BIOPURE PORTABLE RO SYSTEM,BIOPURE 4400 SERIES RO SYSTEM, BIOPURE 8400 SERIES RO SYSTEM,BIOPURE WATER PURFICATION PRETRE
Applicant
Biolab Equipment Canada , Ltd.
Product Code
FIP · Gastroenterology, Urology
Decision Date
Oct 21, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5665
Device Class
Class 2
Indications for Use
The device is intended to remove organic and inorganic substances and microbial contaminants from water that is used to dilute dialysis concentrate to form dialysate, and to produce purified water for dialyzer reprocessing, and equipment rinse and disinfection.
Device Story
Biopure Series RO systems purify feed water via reverse osmosis; pressure forces water through semipermeable membrane, retaining impurities (dissolved inorganics, organics, bacteria, pyrogens) on feed side. Pretreatment components (boost pumps, filters, softeners, carbon filters) condition feed water by removing chlorine/chloramines. Product water distribution components (storage tanks, pumps, UV disinfection, hot water sanitization) deliver purified water meeting AAMI/ASAIO standards for hemodialysis. Systems range from 300 GPD to 100 GPM capacity. Used in home, hospital, or clinic settings by healthcare providers or patients (portable units). Output water used to dilute dialysis concentrate, reprocess dialyzers, and disinfect equipment. Benefits include consistent water quality for dialysis, reducing patient exposure to contaminants.
Clinical Evidence
Bench testing only. No clinical data provided. Performance metrics include rejection rates of >99% for microorganisms, pyrogens, particles, and organics (>300 MW), and 95%-99% for dissolved inorganics.
Technological Characteristics
Reverse osmosis membrane separation; pretreatment includes sediment/carbon filters, softeners, multimedia filters, dealkalizers. Product water distribution includes UV disinfection and hot water sanitization. Capacity: 300 GPD to 100 GPM. Meets ANSI/AAMI RD5-1992 standards.
Indications for Use
Indicated for water purification in hemodialysis applications. Supports patient populations ranging from 1 to 600 patients depending on system model (Portable, 4400, or 8400 series), used in home, hospital, or dialysis clinic settings.
Regulatory Classification
Identification
A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.
Special Controls
*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
## 4. 510 (k) Summary of Safety and Effectiveness
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990 and 21CFR 807.92
#### Applicant
Biolab Equipment Canada Limited 505 Iroquois Shore Road, Unit 14 Oakville, Ontario L6H 2R3 David Weatherill - President Tel: 905 844-2062 Fax: 905 844-0023
## Prepared by
J.P. Consulting Jerry Ponikvar --
128 Grandview Avenue Thornhill, Ontario L3T 1H6 jponikvar@sympatico.ca E-Mail Telephone (905) 882-2068 (905) 333-0601 Fax
Jerry Ponikvar, will respond to any requests for additional information.
#### Device Name
Trade Names Biopure Portable RO System Biopure 4400 Series RO System Biopure 8400 Series RO System Biopure Water Purification Pretreatment Components Softeners Biopure Performa Softener
| Backwash Filters | Biopure Magnum Series Water Softener |
|------------------|-----------------------------------------|
| | Biopure Performa Filter |
| | Biopure Magnum Series Carbon Filter |
| | Biopure Magnum Series Multimedia Filter |
Service Deionization Biopure Service DI
Biopure Premium Grade Service Di BiopureCation Bed DI Biopure Anion Bed DI Biopure Auto DI
Organic Bed Service Carbon
Biopure Carbon Biopure Product Water Distribution Components
Biopure CIP System Biopure Hot Water Sanitization System
Common or Usual Names
Reverse Osmosis systems with pretreatment and product water distribution components
Classification Name
Water Purification System for Hemodialysis Regulatory Classification - Class II 21 CFR 876.5665 Product Code: 78 FIP
Biolab510KPremarketSubmittalA.doc
{1}------------------------------------------------
510(K)Premarket Notification
#### Intended Use
The device is intended to remove organic and inorganic substances and microbial contaminants from water that is used to dilute dialysis concentrate to form dialysate, and to produce purified water for dialyzer reprocessing, and equipment rinse and disinfection.
## Device Description
The Biopure Series RO systems purify feed water through reverse osmosis a membrane separation process. In reverse osmosis, pressure is applied to force water through a semipermeable membrane, while leaving impurities on the feed side of the membrane. The membrane allows the solvent (water) to pass through but retains a large percentage of impurities such as dissolved inorganics, organics, bacteria, and pyrogens. The water that passes through the membrane (permeate), has significantly less contamination than the feed water. while the reject stream, which remains upstream of the membrane, has a much higher concentration than the feed stream.
