration set for peritoneal dialysis, a source of dialysate, and, in some cases, a water purification mechanism. After the dialysate is instilled into the patient's peritoneal cavity, it is allowed to dwell there so that undesirable substances from the patient's blood pass through the lining membrane of the peritoneal cavity into this dialysate. These substances are then removed when the dialysate is drained from the patient. The peritoneal dialysis system may regulate and monitor the dialysate temperature, volume, and delivery rate together with the time course of each cycle of filling, dwell time, and draining of the peritoneal cavity or manual controls may be used. This generic device includes the semiautomatic and the automatic peritoneal delivery system.(2) The peritoneal access device is a flexible tube that is implanted through the abdominal wall into the peritoneal cavity and that may have attached cuffs to provide anchoring and a skin seal. The device is either a single use peritioneal catheter, intended to remain in the peritoneal cavity for less than 30 days, or a long term peritoneal catheter. Accessories include stylets and trocars to aid in the insertion of the catheter and an obturator to maintain the patency of the surgical fistula in the abdominal wall between treatments. (3) The disposable administration set for peritoneal dialysis consists of tubing, an optional reservoir bag, and appropriate connectors. It may include a peritoneal dialysate filter to trap and remove contaminating particles. (4) The source of dialysate may be sterile prepackaged dialysate (for semiautomatic peritoneal dialysate delivery systems or “cycler systems”) or dialysate prepared from dialysate concentrate and sterile purified water (for automatic peritoneal dialysate delivery systems or “reverse osmosis” systems). Prepackaged dialysate intended for use with either of the peritoneal dialysate delivery systems is regulated by FDA as a drug.
Special Controls
*Classification.* Class II (special controls). The following accessories are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9: A catheter finger grip that is non-patient contacting and intended for single use with a peritoneal catheter; a continuous ambulatory peritoneal dialysis (CAPD) belt; and a catheter stand that does not include weigh scales.
Predicate Devices
Baxter's Easy Lock™ Disconnect Cap with Povidone-Iodine Solution (K895631)
Baxter's disconnect MiniCap with Povidone-Iodine Solution (K895631)
{0}------------------------------------------------
Page 1/2
K972579
## 510(k) SUMMARY
. D
్లు
| Submitter's name and<br>Address: | David C. Ross, Pharm.D., MBA, RAC<br>Associate Director, Regulatory Affairs<br>Baxter Healthcare Corporation, Renal Division<br>1620 Waukegan Road<br>McGaw Park, IL 60085 | OCT - 1 1997 |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Telephone Number:<br>Fax Number: | (847) 473-6081<br>(847) 473-6952 | |
| Contact: | David Ross | |
| Trade Name: | High Dose Disconnect Cap and MiniCap with Povidone-<br>Iodine Solution (5C4486) and High Dose Disconnect Cap<br>with Povidone-Iodine Solution (5C4456) | |
| Common Name: | Disconnect caps for Peritoneal Dialysis (PD) therapy | |
| Classification Name: | Peritoneal dialysis system and accessories per 21 CFR<br>876.5630 | |
| Equivalent Predicate: | Baxter's Easy Lock™ Disconnect Cap with Povidone-Iodine<br>Solution ( K895631), Baxter's disconnect MiniCap with<br>Povidone-Iodine Solution ( K895631), and the Medionics QC<br>Cap (K945567) | |
| Device Description: | The High Dose Disconnect Cap and MiniCap with Povidone-<br>Iodine Solution (5C4486) consists of the High Dose<br>Disconnect Cap, which contains a povidone iodine<br>impregnated polyurethane sponge, a spacer, and a MiniCap.<br>The MiniCap also contains a povidone iodine impregnated<br>polyurethane sponge. The povidone iodine sponge is used to<br>disinfect the mating interface between the disconnect caps<br>(High Dose Cap and MiniCap) and the patient connector of<br>the PD disposable set and the patient transfer set,<br>respectively. The High Dose Cap and the MiniCap are<br>packaged together to conveniently allow the patient to<br>disconnect from the peritoneal cycler therapy when desired.<br>The spacer aids in the convenience packaging of the two caps<br>into one product and has no other functional use. The<br>MiniCap product is currently cleared for marketing under<br>K895631 and is sold separately. | |
41
:
{1}------------------------------------------------
## K972579
Similarly, the High Dose Disconnect Cap with Povidone-Iodine Solution (5C4456) simply consists of the High Dose Disconnect Cap and a povidone iodine impregnated polyurethane sponge. The povidone iodine sponge is used to disinfect the mating interface between the disconnect cap and the patient connector of the PD disposable set.
