RECOVERY FILTER SYSTEM, MODEL RF-048F

K031328 · C.R. Bard, Inc. · DTK · Jul 25, 2003 · Cardiovascular

Device Facts

Record IDK031328
Device NameRECOVERY FILTER SYSTEM, MODEL RF-048F
ApplicantC.R. Bard, Inc.
Product CodeDTK · Cardiovascular
Decision DateJul 25, 2003
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 870.3375
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Recovery Filter System is indicated for use in the prevention of recurrent pulmonary embolism via permanent placement in the vena cava in the following situations: -Pulmonary thromboembolism when anticoagulants are contraindicated. -Failure of anticoagulant therapy for thromboembolic disease -Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced. -Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated. -Recovery filter may be removed according to the instructions supplied in the Section labeled: Optional Procedure for Filter Removal. Data from removals in a 58 patient study suggests that the device can be safely removed (mean of 60 days; range 1-161 days).

Device Story

Recovery Filter System consists of nitinol vena cava filter and delivery system (7 Fr ID introducer sheath, dilator, pusher). Designed for permanent placement in vena cava to prevent recurrent pulmonary embolism; optional retrieval procedure allows removal. Used in clinical settings by physicians. Filter features two levels of filtration. Device provides mechanical filtration of emboli; retrieval capability allows removal after clinical resolution of thromboembolic risk. Benefits include prevention of pulmonary embolism in patients where anticoagulation is contraindicated or ineffective, with option for removal to avoid long-term indwelling risks.

Clinical Evidence

Clinical evidence based on a 58-patient study evaluating the safety of filter retrieval. Mean time to removal was 60 days (range 1-161 days). Results support the safety of the retrieval procedure. Animal testing, including histology, was also conducted to support the safety profile.

Technological Characteristics

Nitinol vena cava filter; 7 Fr ID introducer sheath, dilator, and pusher delivery system. Two-level filtration design. Packaged in Tyvek/film pouches. Non-electronic, mechanical device.

Indications for Use

Indicated for patients requiring prevention of recurrent pulmonary embolism via vena cava placement, specifically those with contraindications to anticoagulants, failure of anticoagulant therapy, or massive pulmonary embolism where conventional therapy benefits are reduced.

Regulatory Classification

Identification

A cardiovascular intravascular filter is an implant that is placed in the inferior vena cava for the purpose of preventing pulmonary thromboemboli (blood clots generated in the lower limbs and broken loose into the blood stream) from flowing into the right side of the heart and the pulmonary circulation.

Special Controls

*Classification.* Class II. The special controls for this device are:(1) “Use of International Standards Organization's ISO 10993 ‘Biological Evaluation of Medical Devices Part I: Evaluation and Testing,’ ” and (2) FDA's: (i) “510(k) Sterility Review Guidance and Revision of 2/12/90 (K90-1)” and (ii) “Guidance for Cardiovascular Intravascular Filter 510(k) Submissions.”

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JUL 2 5 2003 K 03 | 328 ## SECTION VIII ### 510(k) Summary of Safety and Effectiveness Information As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a summary of the safety and effectiveness information upon which substantial equivalence determination is based follows: - Submitter Information: A. | Applicant: | Bard Peripheral Vascular, a division of C.R. Bard<br>Inc.<br>1625 West 3rd Street<br>Tempe, Arizona 85280 | |------------|-----------------------------------------------------------------------------------------------------------| | Phone: | 480-303-2640 | | Fax: | 480-449-2546 | | Contact: | Mary Edwards, Vice-President | - B. Device Name: | Trade Name: | Recovery Filter System | |--------------------------|-------------------------------------| | Common or<br>Usual Name: | Percutaneous Vena Cava Filter | | Classification<br>Name: | Cardiovascular Intravascular Filter | - C. Predicate Device Name(s): Recovery Filter System - D. Device Description: The Recovery Filter System consists of a nitinol vena cava filter and a delivery system. The filter has two levels of filtration and is prepackaged in a storage tube. The delivery system consists of a 7 Fr ID introducer sheath and dilator and a pusher system. Both components of the system are packaged in Tyvek/film pouches. {1}------------------------------------------------ - Statement of Intended Use: E. The Recovery Filter System is indicated for use in the prevention of recurrent pulmonary embolism via permanent placement in the vena cava in the following situations: - -Pulmonary thromboembolism when anticoagulants are contraindicated. - -Failure of anticoagulant therapy for thromboembolic disease - Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced. - -Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated. - -Recovery filter may be removed according to the instructions supplied in the Section labeled: Optional Procedure for Filter Removal. Data from removals in a 58 patient study suggests that the device can be safely removed (mean of 60 days; range 1-161 days). - Substantial Equivalence: F. Recovery Filter is identical to the cited predicate device with the exception of the removal of labeling limitations and the addition of specific instructions to allow for safe removal of the device. The safety of removal was addressed in a series of animal and clinical testing. The results of animal testing (including histology) and the confirmatory experience of 58 patients (mean 60 days to removal; range 1-161 days) show that the Recovery Filter may be safely retrieved and that the benefit of this procedure outweighs the potential associated risks. {2}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized eagle emblem, with three curved lines representing the eagle's wings. The seal is black and white. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL 2 5 2003 Bard Peripheral Vascular c/o Ms. Mary Edwards Vice President, Regulatory & Clinical Affairs 1625 West 3rd Street P.O. Box 1740 Tempe, AZ 85280-1740 Re: K031328 Bard® Recovery® Filter System RF-048F Regulation Number: 21 CFR 870.3375 Regulation Name: Filter, Intravascular, Cardiovascular Regulatory Class: Class II Product Code: DTK Dated: April 25, 2003 Received: April 28, 2003 Dear Ms. Edwards: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ #### Page 2 - Ms. Mary Edwards Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4692. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Smail Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Brom D. Zuckerman, M.D. Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # ATTACHMENT A ## INDICATIONS FOR USE STATEMENT Device Name: Recovery Filter System Indications for Use: The Recovery Filter System is indicated for use in the prevention of recurrent pulmonary embolism via permanent placement in the vena cava in the following situations: - Pulmonary thromboembolism when anticoagulants are contraindicated. . - Failure of anticoagulant therapy for thromboembolic disease . - Emergency treatment following massive pulmonary embolism where anticipated . benefits of conventional therapy are reduced. - Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is . contraindicated - Recovery filter may be removed according to the instructions supplied in the Section . labeled: Optional Procedure for Filter Removal. ## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED. Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use **__** (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number K031328
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...