The purpose of a pretreatment to reverse osmosis is to remove chlorinel chloramines and condition the feed water supplying the RO machine. The pretreatment section may include feed water boost pumps, a temperature blend valve, chemical feed unit, sediment/carbon cartridge filters, water softeners, multi media filters, carbon filters and dealkalizers.
The RO unit removes more than 99% of all micro-organisms, pyrogens, particles, and organics with a molecular weight greater than 300, and up to 95% of all dissolved inorganics. The elements of an RO unit include a prefilter, RO pressure pump, and RO membranes.
The purpose of a product water distribution system is to deliver water that meets or exceeds AAMI/ASAIO standards for water treatment equipment and water quality requirements for hemodialysis point of use. Additional product water treatment is intended to control or eliminate bacteria prior to point of use. The product water treatment system may include a storage tank, distribution pump, a distribution loop, UV disinfection, and hot water sanitization.
Depending on the series and model, Biopure water purification systems produce from 300 GPD (gallons per day) to 100 GPM (gallons per minute) at 75% recovery rate. Rejection of total dissolved solids and monovalent ions is in the range of 95%-99% depending on feed water characteristics. Colloids, bacteria, and organics larger than a 200 molecular weight are rejected at greater than 99%.
Biopure water purification systems product water that meets the requirements of ANSI / AAMI RD5-1992, Hemodialysis Systems.
#### Predicate Devices
The Biopure Series RO systems, pre-treatment components, and product water components are substantially equivalent to the following legally marketed devices: Zyzatech Series RO systems, K964539 US Filter Water Purification System, K980182 These devices use reverse osmosis technology to purify water for hemodialysis applications.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three wavy lines or stripes, which are meant to represent the human form. The logo is in black and white.
OCT 2 1 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Biolab Equipment Canada, Ltd. c/o Mr. Jerry Ponikvar, P.Eng J. P. Consulting 128 Grandview Avenue Thornhill, Ontario CANADA L3T 1H6
Re: K030348
Trade/Device Name: Biopure Portable RO System Biopure 4400 Series RO System Biopure 8400 Series RO System Biopure Water Purification Pretreatment Components Biopure Product Water Distribution Components Regulation Number: 21 CFR §876.5665
Regulation Name: Water purification system for hemodialysis Regulatory Class: II Product Code: 78 FIP Dated: July 22, 2003 Received: July 23, 2003
Dear Mr. Ponikvar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. Ixxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Page 2
{4}------------------------------------------------
# 3. Statement of Indications for Use
Applicant: Biolab Equipment Canada Limited
Establishment Registration # 9710313
510 (k) Number (if Known): K030348
Device Name:
Biopure Portable RO System Biopure 4400 Series RO System Biopure 8400 Series RO System Biopure Water Purification Pretreatment System Components Biopure Product Water System Components
Classification Name:
Water Purification System for Hemodialysis Regulatory Classification - Class II 21 CFR 876.5665 Product Code: 78 FIP
Indications for Use:
The device is intended to remove organic and inorganic substances and microbial contaminants from water that is used to dilute dialysis concentrate to form dialysate and to produce purified water for dialyzer reprocessing, and equipment rinse and disinfection. Enclosed below is a chart indicating the physical differences of the different systems, areas of use, and patient population.
| System Model | Biopure Portable | Biopure 4400 | Biopure 8400 |
|---------------------|--------------------------------------------------------------------------------------------|--------------------------------|---------------------------------|
| Area of Use | Home or hospital use, this systems allows dialysis treatment to patients not near a clinic | Hospital or Dialysis Clinics | Large Dialysis Clinics |
| Patient Population | 1 to 3 | 3 to 50 | 65 to 600 |
| Physical Dimensions | 39" x 12.5" x 28" | 72"x24"x32" to<br>72"x18"x144" | 76"x32"x122" to<br>76"x44"x162" |
| Weight | 120 to 135 lbs | 220 to 1610 lbs | 1200 - 8000 lbs |
Prescription Use (Per 21 CFR 801.109)
Nancy C. Brogdon
(Division Sign-Off) (Division of Reproductive, Abdominal, and Radiological D 510(k) Number
BiolabS10KPremarketSubmittalB
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.