K031401 · Disco-O-Tech Medical Technologies, Ltd. · KTT · Aug 6, 2003 · Orthopedic
Device Facts
Record ID
K031401
Device Name
FIXION DYNAMIC HIP SCREW SYSTEM (FIXION DHS)
Applicant
Disco-O-Tech Medical Technologies, Ltd.
Product Code
KTT · Orthopedic
Decision Date
Aug 6, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Fixion Dynamic Hip Screw System (Fixion DHS) is intended for use in fixation of fractures in the proximal femur. The Fixion DHS is indicated for use in pertrochanteric, intertrochanteric, and base of femur neck fractures. The Fixion DHS can be implanted in an open or in a minimally invasive approach.
Device Story
Fixion DHS is a metallic orthopedic fixation system for proximal femur fractures. Components include a titanium plate, a hip peg with a distal expandable section, a locking pin, and screws. The hip peg is expanded in situ using a manual inflation device (pump) to enhance abutment within the femoral head. The system is implanted via open or minimally invasive surgery by orthopedic surgeons. The device provides mechanical stabilization of bone fragments. The expandable peg mechanism is designed to improve fixation in the femoral head, potentially benefiting patients by providing secure fracture reduction and stabilization.
Clinical Evidence
Bench testing only. The device underwent mechanical testing per ASTM F 382-99 and ASTM F 384-00, and testing for non-articulating, mechanically locked modular implant components per FDA guidance.
Technological Characteristics
Materials: Titanium and Titanium alloy (ASTM F 1472-02, ASTM F 136-02, ASTM F 67-00); surgical grade Stainless Steel (ASTM F 899-02, ASTM F 1586-95); Silicone and Celeron (ISO 16061). Principle: Mechanical fixation via plate, screws, and an expandable hip peg. Energy source: Manual (manual pump for peg expansion). Form factor: Modular plate and screw system. Connectivity: None.
Indications for Use
Indicated for patients with pertrochanteric, intertrochanteric, and base of femur neck fractures requiring proximal femur fixation.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Fixion Interlocking Proximal Femoral Intramedullary Nailing System (K010988, K012967, K023437)
{0}------------------------------------------------
## 510(K) Summary
# Disc-O-Tech Medical Technologies Ltd. Fixion® Dynamic Hip Screw System (Fixion® DHS)
#### Company Name
Disc-O-Tech Medical Technologies, Ltd. 3 Hasadnaot St., Herzliya srael, 46728
#### Submitter's Name and Contact Person
Yael Rubin Disc-C)-Tech Medical Technologies, Ltd. 3 Hasadnaot St. Herzliya Ísrael, 46728 Tel: + 972-9-9511511 Fax: + 972-9-9548939
## Date Prepared
April 2003
### Trade/Proprietary Name
Fixion® Dynamic Hip Screw System (Fixion® DHS)
## Classification Name
Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component 21 CFR 888.3030 Class II
{1}------------------------------------------------
## 'Predicate Devices
- Fixion Interlocking Proximal Femoral Intramedullary Nailing System (K010988, D K012967, K023437) by Disc-O-Tech Medical Technologies, Ltd.
K031401 2/2
- > Gotfried PC.C.P (K983814) by Efratgo.
- A Vari-Angle Hip Screw System - VHS (K964880) by Biomet.
- ア TK2"M Compression Hip Screw System (K972629) by DePuy ACE
- Fixion Interlocking Intramedullary Nailing System (K002873, K013449, K023437) A by Disc-O-Tech Medical Technologies, Ltd.
#### Performance Standards
The following standards were referred:
- A ASTM F 382 - 99: Standard Specification and Test Method for Metallic Bone Plates
- > ASTM F 384 - 00: Standard Specification and Test Method for Metallic Angled Orthopedic Fracture Fixation Devices
- A Guidance Document for Testing Non-Articulating, "Mechanically Locked", Modular Implant Components (Draft), ORDB/DGRD/CDRH/FDA, May 1, 1995
- ア The Fixion DHS implants are manufactured from Titanium and Titanium alloy, which meet the requirements of ASTM F 1472 - 02, ASTM F 136 - 02, and ASTM F 67 -- 00.
- A The Fixion DHS System accessories incorporate surgical grade Stainless Steel (complying with ASTM F 899 - 02, ASTM F 1586 - 95), Silicone and Celeron (complying with ISO 16061).
#### Intended Use
The Fixion Dynamic Hip Screw System (Fixion DHS) is intended for use in fixation of fractures in the proximal femur. The Fixion DHS is indicated for use in pertrochanteric, intertrochanteric, and base of femur neck fractures.
The Fixion DHS can be implanted in an open or in a minimally invasive approach.
{2}------------------------------------------------
#### System Description
The Fixion DHS System consists of the following components:
- Plate a solid, titanium-made, component, incorporating holes for the insertion of A the Hip Peg, Locking Pin and Screws. The Plate Body is shaped to comply with the hone curvature and the Barrel is located at 135° - 145° relative to the Plate Body.
- Hip Peg a titanium-made tube with a distal expandable section, to enhance the A abutment of the Hip Peg inside the femoral head. The expandable body ends with a conical shaped distal end.
- Locking Pin a titanium-made pin. It is used to provide additional fixation of the A fragment, if needed, and for stabilizing bone fragments in case of torsional instability
- Screws titanium screws. Maximum of 3 7 Screws (Plate length dependant) can A be used to attach the plate to the bone (the femur shaft) at its distal section.
- Instrumentation Set ~ a set of accessories to be used with the Fixion DHS implants. A Accessories for both open and minimally invasive surgical approaches are available.
- A Inflation Device (Pump) - a manual pump used to expand the Hip Peg.
#### Substantial Equivalence
The Fixion DHS intended use, design, matcrials, technological characteristics and principles of operation are substantially equivalent to those of Disc-O-Tech's Fixion Interlocking PF Intramedullary Nailing System (K010988, K012967, K023437) and Fixion Interlocking Intramedullary Nailing System (K002783, K013449, K023437), Gotfied PC.C.P (K983814), Biomet VHS (K964880) and DePuy ACE TK2 Compression Hip Screw System (K972629).
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white circular seal. The seal contains the logo for the U.S. Department of Health and Human Services. The logo is a stylized image of an eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the top half of the circle.
AUG - 6 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Yael Rubin Director of Regulatory Affairs Disc-O-Tech Medical Technologies, Ltd. 3 Hasadnaot St. Herzliya Israel, 46728
Re: K031401 Trade/Device Name: Fixion Dynamic Hip Screw System (Fixion DHS) Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: KTT Dated: July 7, 2003 Received: July 9, 2003
Dear Mr. Rubin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. Yael Rubin
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Muriam C. Provost
for Celia Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Disc-O-Tech Medical Technologies, Ltd. Fixion Dynamic Hip Screw System 510(k)
#### Indication for Use
| 510(K) Number (if known): | K031401 |
|---------------------------|---------|
|---------------------------|---------|
Fixion® Dynamic Hip Screw System (Fixion® DHS) Device Name:
#### Indication for Use:
The Fixion Dynamic Hip Screw System ("Fixion DHS") is intended for use in fixation of fractures in the proximal femur. The Fixion DHS is indicated for use in pertrochanteric, intert:ochanteric, and base of femur neck fractures.
The Fixion DHS can be implanted in either open or minimal invasive surgical approach.
# (PLEASI) DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
P : escription Use (per 21 CFR 801.109)
Over the Counter Use
Muriam C. Provost
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number K031461
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.