K981341 · OraSure Technologies, Inc. · DJG · May 22, 1998 · Clinical Toxicology
Device Facts
Record ID
K981341
Device Name
STC OPIATES MICRO-PLATE EIA (ORASURE APPLICATION)
Applicant
OraSure Technologies, Inc.
Product Code
DJG · Clinical Toxicology
Decision Date
May 22, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3650
Device Class
Class 2
Indications for Use
The STC Opiates Micro-Plate EIA is intended for use in the qualitative determination of opiates in oral fluid collected with the OraSure Specimen Collection Device. For In Vitro Diagnostic Use.
Device Story
The STC Opiates Micro-Plate EIA is a competitive enzyme immunoassay (EIA) used to detect opiates in oral fluid. The process begins with the collection of oral fluid using the OraSure Oral Specimen Collection Device, which utilizes a treated absorbent pad to collect mucosal transudate. In the laboratory, the specimen is added to an EIA micro-plate well coated with rabbit anti-morphine antibodies, along with an enzyme-labeled morphine hapten. A competitive binding reaction occurs between the drug in the specimen and the enzyme-labeled hapten for the fixed antibody sites. After washing, a substrate (TMB) is added, producing a color change. The absorbance at 450 nm is measured; it is inversely proportional to the opiate concentration in the sample. Results are compared against a 10 ng/mL cutoff to determine if the sample is positive or negative. This assay assists healthcare providers in identifying opiate presence, though results should be interpreted alongside clinical context due to differences in opiate metabolism and partitioning between oral fluid and urine.
Clinical Evidence
Clinical evidence includes a study of 60 samples (10 self-reported opiate users, 50 presumed negative) and a controlled dose study with 5 volunteers. In the chronic user study, sensitivity and specificity compared to GC/MS were 100% and 87.5%, respectively. The controlled dose study showed the assay could detect opiates at 12 mg intravenous doses up to 2 hours post-dose, but not at 3 or 6 mg doses. Analytical sensitivity (LOD) is 0.2 ng/mL. Precision studies (intra-assay and inter-assay) were conducted at 0, 5, 10, and 20 ng/mL levels.
Indicated for qualitative detection of opiates in oral fluid collected via OraSure device. Intended for use in clinical settings for diagnostic testing of individuals suspected of opiate use.
Regulatory Classification
Identification
An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
{0}------------------------------------------------
K98134|
MAY 2 2 1998
### PREMARKET NOTIFICATION 510(k) SUMMARY (As Required By 21 CFR 807.92)
807.92 (a):
- STC Technologies, Inc. 1. Submitter's Name: 1745 Eaton Avenue, Bethlehem, PA 18018 Address: (610) 882-1820 Telephone Number: R. Sam Niedbala, Ph.D., BCFE Contact Person: April 8, 1998 Date Prepared:
- Device Name: 2. Opiates Micro-Plate EIA (OraSure® Application) Proprietary Name: Usual Name: Opiates EIA Enzyme Immunoassay, Opiates Classification Name:
- Device to Which Substantial Equivalence Is Claimed: 3. STC Opiates Micro-Plate EIA (Urine); K924446
- Description of Device: 4.
#### Principle of the Assay
The STC Opiates EIA is a competitive micro-plate immunoassay for the detection of opiates in oral fluid collected with the OraSure® Oral Specimen Collection Device. Specimen or standard is added to an EIA well in combination with an enzyme-labeled hapten derivative. In an EIA well containing an OraSure® specimen positive for opiates, there is a competition between the drug and the enzyme-labeled hapten to bind the antibody fixed onto the EIA well. EIA wells are then washed, substrate is added, and color is produced. The absorbance measured for each well at 450 nm is inversely proportional to the amount of opiates present in the specimen or calibrator/control, Because currently there are no SAMHSA assigned cutoffs for opiates testing using oral fluid, STC recommends a cutoff of 10 ng/mL when testing oral fluid collected with the OraSure® Oral Specimen Collection Device. This cutoff is within the limit of detection by the STC Opiates Micro-Plate EIA.
