The Cozart EIA Cocaine Oral Fluid Kit is intended for use in clinical and analytical laboratories when used in conjunction with the Cozart RapiScan Oral Fluid collection system. Using this collection system it provides qualitative screening results for cocaine and metabolites in human oral fluid at a cutoff concentration of 10ng/ml. This is equal to 30ng/ml. in undiluted oral fluid as the collection system involves a 1:3 dilution of the sample. This assay is for professional use only and provides only a preliminary analytical test result. Clinical consideration and professional judgement must be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. In order to obtain a more confirmed analytical result a more specific alternative chemical method is needed. Gas Chromatography/Mass Spectrometry (GC/MS) is the preferred confirmatory method.
Device Story
Competitive ELISA assay for detection of cocaine and metabolites in human oral fluid; utilizes Cozart RapiScan collection system (1:3 dilution). Input: oral fluid sample; reagents include antibody-coated microtitre plate, enzyme conjugate (Horse Radish Peroxidase), substrate, and stop solution. Principle: competitive binding between drug in sample and enzyme-labeled drug for antibody sites. Output: spectrophotometric absorbance at 450nm. Used in clinical/forensic laboratories by professional staff. Results provide preliminary screening; clinical judgment required; positive results necessitate GC/MS confirmation. Benefits: rapid, qualitative assessment of drug presence to guide further confirmatory testing.
Clinical Evidence
Bench testing only. Method comparison study of 177 samples against GC/MS showed 100% sensitivity (100/100 positives) and 92% specificity (71/77 negatives), with 96% overall agreement. Precision studies reported CVs < 11%. Interference testing performed on 29 substances; no cross-reactivity observed.
Technological Characteristics
Competitive ELISA; Horse Radish Peroxidase enzyme conjugate; microtitre plate format. Calibrators: 0, 5, 10, 50ng/mL benzoylecgonine. Readout: spectrophotometric at 450nm. Manual assay. Standards: BS EN ISO 9001:1994; EN 46001:1996.
Indications for Use
Indicated for qualitative screening of cocaine and metabolites in human oral fluid for clinical and forensic populations. Requires professional use in laboratory settings. Not for diagnostic confirmation; positive results require GC/MS confirmation.
Regulatory Classification
Identification
A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.
Special Controls
*Classification.* Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
{0}------------------------------------------------
K030234
# PREMARKET NOTIFICATION [510(k)] SUMMARY
# APR 2 8 2003
| Submitter | Cozart Bioscience Ltd<br>45 Milton Park<br>Abingdon<br>Oxfordshire OX14 4RU<br>UK<br>Tel No: 01235 861483<br>Fax No: 01235 835607 |
|----------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Dr Roberto Liddi<br>Quality Assurance & Regulatory Affairs Manager |
| Date | 20th January 2003 |
| Device Name | Cozart EIA Cocaine Oral Fluid Kit |
Trade Name Cozart EIA Cocaine Oral Fluid Kit
Classification Name Cocaine Test System
Classification
Class II
Code of Federal Regulations Title 21 Food and Drugs Part 862 Clinical Chemistry and Clinical Toxicology Devices Subpart D Clinical Toxicology Test Systems 862.3250 Cocaine and cocaine metabolite test system
Establishment Registration No
3002336046
Performance Standards
BS EN ISO 9001:1994; EN 46001:1996
Substantial Equivalence
STC Opiates Micro-Plate EIA (Orasure® Application), 510(k) no. K981341
{1}------------------------------------------------
| Parameter | Cozart EIA Cocaine Oral Fluid Kit | Cocaine Metabolite Micro-Plate EIA<br>(Orasure® Application), 510(k) no.<br>K982061 |
|-------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Qualitative test for Cocaine in human oral<br>fluid with a 30ng/ml cutoff. Recommend<br>confirmation of positive results by GC/MS. | Qualitative test for Cocaine in human oral<br>fluid with a 10ng/ml cutoff. Recommend<br>confirmation of positive results by GC/MS. |
