PORTABLE SERIES REVERSE OSMOSIS SYSTEMS 700, 701, 702, 703, 704, 750, MRE-NF, F800-F802, RO SECURA, REVERSE OSMOSIS SYST
K964539 · Zyzatech Water Systems, Inc. · FIP · Nov 26, 1997 · Gastroenterology, Urology
Device Facts
Record ID
K964539
Device Name
PORTABLE SERIES REVERSE OSMOSIS SYSTEMS 700, 701, 702, 703, 704, 750, MRE-NF, F800-F802, RO SECURA, REVERSE OSMOSIS SYST
Applicant
Zyzatech Water Systems, Inc.
Product Code
FIP · Gastroenterology, Urology
Decision Date
Nov 26, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5665
Device Class
Class 2
Indications for Use
ZyzaTech Reverse Osmosis Systems and their pretreatment and product water distribution components are intended for use with a hemodialysis system to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysis concentrate to form dialysate, reprocessing of hemodialyzers and equipment rinse and disinfection.
Device Story
System uses reverse osmosis to purify feed water for hemodialysis applications; high hydrostatic pressure forces water through semi-permeable membrane to separate solutes. Pretreatment components (pumps, filters, softeners, carbon tanks) condition feed water; RO unit (pumps, membranes) removes dissolved inorganic/organic solutes (>300 Dalton), bacteria, viruses, and endotoxins. Product water distribution section (storage tanks, UV irradiators, sub-micron filters) delivers water meeting ANSI/AAMI RD5-1992 standards to points of use. Used in clinical hemodialysis settings; operated by trained personnel. Output is purified water for dialysate dilution, dialyzer reprocessing, and equipment disinfection. Benefits include reduction of contaminants in dialysis water, ensuring compliance with AAMI standards for patient safety.
Clinical Evidence
No clinical data. Substantial equivalence is based on bench testing and performance specifications demonstrating that the systems meet ANSI/AAMI RD5-1992 standards for hemodialysis water quality, including rejection rates for TDS (95-99%) and bacteria/pyrogens (99%).
Technological Characteristics
Reverse osmosis purification; pretreatment includes booster pumps, temperature blend valves, cartridge/multi-media filters, water softeners, dealkalizers, carbon filtration. RO unit includes prefilter, pump, and membranes in pressure vessels. Distribution includes storage tanks, UV irradiators, ultra/sub-micron filters. Performance: 250-42,000 gpd capacity; 17-75% recovery rate; 95-99% TDS rejection; >300 Dalton organic cut-off. Complies with ANSI/AAMI RD5-1992.
Indications for Use
Indicated for use with hemodialysis systems to purify water for dialysate preparation, dialyzer reprocessing, and equipment disinfection by removing organic/inorganic substances and microbial contaminants.
Regulatory Classification
Identification
A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.
Special Controls
*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
Water Purification System (Better Water Inc.)
HD-100 Hemodialysis Systems & Accessories (Water Pure Corporation)
Hemo-RO/Milli RO Water Purification Systems (Millipore Corporation)
Submission Summary (Full Text)
{0}
11-20-1997 8:35AM FROM ZYZATECH 253 395 0928 P.3
K964539 p 1/2
ZyzaTech Water Purification 510(k) 9/11/96
SECTION 8 510(k) Summary of Safety and Effectiveness
This 510(k) Summary of Safety and Effectiveness is provided as part of this Premarket Notification to comply with the provisions of the Safe Medical Device Act of 1990 requiring that either a summary be included in a submission or a statement that a summary is available upon request.
NOV 26 1997
Submitter
Emanuel Amato
ZyzaTech Water Systems, Inc.
7848 South 202nd Street
Kent, WA 98032
Phone: (206) 395-2200 Fax: (206) 395-2363
September 16, 1996
Device Names
- Trade names
ZyzaTech Portable Series Reverse Osmosis Systems:
-700 Series
-System 750
-MRE-NF
-F800-F801
-RO-Secura
ZyzaTech V-Series Reverse Osmosis Systems
ZyzaTech Z-Series Reverse Osmosis Systems
ZyzaTech T-Series Reverse Osmosis Systems
ZyzaTech Water Purification Pretreatment System Components
ZyzaTech Water Product Water Distribution System Components
- Common or usual name
Reverse Osmosis systems with pretreatment and product water distribution components.
- Classification name
Water purification system for hemodialysis (21 CFR 876.5665)
Intended Use
ZyzaTech Reverse Osmosis Systems and their pretreatment and product water distribution components are intended for use with a hemodialysis system to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysis concentrate to form dialysate, reprocessing of hemodialyzers and equipment rinse and disinfection.