{1}------------------------------------------------
| KIT COMPONENTS |
|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Micro-Plate -- Rabbit anti-morphine antibody immobilized on a polystyrene plate supplied<br>n dry form. |
| Enzyme Conjugate -- Horseradish peroxidase labeled with a morphine hapten diluted in a<br>rotein matrix of bovine serum with protein stabilizers. |
| Substrate Reagent -- One bottle containing 3,3', 5,5' tetramethylbenzidine. |
| Stopping Reagent -- Each bottle contains 2 N sulfuric acid. |
| STC Negative Calibrator -- OraSure® Control Matrix negative for morphine. |
| STC Opiates Negative Control -- OraSure® Control Matrix containing 5 ng/mL<br>(± 3 ng/mL) morphine. |
| STC Opiates Cutoff Calibrator -- OraSure® Control Matrix containing 10 ng/mL (± 3<br>g/mL) morphine. |
| STC Opiates Positive Control -- OraSure® Control Matrix containing 20 ng/mL (± 3<br>g/mL) morphine. |
### Principle of the OraSure® Oral Specimen Collection Device
Saliva is a complex mixture of parotid, submandibular, sublingual and minor salivary gland secretions mixed with mucin, bacteria, leukocytes, sloughed epithelial cells and gingival crevicular fluid. The fact that opiates are present in oral fluid following human use is well documented. (2)
OraSure was developed for the purpose of collecting oral fluid for diagnostic testing. The collection device consists of a treated absorbent cotton fiber pad affixed to a nylon stick (Collection Pad) and a preservative solution in a plastic container (Specimen Vial). The Collection Pad is impregnated with a mixture of common salts and gelatin which creates a hypertonic environment and an increased osmotic pressure wherever it contacts oral mucosal cells. The pad is placed in contact with the gingival mucosa (between the lower gum and cheek) which enhances the flow of mucosal transudate across the mucosal surfaces onto the absorptive cotton fibers of the pad. Following the collection period, the Collection Pad is placed into a vial containing a preservative solution which serves to inhibit the growth of oral micro-organisms recovered on the Collection Pad. The vial is sealed with a plastic cap and transported to a laboratory for processing and testing. Following processing, a fluid containing a mixture of saliva components and the preservative solution is recovered which is suitable for testing for the presence of opiates in the STC Opiates Micro-Plate EIA manufactured by STC Technologies, Bethlehem, PA. Refer to the OraSure® Oral Specimen Collection Device product insert from Epitope for specific instructions on the proper collection, handling, and adequacy of saliva samples
- રું . Intended Use Statement
The STC Opiates Micro-Plate EIA is intended for use in the qualitative determination of opiates in oral fluid collected with the OraSure Specimen Collection Device. For In Vitro Diagnostic Use.
{2}------------------------------------------------
- Summary of Technological Characteristics 6.
The STC Opiates Micro-Plate EIA is based on the principle of solid phase competitive enzyme immunoassay . This application is for the modified use of the STC Opiates ELA urine kit for the detection of opiates using specimens collected with the OraSure® Oral Specimen Collection Device manufactured by Epitope, Inc., Beaverton, Oregon. The kit is essentially the same as the predicate urine kit. However, additional standards (supplied with the kit) have been prepared from the OraSure® Specimen Diluent in the OraSure® Collection Device.
#### 807.92 (b):
- Non Clinical Data 1.
Analytical Sensitivity/Limit Of Detection - The Limit of Detection (LOD) was defined from the signal-to-noise ratio at the zero-drug concentration as the mean zero absorbance (Ao) minus the noise times three (LOD = Ao - 3SD). The LOD was determined by obtaining the absorbance value for twelve blank OraSure® devices and calculating the standard deviation (SD) of the absorbance values. The absorbance value minus 3SD was then extrapolated from the curve and represents the sensitivity of the assay. The LOD was calculated to be 0.2 ng/mL.