The High Dose Disconnect Cap and MiniCap with Povidone-Intended Use: Iodine Solution (5C4486) are single use caps that are intended for use with appropriate Peritoneal Dialysis (PD) products. The High Dose Disconnect Cap is designed to isolate (cap off) the Easy-Lock™ connector (patient line) of the PD disposable set. The MiniCap is designed to isolate (cap off) the Easy-LockTM connector of the patient's transfer set.
> The High Dose Disconnect Cap with Povidone-Iodine Solution (5C4456) is a single use cap that is intended for use with appropriate Peritoneal Dialysis (PD) disposable products. The High Dose Cap is designed to isolate (cap off) the Easy-Lock™ connector (patient line) of the PD disposable set.
In general, the design and materials of the subject disposable disconnect caps are similar to the Baxter predicate devices. The caps provide an integral seal to the mating part, have exterior ribs to facilitate use, and contain a iodine impregnated sponge for disinfecting purposes.
Not applicable
None to date.
The performance/functional testing included microbiological, cap/connecter seal, and ease of use (torque) testing. The data demonstrate that the High Dose Disconnect Cap was able to deliver the desired kill of the test organism. Additionally, the cap/connector seal passed the specified pressure test and the torque to remove the cap is not greater than currently available disconnect caps.
Additional information requested by FDA:
Summary of the technological
Clinical Data:
tests:
characteristics compared
Conclusions drawn from
to the predicate device:
David C. Ross
David C. Ross, Pharm.D., MBA, RAC Associate Director, Regulatory Affairs
July 8, 1997
Date
2|2
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized abstract symbol to the right, consisting of three curved lines that resemble human figures. To the left of the symbol, the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Robert L. Wilkinson, RAC Director, Regulatory Affairs Baxter Healthcare Corporation 1620 Waukegan Road McGaw Park, Illinois 60085-6730 1997
Re: K972579
High Dose Disconnect Cap and MiniCap with Povidone-Iodine Solution (5C4486) and High Dose Disconnect Cap with Povidone-Iodine Solution (5C4456) Dated: July 8, 1997 Received: July 10, 1997 Regulatory class: II 21 CFR §876.5630/Product code: 78 KDJ
Dear Mr. Wilkinson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate.commence. prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours.
William Yin
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{3}------------------------------------------------
## 510(k) Premarket Notification
High Dose Disconnect Cap and MiniCap with Povidone-Iodine Solution (5C4486) and High Dose Disconnect Cap with Povidone-Iodine Solution (5C4456)
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
2 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1
Device Name: High Dose Disconnect Cap and MiniCap with Povidone-Iodine Solution (5C4486) and High Dose Disconnect Cap with Povidone-Iodine Solution (5C4456)
Indication for Use: The High Dose Disconnect Cap and MiniCap with Povidone-Iodine Solution (5C4486) are single use caps that are intended for use with appropriate Peritoneal Dialysis (PD) products. The High Dose Cap is designed to isolate (cap off) the Easy-Lock™ connector (patient line) of the PD disposable set. The MiniCap is designed to isolate (cap off) the Easy-Lock™ connector of the patient's transfer set.
The High Dose Disconnect Cap with Povidone-Iodine Solution (5C4456) is a single use cap that is intended for use with appropriate Peritoneal Dialysis (PD) disposable products. The High Dose Cap is designed to isolate (cap off) the Easy-Lock™ connector (patient line) of the PD disposable set.
· (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | Robert D. Sather |
|---------------|-------------------------------------------|
| | (Division Sign-Off) |
| | Division of Reproductive, Abdominal, ENT, |
| | and Radiological Devices |
| 510(k) Number | K972579 |
| Prescription Use | OR | Over-The-Counter Use ______ |
|----------------------|----|-----------------------------|
| (Per 21 CFR 801.109) | | |
0 ਰੇ
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.