| Target Population | Clinical and forensic samples. | Clinical samples. |
| Design | Competitive ELISA | Competitive ELISA |
| Enzyme | Horse Radish Peroxidase | Horse Radish Peroxidase |
| Results | Read spectrophotometrically at 450nm. | Read spectrophotometrically at 450nm. |
| Calibrators | 0, 5, 10, 50ng/mL | 0, 5, 10, 50ng/mL |
| Matrix | Human Oral Fluid | Human Oral Fluid |
| Controls | None supplied but Cozart recommends<br>using external controls. | Unknown |
| Method Comparison | 177 samples were tested, 100 screened<br>positive for cocaine and metabolites, of<br>which 100 were confirmed positive by<br>GC/MS. 77 samples screened negative for<br>cocaine and metabolites and 71 were<br>confirmed negative by GC/MS. | 92% Agreement as compared to GC/MS |
| Precision | CV (%) of 1.9 – 10.7% | CV (%) of 5 – 11.2% |
| Sensitivity | 1.8ng/mL | <5ng/mL |
| Specificity | 29 potential interferents tested – none<br>cross-reacted | Unknown |
# Introduction
The Cozart EIA Cocaine Oral Fluid Kit is a laboratory based test for the detection of cocaine and metabolites in human oral fluid using a cutoff equivalent to 30ng/mL. The device detailed above was compared to Gas Chromatography/Mass Spectrometry (GC/MS).
Cozart Bioscience Ltd is the manufacturer of the Opiates Oral Fluid Kit. We have not purchased this device from another manufacturer and the device is not marketed under another product name.
# Intended Use
The Cozart EIA Cocaine Oral Fluid Kit is intended for use in clinical and forensic laboratories when used in conjunction with the Cozart RapiScan Oral Fluid collection system. Using this collection system it provides qualitative screening results for cocaine and metabolites in human oral fluid at a cutoff concentration of 10ng/ml. This is equal to 30ng/mL in undiluted oral fluid as the collection system involves a 1:3 dilution of the sample.
The remainder of this document will refer to the 10ng/mL cutoff concentration.
This assay is for professional use only and provides only a preliminary analytical test result. Clinical consideration and professional judgement must be applied to any drug of abuse test result, particularly in evaluating a preliminary positive
{2}------------------------------------------------
result. In order to obtain a more confirmed analytical result a more specific alternative chemical method is needed. Gas Chromatography/Mass Spectrometry (GC/MS) is the preferred confirmatory method.
#### Target Population
The target population for the Cozart EIA Cocaine Oral Fluid Kit is clinical and forensic samples.
#### Where Used
The Cozart EIA Cocaine Oral Fluid Kit is designed for use in clinical and forensic laboratories. For professional use only.
# Design
As can be seen from the Principle of the Test section in the pack insert, the Cozart EIA Cocaine Oral Fluid Kit is a competitive ELISA for the detection of cocaine and its metabolites in human oral fluid.
# Materials
The Cozart EIA Cocaine Oral Fluid kit supplies the following reagents - a microtitre plate coated with antibody, enzyme conjugate reagent, wash buffer, substrate solution, stop solution and four calibrators (0, 5, 10 and 50ng/ml benzoylecgonine in oral fluid matrix).
# Performance
#### Method Comparison
Cocaine Oral Fluid The Cozart EIA Kit was compared to Gas Chromatography/Mass Spectrometry (GC/MS). All the samples were tested with the Cozart EIA Cocaine Oral Fluid Kit according to the pack insert enclosed.
177 samples were tested through the Cozart EIA Cocaine Oral Fluid Kit, 100 screened positive and 100 were confirmed positive by GC/MS. 77 samples screened negative and 71 were confirmed negative by GC/MS. Of the 177 samples tested 15 were between -50% cutoff and +50% cutoff.
| New Device | GC/MS<br>Negs | Near Cutoff<br>GC/MS<br>Negs * | Near Cutoff<br>GC/MS<br>Pos ** | GC/MS<br>Pos*** | Percent Agreement<br>with<br>GC/MS |
|------------|---------------|--------------------------------|--------------------------------|-----------------|------------------------------------|
| Pos<br>100 | 0 | 0 | 4 | 100 | 100 |
| Neg<br>77 | 71 | 6 | 5 | 6 | 92 |
* Between -50% Cutoff and the Cutoff.