Precision - The precision of the STC Opiates Micro-Plate EIA was assessed by testing OraSure® Control Matrix containing 0, 5, 10, or 20 ng/mL morphine. The intra-assay precision was determined by analyzing each level 16 times per run for 4 runs. Inter-assay precision was determined by analyzing 2 samples at each level twice per day for 20 days. The results of this testing are described in the following table:
| Company Controller Children Children Children<br> | <br>A Laboration Company<br> | <br>. |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|--------------------|
| A FINAL COLLEGION<br>--------- | 1<br>--------- | |
| | | |
| | | |
| Part College of Canadian Cart I provident I laminen leaster and<br>LARRIERA In REENWORD LEASTER BROMERS IN LE MUNICIPALIES LE MUNICIPALIES LE MUNICIPALIES LE MUNICIPALIES LEA MULTION | SELECT COLLECT CLAIM OF & MILE OF STORE SECT COLLECT COLLEGION | CANTON INSTITUTION |
Analytical Specificity/Cross-Reactivity - The analytical specificity of an immunoassay is defined as the cross-reactivity of substances in the assay which are structurally related to the target compound. The percent cross-reactivity of a compound in the STC Opiates Micro-Plate EIA is defined as the apparent morphine concentration divided by the spiked concentration times 100.
The cross-reactivity of structurally related compounds was calculated at several spiked concentrations in normal human urine. The following table indicates the apparent concentration of morphine for each substance at a concentration which cross-reacted in the assay.
{3}------------------------------------------------
| Compound | Tested Concentration<br>(ng/mL) | Morphine Equivalents<br>(ng/mL) | Cross-Reactivity<br>(%) |
|------------------------------|---------------------------------|---------------------------------|-------------------------|
| 6-Acetylmorphine | 100 | 14.1 | 14 |
| Codeine | 10 | >100 | >100 |
| Dextromethorphan | 10000 | <5 | n.d. |
| Diacetylmorphine | 5 | 10 | >100 |
| Hydrocodone | 5 | 33 | >100 |
| Hydromorphone | 100 | 79 | 79 |
| Levorphanol | 100 | 24 | 24 |
| Meperidine | 10000 | 10 | 0.1 |
| Morphine | 10 | 10 | 100 |
| Morphine-3-β-<br>Glucuronide | 100 | 21 | 21 |
| Nalorphine | 1000 | 11 | 1.2 |
| Normorphine | 1000 | 7 | 0.7 |
| Oxycodone | 1000 | 19 | 1.9 |
| Oxymorphone | 10000 | 51 | 0.5 |
The following compounds cross-react in the assay at the levels shown:
*n.d. = not detectable
The following compounds were spiked into OraSure® Control Matrix at a target concentration of 10,000 ng/mL and tested for cross-reactivity. None were found to produce a signal less than or equal to that of the Opiates Cutoff Calibrator.
| Acetylsalicylic Acid | Clorazepate | L-Ephedrine | Pentobarbital |
|----------------------|-------------------|---------------------|---------------------|
| Alprazolam | Cocaethylene | L-Methamphetamine | Phenobarbital |
| Amobarbital | Cocaine | Lidocaine | Phenylephrine |
| Ampicillin | Cotinine | Medazepam | Phenylpropanolamine |
| β-Phenethylamine | D-Amphetamine | Methadone | Procainamide |
| Benzoylecgonine | D-Methamphetamine | Metroprolol | Procaine |
| Butabarbital | Fenoprofen | Naproxen | Pseudoephedrine |
| Butalbital | Gemfibrozil | Niacinamide | Quinidine |
| Caffeine | Gentisic Acid | Norchlordiazepoxide | Temazepam |
| Chlordiazepoxide | Glipizide | Nordiazepam | Δ9-THC |
| Chlorpromazine | Ibuprofen | PCP | Theophylline |
| Clonazepam | Imipramine | Penicillin | Zomepirac |
It is possible that other substances and/or factors not listed above may interfere with the test and cause false results, e.g., technical or procedural errors.
