K021487 · Sterling Medivations, Inc. · FPA · May 17, 2002 · General Hospital
Device Facts
Record ID
K021487
Device Name
SIMPLICITY EASY ACCESS P INFUSION SET
Applicant
Sterling Medivations, Inc.
Product Code
FPA · General Hospital
Decision Date
May 17, 2002
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the Simplicity Easy Access P Infusion Set is to provide a means for infusion and/or injection of fluids into the body below the surface of the skin when attached to a MiniMed Paradigm pump or syringe.
Device Story
Simplicity Easy Access P Infusion Set is an intravascular administration set for subcutaneous fluid delivery. Device components: AISI 304 stainless steel indwelling catheter/needle, infusion hub, connecting tube, and threaded YP connector. Operation: Indwelling needle inserted below skin surface; connector needle mates with catheter hub to form leak-proof seal; YP connector attaches to MiniMed Paradigm pump reservoir or syringe. Used by patients with diabetes for continuous or intermittent medication delivery. Features 360-degree rotating infusion hub for improved mobility. Available in multiple package configurations (10-pack or 5/10-pack) to accommodate varying catheter change frequencies. Benefits include secure, reliable subcutaneous access for insulin or other fluids.
Clinical Evidence
Bench testing only. Device performance validated against CDRH 21 C.F.R. Section 880.5440 and multiple ISO standards, including ISO 10555 (catheter requirements), ISO 11135/11138 (ethylene oxide sterilization), ISO 9626 (stainless steel tubing), ISO 11607 (packaging), ISO 8535 (syringes), and ISO 10993-1 (biocompatibility). Design controls adhered to 21 CFR 820.30 and ISO 9001.
Technological Characteristics
Materials: AISI 304 stainless steel (needles/catheters), polyethylene (tubing). Principle: Subcutaneous infusion via indwelling needle. Form factor: Infusion set with 360-degree rotating hub and threaded YP connector. Sterilization: Ethylene oxide. Connectivity: Mechanical interface to external infusion pump/syringe. Standards: ISO 10555, ISO 9626, ISO 11135, ISO 11138, ISO 11607, ISO 8535, ISO 10993-1.
Indications for Use
Indicated for people with diabetes requiring subcutaneous infusion or injection of fluids via an external MiniMed Paradigm pump or syringe.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
# Sterling Medivations, Inc. 66 Neptune Drive Groton, CT 06340 650-814-4083 (voice) 770-242-8639 (fax)
## 510(k) SUMMARY
Date Submitted: May 7, 2002
| Submitter: | Sterling Medivations, Inc. 66 Neptune Drive, Groton, CT 94022<br>Company Phone 650-814-4083, Company Fax 770-242-8639 |
|------------|-----------------------------------------------------------------------------------------------------------------------|
| Contact: | Joel Douglas, Chief Technology Officer<br>Sterling Medivations, Inc.<br>Applicant Phone 650-814-4083 or 770-242-8639 |
Trade Name of Device: Simplicity™ Easy Access P Infusion Set for infusion and/or injection of fluids into the body below the surface of the skin when attached to an external pump or syringe
| Common Name of Device: | Intravascular administration set. |
|------------------------|--------------------------------------|
| Classification Name: | Percutaneous intravascular catheter. |
FDA/CORH/00MC
RECEIVE MAY 8 3 1802Simplicity Easy Access Infusion Set, FDA 510 (k) K014062. Predicate Device:
Description of the New Device: Sterling Medivations Inc.'s ("SMI") Simplicity Easy Access P designed for use by people with diabetes for the infusion and/or injection of fluids into the body below the surface of the skin when attached to a MiniMed Paradigm pump or syringe. It is inserted into thssue of a patient and the catheter is attached to the infusion device. It is substantially equivalent to the Simplicity ™ Easy Access Infusion Set, K014062 and Soft YP Infusion Set, FDA 510 (k) K011187.
The Simplicity Easy Access P Infusion Set proposed for commercial distribution is similar in all significant respects to the existing Simplicity Easy Access Infusion Set, FDA 510 (k) K014062 and it has the same intended use.
The device consists of four main parts: (1) infusion catheters made from AISI 304 stainless steel, (2) an infusion hub that provides the patient the capability of disconnecting tube from the infusion catheter, (3) a connecting tube and (4) a threaded YP connector.
The Simplicity Easy Access P Infusion Set is an infusion administration set, connecting to a medicine reservoir or syringe that is placed in an external infusion device and inserted below the surface of the skin of a patient. The SMI Simplicity Easy Access Infusion Set may be used with any Paradigm infusion device that delivers continuous or intermittent flow.
The administration set attaches to the reservoir/syringe by means of a threaded YP connector, and under the surface of the skin in the patient through an indwelling needles formed from AISI 304 stainless steel. The connecting tubing is made from a polyethylene tube.