** Between +50% Cutoff and the Cutoff.
*** Total number of positives (includes near cutoff samples).
96% overall agreement as compared with GC/MS
Cozart EIA Cocaine Oral Fluid Kit
{3}------------------------------------------------
# Precision
The precision obtained for the Cozart EIA Cocaine Oral Fluid Kit produced CVs less than 9%. The total precision for the kit produced CVs less than 11%. The Cozart EIA Cocaine Oral Fluid Kit is a qualitative manual ELISA assay and CVs of less than 10% are acceptable for this assay type.
#### Sensitivity
The sensitivity of the Cozart EIA Cocaine Oral Fluid Kit is 1.8ng/ml.
# Specificity
Twenty-nine potentially interfering unrelated substances were tested for cross reactivity in the Cozart Cocaine Oral Fluid Kit and none were found to cross react. Four related compounds were tested and three showed a level of cross reactivity.
#### Cutoff Concentration
Testing samples at the cutoff concentration, 50% above and 50% below were carried out to validate the cutoff concentration. The absorbances obtained for the 5ng/ml sample were all higher than the 10ng/ml cutoff calibrator. Similarly the absorbances obtained for the 15ng/ml sample were all lower than the 10ng/ml cutoff calibrator.
#### Interference Studies
A range of parameters including alcohol, sample adequacy indicator dye, haemoglobin, smoking, coffee, tea, water, food, orange juice, hard candy, chewing gum and mouthwash were tested for interference in the Cozart Cocaine Oral Fluid Kit. No interference was observed with any of the parameters.
#### Stopped Assay Stability
The stability of the stopped assay was investigated by reading the absorbance at 450nm at times 0, 5, 10, 15, 30, 45 and 60 minutes. The Cozart EIA Cocaine Oral Fluid Kit must be read within 15 minutes at 450mm.
#### Assay Drift
Sample addition at time 0, 2.5, 5, 7.5, 10, 12.5, 15, 17.5, 20, 22.5 and 25 minutes was investigated. Little change was observed across the plate and therefore sample addition to a Cozart EIA Cocaine Oral Fluid Kit must take place within 25 minutes.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The logo is black and white.
APR 2 8 2003
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Dr. Roberto Liddi Quality Assurance & Regulatory Affairs Manager Cozart Bioscience Ltd 45 Milton Park Abingdon Oxfordshire OX14 4RU UK
Re: k030234
Trade/Device Name: Cozart EIA Cocaine Oral Fluid Kit Regulation Number: 21 CFR 862.3250 Regulation Name: Specimen transport and storage container Regulatory Class: Class II Product Code: DIO Dated: April 15, 2003 Received: April 18 2003
Dear Dr. Liddi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{5}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page 1 of
510(k) Number (if known): K 030 Z3 4
Device Name:__________________________________________________________________________________________________________________________________________________________________ Cozart EIA Cocaine Oral Fluid Kit
Indications For Use:
The Cozart EIA Cocaine Oral Fluid Kit is intended for use in clinical and analytical laboratories when used in conjunction with the Cozart RapiScan Oral Fluid collection system. Using this collection system it provides qualitative screening results for cocaine and metabolites in human oral fluid at a cutoff concentration of 10ng/ml. This is equal to 30ng/ml. in undiluted oral fluid as the collection system involves a 1:3 dilution of the sample.
This assay is for professional use only and provides only a preliminary analytical test result. Clinical consideration and professional judgement must be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. In order to obtain a more confirmed analytical result a more specific alternative chemical method is needed. Gas Chromatography/Mass Spectrometry (GC/MS) is the preferred confirmatory method.
Sean Cooper
Division Sign-Off)
Clinical Labora
510(k) Number K 030234
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ry
(Optional Format 3-10-98)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.