#### 2. Clinical Data
#### Accuracy
The clinical sensitivity and specificity of the STC Micro-Plate assay were determined by testing specimens from chronic opiate users. A controlled dose study was also conducted to determine a minimum dose to produce a positive response in the STC EIA. Both studies are described below. The cutoffs for EIA and GC/MS were 10 ng/mL and 15 ng/mL, respectively, for OraSure® specimens. For urine specimens, the EIA cutoff was 300 ng/mL and the GC/MS cutoff was 150 ng/mL (based on SAMHSA recommendations).
{4}------------------------------------------------
### Study 1 - Chronic Opiates Users
Matched urine and OraSure® specimens were collected from ten (10) self-reported opiates users. An additional fifty (50) presumed negative samples were also included in the analysis. All sixty (60) samples were tested in the STC Opiates Micro-Plate EIA and presumed positives (10) were also tested by gas chromatography/mass spectrometry (GC/MS). Results are presented in the following tables:
| | ------------- | VMS*<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>---------------------------------------------------------------------------------------------------------------- |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>ANNUAL CALL CONNECT COLLEGION A | ------------------<br>Acres of the |
| ------------------<br>---------------<br>------------------------------------<br>I | Acres an<br>Status Callery Concession College<br>------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>An a case<br>.<br>--------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ----------<br>1<br><br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br><br><br>------------------ |
*Of the 10 samples presumed positive, the % sensitivity and % specificity of the OraSure® STC EIA as compared to GC/MS were 100.0% and 87.5%, respectively.
Image /page/4/Figure/4 description: This image shows a table comparing Urine-STC EIA (300 ng/mL Cutoff) and OraSure-STC EIA (10 ng/mL Cutoff). The table shows the number of positive and negative results for each test. For Urine-STC EIA, there are 3 positive and 0 negative results, while for OraSure-STC EIA, there are 4 positive and 53 negative results. The relative sensitivity is 42.9% and the relative specificity is 100.0%.
| <br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>CARRELLE COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION<br><br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br> | <br>.<br>--------<br>-<br> | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>1000 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 -<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br><br>--------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>Commended on the commender of the different of the closed on the closed to the closed to the closes of the closes of the closes of the closes of the closes of the closes of t<br>LANDS<br>----- |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Carlos and Concession of Children Company of Children Company of Children<br>-----------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | 1<br><br><br>-----------<br>- | I<br>CANADA CALLERY<br>Controllers of the contributed to the country of the country of the country of the country of the country of the country of the country of the country of the country of the c<br>CARATING<br>------------------<br><br>None of Children Area<br>1<br>Commons of Canadian Canadian Canada California Carolina Carolina Comercial Construction of |
| <br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>1<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br> | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>--------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br><br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br><br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>Call College College College College College Concession Comparis Comparis<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ------------<br>-----<br>--------------<br>any of the program and company to the control of the collection of the collection of the contribution of the contribution of the contribution of the contribution of the contr<br><br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>--------------------------------------<br>-----<br>--------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>1 |
** Of the 10 samples presumed positive, the % sensitivity and % specificity of the urine STC EIA as compared to GC/MS were 87.5% and 100.0% respectively.
It is believed that the poor agreement between the STC Opiates Micro-Plate EIA (OraSure® application) and the STC Opiates Micro-Plate EIA (urine) is driven by differences in the metabolism of heroin and partitioning in oral and urine. Opiates generally appear later in urine and persist for up to 3-6 days. " Conversely, opiates in saliva is not detectable after 2-8 hours. 10 These differences do not diminish the clinical utility of either fluid. However, users should be aware of these differences.
### Study 2 - Controlled Dose Study
A total of 5 volunteers in a controlled environment were administered an initial 3 mg of intravenous heroin HCl, followed by a second 6 mg or 12 mg heroin dose 4 to 12 days later. Saliva and urine samples were collected before dosing and periodically for up to 74 hours after dosing. Saliva specimens (400 uL) were added to the collection pad of the OraSure Oral Fluid Collection Device and processed according to the package insert. Samples were then tested using the STC assay and GC/MS.