The indwelling needles formed from AISI 304 stainless steel are introduced below the surface of the skin. The connector needle is attached to the indwelling catheter. This seal on the connector needle mates with the indwelling catheter hub that forms a seal that permits the infusion of medication without leakage. The connector needle is made from AISI 304 stainless steel and it is connecting tubing. The connector tubing proximal end is attached to a threaded YP connector for attachment to the Paradigm medicine reservoir.
Intended Use of the New Device: The intended use of the Simplicity Easy Access P Infusion Set is to provide a means to for infusion and/or injection of fluids into the body below the surface of the skin when attached to a MiniMed Paradigm pump or syringe.
SK시
Sterling Medivations, Inc. Simplicity Easy Access P Infusion Set 510(k) Summary
Page 1 of 2
{1}------------------------------------------------
# Comparison of the Technological Features of the New Device and Predicate Device:
The Simplicity Easy Access Infusion Set proposed for commercial distribution is similar in all significant respects to the existing Simplicity Easy Access Infusion Set, FDA 510 (k) K014062 and it has the same intended use.
The materials and manufacturing processes are substantially equivalent, the labeling is substantially equivalent and it has the same intended use as the Simplicity Easy Access Infusion Set.
The differences that exist between the new and predicate device are as follows:
1) The Simplicity Easy Access P Infusion Set has a 360 degree rotating infusion hub attached to the connecting tube and the predicate device has a fixed infusion hub attached to the connecting tube. The 360 degree rotating hub and connector is equivalent to the hub and connector on the Sterling Medivations Soft QD II Infusion Set, K020990.
2) The Simplicity Easy Access P Infusion Set has is sold in two package configurations a complete 10 pack and a 5/10 package consisting of 10 indwell housing hubs and 5 connecting tubes for patient that change their indwell housing hubs more frequently than every 72 hours. This is similar to the packaging provided by the Maersk Medical Contour™ Infusion Set, K991759 sold as the MiniMed Silhouette.
3) The Simplicity Easy Access P Infusion Set indwelling catheter connector is designed for use with the MiniMed Paradigm pump which is equivalent to the YP connector on the Sterling Medivations Simplicity Soft YP Infusion Set K011187.
Performance Data Supporting Substantial Equivalence: To prove substantial equivalence the Simplicity Easy Access Infusion Set meets the catheter requirements of:
- CDRH 21 C.F.R. Section 880.5440 Intravascular administration set .
- ISO 10555 Sterile, single use intravascular catheters (Part 1: General Requirements) .
- ISO 10555 Sterile, single use intravascular catheters (Part 5: Peripheral Catheters). .
- ISO 11135:1994 Medical devices -- Validation and routine control of ethylene oxide sterilization ●
- ISO 11138-2:1994 Sterilization of health care products -- Biological indicators -- Part 2: Biological indicators . for ethylene oxide sterilization.
- . ISO 9626: 1991 Stainless Steel needle tubing for the manufacture of medical devices.
- ISO 11607: 1997 Packaging for terminally sterilized medical devices. ●
- ISO 8535: 1991 Sterile single use syringes, with or without needle, for insulin. .
- FDA Guidelines on validation of the Limulus Amebocyte Lysate (LAL) Test as an end product endotoxin test . for human and animal parenteral drugs, biological products, and medical devices.
- . ODE Blue Book Memorandum #K90-1.
- ISO 10993-1. Biological evaluation of medical devices Part 1: Evaluation and testing. .
And the design process adhered to is the Center for Devices and Radiological Health. DESIGN CONTROL GUIDANCE FOR MEDICAL DEVICE MANUFACTURERS. This Guidance relates to FDA 21 CFR 820.30 and Sub-clause 4.4 of ISO 9001. This is substantially equivalent to the predicate device.
Signed,
J.E.E.Z
Joel S. Douglas Chief Technology Officer
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three horizontal lines above it, representing the department's commitment to health, human services, and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol. The logo is black and white.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# MAY 1 7 2002
Mr. Joel Douglas Chief Technology Officer Sterling Medivations, Incorporated 66 Neptune Drive Groton, Connecticut 06340
Re: K021487
Trade/Device Name: Simplicity Easy Access P Infusion Set Regulation Number: 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: May 7, 2002 Received: May 8, 2002
Dear Mr. Douglas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{3}------------------------------------------------
# Page 2 - Mr. Douglas
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html
Sincerely yours,
for Susan Runner
Timothy A. Ulatowski Director Division of Dental. Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Ko21487
510(k) Number (if known):
Device Name: Simplicity Easy Access P Infusion Set
Indications For Use:
The intended use of the Simplicity Easy Access P Infusion Set is to provide a means @ for infusion and/or injection of fluids into the body below the surface of the skin when attached to a MiniMed Paradigm pump or syringe.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (PER 21 CFR 801.109) OR
Over-The-Counter Use
(Optional Format 1-2-96)
Patrucci Curente
(Division Sign-Off)
Division Of Dental, Infection Control, and General Hospital Hospital Device 510(k) Number ________________________________________________________________________________________________________________________________________________________________
Page 7 of 88
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.