Device Description
The ZyzaTech series RO systems purify feed water through reverse osmosis which is the opposite of osmosis. Osmosis is the natural process wherein concentrated and dilute solutions, separated by a semi-permeable membrane, flow to equalize concentrations. Solvent will flow from the dilute to the concentrated side of the membrane creating equal levels of concentration. Reverse osmosis is the application of high hydrostatic pressure on the concentrated solution; the direction of flow is reversed so that high solute concentrations flow across a semi-permeable membrane and become dilute concentrations. This results in two streams of water; one that is purer than the feed water, and the other
D-50
{1}
11-20-1997 8:35AM FROM ZYZATECH 253 395 0928
P. 4
K964539 p. 2/2
9/11/96
ZyzaTech Water Purification 510(k)
that is higher in concentrates than the feed water. The pure flow is product water. The high concentrate flow is waste water.
The purpose of a pretreatment section of the system is to remove chlorine/chloramines and to condition the feed water supplying the RO unit. Conditioning the feed water can include: boosting the pressure, tempering, removal of particulates and, often, membrane scaling factors, along with volatile inorganic and organic solutes. The pretreatment section can include: feed water booster pumps, a temperature blend valve, chemical feed unit, cartridge filters, multi-media (sediment) filters, water softeners, dealkalizers and carbon filtration tanks.
The RO unit removes dissolved inorganic solutes, organic solutes with a molecular weight greater than 300, bacteria, viruses, endotoxins and particles. The primary elements of a RO unit are: a prefilter, RO pump and RO membranes housed in pressure vessels.
The purpose of a product water distribution section of a system is to deliver product water that meets AAMI standards to the points of use. Additional post-treatment measures are intended to control or eliminate bacteria prior to the points of use, especially in an indirect feed type of system. The product water distribution section can include: a storage tank, distribution repressurization pumps, distribution loop, UV irradiators, ultra and sub-micron filters.
Depending upon the series and model, ZyzaTech water purification systems produce typically from 250 gpd to 42,000 gpd of product water with a range of 17-75 percent recovery rate. Rejection of total dissolved solids (TDS) and monovalent and polyvalent ions is in the range of 95-99% depending upon feed water characteristics. Bacteria and pyrogens are rejected 99%. Organic size cut-off is a molecular weight greater than 300 Dalton.
ZyzaTech's water purification systems produce product water that meets the requirements of the voluntary standard issued by the American National Standard Institute and the Association for the Advancement of Medical Instrumentation: ANSI/AAMI RD5-1992, Hemodialysis Systems.
## Predicate Devices
The ZyzaTech series RO systems and pretreatment and product water distribution components are substantially equivalent to the following legally marketed devices: Water Purification System manufactured by Better Water Inc., HD-100 Hemodialysis Systems & Accessories manufactured by Water Pure Corporation, and Hemo-RO/Milli RO Water Purification Systems manufactured by Millipore Corporation. All these devices use reverse osmosis technology to purify water for hemodialysis applications.
Page 60
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
NOV 26 1997
Mr. Emanual J. Amato
President
ZyzaTech Water System, Inc.
7848 S. 202nd Street
Kent, Washington 98032
Re: K964539
Water Purification Systems and Components and
Portable Reverse Osmosis Systems
Dated: August 25, 1997
Received: September 9, 1997
Regulatory class: II
21 CFR §876.5665/Product code: 78 FIP
Dear Mr. Amato:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Lillian Yin, Ph.D.
Director, Division of Reproductive,
Abdominal, Ear, Nose and Throat,
and Radiological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{3}
11-20-1997 8:34AM -RUM 2Y2A 2U- 200 300 200
P. 2
# INDICATIONS FOR USE
510(k) Number: K964539
Device Name: Reverse Osmosis Systems
Trade Names:
ZyzaTech Portable Series, 700 Series, System 750, MRE-NF, F800-F801, RO-Sccura
Central Systems
ZyzaTech V-Series Reverse Osmosis Systems
ZyzaTech Z-Series Reverse Osmosis Systems
ZyzaTech T-Series Reverse Osmosis Systems
ZyzaTech Water Purification Pretreatment System Components
ZyzaTech Product Water Distribution System Components
Classification name
Water purification system for hemodialysis (21 CFR 876.5665)
Indications for use:
A ZyzaTech Portable, V, Z or T Series Reverse Osmosis System and its pretreatment and product water distribution components is intended for use with a hemodialysis system to remove organic and inorganic substances and microbial contaminants from water that is used to dilute dialysis concentrate to form dialysate, and to produce purified water for other purposes such as dialyzer reprocessing and equipment rinse and disinfection.
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign/Off)
Division of Reproductive, Abdominal, ENT,
Radiological Devices
510(k) Number K964539
Prescription Use ☑
(Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.