Using a 10 ng/mL cutoff for OraSure® samples, opiates were detectable by EIA for up to 2 hours after a 12 mg heroin dose and not detectable at any time for a 3 or 6 mg dose. Of the GC/MS saliva results, only the initial post-12 mg dose samples contained drug concentrations above the 15 ng/mL LOD/LOQ. This demonstrates the ability of the STC
{5}------------------------------------------------
GC/MS saliva results, only the initial post-12 mg dose samples contained drug concentrations above the 15 ng/mL LOD/LOQ. This demonstrates the ability of the STC EIA to detect opiate concentrations at intravenous doses as low as 12 mg heroin HCl. The ability of the kit to detect opiates at doses between 6 mg-12 mg is unknown.
- Conclusions 3.
A comparison of the performance data for the new device vs. the predicate device is given on the following page:
{6}------------------------------------------------
# STC OPIATES MICRO-PLATE EIA (ORASURE® DEVICE)
## PERFORMANCE DATA COMPARISON
## 1. Limit of Detection
| because of the manage of Canadian and the manufacturer specificantly spiralized consistence of elections of election<br>Property of the program of the first for the program and any and<br>. In the call of the mand the same of the security would be come of the security of<br>The same of the consisted on the commend of the country of the count of the county of the receive<br>and the contribution of the commended to the commended to the comments of the comments of the comments of the comments of the comments of the comments of the comments of the<br>BEATHER FOR SEPART AND RESEARCE OF COLLECTION OF CHARRONAL CONSULTER CONSTITUTION CONTENTION COLLEGIONAL COLLEGIONAL COLLECTION COLLEGIONAL COLLECTION COLLEGIONAL COLLECTION<br>The first and the first of the country of the country of the county of the county of<br>CONTRACTORS, AND CARTERS COLLECTION CONTRACTOR CONTRACTOR CONTRACTOR<br>дава (для ава чуртая баздан компания в семь в как в<br><br><br><br><br><br><br>CARTER CARDER CONTRACT CONTRACT CONTENT CONTRACTOR CONTRACTER CONTRACT CONTACT CONTACT CANAL<br> | STATE THE TELE THE THE SECTION CONSULTERS CONSULTERS CONSULTERS AND THE LEAST THE LEAST THE LEAST THE LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST<br>COLL HIT INTERNATIONAL PRODUCTION COLLEGION CONSULTERS CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTO<br>страние продавания продаващие собразования со приведения в присание приват по полнето в поставительного<br><br>THE LAND LEE FEE T ST ST P L 1 FEED .<br>LEAR<br><br>Company of Antal Antage of Carl and Party of Canada Carlos Carder of<br><br><br><br><br>Ann and and<br> |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | |
| CANAL<br>Comments of Children Art Children and Children Children Children Child Children Child Children Children Children Children Children Children | |
and the control of the comments of the comments of the comments of the comments of
### 2. Precision
| Assay | Intra-Assay % CV<br>Range for Levels<br>Tested | Inter-Assay % CV<br>Range for Levels<br>Tested |
|----------------------------------------|------------------------------------------------|------------------------------------------------|
| STC Opiates Micro-Plate EIA (Urine) | 5.5 | 6.3 |
| STC Opiates Micro-Plate EIA (OraSure®) | 8-17 | 9-13 |
### 3. Sample pH Effect
| CHILLE COLLECT COLLECTION COLLECT COLLECT COLLECT COLLECTION CONSULTER CONSULTER CONSULTER CONSULTER CONSULTER CONSULTER CONSULTION CONSULTER CONSULTER CONSULTER CONSULTER CO<br>. Hill be and the country of the comments of the comments of the country of the count<br>MITHAN HERE THE FOR THE CONTRACTOR COLOR COLOR COLOR COLLECTION CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR<br>a manager and a comprehensive of the contribution of 2000 compression of the comments of the comments of the comments of the comments of the comments of the comments of the c<br>A B L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L W L C O W L C O W L C O W L C O C<br>S CONSEL<br><br><br>ALT 240 .<br><br>and the same of the country of the comments of the comments of the comments of the comments of<br>.<br>STEERS A F REE E B S C F F C S C F C F . C | 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1<br>Comments of the local status and the result and a sure well the receive as result and a firm and<br>да се продавата и подарительные продавания продавания пристории по полнения по полнения и политические по получается в политические по политические по политические по подател<br>and the controlled on the commend of the comments of the comments of the comments of the<br><br>and the commend of the state of the been and the comments of the comments of<br>THE CHILD STATE CONTRACT COLLECTION CONTRACTOR COLLECTION CONTRACTOR COLLECTION CONTRACTOR COLLECTION CONTRACTOR COLLECTION CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRA<br><br>W. M. L. L. L. B. L. F. C. .<br><br><br><br>and and the contract and the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the c<br><br>.<br> |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| BULLE LEASE & LAND &<br>AL RAS LEMENTER AND A | |
| Part 1 11 112 . 114 . 114 . 11 | CONTRACTOR PROPERTY PROLL PALLASSES COLLECTION COLLECTION CONSULT COLLEGION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION CONSULTION CONTRACTOR COLLECTION |
### 4. Effect of Common Materials
| Assay | Adulterants Which Affect Assay Results |
|----------------------------------------|----------------------------------------|
| STC Opiates Micro-Plate EIA (Urine) | Not Tested |
| STC Opiates Micro-Plate EIA (OraSure®) | Cranberry Juice |
| | Orange Juice |
| | Cola |
| | Cough Syrup |
{7}------------------------------------------------
### 5. Cross-Reactivity
| Compound | Tested Conc.<br>(ng/mL) | OraSure® Assay | % Cross-Reactivity in the STC Opiates Micro-<br>Plate EIA<br>Urine Assay, Positive @ |
|------------------------------|-------------------------|----------------|--------------------------------------------------------------------------------------|
| 6-Acetylmorphine | 100 | 14 | 500 ng/mL |
| Codeine | 10 | >100 | 75 ng/mL |
| Dextromethorphan | 10000 | >0.1 | Not Tested |
| Diacetylmorphine | 5 | >100 | 600 ng/mL |
| Hydrocodone | 5 | >100 | 300 ng/mL |
| Hydromorphone | 100 | 79 | 600 ng/mL |
| Levorphanol | 100 | 24 | 4500 ng/mL |
| Meperidine | 10000 | 0.1 | Not Tested |
| Morphine-3-β-<br>Glucuronide | 100 | 21 | 125,000 ng/mL |
| Nalorphine | 1000 | 1.2 | 125,000 ng/mL |
| Normorphine | 1000 | 0.71 | 125,000 ng/mL |
| Oxycodone | 1000 | 1.9 | 125,000 ng/mL |
| Oxymorphone | 10000 | 0.51 | 125,000 ng/mL |
Rs. Nudlah
R. Sam Niedbala, Ph.D., BCFE Executive Vice President
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle with its wings spread, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle, indicating the department's name and national affiliation.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
MAY 2 2 1998
R. Sam Niedbala, Ph.D. . Executive Vice President STC Technologies, Inc. 1745 Eaton Avenue Bethlehem, Pennsylvania 18018-1799
Re : K981341 STC Opiates Micro-Plate EIA Requlatory Class: II Product Code: DJG Dated: April 8, 1998 Received: April 13, 1998
Dear Dr. Niedbala:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate abe Beaced in the eny 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{9}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807,97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html".
Sincerely yours,
Steven Litman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
### STATEMENT OF INDICATIONS FOR USE
981341 510(k) Number (if known):
Device Name: STC Opiates Micro-Plate EIA
Indications For Use: The STC Opiates Micro-Plate EIA is intended for use in the qualitative determination of opiates in oral fluid collected with the OraSure® Oral Collection Device. For In Vitro Diagnostic Use.
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
**Prescription Use**
(Per 21 CFR 801.109)
c:\fda\indica.doc
OR
Over-The-Counter Use - Over
---
Division Sign-Off)
ivision of Clinical Laboratory Devices
10(k) Number. K941341
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.